Fourier Transform Infrared Spectrophotometer: Parts and Specs

How an FTIR spectrophotometer is built, how detector types differ, and how to read range, resolution and beamsplitter lines on a datasheet.

Written and maintained by CASRAI Editorial Board

Last updated

A Fourier transform infrared (FTIR) spectrophotometer is a laboratory instrument that identifies materials by how they absorb infrared light. This page covers what is inside one, how the detector choice changes what it can do, and how to read the specification lines on a datasheet. It is aimed at lab staff, students and anyone writing an instrument specification.

What an FTIR spectrophotometer is

An FTIR instrument uses interferometry to record information about a material placed in the infrared beam, then applies a Fourier transform to turn the detector output into a spectrum a person can interpret (Thermo Fisher Scientific, FTIR basics). Each molecule gives a distinct spectral pattern, which is what allows a sample to be identified.

Who uses it: analytical and quality-control laboratories that need to identify or compare materials from their infrared spectra. The instrument is laboratory equipment, not a clinical device.

How it is built

Three ideas describe the signal path. The interferometer modulates the infrared beam, the sample sits in that beam, and a detector records the result as an interferogram. Software then applies the Fourier transform to convert the interferogram into the spectrum (Thermo Fisher Scientific, FTIR basics).

  • Interferogram: the raw detector signal, before the transform. Saving it lets you reprocess the data later with different settings.
  • Apodization: a processing step applied to the interferogram. At a resolution of 4 cm-1 or coarser, heavy apodization does not badly distort spectra, but for narrow gas-phase lines its effect can be significant (Thermo Fisher Scientific).
  • Optics and windows: inorganic peaks below 400 cm-1 need far-infrared optics. For protein work the useful region is roughly 1400 to 1750 cm-1, where BaF2 windows transmit (Thermo Fisher Scientific).

Detector types

The detector is the specification choice that most changes how an instrument behaves. Thermo Fisher describes two common types:

Detector Cooling What the source says
MCT-A (mercury cadmium telluride) Liquid nitrogen High sensitivity
DLaTGS Room temperature Wider spectral range, lower sensitivity

Source: Thermo Fisher Scientific, FTIR basics, read 2026-10-29.

Reading the specifications on a datasheet

As a worked example, here are the lines published for one manufacturer’s instrument, the Shimadzu IRXross. They are shown to illustrate what a datasheet lists, not as a recommendation.

Specification Published value (Shimadzu IRXross) What to look for
Wavenumber range 7,800 to 350 cm-1 The span of the spectrum the instrument records
Resolution settings 0.25, 0.5, 1, 2, 4, 8, 16 cm-1 The selectable steps; the right setting depends on the sample, since gas-phase lines are narrow
Beamsplitter Germanium-coated KBr The material in the interferometer
Standard detector DLATGS with temperature control Matches the room-temperature type in the table above
Optional detector MCT (Hg-Cd-Te), liquid nitrogen cooled Matches the high-sensitivity type above

Source: Shimadzu IRXross product page, read 2026-10-29. Values are the manufacturer’s own.

Shimadzu also states that a KRS-5 window option is compliant with pharmacopoeia wavenumber range requirements (350 to 7,800 cm-1) and that its software enables easy compliance with USP, EP, JP and ChP regulations. These are the manufacturer’s statements; if your lab works to a pharmacopoeia, check the requirement in the pharmacopoeia itself.

Sampling options

Thermo Fisher notes that attenuated total reflectance (ATR) needs minimal sample preparation, while transmission sampling is economical. The choice between them is covered in detail in ATR-FTIR vs transmission FTIR sampling.

How to choose

Only criteria stated in the sources above can be offered:

  • Match the detector to the job: MCT-A for sensitivity, DLaTGS where a wider range and no liquid nitrogen matter more (Thermo Fisher).
  • Check that the wavenumber range covers the region you need. Far-infrared optics are required below 400 cm-1 (Thermo Fisher).
  • Check the window material against your region of interest, for example BaF2 for protein work around 1400 to 1750 cm-1 (Thermo Fisher).
  • Check the resolution settings suit your sample type (see apodization above).

Common mistakes

  • Discarding the interferogram, which removes the option to reprocess the data with different settings.
  • Applying heavy apodization to narrow gas-phase lines without checking its effect.
  • Using a window or optics that do not transmit in the region of interest.

Where to go next

Once you have a spectrum, the IR functional-group frequency table shows how to read it. For how infrared absorption compares with another vibrational technique, see Raman vs FTIR spectroscopy. Broader context: what is analytical chemistry.

Sources

  • Thermo Fisher Scientific, FTIR basics (spectroscopy learning center): thermofisher.com, read 2026-10-29.
  • Shimadzu, IRXross FTIR product page: ssi.shimadzu.com, read 2026-10-29.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Fourier Transform Infrared Spectrophotometer: Parts and Specs

Ask your first 2 questions free below. Subscribers get 150 a day for $29 a month.

An AI assistant specialized in research administration. It cites the sources behind every answer, labels web answers and says when it can't answer.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays free to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →

Ask CASRAI · Regulatory Radar

Research-admin question? Get an answer that links its sources.

An AI assistant specialized in research administration. Every answer links its sources to check before you act. 2 questions free, no account. $29/month after.

  • Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.
  • Every answer numbers its sources and links each one, so you can check the source yourself.