A lancet’s gauge number describes how thick its point is: the higher the number, the thinner the point. This page lists the gauge and penetration depth pairs that manufacturers declare for single-use safety lancets in the U.S. FDA’s device database, so you can see which combinations exist and read a product label against them.
Lancet gauge and depth pairs declared in FDA records
The table comes from the openFDA UDI records for FDA product code FMK, checked on 2026-10-30. Each row is a gauge and depth pair that at least one record declares; the count is the number of records with that pair. Only pairs where a record states both values are listed, and depth is shown as declared, in millimetres.
| Declared gauge | Declared depth | Records | Products declaring it |
|---|---|---|---|
| 18G | 1.8 mm | 3 | Unistik 3, Unistik TinyTouch (Owen Mumford) |
| 18G | 3.0 mm | 1 | Unistik 2 Super 18G (Owen Mumford) |
| 21G | 2.0 mm | 10 | Unistik, Unistik 3, Unistik Touch (Owen Mumford); Cardinal Health |
| 23G | 1.8 mm | 7 | Unistik, Unistik 3 (Owen Mumford) |
| 23G | 2.0 mm | 2 | Unistik Touch (Owen Mumford) |
| 23G | 2.2 mm | 1 | Comfort EZ Safety Lancet (Simple Diagnostics) |
| 25G | 1.6 mm | 2 | Unistik Pro (Owen Mumford); Cardinal Health |
| 26G | 1.5 mm | 1 | Comfort EZ Safety Lancet (Simple Diagnostics) |
| 28G | 1.2 mm | 2 | Unistik Pro (Owen Mumford); Cardinal Health |
| 28G | 1.5 mm | 1 | SafetiCet (NSP Tech) |
| 28G | 1.8 mm | 8 | Unistik, Unistik 3, Unistik Touch (Owen Mumford) |
| 30G | 1.5 mm | 5 | Unistik 3, Unistik Touch (Owen Mumford); Discount Drug Mart |
These are manufacturer-declared values for specific products. They are not a standard, and the FDA records do not state outer diameters in millimetres, so none are listed. Check your product’s labelling for its own gauge and depth.
How to read the table
Read gauge and depth as two separate specifications. Gauge is a point-thickness number, and a higher number means a thinner point. Depth is how far the point is declared to penetrate. Across the 206 FMK records that declare a gauge, the most common are 30G (57 records), 28G (51), 21G (45) and 23G (20); 25G, 26G, 18G, 33G, 31G and 16G appear less often (9, 8, 7, 4, 3 and 2 records).
Gauge is not the whole specification
The same gauge is declared with different depths. 28G appears at 1.2 mm, 1.5 mm and 1.8 mm, and 23G at 1.8 mm, 2.0 mm and 2.2 mm. Two lancets sharing a gauge number can therefore differ in depth, and a depth figure on its own does not tell you the gauge. When comparing products, compare both values.
What the records cover
FDA classifies product code FMK as a “Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature”: Class 2, regulation number 21 CFR 878.4850. FDA’s definition describes a disposable lancet with a single-use blade on a non-reusable base, used to puncture the skin to obtain a drop of blood for diagnostic purposes, whose sharps injury prevention feature makes it inoperable after one use. Reusable lancets for lancing devices fall outside this code, so the table does not describe them. For the item itself, see what a lancet is.
Common mix-ups
- Gauge versus depth. They are different numbers on the same label; neither one implies the other.
- Gauge versus a device setting. A number on an adjustable lancing device is not the lancet’s gauge unless the product’s labelling says so.
- Lancet gauge versus needle gauge. Both use the higher-is-thinner convention, but a lancet is a blade or point for a skin prick, not a hollow needle. For the hollow device, see what a hypodermic needle is.
Check your product’s labelling
Gauge and depth are declared by each manufacturer for each product, so the packaging and instructions for use are the authority for the lancet in your hand. A product’s record can also be looked up in AccessGUDID by brand name or UDI. Related pages: a diabetes clinic supply checklist and FDA-cleared sharps containers. For wider supply categories, see diagnostic and monitoring supplies.
FAQ
What are lancet sizes?
Lancet size is usually given as a gauge number plus a penetration depth. In FDA’s records for single-use safety lancets, declared gauges range from 16G to 33G and the declared depths in the table run from 1.2 mm to 3.0 mm, as checked on 2026-10-30.
Is a higher lancet gauge thinner?
Yes. By the usual convention a higher gauge number means a thinner point, so a 30G lancet is thinner than a 21G lancet.
Does the same gauge always have the same depth?
No. In the records above, 28G is declared at 1.2, 1.5 and 1.8 mm. Read both values on the label.
How many millimetres is a lancet gauge?
The FDA records used here do not state outer diameters in millimetres, so this page gives none. The millimetre figures in the table are penetration depths.
Sources
- openFDA device UDI API, product code FMK: https://api.fda.gov/device/udi.json?search=product_codes.code:FMK (manufacturer-submitted UDI data; checked 2026-10-30). Human-readable lookup: https://accessgudid.nlm.nih.gov/
- openFDA device classification, product code FMK: https://api.fda.gov/device/classification.json?search=product_code:FMK (checked 2026-10-30)