Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
A phoropter is a manual refractor: a set of lenses of various powers used to measure the refractive error of the eye. In the United States the FDA regulates it as a Class I device, exempt from premarket notification (510(k)), under product code HKN. Eye-care clinicians use it during a refraction exam, and clinic staff and buyers meet it as a line item on equipment lists.
This page describes products and general use; it is not medical advice.
What a phoropter is
The U.S. regulation that covers the device does not use the word “phoropter”. It defines a “manual refractor” in 21 CFR 886.1770 as a set of lenses of various dioptric powers intended to measure the refractive error of the eye. The FDA product classification database lists the device under the name “refractor, manual, non-powered, including phoropter” (product code HKN).
The University of Waterloo School of Optometry and Vision Science describes it in plain terms as “essentially a box full of lenses” with many dials and buttons that adjust lens power. The purpose is to assess refractive error such as myopia, hyperopia and astigmatism, which glasses or contact lenses can correct.
How it is built and used
Per the Waterloo description, the clinician changes the lenses in front of the patient’s eyes with the dials and buttons while the patient reads an eye chart, and the combination that gives the clearest view informs the correction. The regulatory definition is consistent: the device is the lens set itself, and it is non-powered. The published descriptions do not break the instrument into a standard list of named components beyond the lenses and the dials and buttons that change them.
Phoropter vs autorefractor
People often ask how a phoropter differs from an autorefractor. The FDA regulation does not use the word “autorefractor”. The nearest regulated description is the ophthalmic refractometer in 21 CFR 886.1760, an automatic AC-powered device; “autorefractor” is the common name many people use for this kind of instrument. The table sets the two regulatory definitions side by side.
| Feature | Manual refractor (phoropter), 21 CFR 886.1770 | Ophthalmic refractometer, 21 CFR 886.1760 |
|---|---|---|
| Power | Non-powered (FDA product name: “refractor, manual, non-powered, including phoropter”) | Automatic, AC-powered |
| What it is | A set of lenses of various dioptric powers | A device with a fixation system, a measurement and recording system, and an alignment system |
| What it measures | Refractive error of the eye | Refractive power of the eye, by measuring light reflexes from the retina |
| FDA class | Class I (general controls) | Class I |
| Premarket notification | Exempt, subject to the limitations in 21 CFR 886.9 | Exempt, subject to the limitations in 21 CFR 886.9 |
Sources: 21 CFR 886.1770 and 21 CFR 886.1760 (Cornell Legal Information Institute, checked 2026-10-31).
How it is regulated
The FDA classification record for product code HKN (checked 2026-10-31) shows class 1, regulation number 886.1770, review panel Ophthalmic and submission type 510(K) Exempt. It also states that the device is exempt from the good manufacturing practice regulation except for the general requirements concerning records and complaint files, as long as the device is not labeled or otherwise represented as sterile. The regulation was first published at 52 FR 33355 (September 2, 1987) and its most recent amendment is 90 FR 55991 (December 4, 2025).
Safe use and handling basics
The FDA’s Global Unique Device Identification Database (AccessGUDID) record for one HKN-coded refractor, the NIDEK RT-600 (checked 2026-10-31), lists the device as not sterile, not single-use and not made with natural rubber latex. That is one manufacturer’s record, not a rule for every refractor. For cleaning, disinfection and use, follow the manufacturer’s instructions for use for the specific model.
Common mistakes
- Treating “phoropter” and “autorefractor” as the same instrument. The regulations define a non-powered lens set and an automatic AC-powered measuring device separately.
- Assuming the device needs a 510(k). The classification record shows a manual refractor as 510(k) exempt, subject to the limits in 21 CFR 886.9.
- Assuming it is a sterile or single-use item. The GUDID record for the NIDEK RT-600 lists it as neither.
Who uses it
Optometrists and other eye-care clinicians use it in refraction exams. For the wider list of instruments and supplies in such a practice, see the optometry clinic supply checklist. A related instrument for examining the inside of the eye is covered in what is an ophthalmoscope. For the regulatory background, see what is a medical device and the Medical Device Amendments of 1976. Related products are listed under diagnostic, monitoring and imaging equipment.
FAQ
What is a phoropter?
A phoropter is a manual, non-powered refractor: a set of lenses of various powers, adjusted with dials and buttons, used to measure the refractive error of the eye. The FDA classifies it as Class I under 21 CFR 886.1770.
What are the parts of a phoropter?
A phoropter is described as a box of lenses of different powers with dials and buttons that change the lens power in front of the patient’s eyes. The regulation and the FDA classification record define it as a set of lenses and do not give a standard list of other named parts.
What is an autorefractor?
“Autorefractor” is the common name for an automatic instrument that measures the eye’s refractive power. The FDA regulation 21 CFR 886.1760 describes the corresponding ophthalmic refractometer as an automatic AC-powered device with fixation, measurement and recording, and alignment systems that measures refractive power from light reflexes from the retina.
Sources
- 21 CFR 886.1770, Manual refractor (checked 2026-10-31)
- 21 CFR 886.1760, Ophthalmic refractometer (checked 2026-10-31)
- FDA product classification, product code HKN (checked 2026-10-31)
- AccessGUDID record, NIDEK RT-600 refractor (checked 2026-10-31)
- University of Waterloo School of Optometry and Vision Science, article on the phoropter (checked 2026-10-31)