FDA Artificial Intelligence Regulation: AI/ML Medical Devices, Pathways, and GMLP

The FDA regulates artificial intelligence and machine learning (AI/ML)-enabled medical devices through risk-based 510(k), De Novo, and PMA pathways, Good Machine Learning Practice (GMLP) principles, and Predetermined Change Control Plans.

Written and maintained by CASRAI Editorial Board

Last updated

FDA artificial intelligence regulation governs how artificial intelligence and machine learning (AI/ML) algorithms used in healthcare—whether deployed as standalone Software as a Medical Device (SaMD) or embedded as Software in a Medical Device (SiMD)—are classified, validated, cleared or approved for marketing, and monitored post-market in the United States. Coordinated by the Center for Devices and Radiological Health (FDA CDRH) and its Digital Health Center of Excellence (DHCoE) alongside CDER and CBER AI regulatory frameworks, the agency regulates clinical AI through risk-based premarket pathways, Good Machine Learning Practice (GMLP) principles, and statutory Clinical Decision Support (CDS) boundaries.

How Does the FDA Regulate AI/ML-Enabled Medical Devices?

Before selecting a premarket submission pathway, developers and clinical investigators must first determine whether an AI algorithm meets the statutory definition of a medical device under Section 201(h) of the Federal Food, Drug, and Cosmetic (FD&C) Act, or whether it qualifies as Non-Device Clinical Decision Support (CDS) under Section 520(o)(1)(E) (enacted by Section 3060 of the 21st Century Cures Act).

1. The Four-Criterion Non-Device CDS Boundary (Section 520(o)(1)(E))

Under the FDA’s final guidance on Clinical Decision Support Software (September 2022), a clinical software function is excluded from FDA device regulation only if it satisfies all four statutory criteria simultaneously:

  1. Criterion 1 (Signal/Image Exclusion): It is not intended to acquire, process, or analyze a medical image (such as DICOM CT, MRI, or pathology slides), a signal from an in vitro diagnostic (IVD) device, or a pattern/signal from a physiological signal acquisition system (such as ECG waveforms or continuous telemetry).
  2. Criterion 2 (Medical Information Input): It is intended for displaying, analyzing, or printing medical information about a patient or other medical information (such as peer-reviewed clinical studies or clinical practice guidelines).
  3. Criterion 3 (HCP Recommendation Scope): It is intended for supporting or providing recommendations to a health care professional (HCP) about prevention, diagnosis, or treatment—rather than outputting a single specific directive or a time-critical emergency alarm that requires immediate action.
  4. Criterion 4 (Independent Review of Basis): It enables the health care professional to independently review the basis for the recommendation so that the clinician does not rely primarily on the software output to make a clinical decision for an individual patient.

Any AI/ML model that analyzes medical images or physiological waveforms (failing Criterion 1), targets patients directly rather than HCPs (failing Criterion 3), or operates as an unexplainable “black-box” urgency triage score without transparent clinical logic (failing Criterion 4) is regulated as an AI/ML-enabled medical device (AIaMD).

2. The 10 Good Machine Learning Practice (GMLP) Guiding Principles

Jointly issued by the FDA, Health Canada, and the UK Medicines and Healthcare products Regulatory Agency (MHRA) in October 2021, the 10 GMLP Guiding Principles establish quality expectations across the Total Product Life Cycle (TPLC):

  • Representativeness & Independence of Datasets (Principles 3 & 4): Training datasets and clinical validation cohorts must reflect the intended target population across age, sex, race, ethnicity, and scanner hardware, and test datasets must be kept strictly independent from training and tuning datasets to prevent data leakage.
  • Reference Standards & Tailored Model Design (Principles 5 & 6): Ground-truth labels must rely on best-available clinical reference standards (such as multi-reader consensus adjudication or biopsy confirmation), and model validation must document subgroup performance and generalizability risks.
  • Human–AI Team Performance (Principle 7): Where the AI model assists a clinician, validation studies must evaluate the combined performance of the human clinician plus the AI system, rather than standalone algorithm accuracy in isolation.
  • Post-Market Drift Monitoring & Change Control (Principles 9 & 10): Deployed models must be monitored for real-world data drift and managed via Predetermined Change Control Plans (PCCPs) authorized under Section 515C of the FD&C Act.

How Do 510(k), De Novo, and PMA Pathways Apply to Clinical AI Algorithms?

Note that the FDA does not “certify” AI algorithms (why FDA does not certify devices); instead, the agency clears Class II devices via 510(k), grants marketing authorization via De Novo, or approves Class III devices via PMA.

Premarket Pathway Statutory / CFR Basis Share of Listed AI/ML Devices Typical Clinical AI/ML Use Case & Evidence Standard
510(k) Premarket Notification FD&C Act § 510(k); 21 CFR Part 807 Subpart E ~95%+ Moderate-risk (Class II) AI software demonstrating substantial equivalence to an existing cleared predicate (e.g., radiology computer-aided triage under product code QAS, quantitative cardiac MRI segmentation, or ECG arrhythmia detection). Typically validated via standalone retrospective multi-site testing and/or multi-reader multi-case (MRMC) studies.
De Novo Classification Request FD&C Act § 513(f)(2); 21 CFR Part 860 Subpart D ~3–4% Novel low-to-moderate-risk AI devices with no legally marketed predicate. Establishes a brand-new Class I or Class II regulation and product code with binding Special Controls (e.g., autonomous diabetic retinopathy diagnostic software DEN180001 / product code PIB, or novel computer-aided detection software). Usually requires prospective clinical trial validation.
Premarket Approval (PMA) FD&C Act § 515; 21 CFR Part 814 <1% High-risk (Class III) AI/ML devices that support or sustain human life or carry high risk of illness or injury without sufficient special controls, requiring prospective pivotal clinical trials proving reasonable assurance of safety and effectiveness.

Submissions for 510(k), De Novo, and PMA AI/ML devices are prepared and transmitted electronically using the FDA eSTAR structured submission template and must include premarket cybersecurity documentation under Section 524B (including a Software Bill of Materials and adversarial ML threat modeling).

What Is Included on the FDA List of AI/ML-Enabled Medical Devices?

To provide public transparency across authorized clinical algorithms, the FDA Digital Health Center of Excellence maintains the Artificial Intelligence and Machine Learning (AI/ML)-Enabled Medical Devices database. Each record on the FDA list includes:

  • Submission Number & Pathway Prefix: The K number (510(k)), DEN number (De Novo), or P number (PMA), linked directly to the public FDA decision summary;
  • Date of Final Decision, Device Name, and Applicant/Company;
  • Lead Review Panel (Medical Specialty) and 3-Letter FDA Product Code: Over 75% of all listed AI/ML authorizations fall under Radiology (such as product codes QAS for radiological computer-aided triage and notification, LLZ for medical image management and processing, and QBS/MYN for computer-aided detection), followed by Cardiovascular (approximately 10%, including ambulatory ECG patch analysis and echocardiogram quantification), Neurology, Hematology, Pathology, and Ophthalmology.

For clinical researchers and hospital AI governance committees, the FDA database serves as the primary verification source to confirm a vendor algorithm’s exact cleared Indications for Use, target patient age group, and locked versus PCCP-enabled update status before clinical deployment.

Sources

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about FDA Artificial Intelligence Regulation: AI/ML Medical Devices, Pathways, and GMLP

Ask your first 2 questions free below. Subscribers get 150 a day for $29 a month.

An AI assistant specialized in research administration. It cites the sources behind every answer, labels web answers and says when it can't answer.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays free to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →