Short answer: ASTM F2503 is the standard practice for marking medical devices and other items for safety in the magnetic resonance (MR) environment. It sets a uniform system of icons and terms (MR Safe, MR Conditional, MR Unsafe) so staff can tell, from the label, whether an item may enter the MR environment. The edition listed as active on the ASTM store is F2503-26, last updated May 13, 2026. This page is general product and labeling information, not medical advice and not a substitute for your site’s MR safety procedures.
What ASTM F2503 covers
ASTM F2503 is maintained by ASTM Committee F04, subcommittee F04.15. Its scope is devices and other items anticipated to enter the MR environment: it specifies how they are marked with terms and icons and recommends what labeling information should accompany them. ASTM states that MR image artifacts are excluded from the mandatory portions because an artifact is not a direct safety issue. ASTM’s stated reason for the standard is that interactions between items and the MR environment have caused serious injuries and deaths, so a uniform system of visual icons and terms is needed (source: ASTM store page for F2503-26, checked 2026-11-01). The full text is sold by ASTM; this page does not reproduce it. For a plain-language walk through one of the terms, see the MR Conditional guide.
Editions and what changed in F2503-26
The ASTM store lists these editions: F2503-26 (active), F2503-23E01, -23, -20, -13, -08 and -05. The F2503-23 page is labelled historical. The British Institute of Radiology (BIR) “MR Safety Advice 2026” note says F2503-26 was published earlier in 2026 with small changes to improve the wording of the definitions, while the intention behind each definition remains unchanged (BIR document dated 21 July 2026, checked 2026-11-01).
That matters when you compare labels or documents from different years: older wording and newer wording describe the same three categories. The BIR note gives the updated definitions as:
- MR Conditional: having demonstrated safety in the MR Environment within defined conditions.
- MR Safe: posing no known hazards resulting from exposure to any MR Environment.
- MR Unsafe: posing unacceptable risk within the MR Environment.
MR Safe symbol and MR Unsafe symbol: terms, symbols and text
People searching for the “MR Safe symbol” or “MR Unsafe symbol” are usually looking at a label that carries an icon. FDA’s guidance on testing and labeling devices for the MR environment recommends that labeling use one of three standardized terms, MR Safe, MR Unsafe or MR Conditional, and that it “include the appropriate symbol from ASTM F2503 and/or the corresponding term”. The same guidance says color versions of the symbols are recommended, but black-and-white versions can be used. So a label may show a colored icon, a black-and-white icon, the written term, or both; all are expected forms, and a missing color does not by itself mean the marking is wrong.
The ASTM definitions and icon artwork are in the paywalled standard. For the exact graphics, use the standard or the manufacturer’s labeling rather than a redrawn copy.
Two further points from the BIR note: the MR Safe term is valid only for items composed entirely of electrically nonconductive, nonmetallic and nonmagnetic materials, and an MR Safe item can still cause image artifacts (the note gives an alcohol wipe as an example).
How FDA guidance uses the standard
The FDA final, nonbinding guidance “Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment” was issued October 10, 2023 (originally issued May 20, 2021; docket FDA-2019-D-2837). It points to F2503 for the symbols and terms and adds these labeling expectations:
- MR safety information goes in a separate labeling section titled “MRI Safety Information”, written so a healthcare professional can determine whether the device can safely enter the MR environment.
- The information should be readily accessible, for instance on the manufacturer’s website and/or by telephone, because the original labeling may not be available for the whole life of the device.
- For some passive implants and devices with no historical MRI information, labeling under the heading “Safety in MRI Not Evaluated” could be used in certain circumstances. That is a statement that evaluation was not done, not a safety category.
Label-check list for supply and lab staff
This checklist is built only from the FDA guidance and BIR note above. It helps you read what a label or product document says; it does not decide whether an item may enter a scanner room.
| Question to ask of the label or document | Why (source) |
|---|---|
| Is one of the three terms (or its symbol) present, or does it say “Safety in MRI Not Evaluated”? | FDA guidance recommends the three terms; the not-evaluated heading is a separate case. |
| Is it a symbol, a term, or both, and is the symbol in color or black and white? | FDA guidance: symbol and/or term; color recommended, black and white permitted. |
| For MR Conditional: are all the stated conditions listed (static field, gradient fields, RF fields, others) and can every one be met? | BIR note: the statement applies only when all stated conditions are met. |
| For MR Safe: is the item entirely nonconductive, nonmetallic and nonmagnetic? | BIR note: MR Safe is valid only for such items. |
| Is there an “MRI Safety Information” section, and a website or phone contact for the manufacturer? | FDA guidance on labeling section and accessibility. |
| Does the wording match an older edition’s phrasing? | BIR note: F2503-26 changed wording, not intent. |
Common misunderstandings
- “MR Safe means no effect on the image.” No. The BIR note says MR Safe items may still cause image artifacts.
- “MR Conditional means safe in any MRI.” No. The label applies only when all stated conditions are met, so the conditions must be found and checked.
- “The standard decides who may enter the scanner room.” No. The BIR note says the underlying ISO 14971 risk framework excludes individual clinical decisions, which need a qualified practitioner, and local procedures follow guidance from bodies such as MHRA and the American College of Radiology. See the guide on MRI safety zones and MR personnel.
- “A black-and-white symbol is non-compliant.” FDA guidance says black-and-white versions can be used.
FAQ
What is ASTM F2503?
The ASTM standard practice for marking medical devices and other items for safety in the MR environment, using standard terms and icons.
What does the MR Safe symbol mean?
It marks an item as MR Safe: posing no known hazards from exposure to any MR environment, per the 2026 wording summarized by BIR. Per the same note, it is valid only for entirely nonconductive, nonmetallic, nonmagnetic items.
What does the MR Unsafe symbol mean?
It marks an item as MR Unsafe: posing unacceptable risk within the MR environment.
Is ASTM F2503 required by FDA?
FDA’s guidance is nonbinding. It recommends labeling that includes the appropriate F2503 symbol and/or the corresponding term.
Which edition is current?
The ASTM store lists F2503-26 as active (last updated May 13, 2026), with F2503-23 labelled historical, as checked on 2026-11-01.
Related reading: radiology imaging center supply checklist, what an MRI scanner is, and what counts as a medical device.
Sources
- ASTM International, F2503-26 store page, https://store.astm.org/f2503-26.html (checked 2026-11-01).
- U.S. FDA, Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, guidance of October 10, 2023, https://www.fda.gov/media/74201/download (checked 2026-11-01).
- British Institute of Radiology, MR Safety Advice 2026, MR safety labelling standard, https://www.bir.org.uk/media/567268/1._mr_safety_labelling_standard.pdf (checked 2026-11-01).