Hospital Crib (Pediatric Medical Crib): FDA Class and Safety Requirements

A hospital crib is an FDA Class II pediatric medical crib (21 CFR 880.5140). See its definition, the eight special controls and the required warning label.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

General product and regulatory information only; this page is not medical advice and does not describe how to care for any patient.

Short answer

In U.S. regulation, a hospital crib is a “pediatric medical crib”: a prescription device for use with a pediatric patient, made of an open crib with fixed end rails and movable, latchable side rail components, and possibly a mattress. FDA classifies it as Class II under 21 CFR 880.5140 (product code FMS) and exempts it from 510(k) premarket notification.

Definition and scope

21 CFR 880.5140(a) identifies a pediatric medical crib as “a prescription device intended for medical purposes for use with a pediatric patient.” The same sentence lists its parts: an open crib, fixed end rails, movable and latchable side rail components, and possibly an accompanying mattress. The contour of the crib surface may be adjustable.

FDA’s product classification database lists the device under the name “Bed, Pediatric Open Hospital”, product code FMS, review panel Hospital (HO), medical specialty General Hospital, regulation number 880.5140, device class 2 (openFDA record, checked 2026-10-30).

The wording matters for supply decisions. The regulation covers a medical device with its own labeling duties, not a household nursery product. Adult and general-purpose beds are a separate topic; see what is a hospital bed.

How it is regulated

Under 880.5140(b), the crib is Class II (special controls). It is exempt from the premarket notification procedures in subpart E of 21 CFR part 807, subject to the limitations in 21 CFR 880.9. The openFDA classification record also shows that the device is not exempt from good manufacturing practice requirements (gmp_exempt_flag “N”) and that it is eligible for summary malfunction reporting.

The rule was published as part of a final rule at 81 FR 91737 (December 19, 2016), which is the citation printed at the end of the section in the eCFR.

Exempt from 510(k) does not mean unregulated. The special controls below still define what a conforming crib must demonstrate or carry. Manufacturers also remain subject to the wider device rules, including adverse-event reporting; for that topic see medical device reporting (MDR) and the background in Medical Device Amendments of 1976.

The eight special controls, as a receiving-inspection checklist

880.5140(b) lists eight special controls. The table restates each one with the measurable value where the regulation gives one. Only the regulation’s own text is used; the “where it is checked” column is an editorial reading of what kind of evidence the control points to, not an FDA requirement.

# Special control (880.5140(b)) Measurable value in the text Where it would be checked
1 Design and performance testing shows mechanical and structural stability under expected conditions of use, including latch and locking-mechanism security when engaged None stated Manufacturer test documentation
2 Materials appropriate for conditions of use, allow proper sanitation, free of surface defects that could cause injury None stated Manufacturer documentation; visual condition of the unit
3 Rail and end panel height, from the top of the rail or panel in its highest position to the top of the mattress support in its lowest position At least 26 inches (66 cm); any mattress not over 6 inches thick Measurement of the unit
4 Hardware and fasteners designed and built to eliminate mechanical hazards to the patient None stated Manufacturer documentation; visual condition of the unit
5 Distance between side rail components (slats, spindles, corner posts) Not greater than 2 3/8 inches (6 cm) at any point Measurement of the unit
6 Mattress fits tightly around all four sides of the crib base so entrapment or impingement is prevented None stated Fit of the supplied mattress in the supplied base
7 Mattress meets the CPSC flammability standards (FF 4-72, amended) and the open-flame standard for mattress sets, 16 CFR parts 1632 and 1633 Compliance with the named standards Manufacturer or mattress documentation
8 Labeling: instructions to care for, maintain and clean the crib, and a warning label on at least two sides of the crib Text at least 9 mm high Label on the unit; supplied instructions

The required warning text reads: “WARNING: Never leave a child unsupervised when the moveable side is open or not secured.”

  • Pediatric medical crib / “Bed, Pediatric Open Hospital”: the regulation title and the FDA product-code name for the same device (FMS).
  • Hospital bed: a broader category; see the hospital bed guide.
  • Other infant and neonatal devices: separate devices with their own classifications, for example the bili blanket. Infant care areas also raise security questions covered in infant abduction prevention programs.

Common misunderstandings

  • “It is just a baby crib.” The regulation defines it as a prescription device for medical purposes, with labeling and special controls specific to it.
  • “510(k) exempt means no FDA rules apply.” The Class II special controls apply, and the classification record shows the device is not exempt from good manufacturing practice.
  • “The side rail gap is a design preference.” 880.5140(b)(5) sets a ceiling of 2 3/8 inches (6 cm) between rail components at any point.
  • “Any mattress will do.” The mattress is addressed twice: thickness (6 inches maximum), tight fit on all four sides, and the CPSC flammability standards.

FAQ

What is a hospital crib?

FDA’s term is “pediatric medical crib”: an open crib with fixed end rails and movable, latchable side rails, possibly with a mattress, that is a prescription device for pediatric patients (21 CFR 880.5140(a)).

What FDA class is a hospital crib?

Class II with special controls, exempt from 510(k) premarket notification subject to 21 CFR 880.9 (product code FMS).

What is the minimum rail height?

At least 26 inches (66 cm), measured from the top of the rail or end panel in its highest position to the top of the mattress support in its lowest position. Any mattress must not exceed 6 inches in thickness.

What warning must a hospital crib carry?

A label on at least two sides, in text at least 9 mm high: “WARNING: Never leave a child unsupervised when the moveable side is open or not secured.”

Sources

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Hospital Crib (Pediatric Medical Crib): FDA Class and Safety Requirements

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →