Short answer
The CAIRE FreeStyle Comfort (model AS200) is a portable, pulse-dose oxygen concentrator made by Caire, Inc. FDA cleared it under 510(k) K250671 on 2025-07-11. Battery runtime depends on the pulse setting: the technical manual’s battery table gives up to 8 h (single pack) and 16 h (double pack) at setting 1, falling to 2 h and 4 h at setting 5, and its specification table separately rates the packs at 3.5 h and 6.0 h. No source read for this page gives a price, a ranking of brands, or a number of hours to use it each day.
What the CAIRE FreeStyle Comfort is
It is a battery-powered unit that concentrates oxygen from room air and delivers it in pulses rather than as a continuous stream. Per the technical manual (MN242 Rev D), it offers pulse settings 1 to 5, runs from AC (100-240 V, 50-60 Hz), DC (11-18 V) or its lithium-ion battery, and can be carried by handle, shoulder strap or backpack. It is a prescription device: the manual states it should only be used when prescribed by a physician, that the flow selection must not be changed unless directed by a licensed clinician, and that it is not for life support.
Regulatory identity
| Item | Value | Source |
|---|---|---|
| 510(k) number | K250671, Traditional 510(k), Substantially Equivalent | openFDA 510(k) database |
| Applicant and decision date | Caire, Inc.; 2025-07-11 | openFDA 510(k) database |
| Product code and name | CAW, Generator, Oxygen, Portable | openFDA classification database |
| Device class and regulation | Class 2; 21 CFR 868.5440 | openFDA classification database |
The same product code and regulation number apply to other portable oxygen generators, so they are a way to find comparable cleared devices, not a statement about performance.
Specifications from the technical manual
| Specification | Stated value (MN242 Rev D, firmware 2.7+, 4-button interface) |
|---|---|
| Delivery mode | Pulse dose only, settings 1-5 |
| Output per minute, settings 1 to 5 | 210 / 420 / 630 / 840 / 1050 mL/min |
| Oxygen concentration | 90% (+5.5/-3%) at 14.7 psi, 70 F |
| Sound level | 39.93 dB(A) +/- 0.42 |
| Battery | Lithium-ion; single 6700 mAh, double 13400 mAh |
| Operating temperature and humidity | 5-40 C; 15-95% non-condensing |
| Warranty | 3 years on the unit; 1 year on battery and power supplies |
The manufacturer’s product page lists a weight of 2.3 kg and dimensions of 10 x 7.3 x 3.1 in, and states the unit meets FAA requirements for commercial air travel. The manual advises contacting the airline at least 48 hours before travel.
Battery runtime by pulse setting
The technical manual (MN242 Rev D) gives battery duration by pulse setting. The product page’s “up to 8 hours” and “up to 16 hours” figures for the 8-cell and 16-cell packs match the manual’s setting 1 maximums, so the two Caire documents agree.
| Pulse setting | Single (8-cell) pack | Double (16-cell) pack |
|---|---|---|
| 1 | up to 8 h | 16 h |
| 2 | up to 4 h | 8 h |
| 3 | up to 3 h | 6 h |
| 4 | up to 2.25 h | 4.5 h |
| 5 | up to 2 h | 4 h |
Source: the manual’s battery pack duration table. Separately, the manual’s specification table lists battery pack durations of 3.5 h (single) and 6.0 h (double), plus “up to 4 hours” and “up to 8 hours” at setting 2, which matches the setting 2 row above. The manual does not state in that table which setting the 3.5 h and 6.0 h ratings apply to, so this page does not attribute them to one. Pulse-dose runtime depends on the setting, so a single figure cannot be applied to every user. The manual states the batteries are designed to last 300 charge/discharge cycles and carry a 1-year warranty; extra battery packs are listed by Caire as separate accessories. For battery behaviour on concentrators in general, see oxygen concentrator batteries.
Accessories and backpack
Caire lists a backpack, a shoulder strap, a desktop charger and extra battery packs, sold separately. The manual describes carrying by handle, shoulder strap or backpack.
Test mode
The technical manual describes a Test Mode used for service and troubleshooting, such as an oxygen concentration test and an hour meter. It is a service function, not an everyday user setting, and this page gives no steps for it.
Where it matters for supply and clinic decisions
Staff handling these units can use three identifiers to match paperwork: the 510(k) number, the product code, and the model name AS200. Equipment billing for portable oxygen is covered separately in HCPCS E1390. Fire and handling precautions are in home oxygen safety, and the choice between concentrators and cylinders in oxygen concentrator vs O2 tank. Power draw is discussed in how much electricity an oxygen concentrator uses, and the wider category in respiratory therapy equipment.
Common misunderstandings
- “Up to 16 hours” is not a guaranteed figure. It is the manual’s setting 1 maximum for the double pack; at higher pulse settings the manual’s durations are shorter (4 h at setting 5).
- Setting numbers are not litres per minute. Pulse settings 1-5 correspond to 210-1050 mL/min of output per the manual and are not directly equivalent to continuous flow.
- It is not an over-the-counter device. Federal law restricts sale or rental to the order of a physician, per the manual.
- Test mode is not a user feature. It is documented for service troubleshooting.
FAQ
How much does the Caire FreeStyle Comfort cost?
None of the manufacturer, FDA or manual sources read for this page states a price, and prices are set by suppliers, so this page gives no figure.
Which brand of oxygen concentrator is considered the best?
No regulator or standards body ranks brands. FDA clears devices individually under product code CAW and similar codes. Which device suits a patient is a prescriber and supplier decision.
How long does the Caire FreeStyle Comfort battery last?
The technical manual’s battery table gives, for the single (8-cell) and double (16-cell) packs, 8 h and 16 h at setting 1, 4 h and 8 h at setting 2, 3 h and 6 h at setting 3, 2.25 h and 4.5 h at setting 4, and 2 h and 4 h at setting 5. The product page’s “up to 8 or 16 hours” matches the setting 1 figures. The manual’s specification table also lists 3.5 h and 6.0 h as battery pack durations. Runtime varies with the setting and conditions.
How many hours a day should I use my oxygen concentrator?
That is a medical question and depends on the individual prescription. The manual says to use the device only as prescribed and not to change the flow selection unless directed by a licensed clinician. Ask the prescribing clinician.
Sources
- openFDA 510(k) database, K250671 (api.fda.gov/device/510k.json), data last updated 2026-10-26, read 2026-11-07.
- openFDA device classification, product code CAW, 21 CFR 868.5440, read 2026-11-07.
- CAIRE FreeStyle Comfort Technical Manual MN242 Rev D (files.caireinc.com/TechMan_PartsLists_ServiceBull/MN242.pdf), read 2026-11-07.
- CAIRE FreeStyle Comfort product page (www.caireinc.com/products/freestyle-comfort/), read 2026-11-07.