An oxygenator is a medical device that exchanges gases between a patient’s blood and a gas supply outside the body, adding oxygen and removing carbon dioxide while the heart-lung machine or an extracorporeal circuit does the work of the lungs. This page covers the medical blood oxygenator, not the aerators sold for fish tanks, which are unrelated products. It is written for clinic staff, buyers and students who need to know what the device is, how FDA classifies it, and where to read the rules.
What an oxygenator is
FDA’s regulation for the cardiopulmonary bypass oxygenator (21 CFR 870.4350) identifies it as “a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.” A related regulation, 21 CFR 870.4100, covers extracorporeal circuits for long-term support in acute respiratory or cardiopulmonary failure, and lists oxygenators among that circuit’s disposable components.
Who uses it: hospital cardiac surgery and perfusion teams use bypass oxygenators in the operating room; critical care teams use long-term circuit oxygenators. Purchasers in these settings buy them as part of a wider circuit. For a broader look at the equipment category, see what a medical device is.
How it is built
FDA’s long-term support regulation (21 CFR 870.4100) describes the circuit as console hardware and software plus disposables such as oxygenators, blood pumps, heat exchangers, cannulae, tubing and filters, with accessories such as monitors, sensors and connectors. The oxygenator is therefore one disposable part of a larger system, not a stand-alone machine.
A 2016 open-access review of membrane oxygenator technology (Melchior et al., Pediatric Health, Medicine and Therapeutics) describes how the design developed:
- The bubble oxygenator introduced by DeWall and Lillehei became the safest and most cost-efficient disposable option and dominated for about 25 years until membrane oxygenators evolved.
- Membrane designs separate blood from gas with a membrane. Hollow fiber technology formed capillaries into a compartment that allowed more efficient gas exchange while reducing the overall surface area of the oxygenator.
- Early microporous membrane oxygenators were investigated using the hydrophobic material Teflon, and silicone became the most practical substance for coating biomaterials.
- As the technology advanced, the heat exchanger, the component that regulates blood temperature, was built into the oxygenator rather than kept as a separate component in the circuit.
Types and regulatory classes
FDA separates oxygenators by intended duration of use. The table below is taken from the Code of Federal Regulations and FDA’s openFDA device classification records (checked 2026-10-30).
| FDA product code | Device name | Regulation | Intended use | Class |
|---|---|---|---|---|
| DTZ | Oxygenator, cardiopulmonary bypass | 21 CFR 870.4350 | Gas exchange during open-heart surgery | II (special controls) |
| BYS | Oxygenator, long term support greater than 6 hours | 21 CFR 870.4100 | Long-term extracorporeal support (more than 6 hours) | II (special controls) |
Both codes are in the cardiovascular review panel, are marked life-sustaining or life-supporting, are not implants, are not exempt from GMP requirements, and reach the market through 510(k) premarket notification. For how that pathway works, read the FDA 510(k) premarket notification guide and the overview of FDA device classification, 510(k) and PMA.
What the special controls require
For bypass oxygenators the special control is FDA’s guidance “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.” For long-term support circuits, 21 CFR 870.4100 sets six special controls: design characteristics, biocompatibility, sterility and shelf life, nonclinical performance testing, in vivo evaluation, and labeling. The labeling must include adequate instructions on anticoagulation, circuit setup, compatibility, and maintenance during a procedure (final rule 81 FR 7451, 12 February 2016).
How to choose
The regulations do not set a buying checklist. The criteria they do state are that a device is cleared for the intended use you need (surgical bypass under 870.4350, or support beyond six hours under 870.4100) and that its labeling carries the required instructions. Compatibility with the rest of the circuit is something the manufacturer’s labeling must address, so read the instructions for use for the specific oxygenator and circuit before any purchasing decision. Biocompatibility background is covered in the FDA biocompatibility guidance guide.
Safe use and handling basics
- Use only within the intended use stated in the manufacturer’s labeling and FDA clearance.
- The manufacturer’s instructions for use are the reference for the product; FDA requires long-term circuit labeling to cover circuit setup, compatibility, anticoagulation and maintenance during a procedure.
- Oxygenators are life-sustaining devices operated by trained clinical teams. Clinical decisions are outside the scope of this page.
- Traceability matters for devices like this; see the unique device identification guide.
Common mistakes
- Treating a bypass oxygenator and a long-term support oxygenator as interchangeable. FDA gives them different product codes, regulations and intended uses.
- Confusing an oxygenator with a ventilator. A ventilator moves gas in and out of the patient’s lungs; an oxygenator exchanges gases with blood outside the body. See the ventilator guide.
- Confusing it with an oxygen concentrator or a fish-tank aerator. Neither is a blood gas exchange device.
- Assuming an oxygenator is a stand-alone device. It is a disposable component within a circuit.
Where to find related supplies
Monitoring and diagnostic equipment used around critical care is listed under diagnostic, monitoring and imaging.
Frequently asked questions
What is an oxygenator?
A device that exchanges gases between blood and a gas supply outside the body, used to meet a patient’s gas exchange needs during open-heart surgery or long-term cardiopulmonary support (21 CFR 870.4350 and 870.4100).
Is an oxygenator the same as a heart-lung machine?
No. It is one component. The long-term support circuit regulation lists oxygenators alongside blood pumps, heat exchangers, cannulae, tubing and filters.
How does FDA classify oxygenators?
Both cardiopulmonary bypass oxygenators (DTZ) and long-term support oxygenators (BYS) are Class II with special controls and are cleared through 510(k).
How is an oxygenator different from a fish-tank oxygenator?
Aquarium products add air or oxygen to water. A medical oxygenator is a regulated device that handles blood.
Sources
- 21 CFR 870.4350, Cardiopulmonary bypass oxygenator: law.cornell.edu/cfr/text/21/870.4350 (checked 2026-10-30)
- 21 CFR 870.4100, Extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary failure: law.cornell.edu/cfr/text/21/870.4100 (checked 2026-10-30)
- FDA openFDA device classification, product codes DTZ and BYS: api.fda.gov/device/classification.json (checked 2026-10-30)
- Melchior RW et al. Evolution of membrane oxygenator technology for utilization during pediatric cardiopulmonary bypass. Pediatr Health Med Ther. 2016;7:45-56. PMC5683297: pmc.ncbi.nlm.nih.gov/articles/PMC5683297 (checked 2026-10-30)