EPA Method 533 is a drinking-water method that measures 25 selected per- and polyfluoroalkyl substances (PFAS) by isotope dilution, anion exchange solid phase extraction (SPE) and liquid chromatography with tandem mass spectrometry (LC-MS/MS). EPA published it in late 2019 as document EPA 815-B-19-020. It focuses on short-chain PFAS and is used alongside EPA Method 537.1, which measures 18 PFAS.
What is EPA Method 533?
The full title is “Determination of Per- and Polyfluoroalkyl Substances in Drinking Water by Isotope Dilution Anion Exchange Solid Phase Extraction and Liquid Chromatography/Tandem Mass Spectrometry.” EPA’s Office of Water issued it as a validated drinking-water method, and the method text requires tandem mass spectrometry in multiple reaction monitoring (MRM) mode. The method document reports accuracy and precision data in both reagent water and drinking water.
EPA’s own summary describes Method 533 as focused on “short chain” PFAS, meaning those with carbon chain lengths of 4 to 12. It complements Method 537.1 (published November 2018) and can test for 11 additional PFAS. Together the two methods measure 29 unique PFAS in drinking water. EPA’s drinking-water laboratory methods page (updated April 24, 2026) states that both methods are approved for monitoring under the fifth Unregulated Contaminant Monitoring Rule (UCMR 5) and for the PFAS National Primary Drinking Water Regulation, and that techniques other than Methods 537.1 and 533 have not been evaluated by EPA and are not approved for UCMR 5 or the PFAS NPDWR.
How many analytes are in EPA Method 533?
Method 533 has 25 target analytes. The data table below this page lists each one with its abbreviation and CAS registry number, and marks which of them are also in Method 537.1. The 25 break down as follows, using EPA’s December 2019 comparison table (EPA 815-B-19-021):
- 14 analytes shared with Method 537.1: 11Cl-PF3OUdS, 9Cl-PF3ONS, ADONA, HFPO-DA, PFBS, PFDA, PFDoA, PFHpA, PFHxA, PFHxS, PFNA, PFOA, PFOS and PFUnA.
- 11 analytes only in Method 533: 4:2FTS, 6:2FTS, 8:2FTS, NFDHA, PFBA, PFEESA, PFHpS, PFMBA, PFMPA, PFPeA and PFPeS.
- 4 analytes only in Method 537.1: NEtFOSAA, NMeFOSAA, PFTA and PFTrDA. A lab that runs only Method 533 does not report these.
The arithmetic matches EPA’s statements: 14 + 11 = 25 analytes for Method 533, 14 + 4 = 18 for Method 537.1, and 14 + 11 + 4 = 29 unique PFAS across both. The method measures every form of an analyte (the acid or its ammonium, sodium or potassium salt) as the anion, and it sums branched and linear isomers.
The UCMR 5 page on EPA’s site (read November 20, 2026) gives the same picture from the monitoring side: PFAS were analysed by Methods 533 and 537.1, 29 PFAS fall within the scope of the two methods, and UCMR 5 monitoring ran from January 2023 to December 2025 with PFAS minimum reporting levels between 0.002 and 0.02 µg/L (2 to 20 ng/L).
How the method works
The workflow in Section 2 of the method is a fixed sequence, and each step explains a later QC check:
- Fortify. The sample is fortified with isotopically labelled analogues of the target analytes (the isotope dilution analogues).
- Extract. The sample passes through an SPE cartridge containing a polystyrene divinylbenzene sorbent with a positively charged diamino ligand, which retains the PFAS anions.
- Rinse and elute. The cartridge is rinsed with aqueous ammonium acetate, then methanol. The analytes are eluted with methanol containing ammonium hydroxide.
- Concentrate. The eluate is taken to dryness under nitrogen in a heated water bath (the method text gives 55 to 60 °C in the procedure and no higher than 60 °C in the equipment section).
- Reconstitute. The extract is adjusted to 1.0 mL with 20% water in methanol, and three isotopically labelled isotope performance standards are added.
- Analyse. The extract is run by LC-MS/MS in MRM mode. Each analyte is quantified by isotope dilution, and the analogue recoveries are calculated against the performance standards for QC.
For background on the detection step, see the CASRAI guides on LC-MS and mass spectrometry types.
EPA Method 533 sampling, preservation and holding times
Section 8 of the method sets the collection and shipping rules. These are the points most often missed when a sampling kit is assembled outside the lab:
- Containers: single-use polypropylene bottles with polypropylene caps, or polyethylene bottles with polypropylene caps.
- Preservative: ammonium acetate added as a solid to 1 g/L, which sequesters free chlorine as chloramine. The method itself flags the ammonium acetate as a possible source of interference, so it is monitored with laboratory reagent blanks.
- Collection: nitrile gloves are worn and the tap is flushed until the water temperature stabilises. Samples do not need to be collected headspace-free.
- Field reagent blank: every sample set (samples from the same site at the same time) includes a field reagent blank. The reagent water used for it is analysed beforehand and must be below one-third of the minimum reporting level (MRL).
- Shipping: samples go on ice. They are valid if any ice remains in the cooler on arrival, or if the bottles arrive within 2 days of collection at below 10 °C.
- Storage and holding times: at or below 6 °C and never frozen until extraction. Samples must be extracted within 28 days of collection, and extracts, which are stored at room temperature, must be analysed within 28 days after extraction.
What a laboratory may and may not change
Section 1.3 (“Method Flexibility”) is the section vendor application notes tend to skip, and it decides whether a laboratory’s result is a Method 533 result. The method states that:
- the laboratory may choose its LC column, LC conditions and MS conditions;
- at a minimum the isotope dilution standards and isotope performance standards named in the method must be used, if available;
- the aqueous sample volume may be anywhere in the range of 100 to 250 mL; method development used 250 mL with 500 mg of sorbent, the sorbent-to-sample ratio may not be decreased, and a 100 mL sample needs at least 200 mg of sorbent;
- preservation, QC requirements and the extraction procedure are not open to change;
- any modification requires an Initial Demonstration of Capability (IDC), met QC criteria and verified performance in a representative matrix.
Quality control requirements in Method 533
Section 9 combines a one-time demonstration with ongoing checks. The IDC covers low system background, precision, accuracy, MRL confirmation and calibration verification. Ongoing QC then includes laboratory reagent blanks (LRB), continuing calibration checks (CCC), laboratory fortified blanks (LFB), isotope performance standard peak areas, isotope dilution analogue recoveries, laboratory fortified sample matrix samples (LFSM) and their duplicates or field duplicates.
One criterion is stated plainly in Section 9.2.5: the percent recovery of each isotope dilution analogue must fall within 50 to 200%. If an analogue fails, the method directs the analyst to evaluate the area of the isotope performance standard it is referenced to and the analogue’s recovery before continuing. The method also defines the lowest concentration minimum reporting level (LCMRL) as the concentration at which future recovery is predicted to fall between 50 and 150% with 99% confidence.
Method 533 compared with related EPA methods
Searches for “EPA method 533” often surface other numbered EPA methods. They are not variants of 533. They answer different questions, as the table below shows. All facts in it come from the EPA pages listed under Sources.
| Method | What it is | Technique and matrix |
|---|---|---|
| Method 533 (2019) | 25 PFAS in drinking water | Isotope dilution, anion exchange SPE, LC-MS/MS |
| Method 537.1 Ver. 2.0 (2020) | 18 PFAS in drinking water, four of which Method 533 does not measure | SPE, LC-MS/MS; drinking water |
| Method 1633 / 1633A | 40 PFAS compounds | Wastewater, surface water, groundwater, soil, biosolids, sediment, landfill leachate and fish tissue |
| SW-846 Method 5030C (Rev. 3, May 2003) | Purge-and-trap sample preparation for volatile organic compounds | Aqueous samples and high-concentration soil or waste extracts prepared by Method 5035 |
| Method 552.2 (Revision 1.0, 1995) | Haloacetic acids and dalapon in drinking water | Liquid-liquid extraction, derivatization, gas chromatography with electron capture detection |
What is EPA Method 5030?
EPA Method 5030C is a purge-and-trap procedure in the SW-846 compendium (Test Methods for Evaluating Solid Waste). It prepares aqueous samples, and high-concentration soil and waste extracts prepared by Method 5035, for analysis of volatile organic compounds. It is a sample-preparation method for volatiles and has no role in PFAS testing, which concerns non-volatile ionic compounds extracted by SPE. The current EPA page was last updated March 2, 2026, and the PDF is dated May 2003 (Revision 3).
What is the EPA 552.2 method used for?
EPA Method 552.2 determines haloacetic acids and dalapon in drinking water. The title in EPA’s HERO reference record (EPA/600/R-95/131, 1995) names the technique: liquid-liquid extraction, derivatization and gas chromatography with electron capture detection. It is not a PFAS method.
What is the EPA 537.1 method used for?
Method 537.1 (Version 2.0, 2020) is the other EPA drinking-water PFAS method. It measures 18 PFAS and shares 14 analytes with Method 533. Use it when a program needs NEtFOSAA, NMeFOSAA, PFTA or PFTrDA, which Method 533 does not include. CASRAI’s Method 537.1 guide covers its workflow and procurement side.
What about Method 1633 and Method 8327?
Method 1633 is a Clean Water Act method, not a drinking-water method. EPA’s CWA PFAS methods page (last updated February 25, 2026) describes Method 1633A as a method to test for 40 PFAS compounds in matrices including wastewater, groundwater, soil, biosolids and fish tissue. It states that 1633A is not nationally required for CWA compliance monitoring until EPA promulgates it through rulemaking, and that EPA recommends it now for individual permits. The same page lists SW-846 Method 8327 among methods for PFAS in non-potable water. A 2025 peer-reviewed study in Environmental Monitoring and Assessment (Eldridge et al.) compared Method 1633 with other soil extractions and commercial laboratory results for 40 PFAS in various soils, which shows the method’s solid-matrix use. Neither method is approved for drinking-water compliance: EPA’s drinking-water page names only Methods 533 and 537.1 for UCMR 5 and the PFAS NPDWR. For the wastewater regulatory framework see the guide to 40 CFR Part 136.
Where Method 533 matters for labs
- Scope decisions. A program that needs only the 25 Method 533 analytes can run one method. A program that also needs the four 537.1-only analytes must run 537.1 as well, because EPA presents the two as complementary.
- Method claims. A report that cites Method 533 while changing preservation, extraction or QC is outside Section 1.3. Only the listed flexibilities are permitted without a new demonstration.
- Background control. Because PFAS are measured at nanogram-per-litre reporting levels, the method’s controls (polypropylene or polyethylene containers, reagent blanks below one-third of the MRL, field reagent blanks) are part of the measurement, not extras. Background on the two best-known analytes is in PFOS and PFOA.
Common misunderstandings
- “Method 533 replaces 537.1.” It does not. EPA describes the two as complementary, and four analytes appear only in 537.1.
- “Any LC-MS/MS PFAS method is acceptable for drinking-water compliance.” EPA’s page states that other techniques have not been evaluated and are not approved for UCMR 5 or the PFAS NPDWR.
- “The method fixes the instrument.” It does not. Column, LC conditions and MS conditions are the laboratory’s choice, while the labelled standards, preservation and extraction are fixed.
- “Methods 5030, 552.2 and 533 are the same family.” They are unrelated: volatile organics sample preparation, haloacetic acids by GC, and PFAS by LC-MS/MS.
Frequently asked questions
How many analytes are in EPA Method 533?
Twenty-five PFAS. Fourteen are shared with Method 537.1 and eleven (4:2FTS, 6:2FTS, 8:2FTS, NFDHA, PFBA, PFEESA, PFHpS, PFMBA, PFMPA, PFPeA, PFPeS) appear only in Method 533, per EPA 815-B-19-021 and EPA 815-B-19-020.
What is EPA Method 5030?
A purge-and-trap sample-preparation method for volatile organic compounds in aqueous samples and in high-concentration soil or waste extracts, from the SW-846 compendium. It is unrelated to Method 533.
What is the EPA 552.2 method used for?
Haloacetic acids and dalapon in drinking water, using liquid-liquid extraction, derivatization and GC with electron capture detection (Revision 1.0, 1995).
What is the EPA 537.1 method used for?
Measuring 18 PFAS in drinking water by SPE and LC-MS/MS. Together with Method 533 it covers 29 unique PFAS, and both are approved for UCMR 5 and the PFAS NPDWR.
How long can samples be held under Method 533?
Samples must be extracted within 28 days of collection, stored at or below 6 °C and not frozen, and extracts must be analysed within 28 days after extraction.
Sources
- U.S. EPA, Method 533: Determination of Per- and Polyfluoroalkyl Substances in Drinking Water by Isotope Dilution Anion Exchange SPE and LC/MS/MS (EPA 815-B-19-020), read November 20, 2026.
- U.S. EPA, EPA Analytical Methods for PFAS in Drinking Water (EPA 815-B-19-021, December 2019), read November 20, 2026.
- U.S. EPA, EPA PFAS Drinking Water Laboratory Methods (page updated April 24, 2026), read November 20, 2026.
- U.S. EPA, Fifth Unregulated Contaminant Monitoring Rule, read November 20, 2026.
- U.S. EPA, CWA Analytical Methods for PFAS (page updated February 25, 2026), read November 20, 2026.
- U.S. EPA, SW-846 Method 5030C (page updated March 2, 2026), read November 20, 2026.
- U.S. EPA HERO, Method 552.2, Revision 1.0 (EPA/600/R-95/131, 1995), read November 20, 2026.
- Eldridge M, LaFond J, Anderson T, Guelfo J, Jackson WA. Comparison of PFAS soil extractions and instrumental analysis: large-volume injection LC-MS, EPA Method 1633, and commercial lab results for 40 PFAS in various soils. Environmental Monitoring and Assessment 197(6):686, 2025, read via PMC November 20, 2026.
Data: PFAS analytes in EPA Methods 533 and 537.1 drinking-water methods
Each of the 29 unique PFAS measurable by EPA Methods 533 and 537.1, with abbreviation, CAS registry number and whether each method includes it. Compiled from EPA's analyte comparison table (EPA 815-B-19-021) and cross-checked against the Method 533 analyte list in EPA 815-B-19-020.
Period covered: 2019/2020. Coverage: United States.
| Abbreviation | Analyte | CASRN | In Method 533 | In Method 537.1 |
|---|---|---|---|---|
| 11Cl-PF3OUdS | 11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid | 763051-92-9 | Yes | Yes |
| 9Cl-PF3ONS | 9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid | 756426-58-1 | Yes | Yes |
| ADONA | 4,8-Dioxa-3H-perfluorononanoic acid | 919005-14-4 | Yes | Yes |
| HFPO-DA | Hexafluoropropylene oxide dimer acid | 13252-13-6 | Yes | Yes |
| PFBS | Perfluorobutanesulfonic acid | 375-73-5 | Yes | Yes |
| PFDA | Perfluorodecanoic acid | 335-76-2 | Yes | Yes |
| PFDoA | Perfluorododecanoic acid | 307-55-1 | Yes | Yes |
| PFHpA | Perfluoroheptanoic acid | 375-85-9 | Yes | Yes |
| PFHxA | Perfluorohexanoic acid | 307-24-4 | Yes | Yes |
| PFHxS | Perfluorohexanesulfonic acid | 355-46-4 | Yes | Yes |
| PFNA | Perfluorononanoic acid | 375-95-1 | Yes | Yes |
| PFOA | Perfluorooctanoic acid | 335-67-1 | Yes | Yes |
| PFOS | Perfluorooctanesulfonic acid | 1763-23-1 | Yes | Yes |
| PFUnA | Perfluoroundecanoic acid | 2058-94-8 | Yes | Yes |
| 4:2FTS | 1H,1H,2H,2H-Perfluorohexane sulfonic acid | 757124-72-4 | Yes | No |
| 6:2FTS | 1H,1H,2H,2H-Perfluorooctane sulfonic acid | 27619-97-2 | Yes | No |
| 8:2FTS | 1H,1H,2H,2H-Perfluorodecane sulfonic acid | 39108-34-4 | Yes | No |
| NFDHA | Nonafluoro-3,6-dioxaheptanoic acid | 151772-58-6 | Yes | No |
| PFBA | Perfluorobutanoic acid | 375-22-4 | Yes | No |
| PFEESA | Perfluoro(2-ethoxyethane)sulfonic acid | 113507-82-7 | Yes | No |
| PFHpS | Perfluoroheptanesulfonic acid | 375-92-8 | Yes | No |
| PFMBA | Perfluoro-4-methoxybutanoic acid | 863090-89-5 | Yes | No |
| PFMPA | Perfluoro-3-methoxypropanoic acid | 377-73-1 | Yes | No |
| PFPeA | Perfluoropentanoic acid | 2706-90-3 | Yes | No |
| PFPeS | Perfluoropentanesulfonic acid | 2706-91-4 | Yes | No |
| NEtFOSAA | N-ethyl perfluorooctanesulfonamidoacetic acid | 2991-50-6 | No | Yes |
| NMeFOSAA | N-methyl perfluorooctanesulfonamidoacetic acid | 2355-31-9 | No | Yes |
| PFTA | Perfluorotetradecanoic acid | 376-06-7 | No | Yes |
| PFTrDA | Perfluorotridecanoic acid | 72629-94-8 | No | Yes |