Bubble CPAP is a way of giving a baby continuous positive airway pressure (CPAP) in which the expiratory tubing is submerged in water, and the depth and the bubbling water help produce the pressure. This page explains the device class for clinic staff, buyers and students: what it is, how the pressure is generated, how it compares with other CPAP systems, and what published studies report. It describes products and general use; it is not medical advice.
Who uses it: neonatal care units, particularly for preterm infants with respiratory distress, and hospitals in low- and middle-income countries where the published trials were run (Gross et al. 2025; Martin et al. 2014).
What a bubble CPAP is
A bubble CPAP is a nasal CPAP system for neonates that generates pressure with an underwater seal. Herrera and colleagues define the underwater seal as “a type of threshold resistor in which the positive pressure generated is associated with the immersion depth of the expiratory tubing in a water container” (Herrera et al. 2016). It is the same therapy family as the nasal CPAP used in neonatal intensive care, which the same authors link to less invasive mechanical ventilation and less exogenous surfactant use.
Bubble CPAP is different from the adult sleep-apnea machines described in our CPAP machine guide, which use a motorised blower and a mask; the tubing used on those machines is covered in the CPAP hose guide. For the oxygen-delivery devices that are not CPAP, see the nasal cannula guide and the respiratory therapy equipment overview.
How a bubble CPAP system works
In the published description, bubble CPAP “generates continuous distending pressure by submerging the distal expiratory limb in water, producing low-amplitude, high-frequency pressure oscillations as airflow escapes and bubbles into the water canister” (Gross et al. 2025).
The water is not just a pop-off valve. Bench work found that the pressure generated was associated with both the immersion depth and the diameter of the bubble tubing (Herrera et al. 2016). Gross et al. add that the oscillations vary with airway and lung mechanics, so the pressure the infant experiences is not a static number set by a dial.
Types and variants
Herrera et al. classify CPAP generating systems as threshold resistors and flow resistors. The table below places bubble CPAP among the options the sources compare. Rows are drawn from the cited studies.
| System | How it is described | Key difference reported | Source |
|---|---|---|---|
| Bubble CPAP (underwater seal) | Threshold resistor; pressure associated with immersion depth of the expiratory tubing | Pressure also associated with bubble-tubing diameter in a lung-model bench study | Herrera et al. 2016 |
| Flow resistor combined with underwater seal | Flow resistor plus bubble tubing in the same bench set-up | With small bubble tubing it showed no important differences in the evaluated pressures and had a stabilizing effect on the generated pressures | Herrera et al. 2016 |
| Bubble CPAP versus oxygen therapy | Systematic review of 3 studies, infants under 28 days, low- and middle-income country hospitals | Reduced the need for mechanical ventilation by 30%-50% | Martin et al. 2014 |
| Bubble CPAP versus ventilator-derived CPAP | Systematic review of 3 trials, same setting | Mortality and complication rates similar; lower CPAP failure in the bubble groups | Martin et al. 2014 |
Why set and measured pressure can differ
The Herrera bench study ran at working pressures of 3-10 cm H2O and flows of 6-18 L/min on a lung model, not on patients. Its authors stress the importance of measuring the pressure actually generated, because they found differences between the working pressures set and the pressures measured. For a buyer or biomedical engineer this means that the water depth alone should not be treated as proof of delivered pressure; measurement is part of the system.
What the evidence says
Martin, Duke and Davis reviewed bubble CPAP in neonatal care in low- and middle-income countries. Against oxygen therapy it reduced the need for mechanical ventilation by 30%-50%; against ventilator CPAP, mortality and complication rates were similar and CPAP failure was lower in the bubble groups (p < 0.003). They concluded there was evidence that bubble CPAP is safe and reduces the need for mechanical ventilation. These findings apply to the populations and settings studied; they are not a statement about any individual product or patient.
Safe use and handling basics
Product-specific set-up, pressure and interface choices belong in the manufacturer’s instructions for use and the hospital’s own protocol, not on this page. A general point that the sources support: because delivered pressure can differ from the set value, the delivered pressure should be measured rather than assumed (Herrera et al. 2016).
Common mistakes
- Assuming water depth alone sets the pressure. Bench data tie pressure to depth and to bubble-tubing diameter, and set and measured values differed.
- Treating the water as only a pop-off valve. The source describes it as the mechanism that produces the pressure and its oscillations.
- Assuming results from one setting transfer everywhere. The trial evidence summarised above comes from low- and middle-income country hospitals and infants under 28 days.
FAQ
Is neonatal nasal CPAP the same as bubble CPAP?
No. Nasal CPAP is the therapy; bubble CPAP is one way of generating the pressure, using an underwater seal. Ventilator-derived CPAP is another, and Martin et al. compared the two.
How is bubble CPAP different from a ventilator?
The sources describe bubble CPAP as non-invasive respiratory support that gives continuous distending pressure. In the review, it was associated with a reduced need for mechanical ventilation compared with oxygen therapy alone.
Does the water do the work?
Yes, in the sense that the submerged expiratory limb produces the pressure and the small oscillations as gas bubbles out. Bench data link the pressure to immersion depth and tubing diameter, so depth is not the only factor.
Sources
- Herrera N, Regnicoli R, Murad M. Evaluation of a continuous-positive pressure generating system. Research and Reports in Neonatology 2016;6:11-15. doi 10.2147/RRN.S91056 (https://www.dovepress.com/evaluation-of-a-continuous-positive-pressure-generating-system-peer-reviewed-fulltext-article-RRN, checked 2026-11-01).
- Gross A, Campbell ER, Oliveira L, Sant’Anna G, Shalish W. Bubble CPAP in neonatal care: mechanisms, evidence, and pathways to optimization. Paediatr Respir Rev, online 27 December 2025. doi 10.1016/j.prrv.2025.12.004 (https://pubmed.ncbi.nlm.nih.gov/41547602/, abstract checked 2026-11-01).
- Martin S, Duke T, Davis P. Efficacy and safety of bubble CPAP in neonatal care in low and middle income countries: a systematic review. Arch Dis Child Fetal Neonatal Ed 2014;99(6):F495-504. doi 10.1136/archdischild-2013-305519 (https://pubmed.ncbi.nlm.nih.gov/25085942/, abstract checked 2026-11-01).