ImageStream: Imaging Flow Cytometry Platform Guide

The ImageStream platform combines multi-channel microscopy with flow cytometry, capturing high-resolution images of cells in flow for quantitative analysis.

Written and maintained by CASRAI Editorial Board

Published

On this page

The ImageStream is a high-throughput multispectral imaging flow cytometry platform that captures high-resolution digital microscopic images of every cell moving in fluidic suspension at rates up to thousands of events per second. By synchronizing hydrodynamic core flow with charge-coupled device detectors operating in Time Delay Integration mode, the platform combines the statistical power and rare-event detection of conventional flow cytometry with the spatial and morphological resolution of fluorescence microscopy.

Definition and scope of the ImageStream platform

Imaging flow cytometry bridges two historically distinct analytical disciplines: flow cytometry and fluorescence microscopy. In conventional flow cytometry, cells hydrodynamically focused in a sheath stream traverse focused laser beams. Light scatter and fluorophore emissions strike photomultiplier tubes (PMTs) or avalanche photodiodes (APDs), generating one-dimensional electronic voltage pulses. While this architecture quantifies tens of thousands of cells per second across dozens of parameters, it discards spatial information: a diffuse cytoplasmic signal, a punctuated membrane receptor, an internalized nanoparticle, and a surface precipitate produce identical integrated intensity values.

Fluorescence microscopy provides diffraction-limited spatial resolution and subcellular detail. However, microscopy is constrained by lower throughput, manual or mechanical stage movement, slide-mounting artifacts, and small sample cohorts that introduce sampling bias when analyzing rare cell subpopulations. The ImageStream platform resolves this compromise by imaging cells directly within a hydrodynamic quartz flow cell, recording multispectral brightfield, darkfield, and fluorescence images for every passing event.

Originally developed by Amnis Corporation (founded by David Basiji and William Ortyn in 1999) and launched commercially with the ImageStream 100 in 2004, the platform progressed through the ImageStreamX to the current Amnis ImageStreamX Mk II. Following corporate stewardship under EMD Millipore, Luminex Corporation, and Cytek Biosciences (which acquired the Amnis line in 2023), the ImageStream represents the benchmark for multispectral imaging in flow across immunology, oncology, cell therapy manufacturing, and toxicology.

How ImageStream works: optics, fluidics, and TDI-CCD detection

The physical operation of the ImageStream system relies on three synchronized subsystems: hydrodynamic fluidics, an optical excitation train, and Time Delay Integration Charge-Coupled Device (TDI-CCD) cameras.

Laminar core stream and fluidic control

The ImageStream transports cell suspensions through a specialized quartz flow cell using hydrodynamic focusing. A high-precision syringe pump meters the sample into a pressurized sheath buffer stream. By adjusting the volumetric ratio between sheath fluid and sample core, the fluidics compress the cell suspension into a centered core stream measuring 5 μm to 10 μm in diameter. This precise focusing ensures that every cell traverses the identical optical focal plane along the longitudinal axis of the flow cell.

Time Delay Integration (TDI) imaging physics

At fluidic velocities between 100 mm/s and 1,000 mm/s, conventional camera exposure causes severe motion blur unless illuminated by brief stroboscopic flashes. Stroboscopic illumination, however, drastically reduces the photon budget available from faint fluorophores. The ImageStream overcomes this limitation through Time Delay Integration (TDI) detector technology.

A TDI-CCD sensor consists of a two-dimensional grid of photosensitive pixels aligned parallel to the fluid stream. As an illuminated cell travels through the laser interrogation zone, the optical train projects its moving image onto the CCD sensor. Instead of holding pixel charges stationary during exposure, the TDI controller continuously shifts accumulated photoelectrons row-by-row down the pixel columns at a clock frequency that exactly matches the physical linear velocity of the cell image. When photons strike consecutive pixels along the column, their generated electrons add constructively to the moving charge packet. Readout occurs at the final row into an analog-to-digital converter, achieving effective integration times of up to 10 milliseconds—providing roughly 1,000-fold greater sensitivity than conventional framing cameras without motion blur.

Multispectral decomposition and optical channels

Following cellular excitation by the laser array, emission light passes through a collection objective and enters a spectral decomposition optical block. Dichroic mirrors, spectral bandpass filters, and dispersion prisms split the composite emission into spatially separated wavelength bands focused onto distinct detector columns. In a dual-camera ImageStreamX Mk II system, the platform records up to 12 simultaneous image channels per cell:

  • Brightfield channels: Channel 1 (Camera 1, 435–480 nm) and Channel 9 (Camera 2, 570–595 nm), illuminated by an external brightfield LED to document cell morphology, volume, and membrane integrity.
  • Side scatter channel: Channel 6 (Camera 1, 745–780 nm), excited by a dedicated 785 nm laser to quantify internal cellular granularity and optical complexity.
  • Fluorescent channels: Up to 10 spectral channels distributed across Camera 1 (Channels 2, 3, 4, 5) and Camera 2 (Channels 7, 8, 10, 11, 12), spanning 420 nm to 800 nm.

Interchangeable objective lenses

The automated turret accommodates three interchangeable objectives, tailoring spatial resolution and throughput to experimental requirements:

  • 20x Objective (0.40 NA): Pixel area 1.0 μm², field of view 120 μm, depth of field 8 μm. Suitable for large targets, multicellular clusters, macroscopic parasites, and high-throughput screens at up to 2,000 cells/sec.
  • 40x Objective (0.75 NA): Pixel area 0.25 μm², field of view 60 μm, depth of field 4 μm. The primary configuration for mammalian immunophenotyping and nuclear translocation assays at roughly 1,000 cells/sec.
  • 60x Objective (0.90 NA): Pixel area 0.10 μm², field of view 40 μm, depth of field 2.5 μm. Delivers highest spatial detail for resolving punctate organelles, viral capsids, extracellular vesicles, and DNA damage foci at 300 to 400 cells/sec.

Software workflow: INSPIRE acquisition and IDEAS quantitative analysis

The operational software pipeline comprises two specialized suites: real-time instrument calibration and acquisition via INSPIRE, followed by offline image compensation, feature extraction, and gating in IDEAS.

Instrument calibration and acquisition in INSPIRE

The INSPIRE software governs fluidic delivery, laser power attenuation, camera clock timing, and objective focus. Before initiating an acquisition run, operators execute the Amnis System Self-Inspection Tool (ASSIST). ASSIST delivers a suspension of SpeedBeads calibration beads into the flow cell to calibrate core operational parameters:

  1. Optical focus: Adjusts objective positioning until bead contrast reaches peak sharpness.
  2. Camera alignment: Verifies spatial co-registration between Camera 1 and Camera 2 within sub-pixel limits.
  3. Core velocity matching: Calculates the exact transit velocity of beads and synchronizes the CCD clock frequency to eliminate image stretch or compression.
  4. Dark current calibration: Measures and subtracts baseline electronic noise across all 12 channels.
  5. Laser power response: Verifies photon emission linearity against calibrated fluorescent bead standards.

INSPIRE writes raw image data continuously to a proprietary Raw Image File (.rif). The .rif file encapsulates uncompensated, 16-bit multi-channel image frames for every acquired event alongside fluidic and optical metadata.

Spectral compensation and IDEAS post-processing

Because fluorophores exhibit overlapping emission spectra, spectral spillover must be corrected. Unlike conventional flow cytometry where compensation operates on integrated scalar pulses, ImageStream compensation requires solving a spatial pixel-by-pixel coefficient matrix.

The IDEAS (Image Data Exploration and Analysis Software) package processes single-stained control files to generate a pixel-level compensation matrix. Applying this matrix to the .rif file generates a Compensated Image File (.cif) containing calibrated, crosstalk-free images. Analysts then build a Data Analysis File (.daf) containing user-defined digital masks, computed morphometric features, and hierarchical bivariate gating templates.

Masking concepts and morphometric feature extraction

A central feature of IDEAS is digital masking. A mask is a binary pixel overlay delineating the image region to measure. IDEAS provides system masks (defining cellular boundaries), morphology masks (encompassing signal above background noise), object masks (isolating punctate spots such as endosomes), and valley masks (separating contact interfaces in doublets). Boolean operators allow combinations, such as subtracting a nuclear mask from a whole-cell mask to isolate cytoplasmic signal.

Using these masks, IDEAS calculates over one hundred quantitative features per cell, including intensity, diameter, circularity, aspect ratio, gradient root-mean-square (RMS) sharpness, and specialized spatial metrics such as Similarity Scores and Internalization Scores. While IDEAS is the native analytical environment, researchers operating multi-instrument cytometry facilities often export numerical parameter files into platforms like FCS Express for integrated laboratory reporting.

Positioning the ImageStream within laboratory instrumentation requires comparing its capabilities with sibling platforms, conventional cytometers, and high-content imagers.

The primary sibling platform within the Amnis product line is the Amnis FlowSight. While the ImageStreamX Mk II specializes in subcellular resolution, the FlowSight is engineered for high-throughput benchtop immunophenotyping. The FlowSight features a fixed 20x objective lens (0.40 NA), a wider 1 mm flow cell, and a pixel size of 1.0 μm², supporting acquisition rates up to 4,000 cells per second. It excels at routine doublet exclusion, cell viability, and general co-localization, but lacks the 60x magnification required for organelle-level resolution.

Compared to conventional non-imaging flow cytometers—such as the full-spectrum Cytek Aurora—the ImageStream trades raw volumetric throughput for spatial data. A conventional spectral cytometer processes 20,000 to 50,000 cells per second across 40 to 64 fluorescent detectors, making it ideal for deep immune profiling. However, as detailed in our comparative analysis of imaging flow cytometry vs conventional flow cytometry, no conventional cytometer can determine whether a fluorescent signal resides inside the cytoplasm or adheres to the outer plasma membrane.

Automated spinning disk and laser scanning confocal microscopes—evaluated in our review of confocal vs fluorescence microscope architectures—excel at multi-plane 3D optical sectioning and live-cell adherence tracking. Yet, confocal microscopes struggle with non-adherent suspension cells and require hours to acquire a few hundred cells, limiting statistical power. The data table below compares the technical specifications, optical limits, and acquisition throughput of these platforms.

Where ImageStream matters for laboratories, core facilities, and protocol workflows

The ImageStream platform converts microscopic visual phenotypes into quantifiable, statistically powered gating parameters across multiple experimental workflows:

  • Transcription factor nuclear translocation: Quantifying the redistribution of signaling molecules (e.g., NF-κB p65, STAT3, IRF3) from the cytoplasm into the cell nucleus. IDEAS computes a pixel-by-pixel cross-correlation (Similarity Score) between the transcription factor fluorophore and a nuclear counterstain, generating an objective metric across thousands of cells.
  • Receptor internalization and antibody-drug conjugate (ADC) kinetics: Measuring endocytosis of surface receptors or therapeutic antibodies. By defining a cellular membrane mask and measuring the fraction of fluorescent intensity located within an eroded internal mask, researchers calculate an automated Internalization Score to track drug delivery kinetics.
  • Phagocytosis and bacterial uptake: Distinguishing between true intracellular engulfment of pathogens by macrophages or neutrophils and extracellular adherence to the cell surface—a classical flaw of non-imaging flow cytometry.
  • Immunological synapse and cell-cell interaction: Visualizing conjugate formation between cytotoxic T lymphocytes (CTLs) or natural killer (NK) cells and target antigen-presenting cells (APCs), quantifying actin accumulation and polarized perforin granules at the contact interface.
  • Extracellular vesicles (EVs) and exosome analysis: Detecting and sizing small membrane vesicles down to 100 nm. Spatial imaging enables exclusion of non-vesicular antibody aggregates and sheath contaminants that confound conventional scatter gates.
  • DNA damage foci and cell cycle kinetics: Counting individual intranuclear repair foci (e.g., γ-H2AX, 53BP1) across distinct phases of the cell cycle, correlating spot counts directly with nuclear DNA content dyes.

Core facility management, procurement, and maintenance considerations

Operating an ImageStream involves distinct logistical requirements compared to standard cytometers:

First, sample preparation protocols require higher cellular concentrations. Because sample core volume is small (tens of nanoliters per second), samples should be prepared at 1 × 10⁷ to 2 × 10⁷ cells/mL in a minimal working volume of 20 μL to 50 μL. Dilute suspensions result in prolonged acquisition times and excessive sheath fluid consumption.

Second, sample filtering is mandatory. The quartz flow cell features a narrow 200 μm interrogation chamber. Unfiltered cell aggregates, genomic DNA strands from lysed cells, or particulate debris can cause complete fluidic clogs. All samples must pass through a 35 μm or 50 μm nylon cell strainer immediately prior to loading.

Third, data storage architecture must accommodate high data density. A typical 12-channel ImageStream acquisition recording 50,000 cells generates approximately 1.5 to 3.0 gigabytes of uncompressed .rif data. When compensated and processed into .cif and .daf files, a single experimental batch can exceed 50 gigabytes, requiring robust network-attached storage (NAS) and systematic archival pipelines.

Worked analytical example: nuclear translocation and internalization scoring

To illustrate the operational mechanics of the ImageStream pipeline, consider an illustrative composite experiment measuring NF-κB p65 nuclear translocation in RAW 264.7 murine macrophages following lipopolysaccharide (LPS) challenge.

Step 1: Staining and sample preparation

Macrophages are cultured, stimulated with 1 μg/mL LPS for 60 minutes (with untreated cells as resting controls), harvested, and fixed with 4% paraformaldehyde. Cells are permeabilized using 0.1% Triton X-100 and stained with an Alexa Fluor 488-conjugated anti-NF-κB p65 monoclonal antibody. To identify the nuclear boundary, cells are counterstained with the cell-permeant bisbenzimide dye Hoechst 33342 (or DRAQ5), which binds to the minor groove of double-stranded DNA. Samples are resuspended at 1.5 × 10⁷ cells/mL in PBS containing 1% bovine serum albumin and filtered through a 35 μm mesh.

Step 2: Acquisition setup in INSPIRE

The operator runs ASSIST bead calibration to verify optical alignment and fluidic velocity matching. The 40x objective lens is engaged. Laser powers are adjusted: 488 nm laser at 100 mW for NF-κB (Channel 2, 505–560 nm), 405 nm laser at 50 mW for Hoechst 33342 (Channel 1, 435–480 nm), and 785 nm side scatter laser at 10 mW (Channel 6). Brightfield illumination is set on Channel 9. A rough acquisition gate is set on Channel 9 brightfield area and aspect ratio to exclude SpeedBead calibration events. Acquisition proceeds until 10,000 raw single-cell events are captured in the .rif file.

Step 3: Gating hierarchy in IDEAS

The raw file is imported into IDEAS along with single-color compensation controls (Alexa Fluor 488 alone, Hoechst alone). Once the spectral compensation matrix is validated and compiled into a .cif file, the operator builds the hierarchical gating template:

  1. Focus gating (Gradient RMS): On Channel 9 (brightfield), a histogram of Gradient RMS sharpness is plotted. Events with low Gradient RMS values represent out-of-focus cells traveling near the periphery of the sheath stream and are excluded. The in-focus gate typically retains events with Gradient RMS values above 45 to 50.
  2. Single-cell gating (Area vs Aspect Ratio): A bivariate dot plot of brightfield Area versus Aspect Ratio is generated. Single spherical macrophages exhibit an aspect ratio near 0.8 to 1.0 with uniform area. Doublets, cell-cell conjugates, and large debris exhibit low aspect ratios and high areas and are gated out.
  3. Positive marker gating: A bivariate plot of NF-κB intensity (Channel 2) versus Hoechst intensity (Channel 1) ensures that analysis is restricted to intact, nucleated, antibody-stained cells.

Step 4: Translocation scoring via the Similarity feature

To quantify translocation objectively, IDEAS calculates the Similarity Feature between Channel 1 (Hoechst nuclear image) and Channel 2 (NF-κB image) over the morphology mask of the nucleus. The Similarity feature measures the degree of spatial pixel co-localization using a log-transformed Pearson correlation coefficient:

When NF-κB is sequestered in the cytoplasm of resting cells, the green fluorescence is distributed around the exterior of the nucleus, producing negative correlation with the nuclear dye (Similarity Score 1.0 to 2.5).

By plotting a histogram of the Similarity Feature, the researcher sets a definitive boundary threshold between untranslocated and translocated populations, calculating the exact percentage of responding macrophages with statistical power derived from thousands of single-cell images.

Common misunderstandings and operational pitfalls

Deploying imaging flow cytometry successfully requires avoiding several widespread operational misconceptions:

  • Assuming the ImageStream can physically sort cells: A frequent misunderstanding among new users is conflating imaging flow cytometry with fluorescence-activated cell sorting (FACS). The ImageStreamX Mk II is an analytical instrument; passing cells are carried directly into the fluidic waste tank after imaging. Standard commercial ImageStream instruments cannot physically isolate sorted subpopulations into collection tubes.
  • Neglecting optical depth of field: Because cells flow through the flow cell in random rotational orientations, fluorophore structures located outside the optical depth of field (2.5 μm at 60x magnification; 4 μm at 40x magnification) appear blurred. IDEAS cannot mathematically deconvolve out-of-focus photons back into crisp coordinates. Strict focus gating using brightfield Gradient RMS is mandatory; failure to gate on focus corrupts morphometric features such as circularity, aspect ratio, and internal spot counts.
  • Using conventional flow cytometry fluorophore concentrations: In standard flow cytometry, samples are stained in dilute suspensions and processed rapidly. In the ImageStream, because cells traverse the laser beam at high velocity and the pixel dwell time on the TDI-CCD is measured in microseconds, fluorophore emission must be exceptionally bright. Antibodies must be carefully titrated, and low-abundance antigens often require high-quantum-yield fluorophores rather than dim conventional dyes.
  • Relying on area and height for doublet discrimination: In conventional cytometry, doublets are excluded by plotting pulse area versus pulse height or pulse width. In an ImageStream, two cells traveling closely in tandem along the vertical axis of the flow cell can produce pulse profiles that mimic large single cells. Examining the actual brightfield image readily unmasks these tandem doublets, allowing precise exclusion via aspect ratio and valley masks.
  • Ignoring brightfield spectral bleed-through: The brightfield LED lamp emits broad-spectrum light that can spill into fluorescent detection channels if the brightfield intensity is set excessively high. In INSPIRE, brightfield intensity must be balanced so that pixel values remain within the linear detection range (typically between 500 and 1,500 counts) without saturating the CCD, which would introduce optical flare and distort fluorescent compensation.

Frequently asked questions

How many cells are required for a standard ImageStream acquisition?

A typical ImageStream protocol requires a minimum of 100,000 to 500,000 cells suspended in 20 μL to 50 μL of buffer, yielding a target concentration of 1 × 10⁷ to 2 × 10⁷ cells/mL. This concentration allows the instrument’s syringe pump to load a concentrated core stream and acquire 10,000 to 50,000 single, in-focus cellular images in approximately three to five minutes without excessive sheath fluid waste.

Can the ImageStream platform sort cells physically after imaging?

No. Standard commercial Amnis ImageStream instruments, including the ImageStreamX Mk II, are dedicated analytical cytometers. The hydrodynamic stream carries interrogated cells directly into the system waste reservoir after optical recording. Researchers requiring physical sorting must use standard droplet-deflection cell sorters or specialized microfluidic sorting platforms.

What is the difference between INSPIRE and IDEAS software?

INSPIRE is the real-time instrument acquisition software that controls hardware subsystems, syringe fluidics, laser attenuation, and ASSIST calibration beads to generate raw, uncompensated data files (.rif). IDEAS is the post-acquisition data exploration and analysis suite used offline to apply single-stained spectral compensation matrices (creating .cif files), build digital segmentation masks, extract quantitative morphometric features, and perform hierarchical statistical gating (saved in .daf files).

How does Time Delay Integration prevent motion blur in flowing cells?

Time Delay Integration (TDI) utilizes a specialized CCD detector that continuously shifts photo-generated electronic charge packets row-by-row along the pixel columns at a clock speed synchronized with the physical velocity of the flowing cells. As a cell travels through the laser beam, its emitted photons continuously strike corresponding rows of the moving charge packet. This provides long effective integration times (up to 10 milliseconds) and high signal-to-noise ratios without motion blur.

Sources

  • Cytek Biosciences. (2024). Amnis ImageStreamX Mk II Imaging Flow Cytometer Specifications and System Overview. Cytek Biosciences, Fremont, CA. https://cytekbio.com/pages/amnis-imagestreamx-mk-ii (Accessed 2026-11-14).
  • Amnis Corporation / Cytek Biosciences. (2021). INSPIRE Software and ImageStreamX Mk II System User’s Guide (Part 09-90001). Cytek Biosciences, Fremont, CA. https://cytekbio.com/ (Accessed 2026-11-14).
  • Amnis Corporation / Cytek Biosciences. (2022). IDEAS Software User’s Guide: Image Data Exploration and Analysis Software v6.3. Cytek Biosciences, Fremont, CA. https://cytekbio.com/ (Accessed 2026-11-14).
  • Rees, P., Summers, H. D., Filby, A., Carpenter, A. E., and Doan, M. (2023). Imaging flow cytometry: a primer. Nature Reviews Methods Primers, 3(1), Article 60. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10468826/ (Accessed 2026-11-14).
  • Barteneva, N. S., Fasler-Santenau, E., and Vorobjev, I. A. (2012). Imaging flow cytometry: coping with heterogeneity in biological systems. Journal of Histochemistry and Cytochemistry, 60(10), 723–733. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3489803/ (Accessed 2026-11-14).
  • Zuba-Surma, E. K., Kucia, M., Abdel-Latif, A., and Ratajczak, M. Z. (2007). The ImageStream System: a key step to a new era in imaging. Folia Histochemica et Cytobiologica, 45(4), 279–290. https://pubmed.ncbi.nlm.nih.gov/18165166/ (Accessed 2026-11-14).
  • University of Warwick Flow Cytometry Facility. (2023). Amnis ImageStreamX Mk II Technical Protocol and Operating Guidelines. Warwick Medical School, Coventry, UK. https://warwick.ac.uk/fac/sci/lifesci/research/facilities/flowcytometry/instrumentation/imagestream/ (Accessed 2026-11-14).
  • Columbia University Flow Cytometry Shared Resource. (2024). Cytek Amnis ImageStreamX Mk II User Guide and Optical Configurations. Columbia University Irving Medical Center, New York, NY. https://www.cancer.columbia.edu/research/shared-resources/flow-cytometry (Accessed 2026-11-14).

Data: Imaging Flow Cytometry vs Conventional Cytometry Specifications

Original compilation comparing optical parameters, magnification, numerical aperture, pixel dimensions, channel capacities, and acquisition throughput across Amnis ImageStreamX Mk II configurations, Amnis FlowSight, and conventional spectral flow cytometry systems.

Period covered: 2024/2026. Coverage: Global.

Imaging Flow Cytometry vs Conventional Cytometry Specifications
Platform and Optical ModeMagnification Power (x)Numerical Aperture (NA)Pixel Resolution (μm²)Maximum Detection Channels (channels)Maximum Acquisition Rate (cells/sec)Primary Detector TechnologyPrimary Image Output
ImageStreamX Mk II (60x Objective)600.90.112400TDI-CCD ArraySubcellular organelles, punctate foci, and bacteria (.rif/.cif)
ImageStreamX Mk II (40x Objective)400.750.25121000TDI-CCD ArrayNuclear translocation and single-cell morphology (.rif/.cif)
ImageStreamX Mk II (20x Objective)200.41122000TDI-CCD ArrayLarge mammalian cells, parasites, and doublets (.rif/.cif)
Amnis FlowSight (20x Fixed)200.41124000TDI-CCD ArrayBenchtop immunophenotyping and basic morphology (.rif/.cif)
Conventional Spectral Cytometry0006430000Avalanche Photodiode (APD)Integrated pulse intensities without spatial imagery (.fcs)

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about ImageStream: Imaging Flow Cytometry Platform Guide

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →