CytoFLEX Flow Cytometer: How It Works vs Attune NxT

What the Beckman Coulter CytoFLEX is, how its APD detection compares with the Attune NxT’s acoustic focusing, with every figure attributed.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

Short answer

The CytoFLEX is Beckman Coulter’s family of flow cytometry analyzers built on avalanche photodiode detectors. The Attune NxT is Thermo Fisher’s analyzer built on acoustic focusing and a syringe pump. Both analyze suspended cells and neither is a cell sorter.

Overview

The CytoFLEX is Beckman Coulter’s family of multicolor flow cytometry analyzers, built around avalanche photodiode (APD) detectors and a set of repositionable band pass filters, while the Thermo Fisher Attune NxT is a separate analyzer that lines cells up with ultrasonic acoustic focusing and measures sample volume with a syringe pump. Both are laboratory analyzers for suspended cells and particles, not cell sorters, and both are sold for research use.

Definition and scope: the CytoFLEX

CytoFLEX is the name of a Beckman Coulter flow cytometry platform, not a single instrument. Beckman’s own 2017 announcement of the CytoFLEX LX describes the platform as expanding its detection range, and its navigation lists further family members: CytoFLEX S, CytoFLEX nano and CytoFLEX mosaic. Laser count, laser wavelengths and detector count differ between models, so a facility’s configuration is the specification that matters, not the family name.

Two design points recur in the primary sources. First, the platform uses avalanche photodiode detectors. Beckman states in its press release that APDs have high and stable quantum efficiency from 400 to 1100 nm, which it presents as the reason the LX can carry an infrared laser, and it says APDs are not available in other flow cytometers (a manufacturer claim, reported here as Beckman’s statement). The University of Copenhagen Flow Cytometry and Single Cell Analysis facility likewise describes its APD detectors as giving unmatched red sensitivity. Second, the optics use an extensive set of repositionable band pass filters, which Beckman says lets a lab upgrade by adding parameters.

CytoFLEX LX

According to Beckman Coulter’s 5 September 2017 release, the CytoFLEX LX offers excitation sources from ultraviolet 355 nm to infrared 808 nm, and Beckman describes it as the first standard-configuration analyzer to include both ultraviolet and infrared lasers. The release also refers to intuitive software for multicolor analysis and to sensitivity, but gives no numerical sensitivity figure.

CytoFLEX S as configured in a core facility

The University of Copenhagen facility publishes the configuration of its four-laser CytoFLEX S: solid-state lasers at 405 nm (80 mW), 488 nm (50 mW), 561 nm (30 mW) and 638 nm (50 mW), with 13 fluorescence parameters plus forward scatter (FSC) and side scatter (SSC). Its example filters include 525/40 for FITC, 585/42 for PE, 660/10 for APC and 450/45 for BV421. The facility lists resolution of particles under 300 nm using the 488 nm side scatter and under 80 nm (polystyrene) using the 405 nm side scatter, a robust CV (rCV) of 3%, and a seven-decade digital display. It also lists a High Throughput Option combining a plate loader with a single-tube switch for 5 mL (12 x 75 mm) or 1.2 mL tubes.

The Attune NxT is Thermo Fisher Scientific’s flow cytometer built on acoustic focusing. Thermo describes acoustic focusing as using ultrasonic waves together with hydrodynamic forces to position cells in a single focused line. This is the main design difference from a conventional hydrodynamic-only system, and it changes what the instrument can do with sample flow rate.

Thermo’s features page gives the following, all of them manufacturer statements: a sample rate adjustable from 12.5 to 1,000 uL/min; acquisition at up to 35,000 events per second; sample throughput up to 10 times faster than traditional hydrodynamic focusing; and a rare-event sample input of up to 1 mL/min compared with 60 to 100 uL/min on conventional systems. Thermo also states that its pressure control runs at 75 PSI against a maximum of 15 PSI on traditional cytometers.

The same page describes up to four spatially separated, solid-state, flat-top lasers and up to 16 parameters. Thermo’s BioProbes issue 71 article on the instrument lists laser lines of 405 nm (50 mW), 488 nm (50 mW), 561 nm (50 mW) and 637 nm (100 mW), a quartz cuvette flow cell gel-coupled to a 1.2 NA lens, an FSC photodiode and an SSC photomultiplier tube (488/10 band pass), and keyed, user-changeable filters. The article gives 14 emission channels where the features page says 16 parameters; the two counts are described differently and should not be added or merged. The red laser appears as 637 nm in BioProbes and as the 638 nm class on other platforms, so compare laser lines by the manufacturer’s own listing for the exact model.

Volumetric counting

Thermo states that the Attune NxT uses a positive-displacement syringe pump, which gives volumetric counts: absolute counts in a known volume of sample. BioProbes gives an analysis volume of 20 uL to 4 mL, tubes from 17 x 100 mm down to 8.5 x 45 mm, a maximum file size of 20 million events, a particle range of 0.5 to 50 um, and an optional autosampler of up to 384 samples. It lists the instrument at 40 x 58 x 43 cm and 29 kg, with startup and shutdown of 15 minutes or less. Thermo’s stated sensitivity is fluorescence resolution CV under 3% and predicted MESF of no more than 80 for FITC, 30 for PE and 70 for APC.

How it works: the two designs in practice

Detection: avalanche photodiodes versus photomultiplier tubes

The CytoFLEX platform is described as APD-based, with the red and infrared end of the spectrum singled out by both Beckman and the Copenhagen facility. The Attune NxT page and article describe a PMT for side scatter and a photodiode for forward scatter, and do not state the fluorescence detector type in its descriptions. A detector comparison beyond what each manufacturer and facility states is therefore not made here.

Sample delivery: hydrodynamic focusing versus acoustic focusing

Acoustic focusing is Thermo’s mechanism for keeping cells in a tight line at high flow rates. The practical question for a lab is whether its samples are rare, dilute or large in volume. Thermo’s own comparison (up to 1 mL/min for rare events against 60 to 100 uL/min conventionally) frames the instrument for those cases. For the CytoFLEX S, the Copenhagen page describes throughput through its plate loader and tube options rather than through focusing, so the two instruments are described in different terms in the primary sources and a like-for-like speed figure does not exist in them.

Compensation workflow

The Copenhagen facility notes that the CytoFLEX S has a compensation library that stores dye spillover values, and that absolute linear gain amplification lets compensation be reused across experiments. That is a statement about workflow, not a replacement for running single-stain controls; see the guide on flow cytometry compensation and single-stain controls for how controls are used to correct spillover.

The data table below sets the facility and manufacturer statements side by side, with each cell tied to a named source. It deliberately keeps vendor claims labelled as vendor claims. Where a source does not state a value, the cell says so instead of filling the gap.

Where it matters for labs and core facilities

The sourced facts point to four decisions a core facility or lab group faces when asked to run one of these instruments. They are framed as questions each design answers, not as a recommendation.

  • Which wavelengths do your panels need? The CytoFLEX LX is described by Beckman as spanning 355 nm to 808 nm; the Copenhagen CytoFLEX S has four lines (405, 488, 561, 638 nm); the Attune NxT is listed with up to four lines (405, 488, 561 and 637 nm in BioProbes). Check the exact laser list of the unit on site, because the family name does not fix it.
  • Do you need absolute counts in the run itself? Thermo states the Attune NxT’s syringe pump provides volumetric counts. The CytoFLEX sources cited here do not describe volumetric counting, so a lab needing absolute counts on a CytoFLEX should ask the facility how it obtains them.
  • How will samples arrive? The Copenhagen CytoFLEX S lists a plate loader with a single-tube switch for 5 mL and 1.2 mL tubes; BioProbes lists Attune NxT tube sizes and an optional 384-sample autosampler.
  • How is compensation handled? A stored compensation library (Copenhagen, CytoFLEX S) helps repeat work with the same dyes, but each experiment still needs its own controls under the facility’s rules.

For the underlying technique, the guide to what a flow cytometer is explains fluidics, optics and electronics in general, and how flow cytometry data are displayed as dot plots and histograms covers reading the output. If your question is about mass-tag detection rather than fluorescence, CyTOF is a different technology altogether.

Common misunderstandings

Acoustic focusing is not cell sorting

Acoustic focusing positions cells for measurement. It does not separate populations. The Attune NxT and the CytoFLEX models discussed here are analyzers; separating cells into collection tubes requires a sorter, a different class of instrument.

Faster flow does not remove the need for good samples

Thermo’s throughput claims concern how quickly cells can be passed through the laser path. They do not change sample preparation, controls or gating, which still determine the quality of the result.

CytoFLEX is not one specification

Quoted laser powers, parameter counts and filter sets belong to a particular model and configuration, such as the four-laser CytoFLEX S at Copenhagen. Applying one unit’s numbers to another is the commonest way to misstate what an instrument can do.

