Bacterial Incubator: Types, Specs, and Operation Guide

A complete technical guide to bacterial incubators: gravity vs forced convection, DIN 12880 standards, temperature mapping, sizing, and daily maintenance.

Written and maintained by CASRAI Editorial Board

Published

On this page

A bacterial incubator is a temperature-controlled laboratory chamber engineered to maintain optimal environmental conditions—primarily precise thermal regulation between ambient plus 5 °C and 80 °C—for cultivating, propagating, and maintaining bacterial and fungal cultures on solid agar media or in liquid broths. Unlike general laboratory ovens that operate at high temperatures for drying or glassware sterilization, a bacterial incubator provides gentle, uniform, and stable heating with tight spatial homogeneity (typically within ±0.3 °C to ±0.6 °C) to ensure reliable microbial growth kinetics.

This page describes products and general use; it is not medical advice.

Microbiologists, clinical laboratory scientists, quality control analysts, and biomedical researchers across diagnostic laboratories, pharmaceutical manufacturing facilities, food safety testing centers, and academic research institutions rely on bacterial incubators daily to perform colony isolation, antibiotic susceptibility testing, sterility assays, and microbial bioburden enumeration.

What a Bacterial Incubator Is and Primary Purpose

In microbiological workflows, temperature is the primary physical parameter governing enzymatic reaction rates, membrane fluidity, and cellular division. A bacterial incubator provides an isolated, insulated microenvironment where temperature remains rigidly constant over hours, days, or weeks. The instrument shields biological samples from ambient room temperature swings caused by facility heating, ventilation, and air conditioning (HVAC) cycling, seasonal shifts, and draft currents.

Standard microbiological incubators focus primarily on thermal management. Unlike mammalian cell culture chambers, a standard bacterial incubator does not incorporate active carbon dioxide (CO2) gas injection or saturated water reservoirs for active relative humidity generation, because typical bacterial media do not rely on a bicarbonate-carbonic acid buffering system to sustain neutral pH. Instead, bacterial cultures grow on nutrient-rich agars (such as tryptic soy agar, nutrient agar, or blood agar) or within buffered liquid broths (such as lysogeny broth or brain heart infusion) that maintain pH stability independently. For a comprehensive overview of how microbiological units fit into the broader landscape of laboratory heating instruments, see our overarching guide on what is an incubator.

Bacterial incubators are calibrated to standard microbiological set points. While the most widely utilized set point in clinical and clinical-adjacent research is 37.0 °C—mimicking the core physiological body temperature of human and mammalian hosts for pathogens such as Escherichia coli, Staphylococcus aureus, and Enterococcus faecalis—microbiological incubators also routinely support environmental and industrial assays. These include water quality testing at 44.5 °C for fecal coliforms, environmental yeast and mold isolation at 25.0 °C to 30.0 °C, and thermophilic spore testing at 55.0 °C to 65.0 °C.

How Bacterial Incubators Work and Chamber Construction

The engineering of a professional bacterial incubator balances thermal transfer efficiency, structural insulation, and hygienic cleanability. The chamber architecture consists of several specialized mechanical and electrical assemblies designed to maintain a uniform thermal envelope.

Chamber Liner and Materials: The internal chamber is constructed from high-grade, corrosion-resistant stainless steel. Premium units utilize electropolished DIN 1.4301 (AISI 304) stainless steel, while standard units often employ AISI 430. The internal liner features smooth, deep-drawn coved corners without sharp joints or external seams. This seamless geometry eliminates crevices where bacterial cells, fungal spores, or media residues could accumulate, while permitting rapid, thorough surface disinfection with liquid antimicrobial agents. Shelving systems utilize perforated stainless steel trays that slide into stamped internal wall runners, avoiding screw threads or removable brackets that harbor bioburden.

Thermal Insulation and Exterior Housing: The outer casing is fabricated from electro-galvanized sheet steel finished with an antimicrobial powder coating. Between the outer shell and the stainless steel interior liner sits a thick layer of high-performance thermal insulation, typically expanded polyurethane foam, high-density mineral wool, or advanced vacuum insulation panels. This insulation minimizes parasitic thermal transfer to the laboratory environment, keeping the exterior surfaces cool to the touch and reducing power consumption.

Dual-Door Architecture: Reliable bacterial incubators feature a dual-door system. The outer door consists of a fully insulated metal panel fitted with a continuous magnetic silicone gasket that seals the chamber perimeter. Behind the outer door sits an inner, independent transparent door made of tempered safety glass. This transparent barrier allows laboratory technicians to visually inspect Petri dishes, colony pigment development, and broth turbidity without disturbing the internal microclimate or collapsing the thermal boundary layer.

Heating Elements and Thermal Jackets: Heat generation occurs through precision electric resistance heating elements. Manufacturers implement either direct jacket heating or preheating air-jacket systems. In direct heating systems (such as perimeter foil heaters), elements surround the four or five internal chamber walls to radiate heat inward across a broad surface area, avoiding localized hot spots. In preheating chamber designs (such as BINDER’s APT.line™ technology), air is heated in an external conditioning plenum before entering the working chamber through symmetrical side walls, ensuring balanced horizontal thermal distribution.

Sensors and Microprocessor Control: Temperature is continuously monitored by high-grade Class A Platinum Resistance Thermometers (Pt100 RTDs) connected in a 4-wire configuration to eliminate lead resistance errors. A digital microprocessor controller running advanced Proportional-Integral-Derivative (PID) algorithms modulates electrical power to the heating elements. The PID controller continuously calculates the error between the measured temperature and the target set point, preventing overshoot during initial warm-up and rapidly compensating for thermal loss following door openings.

Types and Airflow Variants: Gravity vs. Mechanical Convection

The method by which air moves through the chamber defines the primary classification of dry bacteriological incubators. Selecting between natural gravity convection and forced mechanical circulation represents the most critical operational decision in incubator procurement.

Gravity Convection (Natural Convection): Gravity convection incubators operate entirely without internal fans. Air circulation is driven strictly by natural thermal buoyancy: as air comes into contact with the heated chamber walls or bottom heating elements, its density decreases, causing it to rise gently through the chamber. As it transfers heat to samples and reaches the cooler ceiling, it cools, increases in density, and circulates downward along opposite walls.

The principal advantage of gravity convection is exceptionally gentle, low-velocity airflow. In routine microbiology, agar plates incubated over 48 to 72 hours are highly vulnerable to desiccation. High-speed air currents strip moisture from agar gels, causing the media to crack, shrink from Petri dish walls, and concentrate selective salts, which can inhibit bacterial colony formation or alter minimum inhibitory concentration (MIC) values. Gravity convection minimizes this evaporation risk and prevents cross-contamination from airborne spores or powders. However, natural convection exhibits slower thermal recovery times (often 10 to 15 minutes after a 30-second door opening) and wider spatial temperature variance (typically ±0.5 °C to ±0.8 °C at 37 °C) when shelves are heavily loaded.

Mechanical Convection (Forced Air Circulation): Mechanical convection incubators incorporate an internal motorized circulation fan that actively forces air through distribution plenums and across the shelves. This active mixing produces exceptional temperature uniformity (typically ≤ ±0.3 °C at 37 °C) and rapid thermal recovery (frequently returning to set point within 2 to 4 minutes after door closure). Mechanical convection is ideal for high-throughput diagnostic facilities where doors are opened frequently, as well as for chambers carrying full sample loads or large-volume liquid culture flasks. To mitigate the risk of sample drying, modern forced-air units feature digitally adjustable fan speeds that allow operators to reduce circulation velocity to 20% to 40% during sensitive agar incubations.

Refrigerated and Cooling Incubators: Standard incubators can only regulate temperatures above the surrounding room ambient level (typically ambient +5 °C). However, many microbiological protocols require incubation at sub-ambient temperatures, such as 20.0 °C or 25.0 °C for environmental monitoring, food shelf-life testing, or Biochemical Oxygen Demand (BOD) assays. Refrigerated incubators integrate a compressor-driven refrigeration circuit or solid-state Peltier thermoelectric elements, allowing precise regulation from 0 °C or 4 °C up to 70 °C. To explore how specialized cooling incubators differ from chemical-oxidation analytical instruments, refer to our comparison of BOD incubator vs. COD reactor.

The following table outlines the operational trade-offs across bacteriological incubator airflow configurations:

Convection Mechanism Air Velocity and Pattern Spatial Uniformity (37 °C) Door Recovery Speed Agar Desiccation Risk Primary Use Case
Natural / Gravity Convection Very low (under 0.1 m/s); vertical thermal buoyancy ±0.5 °C to ±0.8 °C Slow (10–15 min) Minimal; preserves agar hydration Open Petri dishes, 48–72 hr cultures, delicate powders, low-throughput labs
Mechanical / Forced Air Moderate to high (0.2–0.5 m/s); horizontal fan plenum ±0.2 °C to ±0.4 °C Rapid (2–5 min) Elevated; requires fan throttling or plate covers High-throughput clinical labs, fully packed shelves, liquid broth flasks
Refrigerated (Peltier / Compressor) Active forced circulation with cooling evaporator ±0.3 °C to ±0.5 °C Moderate (4–8 min) Moderate; depends on fan set point BOD 20 °C testing, mycology at 25 °C, water quality, sub-ambient bacterial storage

Sizes and Specifications That Matter

Bacterial incubators are manufactured across a wide volume spectrum, from compact benchtop units to multi-door reach-in floor models. Selecting the proper chamber size requires calculating current specimen container dimensions, peak daily plate volume, and future expansion requirements.

Chamber Capacity Classes:

  • Compact Benchtop (30 L to 75 L): Accommodates 30 to 120 standard 90 mm Petri dishes. Ideal for satellite clinic workstations, point-of-care microbiology, teaching laboratories, and dedicated isolation protocols where cross-contamination between projects must be avoided.
  • Standard Mid-Size (100 L to 250 L): The workhorse configuration for hospital diagnostic microbiology and pharmaceutical QC, holding 200 to 500 Petri dishes across 3 to 6 shelves. Many mid-size models can be stacked two units high without special adapter frames to double capacity on a single bench footprint.
  • High-Capacity Reach-In (400 L to 750+ L): Free-standing floor units holding 800 to 2,000+ plates or large carboys of broth. Utilized in central industrial quality assurance, vaccine manufacturing, and contract testing facilities.

Technical Standards and Performance Specifications: In regulated laboratory environments, performance claims must conform to standardized testing methodologies. The premier international standard governing laboratory incubator testing is DIN 12880:2007-05 (‘Electrical laboratory devices – Heating ovens and incubators’). DIN 12880 defines exact testing procedures for spatial temperature variation (homogeneity across the chamber) and temporal temperature fluctuation (stability over time at the chamber center).

The table below summarizes verifiable engineering benchmarks from leading manufacturer series tested under DIN 12880 at 37.0 °C:

Manufacturer and Model Line Convection Architecture Internal Volume Operating Range Spatial Uniformity (DIN 12880) Temporal Fluctuation
Memmert IN55 Natural Convection (Gravity) 53 L (1.9 cu ft) Ambient +5 °C to +80 °C ±0.3 °C at 37 °C ±0.1 °C at 37 °C
Memmert IF55 Forced Convection (Mechanical) 53 L (1.9 cu ft) Ambient +10 °C to +80 °C ±0.3 °C at 37 °C ±0.1 °C at 37 °C
BINDER BD 56 Natural Convection (Gravity) 57 L (2.0 cu ft) Ambient +5 °C to +100 °C ±0.4 °C at 37 °C ±0.1 °C at 37 °C
BINDER BF 56 Forced Convection (Mechanical) 59 L (2.1 cu ft) Ambient +7 °C to +100 °C ±0.3 °C at 37 °C ±0.1 °C at 37 °C
Thermo Scientific Heratherm IGS60 Natural Convection (Gravity) 75 L (2.6 cu ft) Ambient +5 °C to +75 °C ±0.6 °C at 37 °C ±0.2 °C at 37 °C
Thermo Scientific Heratherm IMH100 Mechanical Convection (Adjustable Fan) 104 L (3.7 cu ft) Ambient +5 °C to +105 °C ±0.3 °C at 37 °C ±0.1 °C at 37 °C

In addition to incubation equipment, laboratory sterilization procedures rely on precise temperature and pressure controls; see our technical analysis on how to operate a lab autoclave: cycle types explained for protocols governing media and waste decontamination.

How to Choose a Bacterial Incubator for Your Facility

Selecting an incubator requires aligning product specifications with the regulatory rigor and workflow dynamics of your specific laboratory. When evaluating potential models, buyers should review four primary technical criteria established in manufacturer documentation and GLP guidelines:

1. Thermal Uniformity Requirements and Sample Vulnerability: If your protocol involves standard 24-hour diagnostic plating where door openings are infrequent, gravity convection offers excellent performance at lower purchase and maintenance costs while safeguarding agar moisture. Conversely, if your laboratory processes hundreds of patient specimens daily, experiences door openings every few minutes, or operates under stringent pharmaceutical release specifications (e.g. USP Chapter 1117), a mechanical convection unit with rapid recovery and verified spatial homogeneity under ±0.4 °C is mandatory.

2. Independent Temperature Safety Devices (DIN 12880 Classes): Under DIN 12880, laboratory heating devices are classified by their integrated safety systems. A Class 2 device provides basic overtemperature limiting to prevent equipment fires. However, for biological applications, facilities should insist on a Class 3.1 independent temperature safety device. Class 3.1 features an adjustable secondary controller with its own dedicated sensor. If the primary controller malfunctions and drives heating elements uncontrollably, the Class 3.1 unit assumes control, maintains temperature at a pre-set upper safety margin (e.g. 40.0 °C), and triggers an alarm, preventing the thermal denaturation and destruction of valuable clinical isolates.

3. Decontamination and Hygiene Routines: Consider the protocol required when contamination occurs. Basic units require manual wipe-down with liquid chemical disinfectants. Advanced microbiological incubators feature automated high-temperature decontamination routines, such as an overnight 140 °C to 180 °C dry heat sterilization cycle. An automated decontamination cycle eliminates bacterial vegetative cells and resistant fungal spores without requiring toxic gaseous vaporizers or aggressive chemical residues that could leach into future culture runs.

4. Data Integrity and Monitoring Connectivity: Facilities operating under ISO 17025, cGMP, or CLIA accreditation must maintain continuous, audit-ready temperature records. Modern incubators should offer external communication ports (Ethernet, USB, or RS-422) and dedicated access ports (typically 18 mm to 30 mm diameter) in the side or rear wall. Access ports allow independent external reference probes and data logger thermocouples to enter the chamber without pinching door gaskets, preserving seal integrity.

Safe Operation, Placement, and Aseptic Handling Basics

Even the most advanced incubator cannot compensate for poor laboratory placement or careless handling. Establishing sound operational practices ensures equipment longevity and microbial culture integrity.

Facility Siting and Environmental Clearance: Bacterial incubators must be positioned on a sturdy, level bench or dedicated floor stand away from environmental turbulence. Never place an incubator in direct sunlight, adjacent to autoclaves or hot plates, or directly beneath an HVAC ceiling vent or air conditioning supply diffuser. External ambient temperature fluctuations degrade the PID controller’s predictive modeling, causing artificial cycling. Maintain at least 10 cm to 15 cm (4 to 6 inches) of clear space behind and beside the unit to allow unimpeded convective ventilation across exterior cooling louvers and electronics housings.

Chamber Loading and Airflow Obstruction: Overcrowding is the most common operational failure in microbiology laboratories. To maintain spatial homogeneity:

  • Never line perforated shelves with aluminum foil, plastic liners, or absorbent paper; doing so completely blocks convective airflow loops.
  • Limit shelf loading to no more than 70% of the usable surface area, leaving open channels between culture vessels.
  • Stack Petri dishes no more than 4 to 5 plates high to prevent tipping and ensure uniform thermal penetration to center plates.
  • Stagger plate stacks across adjacent shelves rather than clustering all samples directly above one another.
  • Maintain a 3 cm perimeter gap between samples and the interior chamber walls to prevent direct conductive hot spots.

Aseptic Practices During Loading: Contamination inside an incubator is almost always user-introduced. Plates must be dry before loading to prevent condensation droplets from bridging lids and bases. Technicians should handle vessels with sanitized gloves and minimize door-opening frequency and duration. For standardized clean handling and barrier protocols, consult our detailed resource on aseptic technique in the lab.

Routine Cleaning and Sanitization: Establish a bi-weekly or monthly maintenance cadence. Switch off and unplug the unit, remove all shelves and runner racks, and clean all interior stainless steel surfaces with a neutral laboratory detergent. Follow with a thorough wipe-down using 70% sterile isopropanol or a validated non-corrosive quaternary ammonium disinfectant. Caution: Never clean stainless steel chamber interiors with bleach (sodium hypochlorite) solutions; chloride ions aggressively attack the chromium oxide passivation layer, initiating irreversible pitting corrosion and micro-cracking.

Temperature Mapping, Validation, and Calibration Protocols

In regulated environments governed by ISO 7218, USP Chapter 1117 (‘Microbiological Best Laboratory Practices’), and USP Chapter 1058 (‘Analytical Instrument Qualification’), proving that an incubator maintains valid temperatures requires formal validation. Relying solely on the incubator’s built-in digital display is insufficient for compliance, as a single-point controller sensor cannot detect chamber-wide thermal variations.

Calibration vs. Temperature Mapping: A critical compliance distinction exists between sensor calibration and chamber mapping. Calibration verifies the electronic accuracy of a specific sensor against a traceable reference standard (such as a NIST-certified thermometer). Temperature mapping, by contrast, evaluates the entire three-dimensional thermal profile of the chamber under real-world operating conditions, identifying geometric hot and cold spots. For specialized methodologies in mapping gas and thermal dynamics, see our related guide on CO2 incubator calibration and temperature uniformity mapping.

DIN 12880 Temperature Mapping Methodology: Under DIN 12880:2007-05, comprehensive temperature mapping requires placing calibrated temperature sensors across an organized grid throughout the chamber volume:

  • For chambers up to 50 liters: a minimum of 9 measurement points (4 corners and 1 center point on a lower plane, and 4 corners on an upper plane).
  • For chambers greater than 50 liters: a minimum of 27 measurement points distributed across three horizontal planes (9 sensors on the bottom plane, 9 on the middle plane, and 9 on the top plane, covering all four corners, four edge midpoints, and the geometric center of each plane).

Mapping Execution: Data loggers record readings at 1- to 5-minute intervals across a minimum 24-hour continuous baseline run. Mapping is conducted first with the chamber empty, followed by a loaded chamber run representing maximum routine biological mass. An opening-recovery stress test is also performed: the outer and inner doors are held open to 90 degrees for exactly 30 seconds, and the logger records the precise time required for all 27 sensors to recover within ±0.5 °C of the set point.

Thermal Buffering of Daily Monitoring Probes: Standard quality standards, including ISO 7218, dictate daily temperature checks. However, a bare sensor suspended in chamber air reacts instantly to transient air displacement whenever the door is opened, causing erroneous alarm triggers and inaccurate log entries. To reflect true sample conditions, daily temperature probes or reference thermometers must be immersed in a thermal buffer—such as a sealed 50 mL polypropylene bottle filled with liquid glycerol, mineral oil, or glass beads. The thermal inertia of the buffer mimics the heat capacity of liquid broths and solid agar plates, providing an accurate representation of microbial specimen temperature.

Common Operational Mistakes and Troubleshooting

Preventing protocol failures requires recognizing the operational pitfalls that commonly compromise incubator performance:

  1. Trusting the Controller Display Without Independent Verification: Many lab operators assume that because the front LCD reads ‘37.0 °C’, the entire chamber is at 37.0 °C. In unmapped or poorly calibrated units, corner shelves or door perimeters may sit at 35.2 °C or 38.6 °C—temperatures sufficient to delay slow-growing pathogens or inactivate heat-sensitive enzymes. Always maintain an independent, calibrated reference thermometer in the chamber center.
  2. Foil-Lining Shelves: Lab workers frequently line wire or perforated shelves with aluminum foil to catch potential media drips. This practice chokes vertical convective heat transfer, creating severe artificial temperature gradients between upper and lower shelves. Use dedicated, autoclavable drip trays that leave peripheral air gaps if spill containment is needed.
  3. High-Velocity Desiccation of Agar Plates: Running mechanical convection incubators at 100% fan speed during multi-day plate incubations rapidly dries out agar. If colonies appear stunted or agar shows edge shrinkage, reduce fan speed to 20% to 30%, utilize breathable Petri dish sleeves, or introduce an unheated water tray to maintain ambient chamber humidity.
  4. Excessive Door Propping: Leaving incubator doors open while organizing or searching for culture plates collapses the internal thermal envelope. In gravity convection units, recovering from an extended 2-minute door opening can take up to 30 minutes, during which cultures remain at sub-optimal growth temperatures. Organize samples by shelf map and open doors only for the minimum time necessary.
  5. Corrosive Chemical Damage: Cleaning chamber walls with hypochlorite bleach or phenolic disinfectants causes catastrophic stainless steel degradation. If bleach must be used following a major biohazard spill, it must be neutralized immediately with sterile water and followed by a 70% ethanol or isopropanol rinse.
  6. Unbuffered Probe Alarms: Wiring an unbuffered environmental monitoring sensor directly into an automated facility monitoring system leads to constant nuisance alarms during normal working hours. Always utilize a glycerol or glycol thermal buffer block on secondary probes.

Frequently asked questions

What is the difference between a bacterial incubator and a CO2 incubator?

A bacterial incubator controls temperature only (typically ambient +5 °C to 80 °C) without active gas management, making it ideal for bacteria and fungi grown on nutrient agar or in buffered broth. A CO2 incubator actively controls carbon dioxide levels (typically 5% CO2) and high relative humidity (above 95%) to maintain physiological pH in bicarbonate-buffered mammalian cell cultures.

Why is gravity convection preferred over mechanical convection for agar plates?

Gravity convection moves air through natural thermal buoyancy without an internal fan, creating low-velocity airflow that prevents open or loose-lidded agar Petri dishes from desiccating during extended 48- to 72-hour incubations. Mechanical convection fans can dry out agar media unless fan speeds are throttled down.

What does DIN 12880 compliance mean for an incubator?

DIN 12880 is the standard testing norm for laboratory heating devices and incubators. Compliance certifies that the manufacturer tested and documented spatial temperature uniformity across 9 to 27 measurement points and temporal stability at 37 °C under rigorous, reproducible conditions, ensuring verified thermal performance.

How often should a microbiological incubator undergo temperature mapping?

Under GLP, GMP, and ISO 17025 standards, incubators should undergo full multi-point temperature mapping annually, as well as following any major component repair (such as replacing heating elements, controllers, or gaskets) or facility relocation.

Why should temperature monitoring probes be placed in a glycerol buffer?

Placing monitoring probes inside a sealed vial of glycerol or mineral oil simulates the thermal mass of biological culture vessels. This prevents false alarm spikes caused by transient room air rushing into the chamber during routine, brief door openings.

Can a bacterial incubator be used as a drying oven?

No. Bacterial incubators generally have a maximum temperature limit of 70 °C or 80 °C and are designed for stable incubation without active exhaust venting. Laboratory drying ovens operate between 150 °C and 250 °C with powered exhaust vents to actively drive off moisture and sterilize glassware.

Sources

  • Memmert GmbH + Co. KG: Technical Data Sheet Incubator IN55 and IF55 (https://www.memmert.com, accessed 2026-11-16)
  • BINDER GmbH: Product Data Sheet Series BD and BF Microbiological Incubators (https://www.binder-world.com, accessed 2026-11-16)
  • Thermo Fisher Scientific Inc.: Heratherm General Protocol Microbiological Incubators Specifications and User Manual (https://www.thermofisher.com, accessed 2026-11-16)
  • Deutsches Institut für Normung: DIN 12880:2007-05 Electrical Laboratory Devices – Heating Ovens and Incubators (https://www.din.de, accessed 2026-11-16)
  • International Organization for Standardization: ISO 7218:2007/Amd 1:2013 Microbiology of Food and Animal Feeding Stuffs – General Requirements and Guidance for Microbiological Examinations (https://www.iso.org, accessed 2026-11-16)
  • United States Pharmacopeial Convention: USP Chapter 1117 Microbiological Best Laboratory Practices (https://www.usp.org, accessed 2026-11-16)
  • ASTM International: ASTM E145-19 Standard Specification for Gravity-Convection and Forced-Ventilation Ovens (https://www.astm.org, accessed 2026-11-16)

Data: Bacterial and Microbiological Incubator Performance Benchmarks

A comparative technical benchmark of commercial microbiological incubators evaluating airflow mechanisms, chamber capacities, operating ranges, spatial temperature uniformity at 37 °C, temporal stability, and DIN 12880 safety classifications, compiled from certified manufacturer engineering specifications and testing documentation.

Period covered: 2024/2026. Coverage: Global.

Bacterial and Microbiological Incubator Performance Benchmarks
Incubator ModelConvection ModeChamber Volume (liters)Operating Range (deg C)Spatial Uniformity at 37 C (+/- deg C)Temporal Stability at 37 C (+/- deg C)Over-Temperature Safety
Memmert IN55Natural Convection (Gravity)53Ambient +5 to 800.30.1DIN 12880 Class 3.1
Memmert IF55Forced Convection (Mechanical)53Ambient +10 to 800.30.1DIN 12880 Class 3.1
BINDER BD 56Natural Convection (Gravity)57Ambient +5 to 1000.40.1DIN 12880 Class 3.1
BINDER BF 56Forced Convection (Mechanical)59Ambient +7 to 1000.30.1DIN 12880 Class 3.1
Thermo Scientific Heratherm IGS60Natural Convection (Gravity)75Ambient +5 to 750.60.2Class 2 Limiter / Audible Alarm
Thermo Scientific Heratherm IGS100Natural Convection (Gravity)117Ambient +5 to 750.60.2Class 2 Limiter / Audible Alarm
Thermo Scientific Heratherm IMH100Mechanical Convection (Adjustable Fan)104Ambient +5 to 1050.30.1DIN 12880 Class 3.1

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Bacterial Incubator: Types, Specs, and Operation Guide

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →