An anion exchange column is a liquid chromatography hardware device packed with positively charged stationary phase resin to bind, separate, and purify negatively charged molecules—including acidic proteins, nucleic acids, peptides, and inorganic anions—based on reversible electrostatic interactions. It retains polyanions at pH values above their isoelectric points and elutes them as salt concentration rises.
What it is
An anion exchange column is a specialized chromatographic hardware device engineered for charge-based fractionation in downstream purification and analytical characterization. Mechanistically, the column operates as an ion exchange column that selectively adsorbs anions while allowing neutral and positively charged species to flow through without retention. The packed bed consists of a rigid or semi-rigid support matrix functionalized with fixed cationic chemical moieties, balanced by exchangeable counterions such as chloride (Cl⁻) or acetate (CH₃COO⁻).
When an aqueous sample enters the column under low-ionic-strength binding conditions, negatively charged functional groups on the sample molecules displace the mobile counterions and electrostatically associate with the stationary ligands. Elution is subsequently triggered by introducing a competing counterion gradient (commonly sodium chloride) or by modulating mobile phase pH, dislodging analytes according to differences in their net surface charge density and charge distribution.
Bioprocess engineers, analytical chemists, structural biologists, and quality control analysts routinely utilize anion exchange columns across academic research, biopharmaceutical manufacturing, and environmental testing. Who uses it in practice spans multiple sectors: bioprocess engineering teams rely on anion exchange chromatography for capture and intermediate polishing in multistep protein purification workflows, gene therapy groups deploy it for full versus empty capsid adeno-associated virus (AAV) separation, and regulatory testing labs employ high-pressure anion exchange columns for trace anion quantification in drinking water and biotherapeutics.
How it works or is built
An anion exchange column integrates precision mechanical hardware with tailored surface chemistry to maintain laminar flow, uniform sample distribution, and bed integrity under operational fluid velocities. The column assembly comprises several structural and chemical components:
- Column Hardware Shell: Analytical and high-performance ion chromatography (HPIC) columns utilize precision-bore electropolished 316L stainless steel or metal-free polyetheretherketone (PEEK) tubing to withstand hydraulic pressures exceeding 200 bar. Preparative and fast protein liquid chromatography (FPLC) columns frequently employ heavy-wall borosilicate glass or medical-grade molded polypropylene (such as Cytiva HiTrap cartridges) rated for low-to-moderate pressures (3 to 5 bar).
- End-fittings and Distribution Frits: Sintered porous titanium, stainless steel, or polyethylene frits (typically 0.5 µm to 10 µm nominal pore size) are seated at the column entrance and exit. These frits retain the packed bed, protect against particulate contamination, and disperse the incoming mobile phase evenly across the entire column cross-sectional area to prevent channel formation.
- Base Support Matrix: The core bead substrate determines physical rigidity, flow-velocity tolerance, and macro-porosity. Common base materials include highly cross-linked spherical agarose (e.g., Sepharose Fast Flow and Capto matrices), hydroxylated polymethacrylate polymers (e.g., Tosoh TSKgel 5PW and Bio-Rad ENrich), cross-linked polystyrene-divinylbenzene (PS-DVB, e.g., Cytiva MonoBeads and Thermo Dionex IonPac), and macroporous silica. Base matrices are selected from specialized porous chromatographic resins engineered to minimize non-specific hydrophobic adsorption.
- Stationary Phase Ligands: Fixed cationic functional groups are covalently grafted to the matrix surface either directly or via hydrophilic polymer extenders (dextran chains). Common ligands include quaternary ammonium (Q), quaternary methylamine (QMA), trimethylaminoethyl (TMAE), and diethylaminoethyl (DEAE) groups.
Depending on internal bead geometry, anion exchange columns are categorized into porous particulate beds, non-porous analytical particles (which eliminate intraparticle diffusion resistance to yield ultra-narrow chromatographic peaks), and continuous monolithic beds (such as Bio-Rad UNO Q columns), where fluid convection flows through interconnected macropores rather than interstitial bead spaces.
Types and variants
Anion exchange columns are classified primarily by the ionization behavior of their functional ligands—distinguishing strong anion exchangers (SAX) from weak anion exchangers (WAX)—as well as their structural resin architecture.
Strong Anion Exchange Columns (SAX)
Strong anion exchange columns feature quaternary amine functional groups (such as –N⁺(CH₃)₃) that remain fully ionized across virtually the entire usable aqueous pH scale (typically pH 2.0 to 12.0, and up to pH 14.0 on polymeric substrates). The chromatographic surface maintains constant charge density regardless of buffer pH variations. Consequently, selectivity and retention are governed entirely by the ionization state and net negative surface charge of the solute molecules, simplifying gradient method development and enabling aggressive alkaline sanitization without ligand loss.
Weak Anion Exchange Columns (WAX)
Weak anion exchange columns utilize primary, secondary, or tertiary amine functional groups, most commonly diethylaminoethyl (DEAE, –CH₂CH₂N⁺H(C₂H₅)₂). Unlike strong exchangers, the functional ligand on a weak anion exchange resin possesses a defined acid dissociation constant (typically pKa 8.5 to 9.5). The ligand becomes progressively deprotonated and loses its positive charge as the mobile phase pH rises above neutral toward basic values. Weak anion exchangers provide an additional dimension of selectivity control: both the target solute and the stationary phase ligand change charge state simultaneously as a function of pH, making WAX columns valuable for resolving closely related protein isoforms or oligonucleotides that co-elute on strong exchangers.
Anion Exchange vs Cation Exchange Columns
Understanding the complementary role of a cation exchange column is essential when designing downstream purification platforms. While an anion exchange column carries fixed positive charges to capture acidic and polyanionic species, a cation exchange column is packed with negative ligands—such as strong sulfopropyl (SP, –CH₂CH₂CH₂SO₃⁻) or weak carboxymethyl (CM, –CH₂COO⁻) functional groups—on a cation exchange resin bed. Cation exchange columns bind basic molecules at mobile phase pH levels below their isoelectric points. In industrial monoclonal antibody downstream processing, cation exchange columns are frequently deployed in bind-and-elute mode for primary monomer capture, whereas anion exchange columns are utilized in flow-through polishing mode to bind host cell proteins, DNA, endotoxins, and adventitious viruses.
Comparison of Anion Exchange Column Chemistries
| Exchanger Category | Typical Ligand Chemistry | Operating pH Range | Surface Charge Stability | Primary Analytical / Process Use |
|---|---|---|---|---|
| Strong Anion Exchanger (SAX) | Quaternary ammonium (–N⁺(CH₃)₃, Q, TMAE) | pH 2.0 – 12.0 (polymeric: 0 – 14) | Completely ionized; charge density remains constant across working pH | General protein purification, nucleic acid capture, standard bioprocess polishing |
| Weak Anion Exchanger (WAX) | Diethylaminoethyl (–CH₂CH₂N⁺H(C₂H₅)₂, DEAE, DMAE) | pH 2.0 – 9.0 | Deprotonates above pH 8.5; charge density decreases as pH rises | Proteins that bind too tenaciously to SAX; resolution of complex charge variants |
| Multimodal Anion Exchanger (MM-AEX) | Quaternary ammonium with phenyl/aromatic residues | pH 2.0 – 12.0 (CIP up to 14) | Combines electrostatic positive charge with hydrophobic interactions | Direct high-salt feedstock capture; aggregate and virus clearance without dilution |
| Non-Porous SAX Analytical | Surface-pellicular quaternary amine | pH 3.0 – 10.0 | Surface-restricted positive charges; zero pore diffusion path | Ultra-high-resolution mAb charge heterogeneity profiling, fast analytical HPLC |
Sizes or specifications that matter
Selecting an appropriate column geometry and resin specification requires matching physical dimensions and bed volume to sample mass, volumetric throughput, and system backpressure limitations. Manufacturer technical specifications across analytical, preparative, and bioprocess scales establish clear operational boundaries:
- Internal Diameter (ID) and Bed Volume: Analytical columns generally range from 0.4 mm ID (capillary formats consuming microliters of eluent) to 4.6 mm ID (standard analytical formats operated at 0.5 to 1.5 mL/min). Laboratory preparative columns span 7.5 mm to 16 mm ID (bed volumes 1 mL to 20 mL, accommodating 10 mg to 500 mg protein), while process development columns exceed 26 mm to 100 mm ID with multi-liter bed volumes.
- Bed Height: Analytical columns typically feature bed heights of 5 cm to 25 cm. For preparative protein separations on high-performance resins, bed heights of 10 cm to 20 cm provide sufficient theoretical plates for sharp resolution while maintaining manageable pressure drops.
- Particle Size (d₅₀): Analytical columns employ 1.7 µm to 10 µm spherical particles to maximize separation efficiency and peak capacity. In contrast, preparative and process columns use larger beads (34 µm High Performance to 90 µm Fast Flow) to minimize hydraulic resistance during high-velocity loading of viscous biological extracts.
- Pore Size: Conventional small-molecule columns use 60 Å to 120 Å pores. Because biopolymers have large hydrodynamic radii, anion exchange columns designed for macromolecules feature pore diameters of 300 Å to 1000 Å (or open-network agarose hydrogels) to allow unhindered access to internal binding surfaces.
- Dynamic Binding Capacity (DBC): Measured at 10% breakthrough (QB10%), DBC quantifies the mass of target molecule that binds under active mobile phase flow before breakthrough occurs. Typical DBC values for bovine serum albumin (BSA) or human serum albumin (HSA) range from 40 mg/mL to over 120 mg/mL of packed medium.
- Pressure Limitations: Column hardware and packed bed pressure ratings dictate compatibility with HPLC vs FPLC systems. Polypropylene prepacked cartridges (e.g., 1 mL and 5 mL formats) carry maximum hardware limits of 0.5 MPa (5 bar / 72.5 psi) with operational packed bed limits of 0.3 MPa (3 bar). Precision glass columns (such as Cytiva Tricorn 10/100 GL) withstand up to 4.0 MPa (40 bar / 580 psi). Stainless steel analytical columns and high-pressure polymer columns (such as Thermo Dionex IonPac AS11-HC) operate safely up to 34.5 MPa (345 bar / 5,000 psi).
| Column Sizing Format | Typical Dimensions (ID × Length) | Particle Diameter | Typical Flow Rate | Rated System Pressure |
|---|---|---|---|---|
| Capillary Analytical | 0.4 mm × 250 mm | 4 µm – 9 µm | 0.01 – 0.03 mL/min | Up to 34.5 MPa (345 bar / 5,000 psi) |
| Standard Analytical HPLC | 4.0 mm – 4.6 mm × 100 – 250 mm | 5 µm – 10 µm | 0.5 – 1.5 mL/min | 15 – 35 MPa (150 – 350 bar) |
| Prepacked FPLC Cartridge | 7 mm × 25 mm (1 mL) / 16 mm × 25 mm (5 mL) | 34 µm – 90 µm | 1.0 – 5.0 mL/min | 0.3 – 0.5 MPa (3 – 5 bar / 72 psi) |
| High-Resolution FPLC Column | 10 mm × 100 mm (8 mL bed) | 10 µm | 1.0 – 4.0 mL/min | 4.0 MPa (40 bar / 580 psi) |
| Preparative Process Column | 26 mm – 50 mm × 100 – 200 mm | 50 µm – 90 µm | 10 – 60 mL/min | 0.3 – 0.7 MPa (3 – 7 bar / 100 psi) |
The structured data table below compiles exact technical specifications, matrix chemistries, and operational limits across prominent commercial anion exchange columns.
How to choose
Selecting the optimal anion exchange column requires balancing analyte physicochemical properties, purification objectives (analytical characterization versus preparative mass recovery), and instrument hydraulic capabilities. Chromatographers should follow a systematic five-step decision process rooted in primary manufacturer guidelines:
- Calculate Analyte Isoelectric Point (pI) and Buffer pH Window: For an analyte to bind an anion exchange column, the mobile phase pH must exceed the molecule’s isoelectric point by at least 0.5 to 1.0 pH unit. This ensures a net negative surface charge. If the target protein is unstable or aggregates at pH levels above its pI, an anion exchange column cannot be used; the operator must switch to a cation exchange column operated at a pH below the pI.
- Select Strong (SAX) vs Weak (WAX) Ligands: Initiate method development using a strong quaternary ammonium exchanger (Q or TMAE). Strong exchangers maintain constant positive surface charge, allowing systematic exploration of elution conditions across pH 5.0 to 9.0 without changing ligand ionization. Transition to a weak exchanger (such as DEAE) if the target molecule binds with excessive affinity and fails to elute under 1.0 M salt, or when subtle selectivity adjustments are needed to separate co-purifying charge variants.
- Match Particle Size to Resolution Requirements: If the goal is analytical purity determination, monitoring monoclonal antibody deamidation variants, or quantifying trace environmental anions, select a small-particle column (1.7 µm to 10 µm) run on an analytical ion exchange HPLC or HPIC system. If the goal is purifying milligram to gram quantities from crude cell lysates, choose larger beads (34 µm to 90 µm) in a cartridge or glass column format to avoid excessive pressure drops and frit clogging.
- Determine Mode: Bind-and-Elute vs Flow-Through Polishing: In ion exchange chromatography protein purification protocols, anion exchange can operate in two distinct modes. In bind-and-elute mode, the target protein binds the resin and elutes via salt gradient while neutral impurities wash out. In flow-through polishing mode (common in recombinant mAb production), buffer conditions are adjusted so that the target mAb (having a high pI of 8.0 to 9.0) remains basic and flows straight through the column unhindered, while acidic impurities—such as host cell DNA, endotoxins, host cell proteins (HCP), and adventitious retroviruses—are tenaciously captured by the column bed.
- Verify System Hardware and Pressure Compatibility: Ensure the column pressure rating matches your instrument configuration. Never connect a low-pressure polypropylene cartridge (0.5 MPa max) to an unthrottled HPLC pump without setting software pressure alarms, as overpressurization will rupture the plastic housing or compact the resin bed.
Safe use and handling basics
Proper column preparation, mobile phase conditioning, and routine sanitization are critical to preserve chromatographic efficiency, prevent biological contamination, and extend column operating life.
Sample Preparation and Mobile Phase Conditioning
Particulates in sample solutions or buffers rapidly deposit onto column entrance frits, causing irreversible backpressure elevation and distorted peak shapes. All mobile phase buffers must be filtered through 0.22 µm or 0.45 µm hydrophilic membrane filters and thoroughly degassed via vacuum sonication or in-line membrane degassers prior to use. Crude biological samples (cell culture supernatants, cell lysates, or tissue homogenates) must undergo clarifying centrifugation at 10,000 × g or higher for 15 minutes, followed by 0.22 µm syringe filtration before injection.
The sample must be prepared in a buffer whose ionic strength is low enough to permit electrostatic adsorption. If the feedstock contains high salt concentrations from prior precipitation or extraction steps, the sample must be desalted via dialysis, buffer exchange chromatography, or diluted at least 3- to 5-fold with binding buffer before loading.
Equilibration and Elution Execution
Equilibrate the column with 5 to 10 column volumes (CV) of starting binding buffer until the effluent pH and electrical conductivity match the inlet buffer exactly. Elution is performed using either linear or step gradient protocols. In salt-gradient elution, sodium chloride concentration is ramped from 0 mM to 500 mM or 1000 mM over 10 to 30 CVs. Higher charge-density species require higher salt concentrations for displacement. Detailed principles governing gradient formation and counterion displacement are documented in guides covering gradient elution strategies.
Cleaning-in-Place (CIP) and Regeneration
Biological feedstocks deposit non-specifically adsorbed lipids, strongly bound proteins, and colloidal debris over repeated chromatographic cycles. A standard cleaning-in-place (CIP) protocol should be executed every 5 to 20 runs, or whenever baseline backpressure rises by 15% to 20%:
- Salt Wash: Flush the column with 2 to 5 CVs of 1.0 M to 2.0 M NaCl in binding buffer to elute tenaciously bound ionic species.
- Alkaline Sanitization: For polymeric and cross-linked agarose resins, wash with 2 to 5 CVs of 0.5 M to 1.0 M NaOH. Maintain contact time for 30 to 60 minutes to solubilize precipitated proteins, strip nucleic acids, and inactivate bacterial endotoxins. Caution: Never apply alkaline solutions to silica-based columns, which dissolve at pH > 8.0.
- Lipid / Hydrophobic Foulant Stripping: If the column has been exposed to membrane proteins or hydrophobic feedstocks, wash with 2 to 4 CVs of 70% ethanol or 30% isopropanol, or 0.1% non-ionic detergent (such as Triton X-100), followed by extensive water washing.
- Re-equilibration: Re-equilibrate with 5 to 10 CVs of neutral buffer before resuming operations or transferring to storage.
Storage Conditions
For short-term storage (under 48 hours), anion exchange columns can remain in neutral equilibration buffer at room temperature or 4 °C. For long-term storage (exceeding two days), flush the column with at least 5 CVs of deionized water followed by 5 CVs of bacteriostatic storage solution—typically 20% ethanol in deionized water or neutral buffer containing 0.02% sodium azide (NaN₃). Store sealed columns at 4 °C to 8 °C. Never allow packed columns to dry out, and never store columns in high-salt solutions (which promote chloride-induced pitting corrosion in stainless steel hardware) or at freezing temperatures (which crack the resin bed).
Common mistakes
Operating errors in anion exchange chromatography compromise separation resolution and can permanently ruin expensive prepacked columns. The most frequent operational mistakes include:
- Operating Buffer pH Too Close to Target Molecule pI: Choosing a mobile phase pH within 0.5 units of the target protein’s pI results in weak, heterogeneous binding. This causes premature leakage during loading, broad elution peaks, and poor batch-to-batch reproducibility. Always maintain buffer pH at least 0.5 to 1.5 units above the target pI.
- Selecting Cationic Buffer Counterions: Buffer salts whose buffering ions carry positive charges (such as Tris or Bis-Tris) are appropriate for anion exchange because the buffering species is either cationic or neutral, avoiding competition with the stationary phase. In contrast, using anionic buffers (such as phosphate or citrate) can cause the buffer counterion to interact competitively with the column ligands, inducing localized pH shifts and baseline drift during salt gradient execution.
- Exceeding Packed Bed Pressure Drop: Conflating system pressure with column net pressure drop is a frequent cause of column failure. Packed bed pressure drop equals total pump pressure minus system pre-column and post-column tubing pressure. Exceeding the packed bed pressure limit compresses agarose and polymeric beads, leading to channel formation, headspace voids at the top of the column, and loss of resolution.
- Injecting Unfiltered or Unclarified Samples: Introducing viscous, turbid, or particulate-laden extracts directly onto the column inlet frit creates immediate pressure spikes. If backpressure increases sharply during sample loading, immediately stop the pump and inspect the inlet frit.
- Ignoring Mobile Phase Viscosity at Low Temperatures: Chromatography conducted in cold rooms (4 °C) experiences an approximate doubling of aqueous mobile phase viscosity relative to 20 °C. Operating at room-temperature flow rates in a cold room can double backpressure, exceeding the column’s safe operating limits. Flow rates must be reduced by 30% to 50% when running at 4 °C.
- Loading High-Ionic-Strength Feedstocks: Loading cell culture media or ammonium sulfate precipitation fractions directly onto an anion exchange column without prior desalting prevents electrostatic binding. Target proteins will flow directly into the void volume along with unbound contaminants.
Frequently asked questions
What is the difference between an anion exchange column and a cation exchange column?
An anion exchange column is packed with positively charged ligands (such as quaternary ammonium or DEAE) that bind negatively charged molecules (anions) when the mobile phase pH is above the analyte’s isoelectric point (pI). In contrast, a cation exchange column contains negatively charged ligands (such as sulfopropyl or carboxymethyl) that bind positively charged molecules (cations) when the buffer pH is below the analyte’s pI. Selecting between them depends on the target molecule’s net charge and stability window.
When should you choose a strong anion exchange resin over a weak anion exchange resin?
Strong anion exchange resin columns (e.g., Q or TMAE chemistries) should be chosen when you need a consistent positive charge across a wide pH range (typically pH 2 to 12), enabling broad method development and high-pH cleaning-in-place. Weak anion exchange resin columns (e.g., DEAE) are chosen when the strong exchanger binds the target molecule too tightly to elute without denaturation, or when the pH-dependent ionization of the weak ligand provides alternative separation selectivity for closely related isoforms.
How does an anion exchange column function in protein purification by ion exchange chromatography?
In protein purification by ion exchange chromatography, an anion exchange column functions as either a capture, intermediate, or polishing step. At an operating pH above its isoelectric point, the target protein carries a net negative charge and adsorbs onto the positively charged resin bed while uncharged or basic impurities pass through. The bound protein is then eluted selectively by applying an increasing sodium chloride gradient (typically 0 to 1.0 M NaCl), which displaces the protein based on its specific net surface charge density.
What is the maximum operating pressure for an anion exchange column?
Maximum operating pressure depends on column hardware and base bead rigidity. Prepacked polypropylene cartridges (like Cytiva HiTrap) are limited to 0.3 to 0.5 MPa (3 to 5 bar; 43 to 72 psi), whereas medium-pressure glass or PEEK columns (like Cytiva Mono Q or Bio-Rad ENrich) tolerate 2.0 to 4.0 MPa (20 to 40 bar; 290 to 580 psi). Analytical stainless steel and HPIC polymer columns (like Thermo Dionex IonPac AS11-HC) operate up to 34.5 MPa (345 bar; 5,000 psi).
Can 1 M sodium hydroxide (NaOH) be used to clean all anion exchange columns?
Most polymer- and cross-linked agarose-based anion exchange columns—such as Cytiva Capto Q, Mono Q, and Thermo Dionex IonPac—are fully stable in 0.5 M to 1.0 M NaOH for cleaning-in-place (CIP) and sanitization. However, silica-based anion exchange columns degrade rapidly at pH values above 8.0 and cannot tolerate sodium hydroxide. Always verify base matrix compatibility in the manufacturer instructions for use before applying alkaline CIP solutions.
Sources
- Cytiva. (2024). Ion Exchange Chromatography: Principles and Methods. Handbook 11-0004-21. Cytiva Life Sciences. https://www.cytivalifesciences.com (Accessed 2026-11-16).
- Cytiva. (2023). HiTrap Q HP and HiTrap SP HP Instructions. Document 71-7006-00. Cytiva Life Sciences. https://www.cytivalifesciences.com (Accessed 2026-11-16).
- Cytiva. (2023). Mono Q and Mono S Tricorn Columns Instructions. Document 71-7053-00. Cytiva Life Sciences. https://www.cytivalifesciences.com (Accessed 2026-11-16).
- Thermo Fisher Scientific. (2022). Dionex IonPac AS11-HC Analytical and Guard Column Product Manual. Document 031333-11. Thermo Fisher Scientific. https://www.thermofisher.com (Accessed 2026-11-16).
- Tosoh Bioscience. (2023). TSKgel Ion Exchange Columns Instruction Manual. Tosoh Bioscience GmbH. https://www.tosohbioscience.com (Accessed 2026-11-16).
- Bio-Rad Laboratories. (2022). ENrich High-Resolution Ion Exchange Columns Instruction Manual. Document 10027798. Bio-Rad Laboratories, Inc. https://www.bio-rad.com (Accessed 2026-11-16).
- Waters Corporation. (2021). Protein-Pak Hi Res IEX Columns Care and Use Manual. Document 720002937EN. Waters Corporation. https://www.waters.com (Accessed 2026-11-16).
Data: Anion Exchange Chromatography Column Technical Specifications and Operating Limits
Comparative specifications and operating parameters for leading analytical and preparative anion exchange chromatography columns compiled from manufacturer instructions for use and technical manuals.
Period covered: 2024/2026. Coverage: Global.
| Column Model | Manufacturer | Exchanger Type | Functional Ligand | Base Matrix | Particle Size (µm) | Operating pH Range | Max Pressure (bar) |
|---|---|---|---|---|---|---|---|
| Cytiva Mono Q 10/100 GL | Cytiva | Strong (SAX) | Quaternary ammonium (–N+(CH3)3) | Polystyrene/divinylbenzene (MonoBeads) | 10 | 2–12 | 40 |
| Cytiva HiTrap Q HP (1 mL) | Cytiva | Strong (SAX) | Quaternary ammonium (–N+(CH3)3) | Spherical agarose (High Performance) | 34 | 2–12 | 3 |
| Cytiva HiTrap Q FF (1 mL) | Cytiva | Strong (SAX) | Quaternary ammonium (–N+(CH3)3) | Highly cross-linked agarose (Fast Flow) | 90 | 2–12 | 3 |
| Cytiva Capto Q | Cytiva | Strong (SAX) | Quaternary ammonium (–N+(CH3)3) | High-flow cross-linked agarose with dextran | 90 | 2–12 | 3 |
| Thermo Dionex IonPac AS11-HC | Thermo Scientific | Strong (SAX) | Alkanol quaternary ammonium latex | Ethylvinylbenzene/55% divinylbenzene | 9 | 0–14 | 345 |
| Tosoh TSKgel SuperQ-5PW | Tosoh Bioscience | Strong (SAX) | Quaternary ammonium / polyamine | Hydroxylated polymethacrylate | 10 | 2–12 | 20 |
| Tosoh TSKgel DEAE-5PW | Tosoh Bioscience | Weak (WAX) | Diethylaminoethyl (–CH2CH2N+(C2H5)2H) | Hydroxylated polymethacrylate | 10 | 2–12 | 20 |
| Bio-Rad ENrich Q (5 x 50 mm) | Bio-Rad | Strong (SAX) | Quaternary ammonium (–N+(CH3)3) | Porous polymethacrylate | 10 | 2–12 | 34 |
| Waters Protein-Pak Hi Res Q | Waters Corporation | Strong (SAX) | Quaternary ammonium (Q) | Non-porous polymethacrylate | 5 | 3–10 | 150 |
Sources
- Cytiva Mono Q Instructions 71-7053-00 (accessed 2026-11-16)
- Cytiva HiTrap Q HP Instructions 71-7006-00 (accessed 2026-11-16)
- Cytiva Capto Q Data File 28-9037-77 (accessed 2026-11-16)
- Thermo Scientific Dionex IonPac AS11-HC Product Manual 031333 (accessed 2026-11-16)
- Tosoh Bioscience TSKgel 5PW Column Instruction Manual (accessed 2026-11-16)
- Bio-Rad ENrich High-Resolution Ion Exchange Columns Instruction Manual 10027798 (accessed 2026-11-16)
- Waters Protein-Pak Hi Res Columns Care and Use Manual 720002937EN (accessed 2026-11-16)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.