Which plastic has the best chemical resistance? Fluoropolymers (PTFE, PFA, FEP) rate excellent against every chemical category in the chart below. HDPE and polypropylene are the economical choice for acids, bases and alcohols, while PVC, polycarbonate and polystyrene are attacked by most solvents. Concentration, temperature and contact time can change a rating, so confirm critical choices against the manufacturer’s chart, and see hydrofluoric acid safety for HF and polypropylene vs polystyrene for the two everyday labware resins.
Scope note: this chart covers chemical resistance of common lab plastics used as containers, tubing, wash bottles, carboys, and general labware — the material question of “what can I safely store or dispense this chemical in.” It is a different decision from selecting wetted materials for process equipment such as pump seals, valve trim, and gasket elastomers, which is covered separately in the chemical compatibility chart for wetted materials of construction. It also does not cover glove selection (see the chemical-resistant glove selection guide) or chemical storage segregation rules. Use this page when you are specifying or purchasing plastic labware; use the wetted-materials chart when you are specifying process equipment.
How to read this chart
Chemical resistance ratings are a starting point for a specification decision, not a guarantee. Whether a plastic holds up against a given chemical depends on the chemical’s concentration, the temperature of contact, the duration of exposure, and the specific resin grade and wall thickness of the part. A rating of “good” at room temperature and typical reagent concentration can fail at elevated temperature or with a concentrated or glacial form of the same chemical. Manufacturers of lab plasticware (for example Thermo Fisher/Nalgene, Cole-Parmer, and VWR) publish their own detailed chemical resistance guides tested under specific conditions — when a decision is safety-critical or involves a chemical not covered here, confirm against the specific product manufacturer’s chart and, where available, the reagent’s safety data sheet (SDS) storage recommendations.
- Excellent — the plastic is a standard, widely used choice for this chemical category; negligible attack under normal lab conditions.
- Good — generally suitable, but resistance narrows at higher temperature, higher concentration, or long-term storage; check the specific grade before committing to long-term use.
- Limited — usable for brief or dilute exposure only (e.g., a quick rinse), not for storage; verify before specifying.
- Poor — documented incompatibility (swelling, softening, cracking, dissolution, permeation, or embrittlement) is common; avoid unless the manufacturer explicitly confirms suitability for your exact formulation.
Chemical resistance chart: common lab plastics by chemical category
The chart translates the manufacturer E/G/F/N key described below (E = Excellent, G = Good, F = Limited, N = Poor) into plain words, using the class-level ratings at 20 °C in the Fisher Scientific/Nalgene reference; where polyethylene grades differ, both are shown. The data table after this article lists the underlying letters together with temperature limits and autoclavability.
| Chemical category | HDPE / LDPE | Polypropylene (PP) | PTFE / PFA / FEP | PVC | Polycarbonate (PC) | Polystyrene (PS) |
|---|---|---|---|---|---|---|
| Weak or dilute acids (e.g., dilute HCl, dilute H₂SO₄) | Excellent | Excellent | Excellent | Excellent | Excellent | Excellent |
| Strong or concentrated acids (class rating at 20 °C) | Excellent | Excellent | Excellent | Excellent | Poor | Limited |
| Bases (NaOH, KOH) | Excellent | Excellent | Excellent | Excellent | Poor | Excellent |
| Aliphatic alcohols (methanol, ethanol, isopropanol) | Excellent | Excellent | Excellent | Excellent | Good | Excellent |
| Ketones (acetone, MEK) | Good | Good | Excellent | Poor | Poor | Poor |
| Halogenated hydrocarbons (chloroform, DCM) | Limited–Poor | Limited | Excellent | Poor | Poor | Poor |
| Aromatic hydrocarbons (toluene, xylene, benzene) | Limited–Good | Limited | Excellent | Poor | Poor | Poor |
| Strong oxidizers (conc. nitric acid, chromic acid, conc. H₂O₂) | Limited | Limited | Excellent | Good | Poor | Poor |
| Esters (ethyl acetate) | Good | Good | Excellent | Poor | Poor | Poor |
General guidance derived from class-level manufacturer ratings, not a substitute for a manufacturer chart or SDS review. “Limited” and “Poor” cells often still mean brief incidental contact is fine (a quick rinse) while storage or continuous exposure is not — the distinction matters for how you use the container, not just what it’s made of.
Notes on each plastic
HDPE and LDPE (polyethylene)
The default choice for general aqueous reagents, dilute acids and bases, and most alcohols. HDPE is more rigid and slightly more chemical- and temperature-resistant than LDPE; LDPE is more common for squeeze wash bottles. Neither holds up well to aromatic or chlorinated solvents, which can cause swelling and slow permeation over time — a polyethylene bottle can “sweat” volatile organics through the wall even when the cap seal looks intact.
Polypropylene (PP)
Similar acid/base resistance to polyethylene but with a higher usable temperature (PP can typically be autoclaved; most PE cannot), which is why PP is the standard material for autoclavable labware, centrifuge tubes, and pipette tips. Reasonable resistance to many ketones and dilute oxidizers, but not to strong oxidizing acids at high concentration or most chlorinated/aromatic solvents.
PTFE, PFA, and FEP (fluoropolymers)
The broadest chemical resistance of any common lab plastic, including strong oxidizers, and most organic solvents. The tradeoff is cost and lower mechanical toughness than HDPE/PP for the same wall thickness, so fluoropolymers are typically reserved for the specific reagents (highly reactive or high-purity chemicals) that need them, rather than used as a default. For examples of corrosive acids, bases and solvents and the first aid for skin and eye contact, see corrosive chemicals in the lab.
PVC
Adequate for dilute aqueous acids and bases and common in tubing applications, but degraded by many solvents (ketones, chlorinated and aromatic hydrocarbons) and unsuitable for concentrated oxidizing acids. Rigid PVC and flexible (plasticized) PVC behave differently — plasticizer can leach into some solvents, which is itself a compatibility failure mode distinct from the base resin degrading.
Polycarbonate (PC)
Valued for clarity and impact strength (autoclavable flasks, safety shields, some carboys) but chemically one of the weaker choices on this list — attacked by many solvents, strong bases, and some amines, which can cause crazing or stress cracking even without full dissolution. Do not assume a polycarbonate container is chemically resistant just because it is durable; those are separate properties.
Polystyrene (PS)
Common for disposable, single-use labware (petri dishes, cuvettes, some pipette bodies) exposed to aqueous solutions only. Polystyrene dissolves or softens in a wide range of organic solvents, including many that other plastics on this chart tolerate well, so it should be treated as single-use/aqueous-only rather than a general storage material.
What is HDPE? What is LDPE?
HDPE (high-density polyethylene) and LDPE (low-density polyethylene) are two forms of the same polymer family. The U.S. food-contact regulation for olefin polymers describes polyethylene as “basic polymers manufactured by the catalytic polymerization of ethylene” (21 CFR 177.1520). What separates the two is chain structure: Thermo Fisher’s Nalgene material pages state that HDPE “has less side branching than LDPE resulting in a denser 3-D structure,” while LDPE “has more side branching than HDPE resulting in a less-dense 3-D structure.”
| Property (Nalgene material pages) | HDPE | LDPE |
|---|---|---|
| Structure | Less side branching, denser | More side branching, less dense |
| Specific gravity | 0.95 | Lighter than water |
| Flexibility | Semi-rigid, less flexible than LDPE | Very flexible without added plasticizers |
| Maximum use temperature | 120 °C | 80 °C |
| Autoclavable | No | No |
| Typical labware | Bottles, carboys, wide-mouth bottles, funnels, buckets | Wash bottles, dropper bottles, sample vials |
For chemical resistance at 20 °C, the manufacturer rates HDPE excellent against dilute or weak acids, alkalis and aliphatic alcohols, good against strong acids, aldehydes, esters and aliphatic hydrocarbons, and reduced to fair or not recommended for continuous contact with aromatic and halogenated hydrocarbons, depending on the grade and the reference table. It also notes that long-term UV exposure can damage HDPE. The two polyethylenes therefore share a column in the chart above; the practical differences are stiffness and heat tolerance, not a different chemical-resistance class. For the wider polyethylene family, see What Is Polyethylene? and Polyolefin vs Polyethylene.
PFA vs PTFE: how the two fluoropolymers differ
PTFE (polytetrafluoroethylene) is a homopolymer of tetrafluoroethylene. PFA (perfluoroalkoxy) is a copolymer of tetrafluoroethylene with a perfluoroalkyl vinyl ether; 21 CFR 177.1550 lists the copolymer of perfluoropropylvinylether and tetrafluoroethylene among the perfluorocarbon resins and addresses the tetrafluoroethylene homopolymer separately in its specifications. Both resist nearly all chemicals in the chart above, so the choice between them rests on how the part is made and used.
- Processing. PFA is melt-processable by conventional injection molding and screw extrusion; PTFE is not (Process Technology). This is why molded PFA bottles, fittings and seamless tubing exist, while PTFE parts are formed by other routes.
- Heat. Process Technology reports a melting point of about 327 °C for PTFE versus about 260 °C for PFA, and describes PTFE as having slightly higher heat resistance. Nalgene lists PFA labware with a maximum use temperature of 260 °C, defined as exposure for minutes to about two hours with little loss of strength. Vendors define temperature limits differently, so compare datasheets using the same criterion.
- Appearance and handling. Nalgene describes PFA labware as translucent and flexible, and states it is autoclavable.
- Chemical resistance. Nalgene rates PFA excellent (no damage after 30 days of constant exposure at 20 °C) for most chemical classes, and notes that temperature, exposure time and stress change results. Process Technology describes PFA as resistant to almost all solvents and chemicals and PTFE as having high chemical resistance, without ranking one above the other.
Where the decision concerns lined pipe, valves or seals rather than a container, use the wetted-materials compatibility chart.
The E/G/F/N rating key manufacturers use
Nalgene’s reference guide and the Fisher Scientific reprint of it grade each plastic against each chemical with four letters, defined by exposure tests rather than opinion:
- E (excellent): no damage after 30 days of constant exposure.
- G (good): little or no damage after 30 days of constant exposure.
- F (fair): some effect after 7 days of exposure, such as cracking, crazing, loss of strength or discoloration. Solvents may soften, swell or permeate PPCO, PP, PMP, LDPE and HDPE, normally reversibly.
- N (not recommended): not recommended for continuous use; immediate damage may occur, including severe cracking, crazing, loss of strength, deformation, dissolution or loss of contents through permeation.
The poster edition gives two letters per chemical, one for 20 °C and one for 50 °C, so a reader can see how much margin a plastic loses when it warms up. The ratings were generated for Nalgene products and the poster states that they do not apply to its centrifugeware or filterware, so a chart is evidence about a resin under test conditions, not a guarantee for another maker’s part.
How chemicals damage plastics: three failure modes
The Nalgene reference separates three ways a plastic can fail, which is why a single “compatible” label hides different risks:
- Chemical attack on the polymer chain: oxidation, reaction with functional groups, or depolymerization. This is permanent and is what strong oxidizers do to polyolefins.
- Physical change: absorption and swelling, permeation through the wall, or dissolution. Solvent swelling of polyethylene and polypropylene is normally reversible, but it softens the part and can carry the contents through the wall.
- Stress cracking: cracking under load that is not a reaction of the chemical with the polymer.
Environmental stress cracking
Environmental stress cracking needs three things at once: mechanical stress, a stress-cracking agent, and a plastic that is inherently susceptible. The reference lists detergents, surface-active chemicals, lubricants, oils, ultrapure water and plating additives such as brighteners and wetting agents as agents, and notes that small concentrations may be enough. This is why a container can survive months of storage and then crack when it is tightened, centrifuged or filled to the neck, and why an apparently harmless wash with detergent can start a crack in a stressed polycarbonate or polyethylene part.
What changes a rating: temperature, concentration, pressure and time
The manufacturer states that resistance falls as temperature rises, and that pressure, mechanical stress (centrifugation is the example given), exposure time and concentration all matter. Two practical consequences follow. First, a chemical rated E at 20 °C may be only G or F at 50 °C on the same chart, so read the second letter on the poster when a solution will be warmed. Second, mixing or diluting chemicals can generate heat and change both temperature and concentration at the wall of the container, so the rating of the starting material is not the rating of the mixture.
Temperature limits and autoclaving by plastic
Resistance is only useful within the temperature range the plastic itself tolerates. From the Nalgene poster (maximum use temperature, brittleness temperature): HDPE 120 °C and −100 °C; LDPE 80 °C and −100 °C; polypropylene 135 °C and 0 °C; PPCO 121 °C and −40 °C; polycarbonate 135 °C and −135 °C; polystyrene 90 °C and 20 °C; FEP 205 °C and −270 °C; PFA 260 °C and −270 °C. The Fisher reprint adds PTFE at 260 °C and −267 °C and PVC at 70 °C and −30 °C. Polystyrene’s brittleness temperature of 20 °C means it is already brittle at normal room temperature, which suits a thin disposable dish and not a vessel that will be knocked or chilled.
The poster defines autoclaving as 121 °C, 15 psig for 20 minutes. It advises rinsing a container with distilled water first, because a chemical that is harmless at room temperature can attack the resin at autoclave temperature. Polypropylene, PPCO and the fluoropolymers are autoclavable; HDPE, LDPE and polystyrene are not. Polycarbonate can be autoclaved but each cycle reduces its strength, and the manufacturer says not to use autoclaved polycarbonate under vacuum. For which resins tolerate steam, see which plastics are autoclavable.
Strong oxidizing agents: the exception to “polypropylene is fine”
The Nalgene poster says not to store strong oxidizing agents in plastic labware other than FEP or PFA (the Fisher reprint adds PTFE). Prolonged exposure embrittles polyolefins, and the poster’s own words are that a forgotten container will fail in time. This matters because oxidizers are easy to overlook: concentrated nitric acid, chromic acid and concentrated hydrogen peroxide all fall in this class, and a polypropylene bottle that survives a week may not survive a quarter in a cupboard. Date the container, record the contents, and move oxidizers to fluoropolymer or a compatible alternative on the SDS.
Inspect and replace aged labware
The poster recommends inspecting plastic labware before each use for cracking, crazing (a network of fine spiderweb cracks), permanent discoloration and embrittlement, and replacing it when aging shows. Crazing in particular is a warning that stress cracking has begun. A bottle that has been in service with aggressive chemicals should be retired on inspection, not on a fixed calendar interval.
Gas and vapour permeation through the wall
Chemical resistance is not the same as impermeability. The Fisher reprint gives gas permeability values (cc-mm per square metre per 24 hours per bar) that show how readily gases cross a plastic wall:
| Plastic | Nitrogen | Oxygen | Carbon dioxide |
|---|---|---|---|
| HDPE | 651 | 2,868 | 8,990 |
| Polypropylene | 744 | 3,720 | 12,400 |
| Polycarbonate | 775 | 4,650 | 16,663 |
| Polystyrene | 853 | 4,650 | 17,825 |
The values explain why oxygen- or CO₂-sensitive reagents are not held in thin-walled plastic indefinitely, and why a volatile solvent can be lost through polyethylene even when the cap is tight. PVC is reported at 31–310 for nitrogen and 62 for oxygen and carbon dioxide in the same table.
Practical selection steps
- Identify the exact chemical, its concentration, and the expected temperature and duration of contact — not just the chemical family.
- Check the reagent’s SDS (Section 7, Handling and Storage) for any manufacturer-specified container material requirements.
- Cross-check the candidate plastic against a manufacturer chemical resistance chart for the specific product line, since resin formulation and wall thickness affect real-world performance.
- For long-term storage of solvents (ketones, chlorinated or aromatic hydrocarbons), default to fluoropolymer or verify the specific PP grade rather than assuming general-purpose PP or PE is adequate.
- Where the decision concerns process equipment (pumps, valves, seals, tubing runs) rather than a storage container, use the wetted-materials compatibility chart instead, since it also covers elastomers and metals not included here.
Frequently asked questions
Can acetone be stored in a plastic bottle?
Not in PVC, polystyrene or polycarbonate: the Nalgene class rating for ketones is not recommended for continuous use (N) for all three. Polypropylene and polyethylene are rated good (G) at 20 °C as a class (little or no damage after 30 days), PTFE/PFA/FEP excellent. The rating is for the ketone class, so check the individual solvent and your container grade, and remember that solvents can swell and permeate polyolefins, normally reversibly.
Is PVC chemically resistant?
PVC has good resistance to dilute aqueous acids, bases, and many salts, which is why it is common in lab tubing and drainage lines, but it is a poor choice for ketones, chlorinated solvents and aromatic solvents,.
What plastic is compatible with concentrated sulfuric acid?
PTFE/PFA/FEP offers excellent resistance. Nalgene rates strong acids excellent for HDPE, PP and PVC at 20 °C as a class, but concentrated sulfuric acid is also an oxidizing agent, and strong oxidizers are rated only fair for polyethylene and polypropylene, so the class rating overstates the margin. For concentrated or fuming sulfuric acid, verify against the manufacturer chart for the specific product before specifying.
Does chemical resistance depend on the container’s wall thickness?
Yes for permeation-driven failures (slow diffusion of a solvent through an intact wall) — a thicker wall slows permeation but does not change whether the resin itself is fundamentally compatible. For attack-driven failures (swelling, cracking, dissolution), wall thickness affects how long the container survives before failing, not whether it is a safe long-term choice.
What is HDPE?
HDPE is high-density polyethylene, a high molecular weight polyolefin with little side branching, so its chains pack densely. In lab plastics it is the rigid polyethylene used for bottles and carboys; Nalgene lists a maximum use temperature of 120 °C and states it is not autoclavable.
What is LDPE?
LDPE is low-density polyethylene, the more branched and more flexible polyethylene, used for wash bottles and dropper bottles. Nalgene lists a maximum use temperature of 80 °C and states it is not autoclavable.
What is the difference between PFA and PTFE?
Both are fluoropolymers with very broad chemical resistance. PFA can be melt-processed by injection molding and extrusion, which allows molded bottles and seamless tubing, while PTFE cannot be melt-processed and has a higher melting point (about 327 °C versus about 260 °C for PFA).
What do E, G, F and N mean on a chemical resistance chart?
E means no damage after 30 days of constant exposure, G means little or no damage after 30 days, F means some effect after 7 days, and N means not recommended for continuous use because immediate damage may occur (Nalgene key).
What plastic is resistant to strong acids?
At 20 °C the Nalgene class ratings for weak or dilute acids are excellent for all six common plastics in the data table, and for strong or concentrated acids they are excellent for HDPE, LDPE, PP, PVC and the fluoropolymers, fair for polystyrene and not recommended for polycarbonate. Oxidizing acids such as concentrated nitric acid fall under the oxidizer class and need fluoropolymer.
Why does polycarbonate crack when it touches detergent?
Detergents and other surface-active chemicals are classic stress-cracking agents. Cracking needs mechanical stress, an agent and a susceptible plastic together, and small concentrations can be enough, so it is a stress effect and not simple chemical attack.
Can I autoclave polycarbonate, polystyrene or HDPE?
Polycarbonate can be autoclaved at 121 °C, 15 psig for 20 minutes but loses strength each cycle and should not be used under vacuum afterwards. Polystyrene and HDPE are not autoclavable.
Can strong oxidizers be stored in polypropylene bottles?
No. The manufacturer’s guidance is to keep strong oxidizing agents out of plastic labware except FEP or PFA, because prolonged exposure embrittles polyolefins.
Sources
- 21 CFR 177.1520, Olefin polymers: https://www.law.cornell.edu/cfr/text/21/177.1520 (checked 2026-10-31)
- 21 CFR 177.1550, Perfluorocarbon resins: https://www.law.cornell.edu/cfr/text/21/177.1550 (checked 2026-10-31)
- Thermo Fisher Scientific, High-Density Polyethylene (HDPE) Labware: https://www.thermofisher.com/us/en/home/life-science/lab-plasticware-supplies/plastic-material-selection/high-density-polyethylene-hdpe-labware.html (checked 2026-10-31)
- Thermo Fisher Scientific, Low-Density Polyethylene (LDPE) Labware: https://www.thermofisher.com/us/en/home/life-science/lab-plasticware-supplies/plastic-material-selection/low-density-polyethylene-ldpe-labware.html (checked 2026-10-31)
- Thermo Fisher Scientific, Teflon (PFA) Labware: https://www.thermofisher.com/us/en/home/life-science/lab-plasticware-supplies/plastic-material-selection/teflon-pfa-labware.html (checked 2026-10-31)
- Process Technology, PFA and PTFE: What is the Difference?: https://www.processtechnology.com/pfa-ptfe-whats-the-difference/ (checked 2026-10-31)
- Thermo Fisher Scientific, Nalgene plastic labware chemical resistance reference guide (poster): https://documents.thermofisher.com/TFS-Assets/LCD/posters/LCD-Chemical-Resistance-Guide-GALSPCHEMRESPOSTER-EN.pdf (accessed 2026-11-17)
- Fisher Scientific, Chemical Resistance and Physical Properties of Plastics (reprinted from Nalge Nunc): https://ehs.web.baylor.edu/sites/g/files/ecbvkj1581/files/2024-02/fisher-chemical-chemical-resistance-of-plastic-technical-bulletin.pdf (accessed 2026-11-17)
Data: Lab plastics: temperature limits and 20 °C chemical resistance ratings by chemical class (Nalgene key)
Compiles maximum use and brittleness temperatures and autoclavability from the Nalgene physical-properties poster and the Fisher Scientific/Nalgene Chemical Resistance and Physical Properties of Plastics reference, and joins them with the class-level chemical resistance ratings at 20 °C from the same Fisher/Nalgene reference. Rating key: E = no damage after 30 days of constant exposure; G = little or no damage after 30 days; F = some effect after 7 days; N = not recommended for continuous use. PP and PPCO, and FEP, PFA and PTFE, share a rating in the source. PTFE and PVC temperatures come from the Fisher reference; other temperatures from the Nalgene poster.
Period covered: 2024.
| Plastic | Maximum use temperature (°C) | Brittleness temperature (°C) | Autoclavable | Weak or dilute acids | Strong or concentrated acids | Aliphatic alcohols | Bases | Ketones | Aromatic hydrocarbons | Halogenated hydrocarbons | Strong oxidizing agents |
|---|---|---|---|---|---|---|---|---|---|---|---|
| HDPE | 120 | -100 | No | E | E | E | E | G | G | F | F |
| LDPE | 80 | -100 | No | E | E | E | E | G | F | N | F |
| PP | 135 | 0 | Yes | E | E | E | E | G | F | F | F |
| PPCO | 121 | -40 | Yes | E | E | E | E | G | F | F | F |
| PC | 135 | -135 | Yes (strength reduced) | E | N | G | N | N | N | N | N |
| PS | 90 | 20 | No | E | F | E | E | N | N | N | N |
| PVC | 70 | -30 | No | E | E | E | E | N | N | N | G |
| PTFE | 260 | -267 | Yes | E | E | E | E | E | E | E | E |
| FEP | 205 | -270 | Yes | E | E | E | E | E | E | E | E |
| PFA | 260 | -270 | Yes | E | E | E | E | E | E | E | E |
Sources
- Thermo Scientific Nalgene plastic labware chemical resistance reference guide (poster) (accessed 2026-11-17)
- Fisher Scientific, Chemical Resistance and Physical Properties of Plastics (reprinted from Nalge Nunc) (accessed 2026-11-17)