Who uses it: hematology laboratory staff, clinic and procurement staff who stock blood-testing consumables, and students learning laboratory hematology. After reading, you should know what a Westergren tube is, how it differs from the other “ESR tube” setups in use, what regulators call it, and which specifications are fixed by the method.
What is a Westergren tube?
A Westergren tube (also called a Westergren pipette) is the long, upright, graduated tube used in the Westergren method for the erythrocyte sedimentation rate (ESR). The method measures how far red blood cells in a citrate-anticoagulated blood sample fall in a standardized vertical tube over one hour. The International Council for Standardization in Haematology (ICSH) describes the Westergren method as the gold standard for determining the ESR (ICSH recommendations, Int J Lab Hematol 2017, checked 2026-10-30).
“ESR tube” is a looser term. It can mean the Westergren pipette itself, a dedicated sedimentation tube, or, on some instruments, an ordinary EDTA blood tube with no separate ESR tube at all (see the types section).
How the tube is built and used
In the manual Westergren method, blood is diluted with sodium citrate at four parts blood to one part citrate, drawn into a 200 mm Westergren pipette (glass in the study cited below), and placed in a vertical stand. One 2024 study describes four volumes of blood to one volume of 3.8% sodium citrate, following the ICSH protocol, with a 200 mm glass Westergren pipette (PMC11324435, checked 2026-10-30). A 2024 multicentre analyzer study describes the same principle: the distance in millimetres that red cells fall in citrate-anticoagulated blood (4:1) in a standardized, upright, elongated tube over 1 hour under gravity, with the blood mixed gently at least 8 times, drawn to the 200 mm mark, and held upright in a rack at room temperature for 1 hour (PMC11431749, checked 2026-10-30).
Because the reading depends on gravity acting along the tube, the vertical stand is part of the device setup, not an accessory. A 2026 study that tested tubes tilted at 45 degrees notes that vertical (90 degrees) is the standard protocol and cites a published finding that a tilt of about 3 degrees from vertical can raise the ESR by roughly 30% (PMC13280657, checked 2026-10-30; the figure is that paper’s citation of earlier work).
Types and variants of ESR tubes
The reference method is not the only one in routine use. An ICSH survey of external quality assessment data from more than 6,000 laboratories on four continents found that only 28% used the unmodified Westergren method and 72% used modified or alternate methods. The same ICSH paper reports that results from newer instruments could differ from the Westergren method by up to 142%, that non-Westergren methods differed from each other by up to 42%, and that newer methods often used standard EDTA tubes, removing the need for a dedicated ESR tube (PubMed 28497537, checked 2026-10-30).
| Type | What it means for the tube | Source |
|---|---|---|
| Westergren method (reference) | Citrate-anticoagulated blood (4:1) in a 200 mm upright Westergren pipette, read after 1 hour | ICSH 2017; PMC11431749; PMC11324435 |
| Modified or alternate methods | Used by 72% of surveyed laboratories; some run on standard EDTA tubes with no dedicated ESR tube | ICSH 2017 (PubMed 28497537) |
| Automated analyzers | Regulated under a separate FDA product code from the sedimentation tube (see below); results can differ from Westergren | openFDA classification; ICSH 2017 |
How the FDA classifies it
In the FDA’s device classification database, product code GHC is “Tube, Sedimentation Rate”: Class 1, hematology specialty, regulation number 21 CFR 864.6700. Related codes under the same regulation include JPH (“Test, Erythrocyte Sedimentation Rate”), while GKB (“Device, Automated Sedimentation Rate”) sits under 21 CFR 864.5800 (openFDA device classification, checked 2026-10-30).
21 CFR 864.6700 defines an erythrocyte sedimentation rate test as a device that measures the length of time required for the red cells in a blood sample to fall a specified distance, or the degree of sedimentation in a given length of time. It is Class I (general controls) and exempt from premarket notification under subpart E of part 807, subject to the limitations in 864.9 (Cornell LII copy of the CFR, checked 2026-10-30).
Specifications that matter
| Specification | Value in the sources | Source |
|---|---|---|
| Tube length / fill mark | 200 mm; blood drawn to the 200 mm mark | PMC11431749; PMC11324435 |
| Blood to citrate ratio | 4:1 (sodium citrate 3.8% in one study) | PMC11324435; PMC13280657 |
| Mixing before filling | Gently, at least 8 times | PMC11431749 |
| Position | Vertical (90 degrees) in a rack or stand | PMC13280657; PMC11431749 |
| Reading time | 1 hour at room temperature | PMC11431749 |
Choosing a setup
The sources fix the method rather than a brand: if a laboratory reports the Westergren method, the tube must suit a 200 mm, vertical, citrate 4:1 setup. If it uses a modified or alternate method, the ICSH paper’s point is that the result belongs to that method and can differ from Westergren, so the method name travels with the tube choice. Confirm product details against the manufacturer’s instructions for use.
Safe use and handling basics
Tubes holding blood are specimens: label them correctly and handle glass with the same care as other laboratory glassware and sharps. See the CASRAI guides on specimen labeling error prevention, sharps disposal regulations for laboratories and laboratory PPE. For how other blood collection tubes are identified, see PST tube color. For the wider discipline, see what is hematology. This page does not interpret any ESR result.
Common mistakes
- Treating “ESR tube” as one product: Westergren pipettes, dedicated sedimentation tubes and EDTA tubes on analyzers are different setups (ICSH 2017).
- Comparing results across methods as if they were interchangeable: the ICSH survey reports differences of up to 142% against Westergren and up to 42% between non-Westergren methods.
- Letting the tube lean: a tilt of about 3 degrees from vertical is reported to raise the ESR by roughly 30% (cited in PMC13280657).
FAQ
What is an ESR tube?
It is the tube in which the erythrocyte sedimentation rate is measured. In the Westergren method it is a 200 mm upright pipette filled with citrate-anticoagulated blood; in some modified methods it is a standard EDTA tube read by an instrument (ICSH 2017).
What are the ESR tube types?
The sources describe the Westergren pipette (reference method), modified or alternate methods, and automated analyzers. See the types table above.
What is a Westergren tube used for?
It is used to measure how far red cells settle in one hour under the Westergren method. Under 21 CFR 864.6700, the test device measures the time red cells take to fall a specified distance or the degree of sedimentation in a given time.
Sources
- 21 CFR 864.6700, Erythrocyte sedimentation rate test: law.cornell.edu/cfr/text/21/864.6700 (checked 2026-10-30)
- openFDA device classification, product code GHC: api.fda.gov (checked 2026-10-30)
- ICSH recommendations for modified and alternate methods measuring the ESR, Int J Lab Hematol 2017: PubMed 28497537 (checked 2026-10-30)
- Multicentre ESR analyzer study, 2024: PMC11431749 (checked 2026-10-30)
- Manual Westergren study, 2024: PMC11324435 (checked 2026-10-30)
- Tube-angle study, 2026: PMC13280657 (checked 2026-10-30)