Ocular Lens on a Microscope: Function, Types, Field Number

The ocular lens (eyepiece) magnifies the objective image. Learn eyepiece types, markings, field number, and sourced eyepiece-objective pairings.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

The ocular lens, also called the eyepiece, is the lens at the top of a microscope that you look through. It magnifies the intermediate image formed by the objective so that specimen detail can be observed. This page explains what it does, how it is built, how to read the markings on the barrel, and how to calculate the field of view from the field number.

Who uses it: anyone who looks through a compound or stereo microscope, including laboratory staff, technicians and students. After reading you should be able to identify what an eyepiece is doing, decode its inscription, and match it sensibly to an objective.

What the ocular lens does on a microscope

According to the Molecular Expressions microscopy primer (Florida State University), eyepieces work in combination with microscope objectives to further magnify the intermediate image so that specimen details can be observed, and “oculars” is simply an alternative name for eyepieces. The objective forms the first magnified image; the ocular enlarges that image for the eye. Total magnification is therefore the product of the two, for example a 10x eyepiece with a 40x objective.

The same source notes that most manufacturers today offer eyepieces in the 10x to 20x range, where older instruments used 6.3x to 25x. Variants of the question, such as “what are the ocular lenses on a microscope” and “what does the ocular lens do”, have the same answer: they are the eyepieces, and they provide the final stage of magnification and the circular view of the field set by the eyepiece diaphragm.

How an eyepiece is built

A typical eyepiece has two lens groups, an eye lens nearest the observer and a field lens below it, with a fixed field diaphragm positioned relative to them. The diaphragm sets the edge of the circular view. Whether the diaphragm sits inside or below the lenses determines the basic family, described next.

Types of eyepiece

Type Construction (per source) Typical use or key difference
Huygenian (negative) Simplest negative eyepiece; internal diaphragm between two lenses Found on most teaching and laboratory microscopes with achromatic objectives; if only the magnification is inscribed it is most likely Huygenian, best used with achromats from 5x to 40x
Ramsden (positive) Diaphragm sits below the lenses; front focal plane at the diaphragm Readily adapted for reticles because the focal plane is accessible
Kellner Modified Ramsden with a cemented eye-lens doublet Higher eye point and larger field than a Ramsden
Compensating Introduces chromatic error equal and opposite to that of the objective Must be used at all magnifications with fluorite, apochromatic and all variations of plan objectives; can help with achromats at 40x and above
High eye point Eye relief extended to about 20 to 25 mm Lets spectacle wearers see the full field; does not correct astigmatism

Source: Molecular Expressions, Eyepieces (Oculars), read 2026-11-07.

Specifications that matter: magnification, field number and eye point

Field number (FN) is the diameter, in millimetres, of the view field at the intermediate image plane; it is usually set by the eyepiece field diaphragm (Nikon MicroscopyU, Field of View). Nikon gives a typical eyepiece range of FN 6 to 28 mm, generally decreasing as eyepiece magnification rises, with wide-field designs at FN 26 and above. Wide-eyefield eyepieces can increase the viewable area by as much as 40 percent (Molecular Expressions).

Eye point (the Ramsden disc) is usually 8 to 10 mm from the eye lens on standard eyepieces, and approaches 20 to 25 mm on high eye point models.

Olympus eyepiece (as listed by source) Magnification Field number
PSWH 10x 26.5
PWH 10x 22
SWH 10x 26.5
WH 15x 14
WH (high eye point) 10x 22

The source lists diopter adjustment of -8 to +2 for these models. Source: Molecular Expressions, Table 1, read 2026-11-07.

Reading the markings on the barrel

An eyepiece inscription such as “10x” with “A/24” means 10x magnification and a field number of 24, the diameter in millimetres of the fixed diaphragm. The letters follow manufacturer conventions listed by the same source:

  • WF: wide-field; UWF: ultra wide-field; SW / SWF: super wide-field; UW: ultra wide viewfield
  • H: high eye point, suited to spectacle wearers
  • K, C or comp: compensating
  • CF: for CF-corrected objectives; Plan-Comp: for flat-field objectives

Calculating the field of view from the field number

The diameter of the area you see on the specimen is the field number divided by the objective magnification (Nikon MicroscopyU). If an auxiliary lens sits between objective and eyepiece, multiply its factor into the objective magnification first. The Molecular Expressions primer gives the worked example: FN 26.5 with a 40x objective and a tube-lens factor of 1.25 gives a viewfield of 0.53 mm (26.5 / (40 x 1.25)).

Objective Viewfield diameter (SWF 10x eyepiece, FN 21.2)
1x 21.2 mm
4x 5.3 mm
10x 2.12 mm
20x 1.06 mm
40x 0.53 mm
100x 0.21 mm

Source: Molecular Expressions, Table 2, read 2026-11-07. A reticle placed at the field-diaphragm plane stays in focus with the specimen and can be used for measurement; see microscope calibration with a stage micrometer for calibrating one.

Pairing the ocular with an objective

The primer gives a useful magnification range of 500 to 1000 times the numerical aperture (NA) of the objective; magnification beyond that adds size without detail (“empty magnification”). Its example: 250x reached with a 25x eyepiece and a 10x objective (NA about 0.25) is inferior to the same 250x reached with a 10x eyepiece and a 25x objective (NA about 0.65). Choose the objective first, then the eyepiece, and use the eyepieces the manufacturer recommends for its objectives.

Objective (NA as listed by source) Eyepiece magnifications within the useful range
10x, NA 0.35 15x, 20x, 25x
40x, NA 0.70 10x, 12.5x, 15x
60x, NA 0.95 10x, 12.5x, 15x
100x, NA 1.42 10x, 12.5x

Source: Molecular Expressions, Table 3, read 2026-11-07. For the objective side of the system, see compound microscope optics.

Setup basics

On binocular heads each eye views through its own ocular, and interpupillary distance and diopter settings are adjusted so both eyes are in focus together; this is covered in microscope ergonomics. High eye point eyepieces help glasses wearers see the whole field but do not correct astigmatism.

Common mistakes

  • Pushing magnification with a high-power eyepiece on a low-NA objective, which gives empty magnification.
  • Using a Huygenian eyepiece with fluorite, apochromatic or plan objectives that require compensating eyepieces.
  • Mixing eyepieces from different makers without checking the recommendation for the objectives in use.
  • Forgetting the tube-lens or auxiliary factor when calculating field of view.

FAQ

What is the ocular on a microscope?

It is the eyepiece, the lens you look through, which further magnifies the image formed by the objective.

What does the ocular lens do on a microscope?

It enlarges the intermediate image from the objective for viewing and defines the circular field you see through its field diaphragm.

How do I find the magnification of an ocular?

Read the inscription on the barrel, for example 10x; the number after a slash, such as 24, is the field number in millimetres.

How is the field of view calculated?

Field number divided by objective magnification, including any auxiliary lens factor. Related instrument pages: microscopes for research and the dissecting microscope.

Sources

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Ocular Lens on a Microscope: Function, Types, Field Number

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →