10-20 System: EEG Electrode Placement, Landmarks and Naming

The 10-20 system places 21 EEG electrodes at 10% or 20% intervals between landmarks. Covers naming, T7/P7 changes and 10-10 vs 10-5, per ACNS.

Written and maintained by CASRAI Editorial Board

Last updated

This page describes products and general use; it is not medical advice.

The 10-20 system is the standard scheme for deciding where scalp electrodes go in a routine EEG (electroencephalogram). The American Clinical Neurophysiology Society (ACNS) guideline on electrode nomenclature describes it as a method proposed in 1958 that records EEG with 21 electrodes placed at relative distances of 10% or 20% between cranial landmarks. It is used by EEG technologists, neurophysiology clinic staff and students, and by buyers who must match caps, electrodes and recording equipment to the labels a lab uses.

What the 10-20 system is

The 10-20 system (also written “10 20 system”, “EEG 10-20” or “10-20 electrode system”) is a head-surface positioning method. Its name comes from the spacing: electrodes sit at 10% or 20% of the distance between fixed landmarks on the skull, so the layout scales to each head instead of using fixed centimetre distances. According to the ACNS guideline, the system was proposed by the International Federation of Societies for Electroencephalography and Clinical Neurophysiology in 1958 and has been the international standard for recording routine scalp EEG for clinical use. The same guideline notes that it has also served as a positioning reference for evoked and event-related potentials and for transcranial brain mapping methods.

How the layout is built: landmarks and 10%/20% spacing

Four landmarks anchor the layout: the nasion, the inion, and the left and right preauricular points. A sleep-research methods article in Current Protocols in Neuroscience (Campbell, 2009, NIH author manuscript) describes the inion as the highest point of the projection of the occipital bone at the lower rear of the skull, and the preauricular points as lying directly in front of the ear canal. Electrodes are then placed at 10% or 20% intervals of the distance between these landmarks.

This page does not give a head-measuring procedure. Measuring and applying electrodes on a person is done by trained technologists following the lab’s own protocol and the equipment instructions for use.

Electrode naming: letters, numbers and the T3/T7 change

In the nomenclature set out in the ACNS guideline, each electrode name encodes its position: all electrodes with the same letter or letters lie on the same coronal line, and, with a few exceptions (Fp1/Fp2 and O1/O2), electrodes on one front-to-back (sagittal) line share the same number. The one deviation from a strict combinatorial scheme is the “AF” (anterior frontal) row.

T3/T4 and T5/T6 versus T7/T8 and P7/P8

The older 10-20 labels T3/T4 and T5/T6 break that logic, so the 10-10 nomenclature replaces them with T7/T8 and P7/P8. This is a frequent source of confusion when reading cap labels or recordings: P7/P8 are posterior temporal positions, not parietal ones. The ACNS guideline states that P with a number of 6 or less is parietal, whereas P with a number of 7 or greater implies posterior temporal. The guideline also says (its recommendation 4) that it is an acceptable alternative to keep using T3/4 and T5/6, or to use both sets of terms, at present. Positions 10% below the frontotemporal electrodes are named F9/F10, T9/T10 and P9/P10.

10-20 vs 10-10 vs 10-5

The 10-20 layout has been extended. The table below states only what the ACNS guideline says about each.

System Electrode positions Status per ACNS guideline
10-20 21 electrodes International standard for routine clinical scalp EEG
10-10 More than 70 positions described Accepted as a standard by ACNS and the International Federation of Clinical Neurophysiology
10-5 Further extension for systems with many channels (the guideline cites 128 or 256 as examples) Proposed, but not accepted by ACNS or the International Federation of Clinical Neurophysiology

Source for the table: ACNS Guideline 2, Guidelines for Standard Electrode Position Nomenclature (checked 2026-10-30).

Equipment tied to the system

EEG electrodes applied to the skin fall under the FDA classification “Electrode, Cutaneous” (product code GXY, 21 CFR 882.1320), a Class 2 device. The regulation describes it as an electrode applied directly to a patient’s skin to record physiological signals, such as the electroencephalogram, or to apply electrical stimulation (openFDA and 21 CFR 882.1320, checked 2026-10-30).

For anyone matching equipment to a layout, the ACNS guideline makes a practical point: additional electrodes must be purchased and take technologist time, and most vendors of commercial EEG machines in the USA continued to provide headboxes with electrode positions and nomenclature limited to the 10-20 system (the guideline’s own statement). If a lab wants to use 10-10 positions, the recording hardware has to provide them. Related supplies, such as Ten20 conductive paste, ECG electrodes and sleep clinic supplies, are covered on separate CASRAI pages, and diagnostic and monitoring supplies are listed in the diagnostic, monitoring and imaging category.

Equipment criteria when choosing between 10-20 and 10-10

Criterion What the ACNS guideline says
Time and cost The extra electrodes of the 10-10 system need more technologist time and effort, and the additional electrodes must be purchased
Recording hardware Most vendors of commercial EEG machines in the USA continue to provide headboxes with electrode positions and nomenclature limited to the 10-20 system; headbox labels would have to change to show T7/T8 and P7/P8
Label set (T3/T4, T5/T6 versus T7/T8, P7/P8) Either set, or both, is an acceptable alternative at present

Which electrode positions suit a given study is a clinical decision made by the clinician who orders and interprets it; this page does not advise on it.

Safe use and handling basics

Electrodes in the Class 2 “Electrode, Cutaneous” category are skin-contact devices, so the manufacturer’s instructions for use govern skin preparation, single-use or reuse status, cleaning and disposal. Follow those instructions and the facility’s infection-control policy rather than general advice.

Common mistakes

  • Reading P7/P8 as parietal. Under the 10-10 nomenclature they are posterior temporal positions.
  • Assuming a recording labeled T3/T4 is a different montage from one labeled T7/T8. The ACNS guideline treats the two label sets as acceptable alternatives for the same positions.
  • Assuming 10-10 or 10-5 positions are available on any EEG system. Recording hardware must provide them.
  • Treating the 10-5 extension as an accepted standard. The ACNS guideline says it has not been accepted by ACNS or the International Federation of Clinical Neurophysiology.

FAQ

What is the 10-20 system in EEG?

It is the standard method for placing 21 scalp electrodes at 10% or 20% distances between cranial landmarks for routine clinical EEG (ACNS guideline, checked 2026-10-30).

What does 10-20 stand for?

The numbers are the percentage spacing of electrodes along lines between landmarks: 10% or 20% of the measured distance.

Which landmarks does the 10-20 system use?

The nasion, the inion, and the left and right preauricular points (Campbell, 2009).

How many electrodes are in the 10-20 system?

21, according to the ACNS guideline. The 10-10 extension describes more than 70 positions.

What is the difference between the 10-20 and 10-10 systems?

10-10 adds electrodes in the anteroposterior, coronal and inferior planes and renames T3/T4 and T5/T6 as T7/T8 and P7/P8. See the table above.

Why do some labels say T3 and others T7?

The 10-10 nomenclature renamed T3/T4 as T7/T8 and T5/T6 as P7/P8 so that letters and numbers follow a consistent grid. The ACNS guideline says continuing to use the older labels, or both, is acceptable at present.

Sources

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about 10-20 System: EEG Electrode Placement, Landmarks and Naming

Ask your first 2 questions free below. Subscribers get 150 a day for $29 a month.

An AI assistant specialized in research administration. It cites the sources behind every answer, labels web answers and says when it can't answer.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays free to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →