A plate centrifuge is a compact benchtop centrifuge built to hold microplates or PCR plates in swinging carriers and give them a short, low-speed spin so that liquid collects at the bottom of the wells. The two dedicated models whose instructions for use were reviewed for this page, the Greiner Bio-One Microplate Centrifuge (REF 846070) and the Benchmark Scientific PlateFuge C2000, are described by their makers as tools for quick spin-downs, not for separating samples at high force.
Searchers use several names for the same instrument: plate centrifuge, microplate centrifuge, centrifuge for plates, or centrifuge plate. This guide uses them interchangeably, compares the two documented models side by side in the data table below, explains why the g-force figure matters more than the rpm figure, and sets out the loading and cleaning rules the manufacturers themselves state.
What a plate centrifuge is
The Greiner Bio-One instructions for use (rev. September 2017) describe the instrument as intended only for indoor use and for separating aqueous solutions and suspensions of various densities, predominantly in research laboratories, and as designed for quick spin downs of approved PCR plates and microtiter plates. The Benchmark Scientific manual uses nearly the same wording: a unit designed for low-speed applications and quick spin downs of PCR plates and microtitre plates.
In practice the purpose is simple. After pipetting reagents into a 96-well plate, droplets can cling to the well walls or sit on a sealing film. A short spin brings that liquid down to the bottom of each well before the plate goes onto a thermal cycler or reader. A plate centrifuge does this for a whole plate at once, which an ordinary tube centrifuge cannot. For the plate formats involved, see the guide to 96-well plate dimensions; the SLAS lists the ANSI/SLAS microplate standards covering footprint, height and bottom outside flange dimensions.
Who uses it: molecular biology, screening and diagnostic-development laboratories that run PCR plates and microtiter plates, and the technicians and researchers who prepare them.
What is a benchtop centrifuge?
A benchtop centrifuge is a centrifuge small enough to stand on a laboratory bench rather than on the floor. Both plate centrifuges reviewed here fit that description: each measures 23 x 26 x 19.7 cm (9.2 x 10.2 x 7.75 in) according to its documentation, and the Greiner unit weighs 4.1 kg (9 lb). Benchtop is a size class, not a function, so benchtop models range from plate spinners like these to larger instruments with different rotors. Our guide to what a centrifuge is covers the family as a whole, and how to choose a centrifuge covers the wider selection question.
How a plate centrifuge works
The rotor carries two plate carriers (buckets). The Benchmark manual states that the carriers swing out automatically to a fully horizontal position during the spin, so the force acts along the long axis of the wells and liquid collects at the bottom of each well rather than on one side. When the run ends, the Benchmark manual says to pull the lid tab, after which the rotor brakes. Both manufacturers state that samples must not be removed until the rotor has come completely to a stop.
Both instruments have a removable rotor. The Greiner rotor is released with a thumb screw, and the Benchmark rotor with the Allen wrench supplied with the unit. A removable rotor matters for cleaning, covered below. For the part itself, see what a centrifuge rotor is.
Types and variants
Dedicated plate centrifuge
The two models described here are dedicated plate centrifuges: two carriers and a low-speed rotor. Each takes two PCR plates or two microplates at a time. The Benchmark manual also lists an adapter carrier that accepts 0.2 mL PCR strips and tubes, and gives its capacity as 192 x 0.2 mL tubes or 24 PCR strips.
Plate carriers in other centrifuges
A plate can also be spun in a general-purpose swing-bucket centrifuge fitted with plate buckets; our guide to the swing-bucket centrifuge explains that rotor design. A mini centrifuge is a different, smaller class aimed at tubes. The documentation reviewed here does not compare plate centrifuges with those instruments, so no performance comparison is made.
Specifications that matter
The data table below sets the published specifications of the two documented models next to each other. Three points stand out. Both are two-plate instruments with the same 8.0 cm maximum rotor radius. Both are rated in the hundreds of times gravity, not thousands. And both sit in the same housing size. The practical conclusion is that these are spin-down tools: the differences between them are small compared with the gap between a spin-down tool and a separation centrifuge.
Other documented points from the Greiner instructions: it runs on 120 V/60 Hz or 230 V/50 Hz, the operating environment is 4 to 45 degrees C, and the warranty is two years. The Benchmark manual lists 120/230 V at 50-60 Hz. The Benchmark manual is not consistent about top speed (its specification table says 2100 rpm / 400 x g while its operating text mentions 2200 rpm), which is one more reason to compare instruments by g-force rather than rpm.
RCF versus RPM: what is the difference?
RPM is how fast the rotor turns, in revolutions per minute. RCF, the relative centrifugal force, is the force the sample actually experiences, expressed in multiples of Earth’s gravitational field, written x g. Beckman Coulter defines RCF as a way to describe and compare the field strength of different rotors and speeds, with the radius measured from the centre of rotation. Because the force depends on both speed and radius, the same rpm gives different g on rotors of different size, so protocols should quote g.
A commonly published form of the relationship, with radius r in centimetres, is RCF = 1.118 x 10-5 x r x rpm2 (when r is in millimetres the constant changes accordingly). As an editorial worked check using the maximum radius of 8.0 cm that both manuals give:
- 2550 rpm: 1.118 x 10-5 x 8.0 x 25502 is about 582 x g. The Greiner instructions state 600 x g, so the calculation agrees to within rounding.
- 2100 rpm: 1.118 x 10-5 x 8.0 x 21002 is about 394 x g. The Benchmark specification table states 400 x g.
The calculation also shows that g rises with the square of speed: raising 2100 rpm to 2550 rpm is about a 21 per cent increase in speed but about a 48 per cent increase in force. Check the figures printed on your own instrument before relying on the arithmetic above, which is rounded.
What goes in a centrifuge like this?
The answer for a plate centrifuge is narrower than for a general centrifuge. Per the manuals, the permitted contents are PCR plates and microtiter plates (and, with the Benchmark adapter, 0.2 mL PCR strips and tubes). Greiner specifies approved plates, and states that plates and PCR strips must be visually inspected for damage and that damaged ones must not be centrifuged. It also states that the unit is not for explosive or highly reactive substances; for toxic, radioactive or pathogenic liquids, national safety rules apply and spills must be decontaminated.
Loading rules stated by the manufacturers
| Topic | Greiner Bio-One REF 846070 | Benchmark PlateFuge C2000 |
|---|---|---|
| Number of plates per run | Never start with one plate or two unequally balanced plates | Never start with one plate or two unbalanced plates |
| Position in carrier | Plates centred in the carrier | Plates loaded in the centre of the carrier |
| Non-skirted plates | Line wells up with the carrier markings; off-centre loading unbalances the rotor and increases vibration | Align wells with the carrier |
| After the run | Never remove samples until the rotor has fully stopped | Never remove samples until the rotor has fully stopped |
| Rotor cleaning | Wipe with 70% isopropanol/water or autoclave at 121 degrees C (250 F) for 20 minutes | Isopropyl alcohol or autoclave at 121 degrees C for 20 minutes |
Safe use and handling
The Greiner instructions give the following general rules for the instrument. Install it on a stable, level surface, away from heat, water and direct sunlight, and keep a safety distance of 30 cm around it during operation (the IFU cites IEC 1010-2-020). The rotor must be securely attached. Load symmetrically and do not move the unit during a run. If the unit has been moved from a cold room, let it warm up unplugged to avoid condensation. Clean the outside with a moist cloth and mild non-corrosive detergent, with the unit disconnected from mains and never immersed; the Greiner IFU adds that after cleaning and drying the rotor, the rotor should be allowed to accelerate once to maximum speed. Follow your own institution’s biosafety rules for what is being spun.
Common mistakes
- Running one plate. Both manufacturers say never to start with a single plate.
- Off-centre plates. Greiner states that this gives an unbalanced load and more vibration. Non-skirted plates in particular need their wells aligned to the carrier.
- Opening before the rotor stops. Both manuals prohibit removing samples until the rotor is completely stopped.
- Spinning damaged plates or strips. Greiner says damaged items must not be centrifuged.
- Comparing instruments by rpm. The same rpm gives different g at different radii, and the Benchmark manual itself quotes two rpm values; compare x g.
- Treating it as a separation centrifuge. At roughly 400 to 600 x g it is a spin-down tool.
Frequently asked questions
What is a benchtop centrifuge?
A benchtop centrifuge is a centrifuge compact enough to stand on a lab bench. The plate centrifuges reviewed here measure 23 x 26 x 19.7 cm, and the Greiner model weighs 4.1 kg according to its instructions for use.
What is RCF and RPM?
RPM is rotor speed in revolutions per minute. RCF is the relative centrifugal force, expressed in multiples of Earth’s gravity (x g), and it depends on both rpm and the radius. Quote x g when you want a result another laboratory can repeat. See the worked check above.
How much does a centrifuge cost?
The sources reviewed for this page do not publish prices, and prices vary by supplier, region and date, so this guide gives no figure. Request a current quotation from a supplier, and compare instruments on the specifications that matter to your plates: g-force, capacity, rotor radius and the plate formats the maker approves.
What goes in a centrifuge?
In a dedicated plate centrifuge, approved PCR plates and microtiter plates, and with an adapter 0.2 mL PCR strips and tubes (Benchmark). Aqueous solutions and suspensions are the stated use; explosive or highly reactive substances are excluded.
Sources
- Greiner Bio-One: Microplate Centrifuge REF 846070, instructions for use (rev. September 2017), checked 2026-11-19
- Benchmark Scientific: PlateFuge C2000 centrifuge instruction manual, checked 2026-11-19
- SLAS: ANSI/SLAS microplate standards, checked 2026-11-19
- Beckman Coulter: Relative Centrifugal Field (RCF), checked 2026-11-19
Data: Published specifications of two dedicated plate centrifuges (Greiner Bio-One REF 846070 and Benchmark PlateFuge C2000)
Side-by-side specifications taken from the two manufacturers' own instructions for use or operation manuals: speed, relative centrifugal force, capacity, maximum rotor radius, dimensions and electrical supply.
Period covered: 2017/2025.
| Specification | Greiner Bio-One REF 846070 | Benchmark PlateFuge C2000 |
|---|---|---|
| Maximum speed (rpm) | 2550 | 2100 |
| Relative centrifugal force (x g) | 600 | 400 |
| Plates per run | 2 PCR plates or microplates | 2 PCR plates or microplates |
| Maximum rotor radius (cm) | 8 | 8 |
| Dimensions W x D x H (cm) | 23 x 26 x 19.7 | 23 x 26 x 19.7 |
| Mains supply | 120 V/60 Hz or 230 V/50 Hz | 120/230 V, 50-60 Hz |
Sources
- Greiner Bio-One Microplate Centrifuge REF 846070 instructions for use (accessed 2026-11-19)
- Benchmark Scientific PlateFuge C2000 instruction manual (accessed 2026-11-19)