PCR Hood: What It Is, How It Works and What UV Does

A PCR hood keeps PCR set-up free of stray DNA with filtered airflow and UV. Types, specs and the limits of UV decontamination.

Written and maintained by CASRAI Editorial Board

Published

On this page

A PCR hood is an enclosed bench-top work area, usually with a filtered laminar airflow and a UV lamp, used to set up polymerase chain reaction (PCR) mixes without introducing stray DNA from the room air, the operator or the bench. Its purpose is to protect the reaction from contamination before it goes into the thermal cycler; it is a clean-work enclosure for the sample, not a containment device for hazardous material.

What a PCR hood is

PCR copies a few template molecules into billions, so a single stray amplified fragment from an earlier run can produce a false positive. A PCR hood gives the person setting up reactions a small, controllable space where the air, the surfaces and the pipetting tools are kept as clean as the enclosure allows. Manufacturers use several names for the same product class: PCR hood, PCR workstation, PCR enclosure, PCR cabinet and PCR laminar flow cabinet.

The manufacturer pages read for this guide describe the product in terms of the work area and the samples. Labconco describes its Purifier Filtered PCR Enclosures as projecting ISO Class 5 air over the work area, and does not classify them as biological safety cabinets. For the difference between sample-protecting and operator-protecting cabinets, see the comparison biosafety cabinet vs laminar flow hood, and for the general airflow principle see what is a laminar flow hood.

Who uses it: molecular biology, diagnostic and forensic laboratories, and any research group that runs PCR or RT-PCR (the same set-up area is used before RT-PCR runs), plus the technicians and students who prepare reaction mixes there.

How a PCR hood is built and how it works

Most filtered PCR hoods share four elements, described below from the manufacturer pages read for this guide.

  • Pre-filter and HEPA filter. Labconco states that room air is drawn through a pre-filter and a 99.99% efficient HEPA filter. Air Science states that its PCR cabinet uses a HEPA filter with 99.97% efficiency for 0.3 micron particulates.
  • Laminar airflow. Labconco states that ISO Class 5 air is continuously projected vertically down over the work area in a laminar flow pattern. Air Science describes the airflow as uniform in both velocity and direction.
  • Positive pressure. Labconco states that the enclosure operates under positive pressure. Air Science likewise describes PCR hoods as maintaining a positive pressure environment.
  • Germicidal UV lamp and timer. Both manufacturers include a UV lamp for decontaminating the work area. Labconco describes the UV light as secondary decontamination between PCR tests, with a digital timer that runs it continuously or in selected increments. Its bi-fold doors close the sash opening while the UV light is on to contain the UV rays.

ISO Class 5 refers to the particle-count classification of cleanroom air; the airflow itself is described in more detail in laminar flow hood certification, which covers how filters and airflow are tested.

Types of PCR hood

Two designs are described by the manufacturers read for this guide. The practical difference is whether air is filtered and moving, or still.

Type Typical use Key difference (as stated by the manufacturer)
Filtered laminar-flow PCR hood (PCR workstation or enclosure) Setting up PCR reactions where airborne particulate contamination must be limited HEPA-filtered air flows downward in a laminar pattern under positive pressure; a UV lamp decontaminates between runs (Air Science; Labconco)
Dead air box Setting up PCR reactions in a still-air enclosure No air circulation and no HEPA filtration, so it does not protect against airborne particulate contamination during PCR amplification; UV is used to decontaminate the workspace after each cycle (Air Science)

Because the second row comes from a manufacturer that sells both designs, treat it as that company’s statement of the difference, not as an independent test. The Labconco page lists its clean benches as separate products from its PCR enclosures, which is a reminder that a PCR hood is a sample-protection product with a defined UV step rather than a general clean bench.

Sizes and specifications

Few manufacturers publish full dimensions on their marketing pages. The values that were stated in the pages read for this guide are compared in the data table below the article body. Labconco offers its Purifier Filtered PCR Enclosures in 2-, 3- and 4-foot widths, UV timer settings of 5, 10, 15, 30, 60, 120 or 240 minutes (or continuous), and an optional airflow monitor. Air Science states its HEPA efficiency as a percentage at 0.3 micron particles.

The two HEPA figures are not directly comparable as quoted: they come from different manufacturers, and each states its own test basis only partly. A buyer or lab manager comparing models should read the filter test method in each manufacturer’s documentation instead of comparing the headline percentages alone.

How UV decontamination in a PCR hood works, and where it stops

The UV lamp is the feature that most distinguishes a PCR hood from an ordinary clean bench, and it is also the feature most often over-trusted. Two peer-reviewed studies give a more precise picture than product literature.

UV works on dried DNA, and cleaning first changes the result

Gršković and colleagues (Croatian Medical Journal, 2013) exposed DNA dried on surfaces to UVC. They compared a laboratory ceiling lamp (Philips TUV 36W, 193 cm above the bench, 4.56 mJ/cm² per minute) with a lamp of the type fitted in laminar hoods (TUV 15W at 20 cm, 153 mJ/cm² per minute). Naked DNA at 0.1 ng/μL lost its profile in 2 minutes whether or not the surface had been cleaned first. DNA in biological fluids was far harder to remove: diluted blood, saliva and semen took about 1 hour, and undiluted saliva took 4 hours with cleaning plus UVC against 54 hours with UVC only. Undiluted semen was not detected after 8 hours when cleaned first, but was still detectable at 250 hours with UVC alone. The authors recommended cleaning with 10% sodium hypochlorite and 20% ethanol followed by a minimum of 16 hours of UVC (about 4,380 mJ/cm²). The study used a single donor, so the exact times are indicative, not universal.

The two lamp figures also show why a hood lamp and a ceiling lamp are not interchangeable. The hood-type lamp delivered 153 mJ/cm² per minute at 20 cm against 4.56 mJ/cm² per minute for the ceiling lamp, about 33 times the dose rate. The recommended 16 hours at the ceiling lamp’s rate corresponds to about 4,380 mJ/cm²; by simple arithmetic, the same dose at the hood lamp’s rate would take roughly 29 minutes. That calculation is an illustration from the two published rates, not a validated hood cycle, and it assumes the same dose-response holds for both lamps.

The practical reading is that a short UV cycle after a run is aimed at the kind of contamination a PCR set-up usually leaves, which is small amounts of DNA on surfaces; it is not a substitute for cleaning, and it does not sterilise heavily soiled material.

UV treatment of reagents trades contamination against assay sensitivity

Humphrey and colleagues (PLoS ONE, 2015) worked in a Labcaire PCR workstation that was cleaned with a nucleic-acid-degrading disinfectant and given half an hour of UV exposure before and after set-up. They then tested UV treatment of the reagents themselves in a crosslinker at 254 nm. Reagents given 2.5 minutes of UV cut contamination from 100% to 5% of reactions (n = 42), but detection of 2 E. coli genome copies fell to 31%. UV also delayed quantification cycle values in a dose-dependent way. Treating primers with ethidium monoazide (EMA) and giving reagents 2 minutes of UV produced 3% contamination while keeping 100% detection of the 2-copy sample (against 75% for UV alone). The paper identified the primers as a contaminant source.

The lesson for hood users is that UV inside a hood cleans surfaces and exposed items, while contamination carried in the reagents themselves needs a separate control. UV applied to the reagents also has a cost to sensitivity.

Safe use and handling basics

  • Not for hazardous material. A PCR hood protects the reaction. The manufacturer pages read for this guide describe protection of the work area and do not describe protection of the operator, so hazardous, infectious or radioactive material belongs in an appropriate containment device as defined by the laboratory’s own biosafety assessment; the comparison linked above explains the distinction.
  • UV with the sash closed. Labconco states that its bi-fold doors close the sash opening while the UV light is on to contain the UV rays; follow the manufacturer’s instructions for use for the UV cycle and for any interlock.
  • Clean before UV. The Gršković study found cleaning before UV changed the time needed to degrade DNA in biological fluids. Humphrey and colleagues cleaned the workstation with a nucleic-acid-degrading disinfectant before and after set-up.
  • Follow the manufacturer’s instructions for airflow and filters. An airflow monitor is an optional accessory on the Labconco enclosures. How filters and airflow are verified is covered in the certification guide linked above.

Common mistakes

  • Treating UV as a complete decontamination step. The Gršković data show that the time needed depends on what the DNA is in and whether the surface was cleaned first.
  • Assuming a PCR hood protects the user. The manufacturer descriptions read for this guide cover sample protection, not operator protection.
  • Using a dead air box where airborne particles matter. Air Science states that dead air boxes lack HEPA filtration and so do not protect against airborne particulate contamination during PCR amplification.
  • Ignoring the reagents. In the Humphrey study the primers were a contaminant source, so a clean hood alone did not remove contamination.
  • Using the hood for the wrong consumables. UV does not reach everything equally, and the Humphrey study found contamination arriving with the reaction components; the choice of PCR tubes and other consumables is a separate contamination question from the hood itself.

Sources

Data: PCR hood features and values stated by manufacturers (Labconco, Air Science)

Compiles the specifications stated on two manufacturer pages for filtered PCR hoods and enclosures: HEPA efficiency, airflow pattern, pressure, UV timer settings and widths, plus the Air Science statement on dead air boxes. Each value is as stated by the named manufacturer; blank values were not stated.

PCR hood features and values stated by manufacturers (Labconco, Air Science)
FeatureLabconco Purifier Filtered PCR EnclosureAir Science PCR cabinet or hood
HEPA filter efficiency99.99% (HEPA, preceded by a pre-filter)99.97% for 0.3 micron particulates
AirflowISO Class 5 air projected vertically down in laminar flowUniform in velocity and direction; air cleanliness meets or exceeds ISO Class 5
PressurePositive pressurePositive pressure
UV decontaminationUV lamp with digital timer: continuous, or 5, 10, 15, 30, 60, 120 or 240 minutesGermicidal UV lamp for decontamination
Widths offered2, 3 and 4 feetNot stated
Airflow monitorOptionalNot stated

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about PCR Hood: What It Is, How It Works and What UV Does

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →