EPA Method 524.2 is a U.S. EPA laboratory method for measuring purgeable volatile organic compounds (VOCs) in surface water, ground water and drinking water at any stage of treatment. The sample is purged with an inert gas, the VOCs are caught on a sorbent trap, and they are then released into a capillary gas chromatograph coupled to a mass spectrometer (GC/MS) for identification and measurement. The current text is Revision 4.1 (1995), and the method also covers the four trihalomethane disinfection by-products.
What EPA Method 524.2 covers
The method’s full title is “Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry.” It was written at EPA’s National Exposure Research Laboratory (Office of Research and Development, Cincinnati). Revision 1.0 dates from 1983, followed by Revisions 2.0 (1986), 3.0 (1989), 4.0 (1992) and 4.1 (1995, edited by J.W. Munch). The Revision 4.0 update added compounds such as acetone, acrylonitrile and methyl tert-butyl ether to the analyte list, which is marked with asterisks in the method.
Section 1.1 describes it as a general-purpose method for the identification and simultaneous measurement of purgeable VOCs. Section 1.2 states that method detection limits (MDLs) depend on the compound, the instrument and especially the matrix, and run from roughly 0.02 to 1.6 µg/L. The working range is about 0.02 to 200 µg/L with a wide-bore thick-film capillary column and about 0.02 to 20 µg/L with a narrow-bore thin-film column.
What are volatile organic compounds in water?
In this method, the target VOCs are compounds with sufficiently high volatility and low water solubility to be removed from a water sample by purge-and-trap (Section 1.1). Benzene, vinyl chloride, trichloroethene and chloroform are among the listed analytes. The same section notes the limits of that definition. Compounds with more than 2% solubility or boiling points above 200 °C purge less efficiently and are more susceptible to matrix effects (Section 1.3), and some structural isomers (two of the three xylenes, two of the three dichlorobenzenes) can co-elute and must then be reported as isomeric pairs.
How the purge-and-trap GC/MS workflow runs
Section 2.1 summarises the workflow in four moves. An inert gas is bubbled through the aqueous sample so the VOCs and added surrogates leave the water. They are collected in a tube of sorbent materials. When purging ends, the tube is heated and backflushed with helium to desorb the compounds into a capillary GC column connected to a mass spectrometer. Compounds are identified by matching measured mass spectra and retention times to reference spectra and retention times, and quantified against an internal standard.
The method fixes the internal standard and surrogates in Section 7.5.1: fluorobenzene is the internal standard, and 1,2-dichlorobenzene-d4 and 4-bromofluorobenzene (BFB) are the surrogates, prepared in methanol. BFB also appears in the mass spectrometer performance check in Section 6.3: the spectrometer must meet the criteria in the method’s Table 3 when 25 ng or less of BFB is introduced.
Consumables and hardware the method specifies
For laboratories planning supplies, the method names its consumable and hardware requirements in plain terms. It does not name brands, and the page does not suggest any.
- Sample vials (Section 6.1): 40 mL to 120 mL screw-cap vials with a Teflon-faced silicone septum. Before use, vials and septa are washed with detergent, rinsed with tap and distilled water, air dried at room temperature, held in a 105 °C oven for 1 hour, and cooled in an area known to be free of organics. For a general look at this consumable, see what GC vials are.
- Purging device (Section 6.2.1): all glass, designed for 25 mL samples with a water column at least 5 cm deep. A smaller 5 mL device is recommended if the GC/MS has enough sensitivity to reach the required MDLs. A glass frit at the base of the sample chamber must deliver the purge gas as fine bubbles (diameter under 3 mm at the origin), although needle spargers may be used with the gas introduced about 5 mm from the base of the water column.
- Trap (Section 6.2.2): at least 25 cm long with an inside diameter of at least 0.105 in. From the inlet, 1.0 cm of methyl silicone coated packing, then one third each of 2,6-diphenylene oxide polymer, silica gel and coconut charcoal. The charcoal may be left out, with the polymer increased to two thirds of the trap, if dichlorodifluoromethane does not need to be determined. A new trap is conditioned overnight at 180 °C by backflushing with at least 20 mL/min of inert gas and conditioned for 10 minutes at 180 °C before daily use.
- Alternative sorbents: acceptable only if the data meet every quality-control criterion in Section 9 and the purge and desorption conditions in Section 11 (purge time, purge gas flow rate, desorption time) are not changed.
- Dechlorinating agents and acid: ascorbic acid (ACS reagent grade, granular, Section 7.2.5), sodium thiosulfate and 1:1 hydrochloric acid, used as described in the preservation table below.
Instrument cost and configuration questions sit outside the method; for those see the gas chromatograph cost guide.
Related EPA drinking water methods compared
Method 524.2 sits alongside several other EPA methods approved for drinking water compliance monitoring of organic contaminants. EPA’s Office of Water table of approved organic methods (EPA 815-B-24-003, January 2024) lists them by contaminant. The CFR is the legal reference and takes precedence over that table. It also states that analysis shall be conducted according to the listed methods “or their equivalent as determined by EPA”, and that the methods and monitoring requirements are specified in 40 CFR 141.24, with additional methods in Appendix A to Subpart C of Part 141. For the first contaminant listed, 1,1,1-trichloroethane, the table gives these methods:
| Method | Title in the EPA table | Date | EPA publication number |
|---|---|---|---|
| 502.2, Rev. 2.1 | In Methods for the Determination of Organic Compounds in Drinking Water, Supplement III | August 1995 | EPA/600/R-95-131 |
| 524.2, Rev. 4.1 | In Methods for the Determination of Organic Compounds in Drinking Water, Supplement III | August 1995 | EPA/600/R-95-131 |
| 524.3 | Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry | June 2009 | EPA 815-B-09-009 |
| 524.4 | Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry using Nitrogen Purge Gas | May 2013 | EPA 815-R-13-002 |
| 551.1, Rev. 1.0 | In Methods for the Determination of Organic Compounds in Drinking Water, Supplement III | August 1995 | EPA/600/R-95-131 |
The practical reading is limited to what the table shows. Methods 524.3 and 524.4 carry the same base title as 524.2 and are later EPA publications, and 524.4 is the one whose title names nitrogen as the purge gas, while Method 524.2 itself describes helium for backflushing the trap and an inert purge gas. Which method a laboratory may use for a given contaminant is decided by 40 CFR 141 and any state or accreditation requirements, not by this page.
Sample collection and preservation under Method 524.2
Most of the method’s day-to-day failure points happen before the sample reaches the instrument. Section 8 covers them, and the data table below the article sets out the rules by situation, with the section for each. The points that tend to be missed:
- Samples are collected in duplicate, and the vials are filled to overflowing with no air bubbles passing through the sample or trapped when the vial is sealed (Sections 8.1.1 and 8.1.4).
- The dechlorinating agent goes into the empty vial before filling, and it must not be mixed with the HCl in the sample bottle before sampling (Section 8.2.1 note).
- When sampling from a tap, the method says to open the tap and flush until water temperature has stabilised (usually about 10 minutes), then adjust the flow to about 500 mL/min (Section 8.1.2).
- Field reagent blanks (FRBs) in duplicate travel with each sample set to the site and back, and use the same batch of ascorbic acid and HCl as the field samples (Section 8.3).
- Samples are stored at 4 °C or below, in an area free of organic solvent vapours and direct or intense light, and analysed within 14 days. Samples not analysed in that period must be discarded and replaced (Section 8.2).
Quality-control batch checklist
Section 9 sets the quality-control (QC) programme. Read as a checklist for each batch, it asks for the following.
- Low background shown first (9.2): before any sample is analysed, a laboratory reagent blank (LRB) must be reasonably free of contamination that would prevent determination of any analyte of concern. Sources named are glassware, purge gas, sorbents, reagent water and equipment.
- Initial accuracy and precision (9.3): four to seven replicates of a laboratory fortified blank (LFB) at 2 to 5 µg/L, with a quality control sample (QCS) from a source different from the calibration standards recommended.
- MDL per analyte (9.3.4): determined from at least 7 LFBs prepared at a low concentration.
- An LRB with each batch within a work shift (9.5) and an LFB with each batch, or one per 20 samples if the shift batch is larger (9.6). The LFB accuracy criterion in that section is ±30%, and the problem must be solved before more samples are analysed if it fails.
- Control charts (9.3.5, 9.6, 9.8): precision and accuracy of analytes and surrogates plotted over time, with surrogate recoveries called an especially valuable record because surrogates are in every sample.
- Internal-standard and surrogate areas (9.4): monitored in all samples, calibration checks and blanks, with a gradual drift of more than 50% in any area treated as a loss of sensitivity that must be corrected.
- Field reagent blanks (9.7): if a sample is contaminated with an analyte, an FRB is analysed to show whether the contamination came from sampling, storage or transport.
- External QCS at least quarterly (9.9).
Contamination sources the method names
- Laboratory volatile materials, impurities in the purge gas and impurities in the sorbent trap (Section 4.1).
- Teflon tubing, Teflon thread sealants and flow controllers with rubber components in the purging device, which out-gas organics that concentrate in the trap (Section 4.1).
- Carryover when a low-concentration sample follows a high-concentration one. The method suggests rinsing the purging apparatus and syringes between samples with two portions of reagent water, and analysing one or more LRBs after a high-concentration sample (Section 4.2).
- Methylene chloride, which needs special precautions: isolate the analytical and storage area from atmospheric sources, use stainless steel or copper for GC carrier gas and purge gas lines because methylene chloride permeates Teflon tubing, and clean laboratory workers’ clothing often (Section 4.3).
- Traces of ketones, methylene chloride and other solvents in even the highest-purity methanol, to be assessed before standards are made up in it (Section 4.4).
Where Method 524.2 matters in a laboratory
Drinking water, ground water and surface water laboratories use purge-and-trap GC/MS methods when they need one run to cover many VOCs and the four trihalomethanes. For labs evaluating how 524.2 relates to other VOC workflows, the EPA 8270, 8260 and 624 comparison covers the wastewater and solid-waste methods, and volatile organic compound detection gives the wider picture of analytical approaches. Because the method specifies vials, traps, reagents and blank handling in such detail, most procurement and quality records for a 524.2 workflow trace back to the sections cited above rather than to a vendor’s application note.
Common misunderstandings
- “A blank can be subtracted from the result.” It cannot. Section 4.1 states that subtracting blank values from sample results is not permitted. If interfering peaks appear in laboratory reagent blanks, the analyst changes the purge gas source and regenerates the molecular sieve purge gas filter.
- “Ascorbic acid and thiosulfate are interchangeable.” Not entirely. Acidification may be omitted for trihalomethane-only analysis if sodium thiosulfate is used, but that exception does not apply when ascorbic acid is the dechlorinating agent (Section 8.2.2).
- “524.2 is the newest EPA method for these compounds.” The EPA table lists 524.3 (2009) and 524.4 (2013) as later publications on the same subject, as shown above.
- “Every compound purges well.” Compounds with higher water solubility or boiling points above 200 °C purge less efficiently and may produce lower-quality data (Sections 1.2 and 1.3).
Frequently asked questions
What is EPA Method 524.2 used for?
It measures purgeable VOCs, including the four trihalomethane disinfection by-products, in surface water, ground water and drinking water at any stage of treatment, using purge-and-trap with capillary GC/MS (Section 1.1, Revision 4.1, 1995).
What is EPA 533?
EPA Method 533 is a different EPA drinking water method, for PFAS rather than VOCs. It does not use purge-and-trap GC/MS. See the CASRAI guide to EPA Method 533 for its analytes and quality controls.
Does the EPA regulate chemicals?
The EPA sets requirements across many programmes. For the methods on this page, the relevant regulation is the federal drinking water rule in 40 CFR 141.24, which EPA’s January 2024 table cites as the place where the approved methods and monitoring requirements are specified. The CFR takes precedence over the table. 40 CFR 141.24, titled “Organic chemicals, sampling and analytical requirements,” identifies Method 524.2 as published in Methods for the Determination of Organic Compounds in Drinking Water, Supplement III (EPA/600/R-95-131, August 1995).
What are volatile organic compounds in water?
For Method 524.2, they are organic compounds with high volatility and low water solubility that can be removed from a water sample by purge-and-trap. Examples named in the method include benzene, vinyl chloride, trichloroethene and chloroform.
How long can samples be stored before 524.2 analysis?
Samples are chilled to about 4 °C, stored at 4 °C or below and analysed within 14 days of collection; otherwise they are discarded and replaced. Samples that foam on acid addition and are left unacidified must be analysed within 24 hours for any compound other than trihalomethanes (Sections 8.2.3 and 8.2.2).
Can blank values be subtracted from sample results?
No. Section 4.1 of the method says subtracting blank values from sample results is not permitted.
Sources
- U.S. EPA, Method 524.2, Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry, Revision 4.1 (1995), epa.gov (read 2026-11-20).
- U.S. EPA Office of Water, Analytical Methods Approved for Drinking Water Compliance Monitoring of Organic Contaminants, EPA 815-B-24-003, January 2024, epa.gov (read 2026-11-20).
- U.S. EPA, Approved Drinking Water Analytical Methods, page listing the organic contaminants table 815-B-24-003, epa.gov (read 2026-11-20).
- 40 CFR 141.24, Organic chemicals, sampling and analytical requirements, Legal Information Institute, Cornell Law School (read 2026-11-20).
Data: EPA Method 524.2 sample preservation rules by situation
Compiled from Section 8 of EPA Method 524.2 Revision 4.1 (1995): the dechlorination, acidification, storage and holding-time instruction that applies in each situation the method describes, with the section where each rule appears.
Period covered: 1995. Coverage: United States.
| Situation | Method instruction | Method 524.2 section |
|---|---|---|
| General case: residual chlorine suspected, e.g. finished drinking water | Collect in duplicate. Add about 25 mg ascorbic acid per 40 mL to the bottle before filling. Then, after dechlorination, adjust pH to below 2 with two drops of 1:1 HCl per 40 mL. | 8.1.1, 8.2.1 |
| Gases such as vinyl chloride, or Table 7 analytes, not to be determined | Sodium thiosulfate is recommended instead: 3 mg per 40 mL of water sample. | 8.1.1 |
| Residual chlorine likely above 5 mg/L | Determine the chlorine amount (DPD test kits are noted); add a further 25 mg ascorbic acid or 3 mg sodium thiosulfate for each 5 mg/L of residual chlorine. | 8.1.1 note |
| Trihalomethane-only analysis, thiosulfate used | Acidification may be omitted. | 8.2.2 |
| Trihalomethane-only analysis, ascorbic acid used | The acidification exception does not apply; acidify as usual. | 8.2.2 |
| Sample foams vigorously when HCl is added | Discard that sample. Collect a duplicate set without acid, flag it as not acidified, store at 4 C or below, and analyse within 24 hours of collection for any compound other than trihalomethanes. | 8.2.3 |
| All samples: temperature, storage and holding time | Chill to about 4 C at collection; store at 4 C or below, free of solvent vapours and direct or intense light; analyse within 14 days of collection or discard and replace. | 8.2.4, 8.2 |