General standards information for laboratory, clinic and supply staff. It is not medical advice and not a substitute for the standard itself or for a manufacturer’s instructions for use.
Short answer
ISO 11137-2 is the part of the ISO 11137 series on radiation sterilization of health care products that deals with the dose. Per ISO’s catalogue abstract, it specifies methods for determining the minimum dose needed to meet a stated sterility requirement, methods to substantiate 25 kGy or 15 kGy as the sterilization dose to reach a sterility assurance level (SAL) of 10⁻⁶, and methods of sterilization dose audit. Edition 3 (2013) is current as of 1 November 2026.
Definition and scope
The full title is Sterilization of health care products — Radiation — Part 2: Establishing the sterilization dose. The scope in the ISO abstract has three parts:
- Determining the minimum dose needed to achieve a specified requirement for sterility.
- Substantiating the use of 25 kGy or 15 kGy as the sterilization dose to achieve an SAL of 10⁻⁶.
- Auditing the sterilization dose, to demonstrate that it remains effective over time.
The standard is about the number on the process specification and the evidence behind it. The requirements for developing, validating and routinely controlling the whole radiation process sit in ISO 11137-1. The ISO abstract for Part 2 describes only the dose-related methods listed above.
How it is used in practice
Radiation sterilization is delivered by sources such as gamma irradiation or electron-beam irradiation. Whatever the source, a sterilization dose is only meaningful if someone has shown why that dose delivers the required sterility assurance level. ISO 11137-2 is the standard that frames that demonstration. It offers two routes, taken straight from the abstract: establish a minimum dose for the product, or substantiate a fixed dose of 25 kGy or 15 kGy. After that, the dose audit method is used to show the chosen dose is still effective.
The normative method text is sold by ISO. This page describes what the standard covers, its status, and the documentation questions that follow from its scope, not the technical procedure for running a dose-setting study. Wider process proof is covered in sterilization validation.
Edition, amendment and FDA recognition status
| Item | Status | Source |
|---|---|---|
| Edition | Edition 3, published 2013-06 (68 pages, ISO/TC 198) | ISO catalogue page |
| Amendment | ISO 11137-2:2013/Amd 1:2022 (one amendment) | ISO catalogue page |
| Review | Last reviewed and confirmed in 2024; ISO states this version remains current | ISO catalogue page |
| Revision | A replacement, ISO/DIS 11137-2, is in development as a Draft International Standard; ISO says it is expected to replace the 2013 edition “within the coming months” | ISO catalogue page |
| FDA recognition | Recognized as “Third edition 2013-06 [Including AMD1:2022]”; recognition number 14-580; recognition list 059; date of entry 12/19/2022; extent of recognition: complete standard | FDA Recognized Consensus Standards database |
Both records were checked on 1 November 2026. Because a revision is under development, the edition a document cites matters more than usual, and “ISO 11137-2” with no year is ambiguous.
How Part 2 fits with the rest of the series
| Question | Where it is answered |
|---|---|
| What must a radiation sterilization process for health care products include, from development to routine control? | Part 1; see ISO 11137-1 |
| How was the minimum dose determined, or how is 25 kGy / 15 kGy substantiated for this product? | Part 2 (this page) |
| Does the dose still work, now and over time? | Part 2, dose audit |
What to ask a sterilization or supply source to document
These are documentation questions, not purchasing criteria. They follow from the scope and status facts above; the standard does not oblige any buyer to ask them.
- Which edition and amendment is cited? The 2013 edition, with or without Amd 1:2022, or a later revision. FDA recognizes the 2013 edition including Amd 1:2022.
- How was the sterilization dose set? By determining a minimum dose for the product, or by substantiating 25 kGy or 15 kGy. These are different claims.
- What SAL does the dose claim? The substantiation methods in the abstract target an SAL of 10⁻⁶.
- Is there dose audit evidence? The standard includes dose audit methods to show continued effectiveness, so a record that audits are performed, and when, is a reasonable thing to request.
Common misunderstandings
“25 kGy is the required dose.”
Not according to the ISO abstract. 25 kGy and 15 kGy are doses whose use can be substantiated; the standard also covers determining a minimum dose for a stated sterility requirement.
“Part 2 covers the whole sterilization process.”
The abstract limits Part 2 to dose determination, substantiation and audit. Process development, validation and routine control belong to Part 1.
“A passing dose audit means a high dose.”
A dose audit is described as a method to demonstrate the continued effectiveness of the sterilization dose. It is evidence about effectiveness, not a measure of how large the dose is. For the logic of microbial kill numbers, see log reduction.
“The standard is outdated because a revision is coming.”
ISO states that the 2013 edition was reviewed and confirmed in 2024 and remains current, and FDA lists it as recognized in full. A draft revision exists but is not the current edition.
FAQ
What is ISO 11137-2?
It is Part 2 of the ISO 11137 series, titled “Establishing the sterilization dose”. It covers determining the minimum dose, substantiating 25 kGy or 15 kGy for an SAL of 10⁻⁶, and dose audit.
Which edition is current?
Edition 3 (2013-06), with Amendment 1 (2022). ISO confirmed it in 2024, and a Draft International Standard, ISO/DIS 11137-2, is in development.
Does the FDA recognize ISO 11137-2?
Yes. The FDA database lists the third edition 2013-06 including Amd 1:2022 as recognized in full under recognition number 14-580, entered 12/19/2022.
Does ISO 11137-2 apply to ethylene oxide sterilization?
The ISO 11137 series title refers to radiation sterilization of health care products, so it is not the standard for other sterilants.