What Are Filtered Pipette Tips?

Learn how filtered pipette tips prevent liquid and aerosol cross-contamination, comparing filter materials, pore sizes, ISO 8655 rules, and volume ranges.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

Filtered pipette tips are single-use laboratory consumables containing an integrated porous barrier—typically made from sintered hydrophobic polyethylene or self-sealing cellulose polymer—mounted inside the upper barrel of the tip. Their primary operational purpose is to block liquid aerosols, microscopic droplets, and vaporous contaminants generated during air-displacement pipetting from entering the pipette shaft or causing sample-to-sample carryover.

This page describes products and general use; it is not medical advice.

Who uses it: Molecular biology researchers, clinical diagnostic technicians, forensic analysts, and pharmaceutical quality-control scientists handling nucleic acid amplifications, infectious pathogens, radioisotopes, or volatile liquid reagents.

What filtered pipette tips are

In quantitative liquid handling, the pipette and tip function together as a unified measurement system. Under international metrological standards—specifically ISO 8655-2:2022 clause 4 and clause 7—the maximum permissible systematic error (inaccuracy) and random error (imprecision) apply strictly to the combined assembly of the mechanical pipette and its fitted tip. Filtered pipette tips, often termed aerosol barrier tips or filter tips, adapt standard disposable polypropylene pipette tips by embedding a porous filtration element into the upper interior collar. This element sits directly between the liquid sample column and the mounting shaft of air-displacement micropipettes.

In standard air-displacement pipetting, an internal piston creates a negative pressure differential to draw fluid into the disposable plastic tip. Because the aspirated fluid never physically enters the mechanical body of the pipette under normal operations, laboratories frequently assume the instrument remains sterile between dispenses. However, high-speed fluid motion, fluid surface agitation, meniscus rupture, and rapid plunger release consistently generate sub-visible liquid aerosols (droplets typically ranging between 0.1 and 10 micrometers). These aerosolized particles travel upward on air currents into the pipette barrel, internal piston cavity, and seal assembly. Once deposited on internal metallic and elastomer surfaces, these residual droplets volatilize or shed into subsequently aspirated samples, creating subtle cross-contamination that can invalidate sensitive enzymatic reactions.

From a regulatory standpoint, filtered pipette tips used in medical testing, pathology, and diagnostic workflows fall under United States Food and Drug Administration (FDA) jurisdiction pursuant to 21 CFR 862.2050 (General purpose laboratory equipment labeled or promoted for a specific medical use). The FDA classifies these precision liquid dispensers and tip consumables as Class I medical devices under Product Code JRC (Micro Pipette / Pipette Tip), which are generally 510(k)-exempt but subject to general controls, facility registration, and labeling compliance under 21 CFR 809.10. Similarly, biosafety authorities including the CDC and NIH in Biosafety in Microbiological and Biomedical Laboratories (BMBL, 6th Edition) classify manual liquid pipetting as an aerosol-generating laboratory procedure. The BMBL guidelines mandate strict containment protocols, barrier tips, and physical enclosure within biological safety cabinets when manipulating infectious materials, bloodborne pathogens, or recombinant genetic elements.

How they work and how they are built

Filtered pipette tips operate through mechanical interception, inertial impaction, and diffusion rather than electrostatic attraction or chemical destruction. A high-grade filtered tip consists of two distinct engineered components: the precision-molded external polypropylene casing and the internal porous barrier plug.

The external tip body is injection-molded from virgin medical-grade polypropylene (PP), selected for its chemical inertness, optical transparency, structural rigidity, and hydrophobic surface tension. High-precision molding ensures consistent concentricity, a smooth internal bore, and a flash-free distal orifice that minimizes capillary sample retention. During fluid intake, liquid occupies only the lower and middle portions of the tip barrel, maintaining a substantial air headspace below the filter.

The barrier plug is manufactured from high-molecular-weight porous polyethylene (PE) or sintered high-density polyethylene (HDPE). Unlike woven textile meshes or porous membrane discs, the filter plug is produced via controlled thermal sintering of calibrated polymer beads. Sintering fuses the polymer spherical surfaces at their contact points without melting them into a solid block, creating an intricate three-dimensional sponge-like labyrinth known as a tortuous path matrix. The nominal pore dimensions within this matrix typically average 8 to 10 micrometers (with tortuous path channels spanning 5 to 20 micrometers). As ambient air flows freely through the interconnected channels during aspiration and dispensing, airborne liquid micro-droplets cannot negotiate the winding matrix geometry. Instead, droplet inertia causes suspended particulates to impact the hydrophobic polyethylene pore walls, where surface tension traps them without permitting aerosol penetration into the pipette barrel.

Crucially, high-performance aerosol barrier filters are fabricated completely free of manufacturing additives, chemical binders, wetting agents, surfactants, biocides, or mold-release plasticizers such as di(2-ethylhexyl) phthalate (DEHP), diisopropyl ethylamine (DiHEMDA), or oleamide. Additive-free construction prevents chemical leachates from dissolving into aspirated liquids or volatile vapors, which could otherwise inhibit Taq polymerase activity, alter spectrophotometric absorbance baselines, or denature delicate protein complexes. The table below delineates the structural anatomy and physical functions of each filtered tip component.

Component Part Material Composition Manufacturing Method Primary Physical Function
Distillate Orifice (Tip Tip) Virgin Polypropylene (PP) High-cavitation diamond-polished injection molding Controls droplet departure geometry, minimizes liquid hangup, and maintains fluid velocity.
Lower Tip Barrel Virgin Polypropylene (PP) Thin-wall precision molding with low-retention surface option Holds calibrated fluid volume; provides clear optical visibility of meniscus and air bubbles.
Expansion Headspace Virgin Polypropylene (PP) Conical interior bore expansion Maintains an isolated air cushion between the maximum fluid meniscus and the filter base.
Porous Aerosol Filter Sintered Porous Polyethylene (PE) Controlled thermal sintering of calibrated PE microbeads Permits clean air exchange while capturing liquid micro-aerosols in a tortuous path matrix.
Mounting Collar (Cone) Virgin Polypropylene (PP) Radial elastomeric or beveled sealing collar Forms an airtight, leak-free mechanical seal against the pipette shaft under low mounting force.

Types and variants

Not all filtered pipette tips function identically. Depending on the assay sensitivity, reagent volatility, and potential for accidental over-aspiration, manufacturers produce several distinct filter architectures and resin formulations. Selecting the correct variant is critical when consulting a broader pipette tip selection guide or designing specialized molecular assays.

The primary technological distinction lies between non-reactive inert porous barriers and active self-sealing barriers. Standard inert barriers rely solely on hydrophobic pore networks to arrest aerosols; if accidental over-aspiration forces gross liquid against the filter, the fluid can slowly wet the matrix under continued vacuum. In contrast, self-sealing barrier tips incorporate hydrogel-forming polymers, such as carboxymethyl cellulose (CMC) or modified cellulose gum, blended into or layered adjacent to the polyethylene granules. Upon direct contact with aqueous liquid, the hydrogel rapidly hydrates and swells, instantly forming an impenetrable physical seal. This automated blockage stops liquid from surging into the pipette shaft, protecting the instrument from corrosive buffers, radioactive tracers, or viral lysates, although the aspirated sample within the sealed tip is permanently immobilized and sacrificed.

Another major technological evolution is the dual-phase or dual-layer filter, exemplified by Eppendorf ep Dualfilter T.I.P.S. systems. These tips integrate two distinct layered zones: a lower, sample-facing white layer composed of wider porous polyethylene that intercepts splashes and gross droplets, and an upper, pipette-facing blue layer engineered with tighter pore geometry that traps fine sub-micron aerosols (yielding particle retention efficiencies exceeding 99.95% when tested under EN 29463 and ISO 29463-5 standards). In parallel, low-retention variants utilize modified polypropylene resins containing hydrophobic fluoropolymer or silicone-free surface modifiers that prevent polar macromolecules (such as genomic DNA, restriction enzymes, and master mix detergents) from binding to the plastic walls.

Tip Variant Typical Laboratory Use Key Structural Difference
Inert Porous Polyethylene (PE) Routine PCR, cell culture passaging, general diagnostic assays Pure hydrophobic sintered PE matrix; non-swelling; allows sample recovery if liquid touches filter.
Self-Sealing (Liquid-Blocking) Radioactive labeling, forensic evidence pipetting, high-consequence viral pathogens Contains water-activated hydrogel additives that swell shut on fluid contact, preventing instrument contamination.
Dual-Phase (Two-Layer) Filter High-sensitivity quantitative PCR (RT-qPCR), next-generation sequencing library preparation Two distinct pore densities: coarse lower tier for droplets, dense upper tier for sub-micron aerosols.
Low-Retention Filter Tips Pipetting viscous master mixes, enzymatic restriction digests, detergent-rich buffers, glycerol Extremely hydrophobic inner resin surface minimizes liquid film retention and increases volumetric recovery.
Extended-Length (Reach) Filter Tips Pipetting from deep microtubes, cryogenic vials, and narrow conical centrifuge tubes Elongated, slender tip shaft prevents the non-sterile pipette nose cone from contacting vessel walls.

Sizes and specifications that matter

Pipette tips are engineered to match standard volumetric pipettor ranges. Tip geometry must correspond precisely to the pipette shaft diameter and taper angle to preserve volumetric delivery accuracy. Using an ill-fitting or generic tip compromises the airtight seal, causing systematic negative delivery bias or fluid leakage during the transfer step.

Volume capacities for filtered tips are generally smaller than their non-filtered counterparts because the sintered filter plug occupies space in the upper portion of the barrel. For instance, an unfiltered 200 µL tip barrel often holds up to 250 µL or 300 µL of total air, whereas a filtered 200 µL tip has its operational capacity strictly limited to 200 µL to preserve essential safety headspace below the filter. Standard volume classes span micro-volumes (0.1–10 µL), intermediate volumes (20 µL, 100 µL, 200 µL, and 300 µL), and macro-volumes (1,000 µL / 1 mL, up to 5 mL and 10 mL).

The table below compiles physical dimensions, standard color-coding, volume capacities, and pipette instrument pairings based on verified manufacturer instructions for use from Eppendorf, Gilson, and Thermo Fisher Scientific.

Volume Capacity Range Nominal Tip Length (mm) Industry Color Code Primary Instrument Pairing Key Geometric Application
0.1 µL – 10 µL (Short) 34 mm Dark Gray / Red P2, P10 single/multichannel pipettes Micro-volume dispensing in 96-well and 384-well PCR plates; low air cushion.
0.1 µL – 10 µL (Medium / Extended) 40 mm – 46 mm Medium Gray P10 pipettes, electronic micro-pipettors Reaches bottom of 0.5 mL and 1.5 mL microcentrifuge tubes without shaft contact.
2 µL – 20 µL 53 mm Yellow P20 single-channel pipettes Intermediate restriction digest and enzyme loading; precise capillary bore.
2 µL – 100 µL 53 mm Yellow P100 pipettes General molecular biology reagent additions and cDNA synthesis protocols.
2 µL – 200 µL 55 mm Yellow P200 single/multichannel pipettes Cell culture media aliquoting, ELISA reagent dispensing, buffer preparation.
20 µL – 300 µL 55 mm – 60 mm Orange / Green P300 multichannel pipettes Multi-well plate washing and high-throughput diagnostic screening assays.
50 µL – 1,000 µL (1 mL) 76 mm – 88 mm Blue P1000 single-channel pipettes Primary sample transfers, extraction lysis buffers, master mix stock distribution.
0.1 mL – 5 mL 120 mm – 175 mm Violet 5 mL macro-pipettors Cell suspension transfers, large-scale media preparation, chromatography loading.
1 mL – 10 mL 165 mm – 240 mm Turquoise 10 mL macro-pipettors Aspirating bulk biological buffers and viscous biological tissue homogenates.

How to choose filtered pipette tips

Selecting the appropriate filtered pipette tip requires evaluating biological safety hazards, sensitivity to trace chemical contaminants, the physical properties of the liquid, and compliance with pipette calibration standards. Laboratories should never treat filtered tips as a universal commodity where all brands perform interchangeably.

The first decision point is assay vulnerability to nucleic acid or enzymatic contamination. In PCR reaction setup protocols, quantitative real-time PCR, and RT-PCR protocols, carryover of even a single copy of target amplicon from a previous pipetting run can produce false-positive results. For these workflows, tips must possess third-party lot-specific certifications verifying that they are free of detectable human genomic DNA, DNase, RNase, and PCR inhibitors. For forensic DNA typing (such as short tandem repeat STR profiling), tips must comply with ISO 18385 forensic DNA-grade cleanliness to prevent human profile drop-in.

The second criterion involves sample rheology and surface tension. When pipetting solutions containing biological detergents (such as Triton X-100, Tween-20, or SDS), concentrated proteins (such as fetal bovine serum or BSA), or viscous glycerol stocks, standard polypropylene tips leave a residual fluid film on the interior walls. This fluid film introduces significant negative delivery errors. In these situations, laboratories must select low-retention filtered tips. However, if the fluid is highly volatile (such as acetone, hexane, or chloroform), vapors can rapidly saturate the air cushion and cause dripping, or even chemically degrade the polyethylene filter plug; in such scenarios, positive-displacement pipettes with capillary pistons must replace air-displacement pipettors entirely.

The third criterion is physical compatibility with the pipette nose cone. Under ISO 8655-2, calibration certificates are only metrologically valid when using the manufacturer’s validated tip system or third-party tips that have been explicitly qualified gravimetrically on that specific pipette model. Mismatched tips require excessive insertion force, which damages pipette shafts and user ergonomics, or fail to form an airtight seal, resulting in insidious volume drift. The decision matrix below summarizes selection pathways across standard laboratory workflows.

Experimental Workflow Primary Risk Factor Required Tip Architecture Mandatory Lot Purity Certification
RT-qPCR / qPCR Amplification Amplicon cross-contamination; RNase degradation of viral/cellular RNA Dual-phase or high-efficiency inert PE filter; low-retention resin Certified RNase-free, DNase-free, PCR inhibitor-free, Pyrogen-free
Forensic STR Genotyping Foreign human DNA allelic drop-in; trace contamination Inert aerosol barrier or self-sealing barrier; extended length ISO 18385 Forensic DNA Grade; free of detectable human gDNA
Radioisotope Labeling (e.g. 32P, 35S) Aerosol vaporization and liquid shaft contamination; radiation hazard Self-sealing liquid-blocking barrier (swells shut on contact) Standard sterile or non-sterile depending on biological assay
Primary Cell Culture Passaging Mycoplasma, bacterial, or viral cross-contamination between flasks Sterile aerosol barrier tips; standard or extended length Sterile (gamma or e-beam irradiated per ISO 11137); Endotoxin under 0.01 EU/mL
Protein Crystallography / NGS Enzymes Enzyme loss on plastic walls; volumetric imprecision in micro-volumes Low-retention filtered micro-tips (0.1–10 µL or 2–20 µL) DNase/RNase-free; protease-free; additive-free sintered PE
Routine Buffer Preparation (Tris, PBS) Non-critical cross-contamination risk; high cost sensitivity Standard non-filtered tips (or non-sterile bulk filtered tips if shared) Standard laboratory grade; particulate-free

Safe use and handling basics

Even the highest-grade aerosol barrier tip cannot protect experimental samples if operators employ poor pipetting mechanics. Operator technique directly governs liquid velocity, aerosol generation rates, and filter longevity.

Tip mounting requires deliberate axial alignment. To seat a tip securely, insert the pipette shaft firmly into the tip collar with a gentle straight-down motion, followed by a slight rotational twist if recommended by the manufacturer. Never bang or hammer the pipette onto tips in the rack; doing so deforms the polypropylene sealing rim, produces micro-cracks that cause air leaks, and accelerates mechanical wear on the pipette’s internal nose cone components.

During aspiration, maintain the pipette within 20 degrees of true vertical. Tilting the pipette at steep angles causes liquid to rise higher along the sidewalls for an equivalent volume, increasing the hazard of fluid touching the filter plug. Immerse the tip only to the depth specified for its volume class (1–2 mm for 0.1–10 µL; 2–3 mm for 20–200 µL; 3–6 mm for 1,000 µL). Immersing the tip too deeply introduces hydraulic drag and coats the exterior barrel with excess liquid that can drop into the destination vessel.

Plunger control is paramount in aerosol mitigation. Always release the pushbutton smoothly and steadily. Abruptly releasing or “snapping” the plunger causes fluid to rush violently into the tip, creating turbulent shear forces that atomize liquid droplets into fine aerosols and drive liquid splashes directly into the filter. When dispensing viscous liquids or detergent-containing buffers, pause for 1 to 2 seconds after the plunger reaches the first stop to allow the liquid column to reach physical equilibrium before executing the blowout stroke.

Finally, inspect tips during and after pipetting. If liquid inadvertently contacts an inert filter, immediately eject the tip into a biohazard waste receptacle and replace it. While inert polyethylene filters do not immediately disintegrate, wetting the filter matrix alters airflow resistance, introduces liquid retention errors, and risks leaching sample residue into subsequent aspirations. For self-sealing tips, liquid contact instantly locks the air pathway; attempting to force the plunger down will generate excessive internal pressure without dispensing the sample.

Common mistakes to avoid

Laboratory audits and quality control investigations consistently reveal several recurring operational errors involving filtered pipette tips:

  • Autoclaving pre-filtered tips contrary to manufacturer specifications: While virgin polypropylene tip bodies withstand standard autoclave cycles (121°C at 15 psi for 15–20 minutes), the integrated porous polyethylene filter plug has a lower thermal tolerance. Autoclaving pre-assembled filter tips can cause the sintered PE microbeads to deform, collapse pore pathways, and alter airflow resistance, which leads to systematic under-delivery of volumes. Furthermore, steam condensation trapped within the tortuous matrix can form liquid plugs that impede air exchange. Always purchase pre-sterilized, gamma- or e-beam-irradiated filter tips rather than autoclaving racked filter tips in-house unless the manufacturer explicitly certifies autoclavability.
  • Reusing single-use filter tips: Filter tips are engineered strictly as disposable single-use consumables. Washing, bleaching, or autoclaving used filter tips compromises the polymer matrix, strips low-retention surface coatings, and risks washing captured nucleic acid contaminants into the tip bore.
  • Assuming an aerosol filter compensates for incorrect pipetting modes: Filter tips trap aerosols, but they do not convert an air-displacement pipette into a positive-displacement device. Using air-displacement pipettes with filter tips for volatile organic solvents (such as ethanol or ether) or high-density viscous liquids (such as 100% glycerol) still results in volumetric errors due to vapor pressure changes and boundary layer friction.
  • Continuing to use a tip after filter wetting: If rapid plunger release or vessel tipping causes sample liquid to splash against an inert polyethylene filter, some operators continue dispensing the volume. This practice introduces substantial volumetric error, risks contaminating the next sample through capillary seepage, and can cause the pipette piston to aspirate fluid droplets. Eject and replace the tip immediately.
  • Using non-filtered tips for master mix preparation while using filter tips for templates: Setting up PCR reactions with non-filtered tips for common master mix reagents and restricting filter tips solely to template DNA is a frequent contamination pathway. Aerosols generated during master mix aliquoting can contaminate the pipette barrel, which subsequently contaminates template-free negative controls in downstream steps.

Frequently asked questions

Can you autoclave filtered pipette tips?

Most manufacturers recommend against autoclaving pre-filtered pipette tips. Although the outer polypropylene casing withstands 121°C, the porous polyethylene filter matrix can soften, warp, or experience partial pore collapse under pressurized steam heat. Autoclaving can also trap steam moisture within the dense labyrinth of the filter, causing airflow restriction and volumetric delivery errors. Laboratories requiring sterile filter tips should purchase factory-certified pre-sterilized tips (treated via electron-beam or gamma irradiation) rather than autoclaving them locally.

What is the difference between filter tips and barrier tips?

The terms “filter tips” and “barrier tips” are frequently used interchangeably in laboratory settings, but they refer to specific technical designs. A standard filter tip incorporates an inert porous polyethylene plug that captures aerosol droplets via a tortuous path without reacting chemically with the liquid. A specialized barrier tip—such as Thermo Scientific ART barrier tips—contains an active self-sealing polymer that physically swells and locks shut upon direct fluid contact, preventing any liquid or vapor from penetrating into the pipette shaft.

What pore size do pipette tip filters use?

Sintered porous polyethylene filters used in pipette tips typically feature nominal pore sizes ranging between 8 and 10 micrometers, with tortuous micro-channels spanning 5 to 20 micrometers. This calibrated pore distribution allows unhindered airflow during standard pipetting rates while forcing aerosol droplets (which typically measure 0.1 to 10 micrometers) to impact and adhere to the hydrophobic pore walls before reaching the pipette shaft.

Do filtered pipette tips affect pipetting volume accuracy?

When properly engineered and matched to compatible pipettors, filtered pipette tips do not diminish volumetric accuracy. However, because the filter creates a small amount of airflow resistance and reduces internal tip headspace, using poorly manufactured or mismatched third-party filter tips can alter pneumatic pressure differentials. Under ISO 8655-2, calibration tolerances apply to the combined system of the pipette and tip; laboratories must verify gravimetric calibration whenever switching filter tip brands.

Can sample liquid be recovered if it contacts the filter?

Sample recovery depends entirely on the filter composition. If an aqueous sample accidentally splashes against an inert, additive-free porous polyethylene filter, the fluid can often be recovered by carefully depressing the plunger to the blowout stop or through low-speed benchtop centrifugation, provided the sample has not been compromised. Conversely, if liquid contacts a self-sealing barrier tip containing cellulose hydrogels, the barrier hydrates and seals permanently, making sample recovery impossible.

Why are filter tips necessary for PCR workflows?

Filter tips are essential in PCR, qPCR, and RT-PCR workflows because air-displacement pipetting generates microscopic aerosols from high-speed meniscus movement and plunger release. These sub-visible droplets enter the pipette shaft and can dislodge into subsequent tubes. Because PCR amplifies target sequences by billions of times, even sub-femtogram aerosol carryover between tubes can generate false-positive results or ruin negative extraction controls.

Sources

  • International Organization for Standardization. (2022). Piston-operated volumetric apparatus — Part 2: Pipettes (ISO Standard No. 8655-2:2022). Geneva, Switzerland: ISO. Retrieved from https://www.iso.org/standard/74205.html (Accessed: 2026-11-16).
  • U.S. Food and Drug Administration. (2026). 21 CFR 862.2050: General purpose laboratory equipment labeled or promoted for a specific medical use (Product Code JRC: Micro Pipette / Pipette Tip). Silver Spring, MD: FDA Center for Devices and Radiological Health. Retrieved from https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/cfrsearch.cfm?fr=862.2050 (Accessed: 2026-11-16).
  • Centers for Disease Control and Prevention and National Institutes of Health. (2020). Biosafety in Microbiological and Biomedical Laboratories (6th ed.). Atlanta, GA: U.S. Department of Health and Human Services. Retrieved from https://www.cdc.gov/labs/BMBL.html (Accessed: 2026-11-16).
  • Eppendorf AG. (2024). ep Dualfilter T.I.P.S.®: Unrivaled Filter Performance and Purity Standards (White Paper No. 102 and Technical Specification Sheet). Hamburg, Germany: Eppendorf SE. Retrieved from https://www.eppendorf.com (Accessed: 2026-11-16).
  • Thermo Fisher Scientific. (2023). Thermo Scientific ART Barrier Pipette Tips: Aerosol and Liquid Contamination Prevention (Technical Specification and Selection Guide). Waltham, MA: Thermo Fisher Scientific. Retrieved from https://www.thermofisher.com (Accessed: 2026-11-16).
  • Gilson Inc. (2024). PIPETMAN® DIAMOND Tips User Guide: Filter Tip Specifications and Aerosol Protection. Middleton, WI: Gilson Inc. Retrieved from https://www.gilson.com (Accessed: 2026-11-16).
  • BRAND GMBH + CO KG. (2024). Liquid Handling Consumables: Pipette Tips, Filter Technology and Cleanroom Standards. Wertheim, Germany: BrandTech Scientific / BRAND GMBH. Retrieved from https://www.brand.de (Accessed: 2026-11-16).

Data: Commercial Filtered Pipette Tip Specifications and Barrier Properties

Comparative technical specifications for aerosol barrier and filtered pipette tips from major liquid handling manufacturers, compiled from published instructions for use, technical white papers, and product specification data sheets.

Commercial Filtered Pipette Tip Specifications and Barrier Properties
Product LineManufacturerBarrier TechnologyFilter MaterialFilter Pore Size (µm)Min Volume Range (µL)Max Volume Range (µL)Liquid Contact Reaction
ep Dualfilter T.I.P.S.Eppendorf AGDual-phase hydrophobic matrixPorous polyethylene (PE)100.110000Permeable to air; sample recoverable without chemical leaching
ART Barrier TipsThermo Fisher ScientificSelf-sealing polymer barrierCellulose-polyethylene composite100.11000Seals permanently upon aqueous contact, trapping sample completely
PIPETMAN DIAMOND FilterGilson Inc.Single-layer inert barrierHigh-density porous polyethylene (HDPE)80.11000Hydrophobic barrier; non-swelling; prevents aerosol bypass
BIO-CERT Filter TipsBRAND GMBHSintered porous barrierPure virgin polyethylene (PE)8.50.11000Additive-free hydrophobic matrix; preserves sample integrity
Safetyspace Filter TipsSartorius AGExpanded headspace barrierHigh-purity porous polyethylene90.11200Extra airspace prevents liquid touch; non-absorbent matrix

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about What Are Filtered Pipette Tips?

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →