EPA Method 21 is the U.S. Environmental Protection Agency’s test method for finding and classifying volatile organic compound (VOC) leaks from process equipment with a portable detector. It is published in 40 CFR Part 60, Appendix A-7. It tells you how to prove the instrument works and how to place the probe on each leak point; it does not set the leak threshold, which each applicable regulation defines, and it does not measure how much is leaking.
Definition and scope
Section 1.2 of the method applies it to VOC leaks from process equipment. The listed sources include valves, flanges and other connections, pumps and compressors, pressure relief devices, process drains, open-ended valves, pump and compressor seal system degassing vents, accumulator vessel vents, agitator seals and access door seals. The list is described as not limited to those items.
The summary in section 2.1 is explicit about the purpose: a portable instrument detects VOC leaks from individual sources, the detector type is not specified so long as it meets the Section 6 specifications, and the leak definition is a concentration based on a reference compound that “is specified in each applicable regulation”. The method is “intended to locate and classify leaks only, and is not to be used as a direct measure of mass emission rate from individual sources.” That sentence is the most common source of confusion: a reading of 10,000 on the meter classifies a component against a threshold, and it is not a rate in grams or kilograms per hour.
The method text on the EPA site is dated 3 August 2017 and carries a note that the web version is not the official one; the official text is the Code of Federal Regulations. For the wider practice of finding and fixing leaks, including programme design, see the CASRAI guide on leak detection and repair. This page covers the method itself.
What is an EPA method?
An EPA test method is a written measurement procedure that a regulation points to when it needs a standard way of determining compliance. Method 21 is one of the test methods collected in the appendices to 40 CFR Part 60. The regulation for a given source category then supplies the numbers: the equipment to monitor, how often, and the leak definition. For example, 40 CFR 60.485a (test methods and procedures for the equipment-leak standards in that part) states that “Method 21 shall be used to determine the presence of leaking sources” and adds daily calibration conditions of its own. The method alone is therefore not a compliance programme; it is the measurement layer under one.
Key terms the method defines
Section 3 of the method fixes the vocabulary, and the rest of the text depends on it.
- Leak definition concentration: the local VOC concentration at the surface of a leak source that indicates a leak is present, expressed as an instrument meter reading based on a reference compound.
- Reference compound: the VOC species chosen as the calibration basis. The method’s own example is a leak definition of 10,000 ppm as methane: any source producing a reading of 10,000 on a methane-calibrated meter is classified as a leak.
- Calibration gas: usually the reference compound at a known concentration approximately equal to the leak definition.
- Response factor: the known concentration of a VOC divided by the meter reading observed when an instrument calibrated with the reference compound measures it. It expresses how differently the detector responds to the compound being monitored.
- Response time: the time from a step change in concentration at the sampling-system input until 90 percent of the final value shows on the meter.
- No detectable emission: a local concentration, adjusted for ambient, of less than 2.5 percent of the leak definition.
The instrument the method requires
Section 6 describes the equipment in performance terms instead of naming a product. The detector must respond to the compounds being processed; the method names catalytic oxidation, flame ionization, infrared absorption and photoionization as detector types that may meet this, and says the list is not exclusive. The wider detector background is in the CASRAI explainer on flame ionization detectors. The other requirements are:
- The instrument must be able to measure the leak definition concentration, and its meter must be readable to ±2.5 percent of that concentration.
- An electrically driven pump must deliver a constant flow, with a nominal rate of 0.10 to 3.0 litres per minute measured at the probe tip when a glass wool plug or filter is fitted.
- The probe or probe extension must not exceed 6.4 mm (1/4 inch) outside diameter, with a single end opening.
- The instrument must be intrinsically safe for explosive atmospheres, at a minimum for Class 1, Division 1 and/or Class 2, Division 1 conditions as appropriate, and must not be operated with a safety device such as an exhaust flame arrestor removed.
The reagents are equally plain. Zero gas is air with less than 10 ppmv VOC. The calibration gas is a known standard of each species to be measured, at a concentration approximately equal to the leak definition. Cylinder gases must be certified by the manufacturer to within 2 percent accuracy and carry a shelf life; gases prepared by the user must be accurate to within 2 percent and are replaced each day of use unless it is shown that they do not degrade in storage. Calibrating with a compound other than the reference compound is allowed, but a conversion factor must then be determined so readings can be converted to reference compound results.
How the method is used in practice
Step 1: prove the instrument before it goes into service
Section 8.1 sets three performance tests, which the data table below sets side by side. The response factor test covers each compound to be measured and uses three alternating measurements of calibration gas and zero gas; the result must be below 10 unless the applicable regulation says otherwise. If published response factors exist for the compound and detector type, the determination can be skipped and the published results referenced; the method cites three EPA reports on analyser response factors for this purpose. The calibration precision test uses three measurements and must come out at or below 10 percent of the calibration gas value. The response time test uses three trials and must come out at 30 seconds or less, with the pump, any dilution probe, the sample probe and the filter that will be used in the survey all in place.
The timing differs between the tests. The response factor test is required before service and does not have to be repeated at later intervals. Calibration precision is required before service and then at 3-month intervals or at the next use, whichever is later. The response time test is required before service and again if a change to the pumping system or flow configuration would alter the response time.
Step 2: calibrate, then survey
Section 10 describes calibration in one short paragraph: after warm-up and the instrument’s zero procedure, introduce the calibration gas to the probe and adjust the meter to the calibration gas value. If the meter cannot be adjusted to the proper value, the method says a malfunction is indicated and corrective action is needed before use.
Type I survey: leak definition based on concentration
Under section 8.3.1 the probe inlet is placed at the surface of the component interface where leakage could occur and moved along the interface periphery while the readout is observed. Where the reading increases, the interface is sampled slowly until the maximum reading is found, and the probe is held there for about two times the instrument response time. If the maximum reading exceeds the leak definition in the applicable regulation, the result is recorded and reported as that regulation requires. The method then gives positioning notes by equipment type; the table below condenses them.
| Equipment | Where the method places the probe inlet (section 8.3.1) |
|---|---|
| Valves | At the interface where the stem exits the packing gland, around the stem circumference; also at the packing gland take-up flange seat, and at all interfaces on multipart valve housings |
| Flanges and other connections | At the outer edge of the flange-gasket interface around the circumference; threaded and other nonpermanent joints with a similar traverse |
| Pumps and compressors | A circumferential traverse at the shaft and seal interface, with the inlet within 1 cm of a rotating shaft-seal interface; all accessible portions if the housing blocks a full traverse |
| Pressure relief devices | Approximately the centre of the exhaust area to atmosphere where an enclosed extension or horn exists |
| Process drains | Approximately the centre of the opening for open drains; the cover interface, with a peripheral traverse, for covered drains |
| Open-ended lines or valves; seal system degassing and accumulator vents | Approximately the centre of the opening to atmosphere |
| Access door seals | At the door seal interface with a peripheral traverse |
Type II survey: no detectable emissions
Where a regulation requires that no detectable emissions exist, section 8.3.2 has the surveyor measure local ambient VOC by moving the probe randomly upwind and downwind at one to two metres from the source. If a nearby emission interferes, the ambient reading may be taken closer, but never nearer than 25 centimetres. The probe is then moved to the source surface and read as in Type I, and the difference between the two concentrations decides whether there are no detectable emissions. Where a regulation requires a specific device installation, or that vents be ducted or piped to a control device, the method requires that to be visually confirmed.
The soap-bubble alternative
Section 8.3.3 allows a soap-solution screening procedure only for sources with none of the following conditions: continuously moving parts, a surface temperature above the boiling point or below the freezing point of the solution, open areas to the atmosphere that the solution cannot bridge, or evidence of liquid leakage. The solution may be a commercial leak detection solution or concentrated detergent in water. If no bubbles form, the source is presumed to have no detectable emissions or leaks; if bubbles form, the instrument technique of section 8.3.1 or 8.3.2 decides whether a leak exists.
Where the method ends and the regulation begins
A frequent mistake is to read Method 21 as if it contained the daily rules. It does not. The method’s own tests are per instrument and per compound; a regulation then adds routine operating requirements. The equipment-leak standards in 40 CFR 60.485a show the split:
| Question | Method 21 itself | 40 CFR 60.485a (equipment-leak test procedures) |
|---|---|---|
| When is calibration done? | Section 10.1 describes the adjustment; precision is checked before service and every 3 months or at next use | Before use each day the instrument is used |
| Zero gas | Air with less than 10 ppmv VOC | Zero air with less than 10 ppm hydrocarbon |
| Calibration gas | Approximately the leak definition, in air | Methane or n-hexane in air at no more than 2,000 ppm above the leak definition; for multiple scales the highest scale approximately 10,000 ppm |
| Drift | Not addressed | Calibration drift assessed at least at the end of each monitoring day; negative drift over 10 percent means re-monitoring equipment checked since the last calibration |
| No-detectable-emission test | Defined as under 2.5 percent of the leak definition | Maximum instrument reading minus background compared with 500 ppm |
The exact leak definitions for particular equipment sit in the sections of that regulation’s standards, not in the method, so a reading is never judged against Method 21 alone. The no-detectable-emission rows differ because they come from different documents: the method defines the term as a fraction of the leak definition, while the regulation sets a fixed 500 ppm difference test for the equipment it covers.
Where it matters for labs and research facilities
Research and pilot facilities that handle VOC-bearing process equipment, from solvent recovery to pilot plants and pressurised gas systems, meet Method 21 when an environmental permit or an equipment-leak standard cites it. The practical consequences follow from the method text. The meter is calibrated to a reference compound, so the reference compound and leak definition in the permit decide what gas to buy and how to document the instrument. Records of the response factor, calibration precision and response time tests belong with the instrument. And because the method excludes mass-rate measurement, a leak count from a Method 21 survey cannot be turned into an emissions inventory without a separate, regulation-approved estimation method. For the detector side of this question, see the CASRAI guides on VOC detection and gas detector calibration.
Safety is one place the method speaks directly. Section 5.2 warns that some compounds whose leaks the method can find are irritating or corrosive (heptane is the example given) or toxic (benzene and methyl alcohol), and that nearly all are fire hazards. Section 6.6 therefore requires an intrinsically safe instrument, and the method says the user must establish appropriate safety practices for the source being surveyed.
Common misunderstandings
- “The reading is the emission rate.” The method says it is not to be used as a direct measure of mass emission rate from individual sources.
- “Method 21 sets the leak threshold.” The leak definition is specified in each applicable regulation; the method gives the example of 10,000 ppm as methane only to explain the idea.
- “Calibration happens every 3 months.” That interval applies to the calibration precision test in the method. A regulation such as 40 CFR 60.485a can require calibration before every day of use.
- “Any detector works for any compound.” The detector must respond to the compounds being processed, and a response factor of 10 or more fails the method’s test unless the regulation says otherwise.
- “The method certifies the operator.” The method text sets instrument tests and survey technique; it does not define an operator qualification. Any training expectation comes from the permit, regulation or employer.
- “Soap bubbles can replace the instrument everywhere.” Only for sources that meet the conditions in section 8.3.3, and bubbles still send you back to the instrument technique.
Where VOC leaks come from, and where VOCs come from more broadly
Method 21 is aimed at one source category: process equipment that leaks VOC vapour, such as valves, flanges and pump seals. VOCs as a class have far wider origins. EPA describes VOCs as chemicals released as gases from certain solids and liquids, and lists common sources in its indoor air guidance: paints, paint strippers and other solvents; wood preservatives; aerosol sprays; cleansers and disinfectants; moth repellents and air fresheners; stored fuels and automotive products; hobby supplies; dry-cleaned clothing; and pesticides. It also names building materials and furnishings, office equipment such as copiers and printers, correction fluids and carbonless copy paper, and graphics and craft materials including glues, adhesives, permanent markers and photographic solutions.
For a laboratory, the practical reading is that solvent stores, aerosol products and cleaning agents all add VOCs to the air, but a hand-held leak detector used under Method 21 answers a narrower question: is a specific piece of regulated equipment leaking above the level the applicable rule sets?
How Method 21 differs from EPA Method 200.8
Method 200.8 is often confused with Method 21 because both carry an EPA number, but they measure different things. EPA’s Selected Analytical Methods (SAM) page lists Method 200.8 as the determination of trace elements in waters and wastes by inductively coupled plasma-mass spectrometry (ICP-MS), cited as Revision 5.4, U.S. EPA 1994, and lists it for preparing and analysing drinking water samples for compounds containing arsenic, thallium and vanadium. It is a laboratory analytical method for metals in water. Method 21 is a field survey method for organic vapour leaks from equipment. The two are not interchangeable, and neither substitutes for the other.
Frequently asked questions
What is an EPA method?
An EPA method is a written test procedure that regulations cite to determine compliance, collected for air-pollution sources in the appendices to 40 CFR Part 60. Method 21 is one of them: it sets how a VOC leak instrument is tested and used, while the applicable regulation sets the leak definition.
Does Method 21 measure how much is leaking?
No. Section 2.1 says the method locates and classifies leaks only and is not a direct measure of mass emission rate from individual sources.
Which instruments can be used?
The method does not specify a detector type. Section 6.1 says catalytic oxidation, flame ionization, infrared absorption and photoionization detectors may meet the requirement, provided the instrument meets the performance criteria on range, flow, probe size, response and intrinsic safety.
How often are the performance tests repeated?
The response factor test is required before service and not repeated at later intervals; calibration precision is checked before service and every 3 months or at the next use, whichever is later; response time is tested before service and again after a change that affects it.
What are common sources of VOCs?
Per EPA indoor air guidance, examples include paints and solvents, aerosol sprays, cleansers and disinfectants, air fresheners, stored fuels and automotive products, dry-cleaned clothing, pesticides, building materials and furnishings, and office equipment such as copiers and printers. Industrial equipment leaks, which Method 21 targets, are a separate source category.
What is the EPA 200.8 test method used for?
Method 200.8 determines trace elements in waters and wastes by ICP-MS; EPA’s SAM page lists it for drinking water analysis for arsenic, thallium and vanadium compounds. It is unrelated to Method 21, which locates VOC leaks from equipment.
Sources
- U.S. EPA: Method 21 – Volatile Organic Compound Leaks, checked 2026-11-20
- U.S. EPA: Method 21, Determination of Volatile Organic Compound Leaks (PDF, 8/3/2017), checked 2026-11-20
- Legal Information Institute (Cornell): 40 CFR 60.485a, Test methods and procedures, checked 2026-11-20
- Legal Information Institute (Cornell): 40 CFR Appendix A-7 to Part 60, Test Methods 19 through 25E, checked 2026-11-20
- U.S. EPA: Volatile Organic Compounds’ Impact on Indoor Air Quality, checked 2026-11-20
- U.S. EPA SAM: Method 200.8, Determination of Trace Elements in Waters and Wastes by ICP-MS, checked 2026-11-20
Data: Method 21 instrument performance tests: requirement, criterion and timing
Compares the three instrument performance evaluations in section 8.1 of EPA Method 21 (response factor, calibration precision, response time) with the daily calibration drift check in 40 CFR 60.485a, giving the pass criterion, number of measurements and when each is required. Compiled from the EPA method text (3 August 2017) and the Cornell LII copy of 40 CFR 60.485a.
Coverage: United States.
| Check | Pass criterion | Measurements | When required | Source |
|---|---|---|---|---|
| Response factor (8.1.1) | Below 10 for each compound unless the regulation says otherwise | 3 alternating calibration gas and zero gas measurements | Before placing the analyzer in service; not repeated at later intervals; published factors may be referenced | EPA Method 21 section 8.1.1 |
| Calibration precision (8.1.2) | At or below 10 percent of the calibration gas value | 3 alternating zero gas and calibration gas measurements | Before service, then at 3-month intervals or at next use, whichever is later | EPA Method 21 section 8.1.2 |
| Response time (8.1.3) | At or below 30 seconds to reach 90 percent of final reading | 3 trials, with pump, probe and filter in place | Before service; again after a pump or flow change that alters response time | EPA Method 21 section 8.1.3 |
| Calibration drift assessment (60.485a(b)(2)) | Negative drift over 10 percent requires re-monitoring equipment checked since last calibration | Check with the same calibration gases used before the survey | At least at the end of each monitoring day | 40 CFR 60.485a |
Sources