Written and maintained by CASRAI Editorial Board
Published
On this page
A dust monitor is an instrument that measures airborne dust, either as mass per volume of air (milligrams or micrograms per cubic metre) or as a count of particles in size bins. Filter-and-balance samplers, wearable mass monitors and optical particle counters are the three working principles covered below.
What a dust monitor is, and who uses it
“Dust monitor” covers several different instruments. What they share is that they sample air and report how much particulate material it holds. What differs is the quantity reported. A gravimetric sampler reports the mass of dust collected on a filter. A tapered element oscillating microbalance reports mass in near real time. An optical particle counter reports how many particles of each size passed through its sensing volume. These are not interchangeable readings, and the page you are reading is organised around that distinction.
The people who use them differ with the principle. Underground coal miners, safety personnel and mine operators use wearable continuous monitors for respirable coal dust. Industrial hygienists use pump-and-filter samplers following published analytical methods. Researchers and facilities staff use optical counters and light-scattering sensors to follow particle levels in a room, a plant or a community. For the indoor and cleanroom end of that range, see the CASRAI comparison of portable and remote cleanroom particle counters.
How dust monitors measure: the three principles
The data table below sets the three principles side by side, with the output each one gives and the size selection it relies on. The sections that follow explain each one using the instrument documentation and agency pages listed under Sources.
Filter and balance (gravimetric)
The reference approach in occupational hygiene is to pull a known volume of air through a size-selective sampler onto a filter, then weigh the filter. NIOSH Method 0600, “Particulates not otherwise regulated, respirable”, describes it. The sampler is a cyclone plus a tared 5-µm pore-size polyvinyl chloride membrane filter. The method lists three cyclones with their pump flow rates: a 10-mm nylon cyclone at 1.7 L/min, a Higgins-Dewell cyclone at 2.2 L/min and an aluminium cyclone at 2.5 L/min, each with a flow tolerance of ±5%. The measurement technique is gravimetric: the mass gained by the filter, divided by the volume of air sampled, gives the concentration of the respirable dust fraction.
The method is explicit about the laboratory side. Filters are equilibrated for at least 2 hours in a controlled weighing area, the analytical balance has a sensitivity of 0.001 mg, the same balance is used before and after sampling, and the method calls for 2 to 10 field blanks per sample set. Its estimated limit of detection is 0.03 mg per sample, and the working range is 0.5 to 10 mg/m³ for a 200-L air sample. The consequence for a buyer or a student is practical: a gravimetric result is only available after the filter has been weighed, so it is a measurement made after sampling, not a live display.
Wearable mass monitor (TEOM)
NIOSH describes the continuous personal dust monitor (CPDM) as a wearable device that gives miners, safety personnel and operators real-time information about respirable coal dust exposure. The commercially available model it names, the PDM3700, is based on the tapered element oscillating microbalance (TEOM), a technology first developed as a fixed-site environmental particulate mass monitor and then, working with the manufacturer and its successor Thermo Fisher Scientific, reduced to a device that can be worn safely underground. The screen shows dust levels for the previous 30 minutes and the cumulative shift-long average, so a worker or supervisor can see a trend during the shift rather than waiting for a laboratory result. The data can be uploaded to the Mine Safety and Health Administration (MSHA) sample collection database.
NIOSH also gives the history behind the change. Federal dust sampling in operating coal mines has been mandated since 1969, and MSHA mandated CPDM use for dust sampling in February 2016. Before the CPDM, filter samples went to a laboratory, which often took weeks; the monitor cut reporting to minutes.
Optical sensors and particle counters
Most low-cost and indoor dust monitors work by light scattering. The US Environmental Protection Agency’s Air Sensor Toolbox separates two designs. In an optical particle counter, particles are individually sized and counted by how they scatter light, and the output is particle counts in size ranges: EPA’s table lists a Dylos unit that reports counts above 0.5 µm and above 2.5 µm, designed for indoor use, and an Alphasense OPC-N2 that counts particles from 0.38 to 17 microns and converts the cumulative counts into estimates of PM2.5 and PM10 mass concentration. In a volume-scattering sensor, particles entering the sensor scatter light from an internal source and the total scattered signal is converted into an estimated mass concentration, which may include particles larger than 2.5 µm. The word “estimated” matters: in both designs, the mass figure is computed from a light signal, not weighed.
How many microns is dust?
Dust has no single size. Airborne particles are grouped by diameter, and the groups that monitors report are PM10 and PM2.5. The EPA defines PM10 as inhalable particles with diameters generally 10 micrometres and smaller, and PM2.5 as fine inhalable particles with diameters generally 2.5 micrometres and smaller. For scale, EPA notes that the average human hair is about 70 micrometres across, which makes it 30 times larger than the largest fine particle. Particles of sand and large dust, larger than 10 micrometres, are not regulated by EPA as particulate matter.
| Term | Diameter | What EPA says |
|---|---|---|
| PM10 | generally 10 µm and smaller | Inhalable particles; some particles under 10 µm can get deep into the lungs |
| PM2.5 | generally 2.5 µm and smaller | Fine inhalable particles, also called fine particles; those under 2.5 µm pose the greatest health risk |
| Human hair | about 70 µm (average) | 30 times larger than the largest fine particle |
| Sand and large dust | larger than 10 µm | Not regulated by EPA |
“Respirable” is a different, sampler-defined fraction used in workplace hygiene. NIOSH Method 0600 defines respirable dust as the aerosol collected by a sampler with a 4-µm median cut point, so a respirable result depends on the cyclone doing its size selection correctly. The method notes that bias depends on the dust size distribution.
What are some techniques for measuring dust concentration?
The three principles in the data table are the main techniques: collect on a filter and weigh it; measure mass continuously with a microbalance; or count and size particles optically. The choice follows from the question being asked. A gravimetric filter result, taken with an approved sampler, is the form of measurement that a regulation can name. A wearable mass monitor adds a live view. An optical counter is suited to following changes in particle levels over time and by size.
Worked example: from filter mass to concentration
The following arithmetic is an illustration only, not a measurement from any real site or a substitute for the full method. Suppose a respirable sampler drew 200 L of air, which at the nylon cyclone’s 1.7 L/min takes about 118 minutes, and the filter gained 0.5 mg. Since 200 L is 0.2 m³, the concentration is 0.5 mg ÷ 0.2 m³ = 2.5 mg/m³. That sits inside the method’s working range of 0.5 to 10 mg/m³ for a 200-L sample. The same filter gain on a 400-L sample, the method’s stated maximum volume, would be 1.25 mg/m³. The example shows why the sampled volume, and therefore the pump flow rate, belongs in every record of a dust result.
What limits apply: the mine dust example
A dust monitor reading means something to an inspector only when it is tied to a standard. For underground coal mines in the United States, 30 CFR 70.100 requires the operator to keep the average concentration of respirable dust in the mine atmosphere during each shift, for each miner in the active workings, at or below 2.0 mg/m³, and at or below 1.5 mg/m³ as of August 1, 2016. The regulation says the concentration is “as measured with an approved sampling device and expressed in terms of an equivalent concentration”. A separate paragraph sets lower limits of 1.0 mg/m³, and 0.5 mg/m³ as of August 1, 2016, for the intake airways within 200 feet outby the working faces of each section.
Two points follow. First, the limit is written in mass concentration, so a count-based optical reading cannot be compared with it directly. Second, the device matters: the regulatory wording ties the limit to an approved sampling device, not to any dust monitor. For a general workplace reference point, the NIOSH method page lists an OSHA limit of 5 mg/m³ and an ACGIH value of 3 mg/m³ for respirable particulates not otherwise regulated; those figures are specific to that dust category and are not a substitute for a substance-specific limit.
When the dust is a hazardous substance rather than nuisance dust, the choice of method changes. For fibre counting, for example, see the CASRAI guide to NIOSH Method 7400 for asbestos and other fibers, which uses a different, microscopy-based technique.
What is an indoor air quality monitor?
An indoor air quality monitor is a sensor device that tracks one or more air-quality measures in a room or building, and for dust it is usually a light-scattering particle sensor of the kind EPA classifies as an optical particle counter or a volume-scattering sensor. The EPA table notes, for example, that its listed Dylos counter is designed for indoor use and outputs counts, while the Shinyei volume-scattering sensor is designed as a stationary miniaturized monitor and outputs micrograms per cubic metre. So two devices sold as indoor monitors can report different quantities, which is why the output column in the data table is worth reading before you compare readings.
How to choose: criteria that the sources state
- Match the quantity to the question. A limit written as mg/m³ needs a mass result; a particle count answers a different question.
- Check whether a regulation names the device. 30 CFR 70.100 refers to an approved sampling device and an equivalent concentration, so a general-purpose monitor is not automatically acceptable for compliance sampling.
- Check the size range. EPA’s table gives size ranges for the optical counters it lists (above 0.5 µm and above 2.5 µm for one, 0.38 to 17 µm for another), and a size-selective cyclone has its own cut point, 4 µm median for respirable dust in Method 0600.
- Know the time resolution you need. A gravimetric result arrives after weighing; the CPDM display shows the previous 30 minutes and the shift average; EPA lists data rates from 1 second to 1 minute for the optical devices in its table.
Safe use and handling basics
The primary sources show where handling decides the result. For cyclone samplers, NIOSH records that over-sized particles in samples are known to be caused by inverting the cyclone assembly, that heavy dust loadings, fibres and water-saturated dusts interfere with the cyclone’s size selection, and that conductive samplers are recommended to limit particle charge effects. Pulsation in pump flow must stay within ±20% of the mean flow. For wearable monitors in mines, the monitor’s job is exposure information that workers and management can use to adjust dust controls or ventilation; the NIOSH page describes that use but not a hazard-control procedure, so for the control measures themselves see the CASRAI term hierarchy of controls and the chemical hygiene plan entry for laboratory settings.
Common mistakes
- Comparing a count with a mass limit. An optical counter’s particle counts do not convert directly to mg/m³; EPA describes the mass figures from counters and scattering sensors as estimates produced by conversion.
- Treating “dust” as one size. PM10, PM2.5 and respirable dust (4-µm median cut point) are different fractions with different instruments.
- Weighing on different balances. NIOSH Method 0600 tells the analyst to use the same balance before and after sample collection.
- Skipping blanks. The method calls for 2 to 10 field blanks per sample set.
- Ignoring bias. The method itself records an approximately +7% bias for non-diesel coal mine dust under the international respirable-dust definition, and says bias depends on the size distribution.
- Assuming larger dust is covered. Particles above 10 µm are outside EPA’s PM10 definition, so a PM10 reading does not describe them.
Frequently asked questions
How many microns is dust?
There is no single size. EPA groups airborne particles as PM10 (generally 10 micrometres and smaller) and PM2.5 (generally 2.5 micrometres and smaller), and notes that the average human hair is about 70 micrometres across. Particles larger than 10 micrometres, such as sand and large dust, are not regulated by EPA.
What are some techniques for measuring dust concentration?
Collecting dust on a filter and weighing it (NIOSH Method 0600), measuring mass continuously with a tapered element oscillating microbalance (the CPDM that NIOSH describes), and sizing and counting particles by light scattering (the optical particle counters in EPA’s Air Sensor Toolbox). Which one fits depends on whether you need mass or counts, and whether a regulation names the device.
What is an indoor air quality monitor?
A sensor device that tracks air-quality measures in a room or building. For particles, EPA lists optical counters and volume-scattering sensors; one listed counter is designed for indoor use and outputs particle counts in two size ranges.
How do I keep dust off a monitor screen?
That is a question about cleaning a display, which is outside what dust-monitoring standards and agency pages cover, so check the instructions supplied with the display itself.
Sources
- US EPA: Particulate Matter (PM) Basics, checked 2026-11-19
- US EPA: Evaluation of Emerging Air Sensor Performance, checked 2026-11-19
- NIOSH (CDC): Continuous personal dust monitor article, 3 February 2017, checked 2026-11-19
- NIOSH Manual of Analytical Methods, Method 0600: Particulates not otherwise regulated, respirable (Issue 3, 15 January 1998), checked 2026-11-19
- Legal Information Institute: 30 CFR 70.100, Respirable dust standards, checked 2026-11-19
Data: Dust monitor measurement principles compared
Compares the three measurement principles used by dust monitors (filter and balance, tapered element oscillating microbalance, optical light scattering) by what each reports, its size selection and when the result is available. Compiled from NIOSH Method 0600, the NIOSH CPDM article and the US EPA Air Sensor Toolbox sensor table.
Period covered: 2026.
| Principle | Quantity reported | Size selection or range | When the result is available | Example named in source |
|---|---|---|---|---|
| Filter and balance (gravimetric) | Mass of respirable dust on a filter, divided by air volume | Cyclone with 4 µm median cut point; 5 µm pore PVC filter; 1.7, 2.2 or 2.5 L/min by cyclone | After the filter is equilibrated and weighed (0.001 mg balance) | NIOSH Method 0600, Issue 3 |
| Tapered element oscillating microbalance (TEOM) | Respirable coal dust mass exposure | Respirable coal dust | Live display of previous 30 minutes and cumulative shift average | PDM3700 continuous personal dust monitor |
| Optical particle counter | Particle counts in size ranges, in some units converted to PM2.5 and PM10 mass estimates | Above 0.5 and above 2.5 µm (one unit); 0.38 to 17 µm (another) | Data as fast as 1 minute (counter) or 1 minute averages (OPC-N2) | Dylos DC1100; Alphasense OPC-N2 |
| Volume-scattering sensor | Estimated mass concentration from total scattered light | May include particles larger than 2.5 µm | Data as fast as 1 second | Shinyei sensor |
Sources
- NIOSH Manual of Analytical Methods, Method 0600: Particulates not otherwise regulated, respirable (accessed 2026-11-19)
- NIOSH: Continuous personal dust monitor (accessed 2026-11-19)
- US EPA: Evaluation of Emerging Air Sensor Performance (accessed 2026-11-19)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.