Gilson PIPETMAN: Classic Models, Volume Ranges and Care

Gilson PIPETMAN Classic explained: eight models from 0.2 uL to 10 mL, error limits against ISO 8655, correct use and which parts to autoclave.

Written and maintained by CASRAI Editorial Board

Published

On this page

The Gilson PIPETMAN is a family of manually operated, adjustable-volume air-displacement pipettes made by Gilson. In its PIPETMAN Classic form, eight models (P2, P10, P20, P100, P200, P1000, P5000 and P10ml) cover 0.2 µL to 10 mL, with the chosen volume shown on a digital volumeter.

This guide takes the Classic line as Gilson documents it in its own user manual: what each model is for, how the two-stop plunger is used, how Gilson’s accuracy limits compare with the ISO 8655 limits it prints beside them, and which parts may be autoclaved. It is a reference for researchers and lab managers and is not a purchasing page.

What the PIPETMAN is and what “Classic” means

Gilson’s 2012 user manual describes the PIPETMAN Classic as a fully adjustable, air-displacement pipette with the selected volume shown on a digital indicator, the volumeter. The 2009 guide for the same eight-model range is titled PIPETMAN P, and the specifications table printed in both documents carries the same figures. Practically, a lab may meet the same instrument under either name depending on the age of its documentation (Gilson 2012 manual; Gilson 2009 guide).

The instrument’s origin is documented by the University of Wisconsin–Madison. Its news service reports that Warren Gilson, a neurologist, designed a variable-output pipette; that a UW–Madison engineering undergraduate, Eric Marteau d’Autry, asked to work on it in the late 1960s; and that he later took it to France to develop the market. The same article says all of Gilson’s instruments are made in Middleton, Wisconsin (UW–Madison News, 21 December 2015). Sources differ on the exact year of the first commercial model, so none is given here.

For the underlying principle of the category, see what a pipette is. The remainder of this page deals only with what is specific to the Gilson Classic line.

How a PIPETMAN is used in practice

The manual describes a two-stop plunger on a push-button. Pressing to the first stop corresponds to the set volume. Holding the pipette vertically, the user immerses the tip, releases the button slowly and smoothly to the top position to draw up the set volume, waits for the time given for the model, and withdraws the tip. To dispense, the tip end is placed against the inside wall of the receiving vessel at an angle of 10° to 40°, the button is pressed slowly to the first stop, and after at least a second it is pressed to the second stop to expel residual liquid. The tip is then drawn along the vessel wall while the button stays fully down.

Setting the volume

The volume is set by turning the thumbwheel or the push-button, which Gilson says makes setting quicker and easier when wearing gloves. For the most accurate setting, the manual says to approach the value slowly when decreasing the setting, and when increasing it to pass the required value by about one third of a turn and then come back down slowly without overshooting the mark. The volumeter’s dial numbers are coloured black or red to show where the decimal point sits, and the colour scheme depends on the model.

Immersion depth and wait time

The manual gives a fixed table of immersion depth and pause per model. The values below are Gilson’s, copied by model.

Model Immersion depth (mm) Wait time (s)
P2 1 1
P10 1 1
P20 2-3 1
P100 2-4 1
P200 2-4 1
P1000 2-4 2-3
P5000 3-6 4-5
P10ml 5-7 4-5

General technique points from the manual

  • Change the tip before aspirating a different liquid, sample or reagent, and when a droplet remains on the tip from the previous operation.
  • Pre-rinse each new tip with the liquid to be pipetted. Gilson singles out protein-containing solutions and organic solvents, which can leave a film on the inside wall of the tip; the first volume is aspirated and dispensed back to waste.
  • For volatile solvents, saturate the air cushion by aspirating and dispensing the solvent repeatedly before taking the sample.
  • Liquid should never enter the tip-holder: operate the button slowly, never turn the pipette upside down and never lay it on its side while liquid is in the tip. Store it vertically.
  • Do not pipette liquids above 70 °C or below 4 °C. The pipette can be used between +4 °C and +40 °C, and specifications may vary with temperature.
  • When the liquid temperature differs from ambient, pre-rinse several times.

For wider technique advice that is not tied to one brand, see pipetting technique for accuracy and precision.

PIPETMAN Classic models and what each is for

Gilson groups the eight models into three use bands. The table reproduces the manual’s grouping with the maximum volume of each model taken from its specification table.

Model Maximum volume Use band stated by Gilson
P2 2 µL Measurement and transfer of microvolumes, DNA sequencing and enzyme assays
P10 10 µL Same band as P2
P20 20 µL General aqueous solutions, acids and bases
P100 100 µL Same band as P20
P200 200 µL Same band as P20
P1000 1000 µL Same band as P20
P5000 5000 µL Large volumes
P10ml 10 mL Large volumes

Gilson adds that with a precise technique the P2 can be used to aspirate volumes as low as 0.1 µL and the P10 as low as 0.5 µL, below the 0.2 µL and 1 µL minimum volumes listed in the accuracy table.

Tips and the P2 and P10 adapter

For best performance Gilson strongly recommends its PIPETMAN DIAMOND tips, made from pure polypropylene with the Gilson logo engraved on the collar, and says plastic tips are for a single application and must not be cleaned for reuse. It also states that the Classic can be used with the main tip brands. The accuracy figures, however, are guaranteed only with genuine DIAMOND tips. The P2 and P10 need a plastic dual-position adapter clipped to the tip-ejector rod to fit long DL10 tips or short D10 tips; they ship with the adapter set for the longer tips. For general tip selection, including filtered tips, see filtered pipette tips.

Accuracy limits: Gilson’s figures against ISO 8655

The manual publishes, for every model, maximum permissible systematic and random errors at the minimum, middle and maximum volume, next to the ISO 8655 limits. Gilson states that its pipettes comply with ISO 8655 by type testing and that adjustment is carried out under the conditions defined in ISO 8655-6. The data table below compares the two sets of limits at each model’s maximum volume and converts Gilson’s limits to a share of that volume.

Reading that table against the manual gives three observations.

  • Random error is tighter than ISO for all eight models. Gilson’s random limit at maximum volume is between about 35% (P2: 0.014 µL against 0.04 µL) and 60% (P20: 0.06 µL against 0.10 µL) of the ISO figure.
  • Systematic error is tighter than ISO on three models. The P2 (0.030 µL against 0.08 µL), the P10 (0.100 µL against 0.12 µL) and the P5000 (30 µL against 40 µL) are stricter than the ISO limit; for P20, P100, P200, P1000 and P10ml the two are equal at maximum volume.
  • Relative error is worst at the bottom of the range. At 0.2 µL on a P2, Gilson’s systematic limit is ±0.024 µL and its random limit 0.012 µL, which is 12% and 6% of the volume. At the P2’s 2 µL maximum the same limits correspond to 1.5% and 0.7%. The same pattern holds across every model’s table, which is why Gilson’s three-point tables matter more than a single headline figure.

The manual’s ISO columns date from 2012. In its customer notice on the 2022 revision, Eppendorf states that ISO 8655:2022 lowered some limits, that the standard now lists limits at 10%, 50% and 100% of nominal volume, that multichannel pipettes now have their own table, and that manufacturer limits are usually stricter than the ISO limits. A lab applying the current edition should therefore take its ISO comparison from the edition its procedure names rather than from the printed 2012 columns (Eppendorf customer information on ISO 8655:2022). Related background is in pipette calibration: how and when.

Care, cleaning and autoclaving

Gilson’s first instruction in the manual is not to lubricate the pipette: its manual states that the use of lubricant cancels the warranty. Beyond that, the manual limits maintenance to cleaning or autoclaving specified parts and replacing the push-button, connecting nut, tip-ejector, tip-holder, seal and O-ring.

Part Autoclavable per the manual
Tip-ejector Yes
Tip-holder Yes
Connecting nut Yes
Upper part (body or handle) No
Piston assembly No
O-ring and seal No; clean or replace

The manual’s autoclave cycle for the permitted parts is 20 minutes at 121 °C and 0.1 MPa, after cleaning the parts, bagging them in an autoclaving sack, and checking they are dry before reassembly. For the general autoclave concept see what an autoclave is.

Cleaning and decontamination

Cleaning comes before decontamination, and Gilson recommends a soap solution. Only the connecting nut, tip-ejector, tip-holder, piston assembly, seal and O-ring may be immersed in cleaning or decontaminant solution. The manual describes an ultrasonic bath of 20 minutes at 50 °C for the lower-part components, followed by a distilled-water rinse and drying. Liquid must never enter the upper part. If a lab chooses chemical decontamination, the manual says to confirm with the decontaminant’s supplier that it is compatible with stainless steel and with PA, PBT, PC, PC/PBT, POM and PVDF, the plastics used in the pipette.

The P2 and P10 contain miniaturised parts. Gilson advises not disassembling them yourself and sending them to an authorised Gilson service centre; only their push-button, tip-ejector, dual-position tip-ejector and adapter are listed as user-replaceable.

Leak test and service

The manual’s leak test fits a DIAMOND tip, sets the maximum volume given in the specifications, pre-rinses, aspirates distilled water and holds the pipette vertically for 20 seconds. A droplet at the tip means a leak. For the P2 to P200 models, the water level inside the tip is also watched after re-immersion and should stay constant. A pipette that fails should have its O-ring and seal replaced and then be retested. The troubleshooting table lists a damaged tip-holder or a worn O-ring or seal as causes of leaking, and a loose connecting nut among the causes of an inaccurate pipette. If the piston assembly is changed, the manual says the pipette must be adjusted and calibrated at a Gilson authorised service centre, and after replacing any part, performance should be verified.

Where it matters for labs and records

Each PIPETMAN has a serial number engraved on the body that gives unique identification and the manufacture date, and the bar code on the box and the certificate of conformity provide traceability; Gilson presents these as its GLP features. For an audit trail, that serial number is the link between a pipette, its calibration records and the experiments it was used for; see Good Laboratory Practice (GLP). The 2012 manual also carries a Gilson EC declaration of conformity under the in vitro diagnostic medical devices directive, self-declared under Annex III, dated 2 May 2012.

The practical consequences are modest but real. Choosing a model whose maximum volume is close to the working volume keeps the relative error near the lower end of the percentages in the data table. Recording which tip brand was used matters because the stated accuracy is guaranteed only with genuine DIAMOND tips. And recording which ISO edition a calibration used matters because limits changed between editions.

Common misunderstandings

  • The whole pipette can be autoclaved. It cannot. Only the tip-ejector, tip-holder and connecting nut are listed; the body, piston assembly, O-ring and seal are not.
  • The second stop is just a harder press. The first stop is the set volume; the second stop is a blow-out to expel residual liquid, used when dispensing.
  • Any tip gives the published accuracy. The manual says the Classic can be used with the main tip brands, but guarantees its figures only with genuine PIPETMAN DIAMOND tips.
  • A little lubricant helps a sticky piston. The manual’s opening notice says not to lubricate and that doing so cancels the warranty.
  • The ISO columns in an old manual are today’s limits. They are the limits printed in a 2012 document; the standard has since been revised.
  • PIPETMAN Classic and PIPETMAN P are different instruments. The 2009 and 2012 guides describe the same eight-model range with the same specification table under two names.

Sources

Data: PIPETMAN Classic maximum permissible errors at maximum volume, Gilson limits against the ISO 8655 limits printed in the manual

For each of the eight PIPETMAN Classic models, the maximum nominal volume with Gilson's maximum permissible systematic and random error and the ISO 8655 limits that Gilson prints beside them. The two percentage columns are computed here by dividing each Gilson limit by the maximum volume. Values come from the Gilson specifications table in the 2012 user manual and match the same table in the 2009 PIPETMAN P guide. Gilson states the figures are guaranteed only with genuine PIPETMAN DIAMOND tips.

Period covered: Manual editions of 2009 and 2012.

PIPETMAN Classic maximum permissible errors at maximum volume, Gilson limits against the ISO 8655 limits printed in the manual
ModelMaximum volume (uL)Gilson systematic error limit (uL)Gilson random error limit (uL)ISO 8655 systematic error limit (as printed by Gilson) (uL)ISO 8655 random error limit (as printed by Gilson) (uL)Gilson systematic limit as share of maximum volume (computed) (%)Gilson random limit as share of maximum volume (computed) (%)
P220.030.0140.080.041.50.7
P10100.10.040.120.0810.4
P20200.20.060.20.110.3
P1001000.80.150.80.30.80.15
P2002001.60.31.60.60.80.15
P1000100081.5830.80.15
P5000500030840150.60.16
P10ml10000601660300.60.16

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Gilson PIPETMAN: Classic Models, Volume Ranges and Care

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →