Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
General product and standards information for laboratory and clinic staff; it is not medical advice and does not replace the manufacturer’s instructions for use or your facility’s sterile-processing policy.
Short answer
An integrating indicator is a Type 5 chemical indicator under ISO 11140-1. It is designed to respond to all the critical variables of a steam cycle (time, temperature and steam/moisture) and to reach its endpoint at stated values set to be equivalent to, or exceed, the calculated performance requirements for biological indicators. A pass shows the pack saw those conditions; it does not prove sterility, and the CDC says chemical indicators should be used together with biological indicators.
Definition and scope
ISO 11140-1:2014 specifies labeling requirements, performance requirements and test methods for chemical indicators used to monitor attainment of one or more process parameters of a sterilization process. It sorts them into types. STERIS notes that the types are not ranked by level of importance and that an indicator should comply with only one type. The full six-type breakdown is in chemical indicators for sterilization; this page covers only Type 5.
What sets Type 5 apart, as described in a 2025 case study published by a sterilization-products manufacturer (it quotes ISO 11140-1:2014), is that it is the only type designed to react to all critical process variables and to have stated values generated to be equivalent to, or exceed, calculated biological-indicator performance. For steam, the critical variables are time, temperature and moisture.
How the stated value works
The standard defines a stated value (SV) as the value or values of a critical process variable at which the indicator is designed to reach its endpoint, as defined by the manufacturer. The endpoint is the point of observed change the manufacturer defines. In moving-front indicators this appears as progressive movement toward a pass or accept region.
According to that manufacturer-authored study, the mechanics for steam Type 5 indicators are:
- SVs must be established at a minimum of three temperatures across the steam range: 121 °C (250 °F), 135 °C (275 °F) and at least one equally spaced point between them.
- Each SV is tested at two test points. At Test Point 1 (the SV conditions) the indicator must reach its endpoint. At Test Point 2, set 1 °C lower with exposure 15% shorter, it must not.
- The SV must be greater than 16.5 minutes at 121 °C and greater than 1.2 minutes at 135 °C. The 16.5-minute floor was derived from calculations based on biological-indicator resistance specifications.
- Testing is done in a steam resistometer compliant with ISO 18472.
How it is used in practice
Type 3, 4, 5 and 6 indicators can all serve as internal indicators, checking that steam reached the inside of containers, wrapped packs or peel pouches. The CDC states that a chemical indicator preferably should also be placed inside each pack to verify sterilant penetration, and that if the internal or external indicator suggests inadequate processing, the item should not be used. Packaging choices are covered in sterilization pouches.
Integrating indicator versus biological indicator
| Question | Type 5 integrating indicator | Biological indicator |
|---|---|---|
| What it reads | Chemical response to time, temperature and steam at stated values | Survival of test organisms after the cycle (see autoclave validation) |
| Link to the other | SVs set to be equivalent to or exceed calculated BI performance | Resistance specifications underlie the 16.5-minute floor at 121 °C |
| CDC position | Chemical indicators do not prove sterilization was achieved and should be used in conjunction with biological indicators. | |
Where it matters for labs and clinics: a label-reading checklist
Each item below follows from the sources above; it is a way to read the box and instructions for use, not a procedure.
- Does the labeling state the indicator complies with ISO 11140-1 as Type 5 (and only one type)?
- At which temperatures are stated values given? A steam Type 5 should state them at 121 °C, 135 °C and a point between.
- How does the manufacturer define the endpoint for this format (for example, a moving front reaching an accept region)?
- Is it intended for internal use inside the pack type you process?
- Does your cycle fall within the temperature range the stated values cover?
The label alone is not proof of performance. In that manufacturer-run study, which tested at 121 °C only, lots of the sponsor’s two products exceeded the 16.5-minute minimum, while three lots of another maker’s integrator did not meet the ISO 11140-1 requirement at that temperature. It is one result from an interested party, useful mainly to show that conformance to stated values is a testable claim, not a ranking of products. The related quantitative concepts are explained in sterility assurance level and log reduction.
Common misunderstandings
- “Type 5 replaces the biological indicator.” The CDC says chemical indicators are used in conjunction with biological indicators, and stated values are set equivalent to or above calculated BI performance rather than measuring spores.
- “Higher type means better.” STERIS states the types are not ranked by level of importance.
- “A pass means the load is sterile.” The CDC says indicators show exposure to the process and do not prove sterilization was achieved.
- “Any indicator labeled Type 5 behaves the same.” Behavior depends on each product’s manufacturer-defined stated values and endpoint.
FAQ
What is an integrating indicator?
A Type 5 chemical indicator under ISO 11140-1 that responds to all critical variables of the process, with stated values equivalent to or exceeding calculated biological-indicator performance.
Is an integrating indicator the same as a biological indicator?
No. It is chemical, reads a physical-chemical response rather than organism survival, and the CDC says chemical indicators should be used in conjunction with biological indicators.
Where is an integrating indicator placed?
Type 5 indicators can be internal indicators. The CDC says a chemical indicator preferably should be placed inside each pack to verify sterilant penetration; follow the manufacturer’s instructions for use for placement.
Sources
- CDC, Sterilizing Practices (guideline for disinfection and sterilization in healthcare facilities): https://www.cdc.gov/infection-control/hcp/disinfection-sterilization/sterilizing-practices.html. Checked 2026-10-30.
- STERIS, 6 ISO Types of Chemical Indicators for Steam Sterilization: https://amer.sterislifesciences.com/Resources/Technical-Learning-Library/Technical-Learning-Library-Folder/Articles/6-ISO-Types-of-Chemical-Indicators-for-Steam-Sterilization. Checked 2026-10-30.
- Solventum (3M) case study comparing Type 5 integrators against ISO 11140-1:2014 stated values at 121 °C (manufacturer-authored; quotes the standard): https://assets.solventum.com/is/content/mmmspinco/Terragene_Comparison_of_Type_5_Integrators_cg3_LowRespdf. Checked 2026-10-30.
- ISO 11140-1:2014, Sterilization of health care products, Chemical indicators, Part 1: General requirements, International Organization for Standardization (the standard itself is sold by ISO and national standards bodies).