General information about a standard for readers in labs, clinics and supply teams. It is not medical advice and not a validation procedure.
Short answer
ISO 11137-1 is the part of the ISO 11137 series that sets requirements for developing, validating and routinely controlling a radiation sterilization process for medical devices. The current version is Edition 2, published in April 2025 as ISO 11137-1:2025, which replaced the 2006 edition and its two amendments. It governs the process, not any single dose or product.
Definition and scope
The full title is “Sterilization of health care products – Radiation – Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices”. It is maintained by ISO/TC 198 and applies to medical devices that need radiation sterilization; other products fall under other standards (ISO 11137-1:2025 record).
Radiation here covers the sources described on our guides to gamma irradiation sterilization and e-beam sterilization; the series also covers X-ray, as summarised on the sterility assurance level (SAL) guide.
How it is used in practice
The standard is used by the manufacturer or sterilization provider as the framework for showing that a radiation process reliably delivers a defined sterility assurance level for a given product. Our gamma and e-beam guides describe the provider-side picture: the dose needed is set per product during validation, and dose delivery depends on product density and packaging size. The e-beam guide notes that one provider lists three validation components: microbiological, dosimetric and application-specific.
In the United States, FDA states that “radiation is an established Category A sterilization method” for devices reviewed through 510(k), and that it encourages manufacturers to use FDA-recognized voluntary consensus standards in submissions because conformity “streamlines regulatory review” (FDA, Sterilization for Medical Devices). FDA also says a change between Category A methods, such as the radiation source, generally does not require a new 510(k) if performance and biocompatibility are not significantly affected.
What changed in the 2025 edition
The ISO record for Edition 2 lists these updates: new normative references (ISO 13004 and ISO/ASTM 52628) to align with current dosimetry practice and terminology; a revised clause structure to align with other ISO/TC 198 documents; increased allowable limits for electron- and X-ray-induced radioactivity; new requirements to prevent control failures and protect the integrity of process-parameter recording; and simplified content on transferring verification or sterilization doses.
| Document | Status on the ISO record |
|---|---|
| ISO 11137-1:2006 | Withdrawn |
| ISO 11137-1:2006/Amd 1:2013 | Withdrawn |
| ISO 11137-1:2006/Amd 2:2018 | Withdrawn |
| ISO 11137-1:2025 (Edition 2) | Published, replaces the three above |
Related parts and concepts
ISO 11137 is a series. Part 1 sets the requirements; the ISO record also lists Part 2 (establishing the sterilization dose), Part 3 (guidance on dose measurement) and the technical specification ISO/TS 11137-4 (process control), and the SAL guide summarises the series. The overall validation logic across steam, EtO and radiation is covered in sterilization validation, and the meaning of a logarithmic microbial reduction in log reduction.
Where it matters for labs, clinics and supply decisions
A sterile single-use product labelled as radiation sterilized relies on this standard family behind the scenes. Three questions follow:
- Which edition does the supplier’s validation cite? The 2006 edition and both amendments are withdrawn, so documentation may refer to either the older or the 2025 text.
- Which radiation source was used? FDA treats a change of radiation source as a change between Category A methods that generally does not need a new 510(k) when performance and biocompatibility are unaffected, so the source is a legitimate question to ask rather than an assumption.
- Was the process validated for that product and packaging? Dose absorption depends on density and packaging size, which is why validation is product-specific.
For how sterility symbols on packaging relate to the method, see medical device labeling symbols.
Common misunderstandings
- It is not a dose. The standard sets requirements for establishing and controlling the process; the dose is set per product through validation.
- It is not a product certificate. It describes how a sterilization process is developed, validated and controlled, not whether a given product is approved.
- Using the standard is not the same as FDA recognition of a specific edition. FDA encourages use of recognized consensus standards; which edition it has recognized is a separate, checkable record.
FAQ
What is ISO 11137-1?
It is Part 1 of the ISO 11137 series on radiation sterilization of health care products, setting requirements for development, validation and routine control of the process for medical devices.
Which edition is current?
ISO 11137-1:2025, Edition 2, published April 2025, which replaced ISO 11137-1:2006 and its Amd 1:2013 and Amd 2:2018.
Does it cover gamma and e-beam?
The ISO 11137 series covers radiation sterilization including gamma, e-beam and X-ray, as summarised on our SAL guide.
Sources
Checked 1 November 2026.
- ISO, ISO 11137-1:2025 (committee mirror of iso.org/standard/81721.html): https://committee.iso.org/standard/81721.html