Sieve Shaker: How It Works, Motion Types and Checks

A sieve shaker oscillates a stack of test sieves to separate particles by size. Motion types, shaker checks and an ASTM E11 sieve-tolerance example.

Written and maintained by CASRAI Editorial Board

Published

On this page

A sieve shaker is a laboratory machine that holds a stack of test sieves and moves it in an oscillating pattern so that particles of a sample fall through successively finer openings. It replaces hand sieving with a more repeatable motion. The sieve shaker is not the part that defines the particle-size result: the openings of the sieves do, and the standards that govern them are separate from the machine.

Who uses it: technicians and core-facility users who run particle-size (sieve) analysis, for example on aggregates, and staff who maintain or buy the equipment.

What is a sieve shaker?

A sieve shaker is a machine that agitates a stack of test sieves, largest openings on top, so a sample is separated into size fractions. Haver & Boecker describe the principle as an oscillating motion that shakes particles down through a series of different test sieves in a stack, designed to give more repeatable results than hand sieving. The same source notes that a sieve shaker can come from any company, brand or model; RO-TAP is one manufacturer’s branded version.

How a sieve shaker works

The operator places the test sieves in a stack with the largest openings at the top and the smallest at the bottom, loads the sample on the top sieve, and starts the machine. The oscillation moves particles down the stack until each is retained on the first sieve whose openings are too small for it. Stacks can hold from 1 to 13 sieves, and most users run 6 full-height sieves (Haver & Boecker). The opening sizes that apply depend on the industry standard and the material being tested, and some standards require a specific type of shaker.

Types of sieve shaker motion

Motion type How it moves the stack Source
Tapping, mechanical Mechanical models use a hammer to tap, combined with circular oscillations Haver & Boecker
Electronic, three-dimensional 3-D motion is used instead of tapping Haver & Boecker
Tapping plus rotating Required by at least one industry standard cited by the source Haver & Boecker

Because some standards specify the shaker type, confirm the motion your method requires before choosing or substituting a machine.

What the shaker must do, and how the sieves are checked

AASHTO re:source states that a mechanical shaker can save time and give more consistent results, as long as it is being used properly. Its guidance on using one includes:

  • Establish agitation periods for each type of aggregate tested.
  • Verify that the shaker timer works.
  • Shake long enough to separate the sample without degrading the particles; this is the evidence that the shaker is adequate.
  • Confirm the shaker is anchored to the floor or counter if it is meant to be.
  • Maintain the shaker as AASHTO R 18 requires.

The sieves themselves are checked separately. Under ASTM E11 Table 1, column 4 (±Y) gives the tolerance on the average opening and column 5 (+X) the maximum for any single opening. AASHTO’s worked example: for a No. 4 sieve (4.75 mm), an average opening of 4.85 mm is acceptable provided no opening exceeds 5.16 mm (4.75 + 0.41 mm); if any single opening is above 5.16 mm the sieve is removed from service. For No. 4 and larger sieves the openings are measured with traceable calipers and recorded; for finer mesh the check is visual with a magnifier, including a look for clogging. Go/no-go gauges are not acceptable for the annual check. ASTM E2427 describes a performance-based alternative using an inspection or calibration sieve, a compliance sieve and reference material, and the annual opening check is still required.

Hand sieving versus a shaker

AASHTO T 27 and ASTM C136 allow hand sieving, but it is less repeatable. AASHTO advises a one-minute hand check on each sieve, and notes that the maximum retained-material limits in Table 1 of T 27 and C136 apply whether sieving is by hand or by machine. For how these methods are applied to aggregates, see ASTM C136 sieve analysis; related particle-size testing of soils is covered in ASTM D6913.

Common mistakes

  • Treating the shaker as the source of accuracy: the sieve openings are what the standards control, so a well-running shaker with an out-of-tolerance sieve still gives a wrong result.
  • Skipping a check of the shaker timer or the agitation period for the material being tested.
  • Shaking so long that particles degrade, which changes the size distribution.
  • Using a shaker motion that the governing standard does not allow.
  • Overloading a sieve beyond the Table 1 limits of T 27 and C136.

FAQ

What is a sieve shaker?

A machine that holds a stack of test sieves and oscillates it so particles in a sample separate by size, more repeatably than hand sieving.

What is a test sieve?

A test sieve is one of the graded-opening sieves placed in the stack. Its openings are checked against ASTM E11 tolerances, as in the worked example above.

Does the sieve shaker need calibration?

AASHTO’s guidance asks that the timer be verified, the agitation period be established per material and the shaker be maintained as AASHTO R 18 requires. The annual opening check applies to the sieves.

Sources

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Sieve Shaker: How It Works, Motion Types and Checks

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →