Written and maintained by CASRAI Editorial Board
Published · Last updated
This page describes products and general use; it is not medical advice.
On this page
An intrauterine pressure catheter (IUPC) is a thin catheter placed through the cervix during labor to measure pressure inside the uterus. In the United States, FDA regulates it under 21 CFR 884.2700 as a Class II device. This page is for clinic and labor-and-delivery staff, supply buyers and students who need to tell the catheter apart from the rest of the intrauterine pressure monitoring system and know which labeling fields to read.
What an intrauterine pressure catheter is
FDA’s regulation 21 CFR 884.2700 identifies an intrauterine pressure monitor as a device “designed to detect and measure intrauterine and amniotic fluid pressure.” The regulation titles the device an intrauterine pressure monitor and accessories, places the catheter transcervically into the uterine cavity, and says the device is used to monitor the intensity, duration and frequency of uterine contractions during labor. It adds that the generic type may include accessories such as signal analysis and display equipment, patient and equipment supports, and component parts. The regulation classifies it as Class II (performance standards).
Who uses it: obstetric and labor-and-delivery units, and the buyers and supply staff who stock them. The external alternative, which measures from the abdomen instead, is covered in the tocodynamometer guide.
How it is built: catheter, transducer, monitor and recorder
The FDA regulation treats the catheter as one part of a pressure-monitoring system, with signal analysis and display equipment listed as accessories. Product listings for a single catheter system can also describe built-in features: the FDA/NLM GUDID record for the INTRAN Plus IUP-450 describes a “Disposable Intrauterine Pressure Catheter System with Amnioinfusion Capability” with a “Zeroing Switch in Disposable Catheter” and a “Small Size Tip.”
Types and variants: the four FDA product codes
FDA’s device classification database lists four product codes under 21 CFR 884.2700. All four are Class II, in the Obstetrics/Gynecology panel, and flagged for third-party review. They separate the catheter from the equipment around it, which matters when you read a catalog or a procurement list.
| Product code | FDA device name | What it denotes | Class |
|---|---|---|---|
| HGS | Catheter, Intrauterine And Introducer | The catheter and its introducer | II |
| HFN | Transducer, Pressure, Intrauterine | The pressure transducer | II |
| KXO | Monitor, Pressure, Intrauterine | The monitor | II |
| HFO | Recorder, Pressure, Intrauterine | The recorder | II |
Source: openFDA device classification database, regulation number 884.2700 (record updated 2026-10-19, checked 2026-10-31). The “what it denotes” column restates each product’s FDA name. Note that the GUDID record below lists HFN, the transducer code, for a catheter system.
Specifications that matter: a worked labeling example
The table below shows one FDA/NLM AccessGUDID record, the Utah Medical INTRAN Plus IUP-450, to show which labeling fields a catheter record carries. It is an example of a record, not an endorsement or a comparison with other products.
| GUDID field | Value in this record |
|---|---|
| Description | Disposable Intrauterine Pressure Catheter System with Amnioinfusion Capability; zeroing switch in disposable catheter; small size tip |
| Brand / version / catalog number | INTRAN / Plus / IUP-450 (Utah Medical Products, Inc.) |
| Package quantity | Box of 10 |
| Tip dimension (small tip) | 0.19 in H x 0.25 in W x 0.90 in L |
| Single-use | Yes |
| Supplied sterile | Yes |
| Requires sterilization before use | No |
| Labeled “Not made with natural rubber latex” | Yes |
| Prescription use | Yes (Rx) |
| GMDN term | 16431, Intrapartum pressure monitoring catheter |
| GUDID product code / FDA submission | HFN / K955443 |
Source: FDA/NLM AccessGUDID record H671IUP4500, public version 4 dated 2026-03-19, checked 2026-10-31. The record states that the labeling does not contain MRI safety information.
How to choose: read the labeling fields
The labeling attributes in the record above are the fields to compare between products:
- Single-use and sterile status. The INTRAN Plus IUP-450 record is single-use, supplied sterile and needs no sterilization before use. Confirm the same fields on any other product.
- Latex labeling. GUDID’s own field explanation states that not all products made without natural rubber latex are marked as such, so an absent label is not the same as a latex-free statement.
- Tip size. The record gives a tip dimension for its small tip. Compare tip dimensions only as printed in each product’s own record.
- Amnioinfusion capability. Some catheter systems describe this feature in their description; others do not.
- Package quantity. The record states the box count (10 here), so the catalog number can be matched to units.
- Prescription status and product code. Rx use and the FDA product code are printed in the record.
Safe use and handling basics
The catheter is a prescription, single-use, sterile-packaged device in this record, so it is handled as a disposable item and not reprocessed. Placement and use are decisions for the clinician and are governed by the manufacturer’s instructions for use, which this page does not replace. For the supplies a birthing unit stocks around monitoring, see the birthing center supply checklist.
Common mistakes
- Treating the catheter and the monitoring equipment as one product. FDA lists the catheter (HGS), transducer (HFN), monitor (KXO) and recorder (HFO) under separate product codes.
- Confusing it with the external toco. The tocodynamometer is the external counterpart; this page covers the catheter inside the uterus.
- Reading a missing latex statement as latex-free. Only an explicit “Not made with natural rubber latex” label is a statement.
- Ignoring the identifier. A catalog number, the device identifier and the production identifier serve different purposes; see the UDI-PI guide.
FAQ
What is an intrauterine pressure catheter?
It is a catheter inserted through the cervix to measure intrauterine and amniotic fluid pressure, used to track contraction strength, length and frequency during labor (21 CFR 884.2700).
How does FDA classify it?
Class II (performance standards) under 21 CFR 884.2700, with product codes HGS, HFN, KXO and HFO in the Obstetrics/Gynecology panel.
Is it single-use?
The AccessGUDID record for the INTRAN Plus IUP-450 is single-use and supplied sterile. Check the labeling of any other product.
Is it available over the counter?
The INTRAN Plus IUP-450 record is marked prescription use (Rx).
Sources
- eCFR, 21 CFR 884.2700, Intrauterine pressure monitor and accessories: ecfr.gov (checked 2026-10-31).
- openFDA device classification API, regulation number 884.2700: api.fda.gov (checked 2026-10-31).
- FDA/NLM AccessGUDID, INTRAN Plus IUP-450: accessgudid.nlm.nih.gov (checked 2026-10-31).