Ellman’s reagent is DTNB, 5,5′-dithiobis(2-nitrobenzoic acid), a yellow powder used to measure free thiol (sulfhydryl, -SH) groups in solution. A thiol cleaves the DTNB disulfide and releases the yellow anion TNB, which absorbs at 412 nm, so the colour of the mixture tracks how much reduced thiol the sample contained. It is a bench assay for research use, not a clinical test.
Short answer
DTNB and Ellman’s reagent are the same compound under two names. Mix it with a sample in a pH 8.0 buffer, wait for the colour to develop, read absorbance at 412 nm, and convert to thiol concentration with either a cysteine standard curve or the TNB extinction coefficient. Two things decide whether the number is trustworthy: which extinction coefficient you use, and whether every thiol in the sample was actually free and reachable. Both are covered below.
What is Ellman’s reagent (DTNB)?
Thermo Fisher’s Pierce product page describes Ellman’s reagent as DTNB, CAS 69-78-3, molecular weight 396.35, supplied as a yellow crystalline powder. The method is named for G. L. Ellman, whose 1959 paper “Tissue sulfhydryl groups” (Archives of Biochemistry and Biophysics 82(1):70-77) introduced it. The page lists the reagent for research use only.
The names you will meet in methods sections are interchangeable for this purpose: Ellman’s reagent, DTNB, and “the Ellman assay” or “Ellman’s test” for the procedure. The coloured product is called TNB (2-nitro-5-thiobenzoate), and the anion form is often written TNB2-. Searching for DTNB and searching for Ellman’s reagent lead to the same chemistry.
How the reaction works
DTNB contains a disulfide bond. The thiolate form of a free thiol attacks that disulfide, which releases one TNB anion per thiol and leaves a mixed disulfide. Because the thiolate (not the protonated thiol) is the reactive species, pH matters: Thermo Fisher notes that reaction rate depends on pH, on the thiol’s own pKa, and on steric and electrostatic effects around the thiol. TNB is yellow with a peak at 412 nm, and the vendor states that TNB absorbance is not affected across pH 7.6 to 8.6, which is why a pH 8.0 buffer is the standard choice.
The same BMG LABTECH application note (AN327) that describes a plate-reader version of the assay also notes an isosbestic point near 356 nm, and reports that spectrophotometric assays of this type are less sensitive than electrochemical or fluorimetric thiol methods. If your thiol is present at very low concentration, that limit matters before you choose DTNB at all.
Buffer, pH and reagent handling
Both the Thermo Fisher instructions and BMG LABTECH’s application note use the same reaction buffer: 0.1 M sodium phosphate, pH 8.0, with 1 mM EDTA. Thermo Fisher also accepts Tris at pH 8.0. EDTA is there because metal ions can oxidise thiols; the vendor advises 1-5 mM EDTA to chelate them, since oxidised thiols no longer react and the sample reads low.
BMG’s example prepares the Ellman solution at 4 mg DTNB per mL of reaction buffer. Thermo Fisher’s storage advice is to keep the dry reagent at 4 °C protected from moisture. The data table below lists how long a made-up working solution lasts: about one week in phosphate or Tris at pH 8.0, but only one to two days in phosphate-buffered saline (PBS) at pH 7.2. The Thermo Fisher page states it has not been tested with DMSO, so do not assume DMSO as a co-solvent is validated by the vendor.
Reduced reagents are a related trap. Dithiothreitol and 2-mercaptoethanol are themselves thiols and will react with DTNB, so a sample that still contains them cannot be read directly. For background on those reagents, see our pages on DTT and b-mercaptoethanol.
Calculating thiol: standard curve or extinction coefficient
Thermo Fisher describes two ways to turn absorbance into concentration: compare the sample against a cysteine standard curve, or apply the Beer-Lambert law with the TNB extinction coefficient. The standard-curve route absorbs differences in cuvette or plate path length and instrument response, which is why a plate-reader protocol uses it.
Standard-curve route (BMG LABTECH AN327)
The BMG note uses cysteine hydrochloride monohydrate (molecular weight 175.6) as the standard: a 1.6 mM stock, diluted two-fold down to 0.1 mM. Its example mixes 50 µL of Ellman solution with 2.5 mL of reaction buffer, adds 250 µL of standard or unknown, incubates 15 minutes at room temperature, transfers 200 µL to a 96-well plate, and reads the endpoint at 412 nm. Plot absorbance against standard concentration and read the unknown off the line.
Extinction-coefficient route
Beer-Lambert gives concentration = absorbance / (ε × path length). The catch is which ε. Thermo Fisher gives 14,150 M-1cm-1 for TNB at 412 nm and notes that Ellman’s original value, 13,600, is still often cited. BMG’s note uses 13,600. Which one a lab reports changes the answer.
Worked arithmetic (illustrative numbers, not a measurement): a corrected absorbance of 0.500 in a 1 cm cuvette gives 0.500 / 14,150 = 3.53 × 10-5 M (35.3 µM) with the 14,150 value, and 0.500 / 13,600 = 3.68 × 10-5 M (36.8 µM) with Ellman’s original value. That is roughly a 4% difference from the coefficient alone, before any dilution factor. Whichever value you choose, state it in the methods section so another lab can reproduce the calculation. The comparison table of values and conditions is in the data table below.
Limits and common misunderstandings
It measures free, reduced, accessible thiol only
Disulfide-bonded cysteines do not react until they are reduced, and oxidised samples read low (Thermo Fisher). A low reading therefore does not mean the protein has few cysteines. It may mean they are oxidised.
Buried thiols can react incompletely
Riener and colleagues (Analytical and Bioanalytical Chemistry 2002; 373(4-5):266-276, PubMed 12110978) reported that many protein sulfhydryl groups react incompletely with Ellman’s reagent even with a prolonged assay. They described using cystamine as a mediator, or the alternative reagent 4,4′-dithiodipyridine (DTDP), and gave a detection limit of 0.2 nmol/mL and a determination limit of 0.6 nmol/mL for a 5-minute assay. If a published cysteine count disagrees with your Ellman result, accessibility is a likelier explanation than a calculation error.
“Same buffer” is not a safe assumption
A coefficient measured in dilute salt is not automatically valid in a different medium, and the pH range where TNB absorbance is stable is 7.6 to 8.6 according to the vendor. Match the buffer you calibrate in to the one you measure in.
It is not specific to one protein or to a clinical sample
DTNB reacts with thiols broadly, so a cross-linking or conjugation experiment that cares about a specific residue needs another readout. For how thiol-reactive chemistry is used downstream, see our overview of cross-linkers.
Alternatives and neighbouring methods
The sources read for this page name three directions when plain DTNB falls short. Riener and colleagues describe cystamine as a mediator that helps otherwise incomplete protein reactions, and 4,4′-dithiodipyridine (DTDP) as an alternative chromogenic reagent. BMG LABTECH’s note points out that electrochemical and fluorimetric methods are more sensitive than spectrophotometric ones. None of these removes the need for a reduced, accessible sample; they change the detection chemistry, not the sample requirement.
What the two documented protocols have in common
The Thermo Fisher instructions and the BMG LABTECH application note differ in format (a vendor product sheet versus a plate-reader note) but agree on the parts that most affect results: a pH 8.0 phosphate reaction buffer containing EDTA, absorbance read at 412 nm, and quantification against cysteine or an extinction coefficient. They differ in the coefficient they print (14,150 versus 13,600), which is the single most visible inconsistency a reader will find when moving between lab protocols. When a methods section gives neither the buffer nor the coefficient, the reported thiol concentration cannot be rechecked.
What to record in a methods section
As a reporting checklist drawn from the parameters above: the reaction buffer and pH, EDTA concentration, how the Ellman solution was made and how old it was, the standard (cysteine hydrochloride monohydrate, stated range) or the extinction coefficient and its source, incubation time and temperature, wavelength, and whether the sample was reduced beforehand and the reductant removed. These are editorial recommendations built from the sourced values on this page, not requirements issued by Thermo Fisher or BMG LABTECH.
Which value should I report? A short decision aid
- If you run a cysteine standard curve on the same plate or in the same cuvette, the extinction coefficient is not used in the calculation; report the standard and its range.
- If you calculate directly, state the coefficient and its source (14,150 per Thermo Fisher, 13,600 per Ellman 1959 as cited by BMG LABTECH) and the buffer.
- If the sample contains DTT or 2-mercaptoethanol, they must be removed or accounted for, because they are thiols too.
- If the result is far below the expected cysteine content, test whether the protein was reduced and whether the thiols are exposed before suspecting the assay.
Sources
- Thermo Fisher Scientific, Pierce Ellman’s Reagent (catalog 22582) product page and instructions, https://www.thermofisher.com/order/catalog/product/22582, read 2026-11-14.
- BMG LABTECH, Application Note 327, https://www.bmglabtech.com/fileadmin/06_Support/Download_Documents/App_Notes/AN327.pdf, read 2026-11-14.
- Ellman GL. Tissue sulfhydryl groups. Arch Biochem Biophys. 1959;82(1):70-77. PubMed 13650640, https://pubmed.ncbi.nlm.nih.gov/13650640/, read 2026-11-14 (citation record).
- Riener CK et al. Anal Bioanal Chem. 2002;373(4-5):266-276. PubMed 12110978, https://pubmed.ncbi.nlm.nih.gov/12110978/, abstract read 2026-11-14.
Data: Ellman's assay reference values: extinction coefficients, buffer, stability and sensitivity
Compiled from the Thermo Fisher Pierce Ellman's Reagent product page, BMG LABTECH Application Note 327 and the Riener et al. 2002 abstract. Each row gives one working parameter of the DTNB thiol assay, the condition it applies to and the source it comes from, so values that differ between sources can be compared in one place.
| Parameter | Value | Unit | Condition | Source |
|---|---|---|---|---|
| TNB extinction coefficient at 412 nm | 14150 | M-1 cm-1 | Value given by the vendor for TNB at 412 nm | Thermo Fisher |
| TNB extinction coefficient at 412 nm (Ellman's original) | 13600 | M-1 cm-1 | pH 8.0; still often cited | Thermo Fisher; BMG LABTECH AN327 |
| Detection wavelength | 412 | nm | TNB absorbance maximum used for endpoint reading | BMG LABTECH AN327; Thermo Fisher |
| pH range with unaffected TNB absorbance | 7.6 to 8.6 | pH | Standard reaction pH is 8.0 | Thermo Fisher |
| Reaction buffer | 0.1 M sodium phosphate, 1 mM EDTA | pH 8.0 | Tris pH 8.0 also acceptable per Thermo Fisher | Thermo Fisher; BMG LABTECH AN327 |
| Working solution stability in phosphate or Tris | About 1 week | pH 8.0 | Thermo Fisher | |
| Working solution stability in PBS | 1 to 2 days | pH 7.2 | Thermo Fisher | |
| Cysteine standard range in plate protocol | 1.6 mM stock, two-fold dilutions to 0.1 mM | Cysteine hydrochloride monohydrate, MW 175.6 | BMG LABTECH AN327 | |
| Detection limit with cystamine mediator | 0.2 | nmol/mL | 5-minute assay for protein sulfhydryls | Riener et al. 2002 |
| Determination limit with cystamine mediator | 0.6 | nmol/mL | 5-minute assay for protein sulfhydryls | Riener et al. 2002 |
Sources
- Thermo Fisher Scientific, Pierce Ellman's Reagent (22582) (accessed 2026-11-14)
- BMG LABTECH Application Note 327 (accessed 2026-11-14)
- Riener et al. 2002, Anal Bioanal Chem 373:266-276 (abstract) (accessed 2026-11-14)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.