Written and maintained by CASRAI Editorial Board
Published
On this page
Tube filters, also called centrifugal filters, spin filters or filtration tubes, are single-use devices in which a filter cup (or sample reservoir) sits inside or on top of a collection tube and centrifugal force drives liquid through a membrane. Depending on the membrane, they either retain molecules above a molecular weight cut-off (ultrafiltration) or retain particles above a pore size (microfiltration), so one spin can clarify, concentrate, desalt or buffer-exchange a small aqueous sample without a syringe or vacuum manifold.
The specifications below come from manufacturer user guides for the Pall Microsep Advance and Macrosep Advance centrifugal devices and the Millipore Ultrafree-MC and Ultrafree-CL filter devices. Values are for those named products; other brands differ, so read the instructions for use that ship with the device you hold.
What tube filters are and who uses them
A tube filter is a filtration device sized to fit a standard centrifuge tube position. The Pall guides describe a screw-on cap, a sample reservoir with a sealed membrane, and a graduated filtrate receiver tube. The Millipore Ultrafree guide describes a filter cup plus a filtrate collection tube with a cap. Centrifugation supplies the driving force, which is why no pump or vacuum line is needed.
They are used by researchers and laboratory staff in molecular biology, protein chemistry and analytical chemistry. The Pall guides list ultrafiltration uses such as concentrating and desalting proteins and nucleic acids, buffer exchange, deproteinizing serum or urine before HPLC, recovering material from supernatants and lysates, and separating primers from PCR product. Microfiltration uses include recovering DNA from agarose gels, clarifying samples before HPLC, removing cells from media, and collecting or washing beads. These are laboratory sample-preparation devices, not patient-care products.
How a tube filter is built
In the Pall Microsep Advance design, the sample reservoir holds a paddle with the membrane sealed on both sides. Liquid and small molecules pass through the membrane into the receiver tube while larger molecules stay in the reservoir. The reservoir, receiver and cap are polypropylene and the paddle is polyethylene. The ultrafiltration membrane is Omega, a modified polyethersulfone; the microfiltration membranes (0.2 and 0.45 µm) are Supor polyethersulfone.
The Millipore Ultrafree-MC and -CL devices are microfiltration units with a cup and tube. Their membranes are Durapore PVDF (0.1, 0.22, 0.45, 0.65 and 5.0 µm) or PTFE (0.2 and 0.45 µm). The cup is PVDF or polypropylene depending on the membrane, and the tube is polypropylene. The Ultrafree guide states these are single-use, disposable filters for removing particulates from aqueous biological solutions.
Two features set practical limits. The hold-up volume is liquid that stays in the device after filtering. The dead stop in the Pall devices is a built-in feature that prevents the sample being spun to dryness; the remaining volume depends on rotor angle (for Microsep Advance, 100 µL at 34°, 80 µL at 45° and 65 µL in a swinging bucket).
Ultrafiltration vs microfiltration tube filters
Per the Pall Microsep guide, ultrafiltration devices operate typically at 3,000 to 7,500 × g and retain molecules larger than the nominal molecular weight cut-off (MWCO), while microfiltration devices run up to 14,000 × g and retain particulate larger than the membrane pore size. See the separate guide on molecular weight cut-off for how nominal cut-offs are defined.
| Feature | Ultrafiltration tube filter | Microfiltration tube filter |
|---|---|---|
| What is retained | Molecules above the nominal MWCO | Particles above the pore size |
| Membrane in the Pall devices | Omega (modified polyethersulfone) | Supor polyethersulfone, 0.2 or 0.45 µm |
| Typical use (Pall) | Concentrate, desalt, buffer exchange, deproteinize | Clarify samples, remove cells, recover DNA from gels, wash beads |
| Maximum g-force (Pall Microsep) | 7,500 × g | 14,000 × g |
| Maximum g-force (Pall Macrosep) | 5,000 × g | 14,000 × g |
Specifications that matter
The data table below compares six device configurations from the four manufacturer guides: maximum sample volume, maximum relative centrifugal force, membrane, membrane area, hold-up volume and the centrifuge fit. Three points in it affect planning most.
Format and centrifuge fit
The Pall Microsep Advance format is 17 mm × 12.0 cm and fits centrifuges that accept 17 × 100 mm conical tubes. Macrosep Advance is 29 mm × 12.0 cm and fits centrifuges accepting 50 mL conical tubes. Ultrafree-MC is for a fixed-angle rotor holding 1.5 mL tubes and Ultrafree-CL for a fixed-angle rotor holding 15 mL tubes. If you are comparing the tube formats themselves, see conical centrifuge tubes.
Operating limits
The Pall guide lists 0 to 40 °C and pH 1 to 14. The Ultrafree guide lists a 50 °C maximum. The Ultrafree guide includes a chemical compatibility table by membrane (for example xylene is listed as not recommended for both the PVDF and PTFE types), so check it before spinning solvents.
RCF is not rpm
Limits are given in × g, not rpm. The Ultrafree guide gives the conversion RCF = 1.118 × 10-5 × radius (cm) × rpm2. For a rotor radius of 10 cm, 5,000 rpm gives 1.118 × 10-5 × 10 × 25,000,000 ≈ 2,795 × g, well under the 7,500 × g ceiling of a Microsep ultrafiltration device. The same rpm on a larger rotor can exceed a device limit, which is why the calculation matters.
Choosing a molecular weight cut-off or pore size
Only criteria stated in the manufacturer guides are listed here. The Pall guide advises that for maximum retention you pick an MWCO 3 to 6 times smaller than the molecular weight of the molecule you want to keep. It also warns that ionic conditions, molecular conformation and protein-protein interactions affect retention, and recommends pretesting retentivity for your application.
| MWCO of device | Typical biomolecule size retained (Pall Table 1) | Suggested spin speed (Microsep, Pall Table 1) |
|---|---|---|
| 1K | 3K to 10K | 5,000 to 7,500 × g |
| 3K | 10K to 20K | 5,000 to 7,500 × g |
| 10K | 30K to 90K | 5,000 to 7,500 × g |
| 30K | 90K to 180K | 5,000 to 7,500 × g |
| 100K | over 300K | 1,000 to 3,000 × g |
Worked example: a 150 kDa protein
To retain a protein of 150,000 Da, the 3 to 6 times rule points to a device of 25,000 to 50,000 Da, so the Pall guide’s example picks the 30K device, which sits in that range.
Worked example: a 2 kb double-stranded DNA fragment
The Pall guide estimates DNA molecular weight as 340 × bases for single-stranded and 680 × base pairs for double-stranded DNA. A 2 kb double-stranded fragment is 680 × 2,000 = 1,360,000 Da, and the guide’s example selects a 100K device for it.
Choosing a pore size for microfiltration
For particle removal, the choice is the membrane pore size and chemistry offered by the device. The Ultrafree guide notes that many factors affect nonspecific protein binding, including membrane, pH, buffer strength, contact time, surface tension, concentration and the protein itself, and states that its Durapore PVDF has the lowest protein binding of its microfiltration membranes (a manufacturer claim).
Safe use and handling basics
- Pipette the sample into the reservoir or filter cup held vertically, and recap to prevent evaporation (Pall and Ultrafree guides).
- Balance the rotor. The Pall guide says to counterbalance with a device holding an equal volume; the Ultrafree guide says to balance a fixed-angle rotor.
- Stay within the maximum g-force printed for the device and the rotor.
- Manufacturer-typical spin durations: Pall lists ultrafiltration at typically 30 to 90 minutes and microfiltration at 1 to 3 minutes; Ultrafree-MC is typically 1 to 4 minutes at 12,000 × g and Ultrafree-CL 1 to 4 minutes at 5,000 × g. Determine the time and g-force for your own application.
- Pall devices are supplied non-sterile and the guide says they may be sanitized by filtering 70% ethanol through before use. The Omega membrane contains trace glycerine, and the guide describes a rinse option as well as a pre-rinse option for nonspecific adsorption.
- Recover the retained sample from the reservoir by pipette, or the filtrate from the receiver tube.
Common mistakes
- Setting rpm instead of × g. Device limits are in × g; convert with the rotor radius.
- Unbalanced rotor. Both manufacturers say to balance the load.
- Choosing an MWCO close to the target size. The Pall guide recommends 3 to 6 times lower for maximum retention and pretesting.
- Ignoring the dead-stop volume. Remaining liquid is a floor set by the device and rotor angle, not a defect.
- Skipping chemical compatibility. Membrane and housing materials differ; use the compatibility table in the user guide.
- Using the wrong tool for the job. A tube filter is not interchangeable with a syringe filter, which is pushed through by hand, and it is not the same as tangential flow filtration, which is a different filtration format.
Frequently asked questions
What is a centrifugal filter?
A centrifugal filter is a tube filter: a membrane device that fits in a centrifuge so that spinning forces liquid through the membrane. Pall describes centrifugation as the driving force, with molecules larger than the nominal MWCO retained in the reservoir.
What is a filtration tube?
The term usually means the collection tube that receives filtrate, or the whole filter-plus-tube assembly. In the Ultrafree guide the device is a filter cup plus a filtrate collection tube; in the Pall guides it is a reservoir plus a graduated filtrate receiver.
How is a centrifugal filter different from a syringe filter?
A tube filter is driven by centrifugal force in a centrifuge tube position, while a syringe filter is driven by hand pressure from a syringe. See the syringe filter guide for that format.
Why use g-force instead of rpm?
Because force depends on rotor radius as well as speed. The Ultrafree guide gives RCF = 1.118 × 10-5 × radius (cm) × rpm2, and device limits are written in × g.
Can I spin the sample to dryness?
The Pall guide states that its built-in dead stop prevents spinning to dryness, leaving a small residual volume that depends on rotor angle.
Sources
- Pall Corporation, Microsep Advance Centrifugal Devices user guide (document 1345_89118C): https://www.pall.com/content/dam/pall/japan/laboratory/literature-library/non-gated/Microsep_PII.pdf, read 2026-11-14.
- Pall Corporation, Macrosep Advance Centrifugal Devices user guide (document 1344_89113C): https://www.pall.com/content/dam/pall/japan/laboratory/literature-library/non-gated/Macrosep_PII.pdf, read 2026-11-14.
- Millipore, Ultrafree-MC and Ultrafree-CL user guide (hosted copy): https://millipore.nt-rt.ru/images/manuals/ultrafree-mc_cl_eng.pdf, read 2026-11-14.
Data: Tube filter (centrifugal filter) device specifications from manufacturer guides
Specification comparison of six tube filter configurations compiled from the Pall Microsep Advance, Pall Macrosep Advance and Millipore Ultrafree-MC/-CL user guides: maximum sample volume, maximum relative centrifugal force, membrane, membrane area, hold-up volume and the centrifuge tube format each fits. Values are manufacturer-stated for these named products only.
| Device configuration | Maximum sample volume (mL) | Maximum RCF (x g) | Membrane | Membrane area (cm2) | Hold-up volume (uL) | Centrifuge fit |
|---|---|---|---|---|---|---|
| Pall Microsep Advance, ultrafiltration | 5 | 7500 | Omega (modified polyethersulfone) | 3.3 | 40 | Centrifuges accepting 17 x 100 mm conical tubes |
| Pall Microsep Advance, microfiltration | 5 | 14000 | Supor polyethersulfone, 0.2 or 0.45 um | 3.3 | 40 | Centrifuges accepting 17 x 100 mm conical tubes |
| Pall Macrosep Advance, ultrafiltration | 20 | 5000 | Omega (modified polyethersulfone) | 7.2 | 80 | Centrifuges accepting 50 mL conical tubes |
| Pall Macrosep Advance, microfiltration | 20 | 14000 | Supor polyethersulfone, 0.2 or 0.45 um | 7.2 | 80 | Centrifuges accepting 50 mL conical tubes |
| Millipore Ultrafree-MC | 0.5 | 12000 | Durapore PVDF or PTFE | 0.2 | 5 | Fixed-angle rotor holding 1.5 mL tubes |
| Millipore Ultrafree-CL | 2 | 5000 | Durapore PVDF or PTFE | 0.8 | 30 | Fixed-angle rotor holding 15 mL tubes |
Sources
- Pall Microsep Advance Centrifugal Devices user guide (accessed 2026-11-14)
- Pall Macrosep Advance Centrifugal Devices user guide (accessed 2026-11-14)
- Millipore Ultrafree-MC and Ultrafree-CL user guide (accessed 2026-11-14)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.