Vendor performance claims are the vendor’s

Statements such as APDs being unavailable in other cytometers, or acoustic focusing being up to 10 times faster, come from the manufacturers’ own materials. Facility pages give configurations and resolution figures for specific units. Neither is an independent head-to-head test.

Research use only

Thermo’s BioProbes article states the Attune NxT is for research use only and not for diagnostic use. Check the labelling of any specific cytometer before using it for anything other than research.

FAQ

Is the CytoFLEX a cell sorter?

No. The sources cited here describe the CytoFLEX platform, including the LX and the S, as analyzers. Separating cells into collection tubes requires a sorter, a different class of instrument.

Is the Attune NxT a cell sorter?

No. Thermo describes acoustic focusing as positioning cells in a line for measurement; it does not separate populations.

Does the CytoFLEX use photomultiplier tubes?

Beckman and the University of Copenhagen facility describe the platform’s detectors as avalanche photodiodes, so the sources cited here describe APDs rather than PMTs.

Does the Attune NxT give absolute counts?

Thermo states that its positive-displacement syringe pump provides volumetric counts, meaning absolute counts in a known sample volume.

Sources

  • Beckman Coulter, “Beckman Coulter Expands Detection Boundaries of CytoFLEX Flow Cytometry Platform” (press release, 5 September 2017): https://www.beckman.com/news/cytoflex-lx-uv-press-release, read 2026-11-15.
  • University of Copenhagen, Flow Cytometry and Single Cell Analysis, CytoFLEX S (4-laser): https://www.fcsc.ku.dk/equipment/flow-cytometers/cytoflex/, read 2026-11-15.
  • Thermo Fisher Scientific, Attune NxT Flow Cytometer features: https://www.thermofisher.com/us/en/home/life-science/cell-analysis/flow-cytometry/flow-cytometers/attune-nxt-flow-cytometer/features, read 2026-11-15.
  • Thermo Fisher Scientific, BioProbes 71, “The Attune NxT Acoustic Focusing Cytometer”: https://www.thermofisher.cn/cn/en/home/references/newsletters-and-journals/bioprobes-journal-of-cell-biology-applications/bioprobes-71/bioprobes-71-attune-nxt-cytometer.html, read 2026-11-15.

Data: CytoFLEX and Attune NxT: sourced specifications side by side

Compiled from Beckman Coulter's CytoFLEX LX press release, the University of Copenhagen core facility page for its four-laser CytoFLEX S, and Thermo Fisher's Attune NxT features page and BioProbes 71 article. Each cell is the source's own statement for that model; vendor claims are attributed to the vendor.

CytoFLEX and Attune NxT: sourced specifications side by side
AttributeCytoFLEX S (Copenhagen facility)CytoFLEX LX (Beckman, 2017)Attune NxT (Thermo Fisher)
Lasers405 nm 80 mW, 488 nm 50 mW, 561 nm 30 mW, 638 nm 50 mW (solid state)Excitation from UV 355 nm to infrared 808 nmUp to 4 flat-top solid-state lasers: 405 nm 50 mW, 488 nm 50 mW, 561 nm 50 mW, 637 nm 100 mW
Detection describedAPD detectors; facility says unmatched red sensitivityAPDs with quantum efficiency stated as high and stable from 400 to 1100 nmFSC photodiode; SSC PMT (488/10 band pass)
Parameters13 fluorescence plus FSC and SSCParameters can be added by upgrade (Beckman)Up to 16 (features page); 14 emission channels (BioProbes)
FiltersExamples: 525/40, 585/42, 660/10, 450/45Extensive set of repositionable band pass filtersKeyed, user-changeable filters
Sample focusingNot stated by the sourceNot stated by the sourceAcoustic focusing: ultrasonic waves plus hydrodynamic forces; 12.5 to 1,000 uL/min (Thermo)
Volumetric countingNot stated by the sourceNot stated by the sourcePositive-displacement syringe pump gives volumetric counts (Thermo)
Sample formats5 mL (12 x 75 mm) or 1.2 mL tubes; optional plate loaderNot stated by the sourceTubes 17 x 100 to 8.5 x 45 mm; optional autosampler up to 384 samples
Stated resolution or sensitivityParticles under 300 nm (488 nm SSC), under 80 nm polystyrene (405 nm SSC); rCV 3%Qualitative only (no numbers given)Fluorescence resolution CV under 3%; predicted MESF 80 FITC, 30 PE, 70 APC or lower
Use statusNot stated by the sourceNot stated by the sourceResearch use only (BioProbes)

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about CytoFLEX Flow Cytometer: How It Works vs Attune NxT

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →