Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
Short answer: yes. A sterile lancet carries an expiry date from its manufacturer, and the manufacturer’s instructions for use say not to use it after that date. Separately, a lancet is a single-use item: the date does not mean it can be used many times before it lapses. This page describes products and general use; it is not medical advice.
Who this is for: clinic staff, buyers and students who handle lancets and need to know what to check before a lancet is used or kept in stock. For what a lancet is and how the classes differ, see what is a lancet; for the same date question on test strips, see glucose blood test strips.
What a lancet is
A blood lancet is a sharp, pointed device used to puncture the skin to obtain a small blood sample, and in the U.S. it is regulated as a medical device under 21 CFR 878.4850. Lancets for sampling are supplied sterile, as the Greiner Bio-One IFU describes for its safety lancet.
Do lancets expire?
Yes. The instructions for use (IFU) of the Greiner Bio-One MiniCollect Lancelino safety lancet state: “Do not use MiniCollect Lancelino Safety Lancets after their expiry date.” The same IFU describes the device as “sterile, single-use” and “sterilized using irradiation”, so the printed date is the manufacturer’s labelled limit for an unused, intact, sterile lancet.
The U.S. regulation for blood lancets (21 CFR 878.4850) does not set a shelf life. It requires sterility testing for any component that breaches the skin. The expiry period itself is therefore a manufacturer’s labelled limit, not a number found in the regulation.
What the expiry date on a lancet means
The date belongs to the sterile pack, not to the metal alone. A lancet punctures skin, so the regulation’s special control requires that “sterility testing must demonstrate the sterility of any device component that breaches the skin.” The date is the manufacturer’s statement of how long the sterile, intact product is labelled for use. Other manufacturers may state a different period, so read the label on the lot in hand.
How many times can you use a lancet?
Once. The labelling that 21 CFR 878.4850 requires for single-use lancets reads: “For use only on a single patient. Discard the entire device after use.” and “Warning: Not intended for more than one use. Do not use on more than one patient.” The Lancelino IFU says the same in shorter terms: the devices are for single-use only, and its label key includes a “Do not re-use” symbol.
The regulation also covers lancet devices with a reusable base. For those, the blade is the single-use part: labelling for multiple-use, single-patient devices states that used blades “must be safely discarded after a single use”. A fresh blade goes in for every use.
Two rules people mix up
The expiry date and the single-use rule answer different questions.
| Rule | What it governs | Source |
|---|---|---|
| Expiry date | Whether an unused, sealed lancet may still be used | Manufacturer IFU (Greiner Bio-One MiniCollect Lancelino) |
| Single use | Whether any lancet may be used a second time, expired or not | 21 CFR 878.4850 labelling; manufacturer IFU |
Types and variants
21 CFR 878.4850 describes several lancet types: single-use lancets with an integral sharps injury prevention feature (Class II), single-use lancets without one (Class II), multiple-use lancets for a single patient with a reusable base (Class II; blade changed with every use) and multiple-use lancets for multiple patients (Class III, premarket approval, 21 CFR 878.4850(d)). The Lancelino IFU also notes that the safety lancet comes in several versions, differing by puncture depth. The regulation sets labelling and sterility-testing controls. It does not give an expiry period.
What to check before using or stocking lancets
These checks come from the Greiner Bio-One MiniCollect Lancelino IFU and apply to that product; other products’ IFUs may add to them.
- Expiry date: not passed. Do not use after it.
- Protective tab: still in place. The IFU says not to use the lancet if the tab has been removed.
- Single-use status: never reuse; the IFU describes the lancet as single-use only and its label key carries a “Do not re-use” symbol.
- Sterilization marking: the label shows the sterilization method (irradiation for this product).
- Version and intended use: the IFU lists several versions of the product and states it is not recommended for heel-sticks. See lancet gauge sizes for how versions differ.
- Disposal: the IFU says to discard used devices in biohazard containers approved for that purpose; see red sharps containers and where to take a sharps container.
For a supplier, the documents that carry these facts are the manufacturer’s IFU and the labelled expiry date and lot on the pack.
Common mistakes
- Treating “single-use” and “expiry date” as one rule. A lancet inside its date is still never reusable.
- Assuming that an expired lancet is fine because it “still works”. The date relates to sterility, which the regulation’s sterility-testing control is about, not to whether the point still pricks.
- Using a lancet whose protective tab or pack is already open.
- Expecting the regulation to state a shelf life. It does not; read the label.
This page does not give individual medical advice. For the wider sampling context, see capillary blood collection.
FAQ
Do lancets expire?
Yes. Manufacturer IFUs, such as the Greiner Bio-One MiniCollect Lancelino IFU, say not to use the lancet after its expiry date. The date reflects the labelled limit of the sterile product.
How many times can you use a lancet?
Once. FDA labelling for single-use lancets says to discard the entire device after use and that it is not intended for more than one use. On reusable-base devices, only the blade is replaced, and it is discarded after a single use.
Sources
- U.S. FDA, 21 CFR 878.4850 Blood lancets, Electronic Code of Federal Regulations, checked 2026-11-01: https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-878/subpart-E/section-878.4850; text also at https://www.law.cornell.edu/cfr/text/21/878.4850.
- Greiner Bio-One GmbH, MiniCollect Lancelino Safety Lancets, instructions for use (manufacturer IFU), checked 2026-11-01; manufacturer site: https://www.gbo.com/preanalytics.