Skip to main content
v2026.11,986 entries · CC-BY 4.0

Bradford Protein Assay: Standard Curve, Dilution Series and Calculating Concentration

How to run a Bradford protein assay: BSA standard series, dye reagent ratios, incubation time, and a worked calculation from A595 absorbance to mg/mL protein concentration.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

The Bradford assay is a colorimetric method for estimating total protein concentration in a solution. It relies on Coomassie Brilliant Blue G-250 dye, which shifts its absorbance maximum from about 465 nm (the free, reddish-brown cationic form of the dye in acidic solution) to about 595 nm (the blue anionic form) when it binds protein — predominantly at arginine residues, with smaller contributions from lysine, histidine and aromatic residues. Because the shift is proportional to the amount of dye-protein complex formed, absorbance at 595 nm can be converted to protein concentration once the reader has been calibrated against a standard curve of known concentrations, almost always bovine serum albumin (BSA).

This guide walks through preparing a BSA standard dilution series, running the assay, building the standard curve, and working back from a sample’s absorbance reading — through the standard curve and any dilution factor — to a final concentration in mg/mL.

To calculate protein concentration from a Bradford standard curve: read A595 for each BSA standard, plot absorbance (y) against BSA concentration (x), and fit a straight line y = mx + b. Subtract the blank, then solve x = (A595 − b) ÷ m for each sample and multiply by its dilution factor. Only readings inside the range of your standards are valid, so dilute and re-read anything above the top standard instead of extrapolating.

When to use Bradford vs. other protein assays

Bradford is one of several common total-protein quantification methods. Which one fits depends on speed, sensitivity needed, and what else is in the sample buffer. The characteristics below are general tendencies, not fixed values — exact sensitivity and tolerance depend on the specific kit, instrument and sample matrix.

Method Typical time Relative sensitivity Known interferences Protein-to-protein uniformity
Bradford (Coomassie) Fast — single reagent addition; Thermo Fisher states colour develops almost immediately, and published microplate protocols allow a 5–60 min read window Moderate; standard and “micro” kit formats cover different ranges Detergent and buffer tolerance depends on the formulation; ionic and nonionic detergents such as SDS and Triton X-100 are common problem additives, so check the compatibility table for your kit Lower — dye response depends heavily on a protein’s arginine/lysine content, so results are reported as BSA-equivalents
BCA (bicinchoninic acid) Slower — typically 30–60 min, often with heat Comparable to or broader than Bradford depending on kit Incompatible with reducing agents (DTT, β-mercaptoethanol) and chelators (EDTA) at meaningful concentrations; generally more detergent-tolerant than Bradford Higher — more consistent response across different proteins
Lowry / Folin-Ciocalteu Slowest — multiple sequential reagent additions and incubations Historically the reference method; largely superseded Sensitive to a wide range of buffer components, detergents and reducing agents Moderate
UV absorbance (A280) Instant — no reagents, non-destructive Requires reasonably concentrated, pure protein Nucleic acid contamination inflates readings; needs a known or estimated extinction coefficient for the specific protein Low unless the extinction coefficient is protein-specific

Bradford’s main advantages are speed and a single-reagent, single-incubation workflow, which is why it remains common for routine tasks like normalizing loading before a Western blot or an ELISA. Its main limitation is that the dye does not respond uniformly to every protein, so a Bradford result is really a BSA-equivalent estimate, not an absolute measurement of every protein species present.

What you need

  • Bradford reagent (commercial dye-based reagent, or dye concentrate diluted per the manufacturer’s instructions)
  • BSA standard stock (commonly supplied as a 2 mg/mL solution, or prepared from lyophilized BSA in the same buffer as your samples)
  • Diluent that matches your sample buffer as closely as possible, to avoid buffer-mismatch artifacts between standards and samples
  • Clean cuvettes or a microplate, and a spectrophotometer or plate reader capable of reading at 595 nm — see UV-Vis spectrophotometer basics and, if it has been a while since the last check, spectrophotometer calibration or microplate reader calibration and maintenance
  • Calibrated pipettes across the volume range you’ll be dispensing — small pipetting errors in the standard series are the single biggest source of a poor standard curve

Procedure

1. Prepare the BSA standard dilution series

Dilute the BSA stock into a series of known concentrations that bracket the concentration range you expect your samples to fall in, plus a zero (blank) standard. Commercial kits state their own working range: the Pierce Bradford kit (23200) and Bradford Plus reagent (23236) from Thermo Fisher Scientific both quote 1–1500 µg/mL, while the 590/450 nm ratio microplate method described below is built around 0–5 µg of BSA per well. Use the range stated for your specific kit, and see the data table below for the formats side by side. A workable general-purpose series, if your kit’s insert doesn’t specify one, is:

Standard BSA concentration (mg/mL)
S0 (blank) 0.0
S1 0.2
S2 0.4
S3 0.6
S4 0.8
S5 1.0
S6 1.2

Use a fresh serial dilution made up the same day the assay is run — BSA standards do not store reliably once diluted to working concentration. If you need to review how much stock and diluent each step requires, see molarity and solution calculations and buffer and solution preparation.

2. Prepare sample dilutions

If you don’t already have an estimate of your sample’s protein concentration, dilute it at more than one ratio (for example 1:5 and 1:20) so that at least one dilution is likely to land inside the standard curve’s linear range. Record every dilution factor — you will need it in the final calculation step.

3. Add the dye reagent

Add Bradford reagent to each standard and sample well or cuvette. Sample-to-reagent ratios vary by format — cuvette-based assays commonly use on the order of 1 part sample to 20–50 parts reagent, while 96-well microplate formats typically use smaller, more concentrated volumes (for example a few microliters of sample to roughly 200–250 µL of reagent). Follow the ratio specified for your kit rather than a generic figure, since deviating from it shifts the standard curve.

4. Mix and incubate

Mix gently (vortexing can introduce bubbles that distort plate-reader absorbance readings). Incubate at room temperature as the kit insert specifies. Thermo Fisher describes colour development in its Bradford reagents as almost immediate (as little as 30 seconds for the Plus reagent), whereas the Ernst and Zor microplate protocol allows a wait of 5 to 60 minutes before reading. Whatever the interval, use the same one for every standard and sample on a plate, and read within the stability window in your kit documentation.

5. Read absorbance at 595 nm

Blank the instrument as your kit or protocol directs (see “Choosing a blank” below for the difference between a reagent blank and a water blank), then read every standard and sample at 595 nm. If any sample reads above your highest standard, don’t extrapolate — dilute it further and re-read, because the Bradford response curve is not perfectly linear once you move past the top of the calibrated range.

Building the standard curve

Plot absorbance (y-axis) against known BSA concentration (x-axis) for each standard and fit a linear regression, y = mx + b, where m is the slope and b is the y-intercept (ideally close to zero if the blank subtraction was done correctly). Check the coefficient of determination (R²); a well-run Bradford standard curve should generally sit at R² ≥ 0.99 across its linear range. If it doesn’t, see the troubleshooting table below before trusting any concentration calculated from it. For the general method behind linear range and back-calculation, see building an analytical calibration curve.

Worked calculation (illustrative example)

The values below are a hypothetical, self-consistent data set built to demonstrate the calculation steps — not measured results from any real assay run. Substitute your own instrument’s readings using the same method.

Standard BSA (mg/mL) A595 (hypothetical)
S0 0.0 0.005
S1 0.2 0.120
S2 0.4 0.235
S3 0.6 0.350
S4 0.8 0.465
S5 1.0 0.580
S6 1.2 0.695

Linear regression on this set gives a slope m = 0.575 A595 per mg/mL and an intercept b = 0.005, i.e. A595 = 0.575 × [protein] + 0.005.

Suppose an unknown sample, diluted 1:10 before the assay was run, gives a reading of A595 = 0.410. Rearranging the regression equation to solve for concentration:

[protein]diluted = (A595 − b) ÷ m = (0.410 − 0.005) ÷ 0.575 = 0.405 ÷ 0.575 = 0.704 mg/mL

That figure is the concentration of the diluted sample that was actually read — and it falls inside the 0–1.2 mg/mL calibrated range, so the reading is valid. Multiply by the dilution factor to recover the concentration of the original, undiluted sample:

[protein]original = 0.704 mg/mL × 10 = 7.04 mg/mL

If a sample’s diluted reading falls above the top standard (here, above A595 ≈ 0.695), the correct response is to dilute it further and re-read — not to extrapolate the regression line past the calibrated range, since the true Bradford response curve tends to flatten (become less sensitive) at higher protein concentrations.

How to make a BSA standard curve for the Bradford assay

With a 2 mg/mL BSA stock, the 0.2–1.2 mg/mL series above can be made directly in 1 mL final volumes. Dilute in the same buffer as your samples, and mix each tube before taking the next volume.

Standard Final BSA (mg/mL) 2 mg/mL BSA stock (µL) Diluent (µL)
S0 (blank) 0.0 0 1000
S1 0.2 100 900
S2 0.4 200 800
S3 0.6 300 700
S4 0.8 400 600
S5 1.0 500 500
S6 1.2 600 400

Each volume follows C1V1 = C2V2: for S3, 0.6 mg/mL × 1000 µL ÷ 2 mg/mL = 300 µL of stock. Run standards and samples in duplicate or triplicate and average the replicates before fitting. If your expected samples are dilute, use a lower-range series suited to your kit rather than this one. The general method for choosing and checking a calibration range is covered in protein assays and the calibration curve guide.

Calculating the Bradford standard curve in Excel

Enter blank-subtracted A595 values in one column and BSA concentrations in another, then use the built-in functions SLOPE(known_y, known_x), INTERCEPT(known_y, known_x) and RSQ(known_y, known_x) to get m, b and R². For each sample, compute (A595 − b) ÷ m, then multiply by the dilution factor. Alternatively, insert a scatter plot of the standards and add a linear trendline with the equation and R² displayed. If the points curve at the top end, trim the highest standards or use a quadratic trendline, and only read samples that fall within the fitted range. The dye and its binding behaviour are described in Coomassie blue, and the older copper-based alternative in the Lowry assay.

Why the Bradford curve is not a straight line

The Coomassie response is only approximately linear. In the JoVE protocol by Ernst and Zor, the explanation given is that the absorbance of the free dye at 590 nm falls as dye binds protein and the dye’s spectrum shifts, so the 590 nm signal does not rise in direct proportion to protein; the same article cites Zor and Selinger (Anal Biochem 1996;236:302–308) on this effect. Practically, that means three options when your points bend at the top end: (1) restrict the fit to the linear part of the standards and dilute samples into it; (2) fit a quadratic (second-order polynomial) trendline and read only samples inside the standards; or (3) linearise the data with the 590/450 nm ratio described below. Sources differ on which is preferable, so pick one approach, apply it to every plate, and do not extrapolate beyond the highest standard.

Choosing a blank

Most kit instructions treat the zero-protein standard (reagent plus the same diluent as the samples) as the blank and subtract it. The Ernst and Zor protocol does something different for the ratio method: it blanks against deionised water, keeps a zero-protein dye well as an ordinary standard point, and calculates the ratio from the water-blanked readings. Subtracting the zero-protein well from a ratio would remove information the linearisation depends on. Follow the convention that goes with your fit: reagent-blank subtraction for a single-wavelength A595 curve as above, water blank plus a zero-protein point for the 590/450 method.

The 590/450 nm ratio method

Instead of reading only near 595 nm, the ratio method reads the plate at both 590 nm and 450 nm and divides the net 590 nm absorbance by the net 450 nm absorbance for each well. Plotting that ratio against BSA mass gives a markedly straighter line. Ernst and Zor report roughly a ten-fold gain in sensitivity compared with the conventional single-wavelength read, with detection down to about 50 ng of BSA. In their microplate format the standards are 0 and 10–50 µL of a 0.1 mg/mL BSA stock (0–5 µg per well, in triplicate) made up to 100 µL, each unknown is added as 100 µL diluted to roughly 5–50 µg/mL, and 100 µL of Bradford reagent diluted 2.5-fold is added for a 200 µL reaction. The protocol waits 5–60 minutes before reading. Calculation is otherwise identical to the worked example above, with the ratio replacing A595 on the y-axis. The data table below compares this format with the two Thermo Fisher kits.

Choosing the standard protein: BSA, BGG or something closer to your sample

BSA is the most common standard because it is cheap, pure and supplied by kit makers (Thermo Fisher ships 2 mg/mL albumin ampules with its Bradford kits), and the Pierce Bradford kit page also refers to BSA or BGG (bovine gamma globulin) dilution series. Different proteins give different slopes because the dye responds to protein composition, so results are expressed as BSA-equivalents. The Mattaini chapter on the Bradford assay in the LibreTexts Encyclopedia of Biological Methods notes that a standard closer in composition to the sample can be used when absolute accuracy matters. If you compare samples against each other, any consistent standard works; if you compare against a published concentration, use the same standard protein.

Matching the standards to the sample buffer

Prepare the standards in the same detergent and buffer as the samples (the JoVE protocol makes this point explicitly), so that any background contribution from the buffer appears equally in the curve and in the unknowns. Thermo Fisher notes that its Bradford Plus reagent is compatible with amine buffers, reducing agents, chaotropes and low concentrations of metal chelators, and its general kit compatibility criterion is a change of no more than 10% in the result; the manufacturer also claims about half the protein-to-protein variability of other commercial Bradford formulations. Treat those as manufacturer statements, and test your own buffer against a zero-protein control.

What the Bradford assay cannot detect

Thermo Fisher states that free amino acids, peptides and proteins below roughly 3,000 Da do not produce colour in its Bradford reagents. A Bradford reading of a digest, a peptide preparation or a low-molecular-weight fraction therefore under-reports the material present; use a method designed for peptides instead.

Scaling from microplate to cuvette

The JoVE protocol notes that cuvette-format volumes are about five times those of the microplate format. The ratios of sample, standard and reagent matter more than the absolute volumes, so scale all components together and keep the instrument pathlength in mind when comparing absorbance values between formats.

Troubleshooting

Symptom Likely cause Fix
Standard curve R² below about 0.99, or visibly curved Degraded or expired reagent; small pipetting errors magnified across the low-volume standard series; standards not made fresh Prepare BSA standards fresh the same day; use calibrated pipettes appropriate to the volumes involved; if you must extend beyond the linear range, consider a quadratic fit rather than forcing a straight line
Sample absorbance reads above the highest standard Sample is more concentrated than the calibrated range Dilute further and re-read; don’t extrapolate past the top standard
Blank (S0) absorbance is high or inconsistent between runs Reagent turbidity, contaminated diluent, dirty cuvettes/plate, or expired reagent Filter the reagent if the kit allows it, use fresh high-purity water, check the reagent’s expiry date, and use dedicated cuvettes/plates for Bradford work
Blue staining builds up on cuvettes or plate wells over repeated use Coomassie dye adsorbs to glass and especially plastic surfaces Use disposable or dedicated Bradford-compatible cuvettes/plates, and read promptly rather than letting the dye sit
Visible precipitate or turbidity in the sample-plus-reagent mixture Sample too concentrated, or high detergent/salt content driving protein-dye aggregation Dilute the sample further before adding reagent; if the sample buffer contains incompatible detergents, clean it up (e.g., dialysis or precipitation-based cleanup) before assaying
Bradford result disagrees notably with a BCA or Lowry result on the same sample Protein-to-protein variability — the Coomassie dye’s response depends on a protein’s arginine/lysine content, which differs from BSA Treat the Bradford number as a BSA-equivalent estimate rather than an absolute value; if precise absolute quantification matters, cross-check with an orthogonal method or a standard protein closer in composition to your sample
Little to no color change across the whole standard series Reagent not added, added in the wrong order or volume, or reagent improperly stored (Coomassie reagent is light-sensitive) Confirm the reagent-addition step, check storage conditions and expiry, and re-run with a fresh reagent aliquot
Readings drift when re-measured later in the same run Plate/cuvettes read outside the reagent’s stated color-stability window Read within the incubation window specified for your kit; don’t leave plates sitting on the bench well past that window before reading

Limitations to keep in mind

  • Protein-to-protein variability. Because the dye binds preferentially to arginine and, to a lesser extent, lysine, histidine and aromatic residues, two proteins at the same true mass concentration can give different Bradford readings. Every Bradford number is technically a “BSA-equivalent” estimate, not an absolute mass concentration, unless you calibrate against a standard closer in composition to your actual protein of interest.
  • Detergent and buffer sensitivity. Tolerance varies by formulation. Detergents such as SDS and Triton X-100 are common causes of interference, so if your samples are lysed or stored in detergent-containing buffer, check the compatibility table for your reagent, match the standards to the sample buffer, or clean the sample up first.
  • Small peptides and amino acids. Proteins below about 3,000 Da and free amino acids give no colour (Thermo Fisher).
  • Non-linearity at the high end. The dye-binding response tends to flatten above the top of the calibrated range rather than staying linear, which is why diluting an out-of-range sample and re-reading is preferred over extrapolating the regression line.

Frequently asked questions

What is the linear range of the Bradford assay?

It depends on the product. Thermo Fisher states a working range of 1–1500 µg/mL for both its Pierce Bradford kit and Bradford Plus reagent, and the 590/450 nm ratio method reaches down to about 50 ng of BSA per well. Always use the range stated for your specific kit.

Why is my Bradford standard curve not linear?

Part of it is the chemistry: the 590 nm background falls as dye binds protein, so the response curves at higher concentrations, and a quadratic fit or the 590/450 nm ratio method handles this better than a straight line. Degraded reagent, pipetting errors in the standards and stale standards add scatter on top. See the sections above and the troubleshooting table.

Can I use the Bradford assay on samples containing SDS or Triton X-100?

It depends on the reagent and concentration. Detergents are a well-known source of interference with Coomassie dye, so check your kit’s compatibility table, prepare the standards in the same buffer, and include a zero-protein control containing the detergent. If the effect is too large, use a detergent-compatible assay format or clean the sample up first.

How long after mixing should I read the plate or cuvette?

Follow your kit insert. Thermo Fisher describes colour development as almost immediate, while the Ernst and Zor protocol allows 5–60 minutes. Use one consistent interval for the whole plate.

Why does Bradford give a different number than BCA for the same sample?

Because the two assays respond to different chemistry — Bradford’s Coomassie dye binds preferentially to arginine and lysine residues, while BCA’s copper-reduction chemistry responds more uniformly across amino acid composition. Neither number is more “true” than the other in an absolute sense; both are estimates relative to whatever standard protein (usually BSA) was used to build the curve.

Is the Bradford standard curve linear or quadratic?

Neither is universally correct. The response is slightly curved, so a straight line works over a restricted range, a quadratic fit works over a wider one, and the 590/450 nm ratio makes the data close to linear. Choose one approach per assay and keep samples within the standards.

What blank should I use for a Bradford assay?

For a single-wavelength A595 curve, the zero-protein standard (reagent plus sample diluent) is the usual blank. For the 590/450 nm ratio method, Ernst and Zor use a deionised-water blank and keep a zero-protein dye point in the curve.

Why use a 590/450 nm ratio?

It linearises the response and, in the published protocol, gives about ten-fold higher sensitivity, down to about 50 ng of BSA.

Can the Bradford assay detect peptides?

Generally not: Thermo Fisher states that free amino acids, peptides and proteins smaller than about 3,000 Da produce no colour.

What standard protein should I use, BSA or BGG?

BSA is the default and is supplied with Thermo Fisher kits; BGG is also used for a dilution series. Use the same standard as any reference value you are comparing against, or one closer in composition to your protein when absolute accuracy matters.

How do I calculate sample concentration from the curve equation?

Rearrange y = mx + b to x = (y − b) ÷ m, using the blank-corrected reading, then multiply by the sample’s dilution factor, as in the worked example above.

Sources

Data: Bradford (Coomassie) assay formats: readout, range and protocol parameters from kit pages and a peer-reviewed method

Compares three Bradford assay formats on working range, absorbance readout, reaction volume and read timing, compiled from two Thermo Fisher Scientific product pages and the Ernst and Zor JoVE linearization protocol. Ranges are as stated by each source and are not directly interchangeable; the 5-50 ug/mL limits for the ratio method are the recommended dilution target for unknowns in that protocol.

Period covered: 2010/2026.

Bradford (Coomassie) assay formats: readout, range and protocol parameters from kit pages and a peer-reviewed method
FormatAbsorbance readoutStated lower limit (ug/mL)Stated upper limit (ug/mL)Total reaction volume per well (uL)Read timingStandard protein
Pierce Bradford Protein Assay Kit (23200)595 nm11500not stated by the manufacturercolour develops almost immediatelyBSA (2 mg/mL ampules supplied); BGG dilution series also referenced
Pierce Bradford Plus (23236)595 nm11500not stated by the manufacturercolour develops almost immediately; as little as 30 sBSA (2 mg/mL ampules supplied)
Ernst and Zor 590/450 ratio microplate method590 and 450 nm (ratio)550200wait 5-60 minBSA (0.1 mg/mL stock, 0-5 ug per well)

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Bradford Protein Assay: Standard Curve, Dilution Series and Calculating Concentration

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →