A medical device recall letter is the notice a recalling firm sends to its affected direct accounts. Under FDA’s recall policy in 21 CFR 7.49 it must tell the recipient that the product is subject to a recall, that further distribution or use of any remaining product should cease immediately, and what to do with the product. FDA also publishes a model recall notification letter for devices; the template below follows its order and ties each part to the regulation text.
Who uses it: regulatory and quality staff at manufacturers and distributors who draft recall letters, and clinic, hospital and laboratory staff who receive them and must act on them. For the wider picture of recalls, corrections and removals, see medical device recall and field safety corrective action (FSCA), which covers the EU field safety notice.
What a recall letter must say
21 CFR 7.49(a) says the format, content and extent of a recall communication should be commensurate with the hazard of the product and the recall strategy. The table maps each element to its source.
| Element | What the source says | Source |
|---|---|---|
| Recall statement | Convey that the product is subject to a recall | 21 CFR 7.49(a)(1) |
| Stop use | Further distribution or use of any remaining product should cease immediately | 21 CFR 7.49(a)(2) |
| Downstream notice | Where appropriate, the direct account should in turn notify its customers who received the product | 21 CFR 7.49(a)(3) |
| Instructions | Provide specific instructions on what should be done with the recalled products | 21 CFR 7.49(a)(4), (c)(1)(iv) |
| Product identification | Identify the product, size, lot numbers, codes or serial numbers and other descriptive information for accurate and immediate identification | 21 CFR 7.49(c)(1)(ii) |
| Reason and hazard | Explain concisely the reason for the recall and the hazard involved, if any | 21 CFR 7.49(c)(1)(iii) |
| Brevity | Be brief and to the point | 21 CFR 7.49(c)(1)(i) |
| Reply route | Provide a ready means for the recipient to report whether it has any of the product, for example a postage-paid, self-addressed postcard | 21 CFR 7.49(c)(1)(v) |
Template structure in FDA model-letter order
FDA’s Medical Device Model Recall Notification Letter uses the sections below. Placeholders are in square brackets; replace them with your own recall facts.
- Title block. “URGENT: MEDICAL DEVICE RECALL”, the date, the recipient’s address and an attention line.
- Purpose. The model wording is: “The purpose of this letter is to advise you that [Company Name] is voluntarily recalling [Product X].”
- Reason for the recall. The model asks the firm to identify the product concerns or problems, actual or potential, in detail, including failure frequency and adverse events.
- Risk to health. How failure affects patients or providers, and how to recognize that the device may fail.
- Actions for the customer. Actions for safe handling of the recalled product, interim steps, replacement, and acknowledging receipt.
- Product and distribution information. A table of product names, identifiers, lot numbers, dates and quantities.
- What the firm is doing. The model asks “What is the firm doing to correct this issue?” and gives a timeline.
- Contact information. Plus a note that adverse events or quality problems can be reported to FDA MedWatch.
- Return response form. Headed “MEDICAL DEVICE RECALL RETURN RESPONSE”, marked “Response is Required”, covering acknowledgment, adverse events, implants and inventory.
Marking and delivery
21 CFR 7.49(b) says a recall communication can be made by telegrams, mailgrams or first class letters conspicuously marked, preferably in bold red type, on the letter and the envelope, with the words “recall” or “correction” for the product type. The letter and envelope should also be marked “urgent” for class I and class II recalls and, when appropriate, for class III recalls. Telephone calls or other personal contacts should ordinarily be confirmed by one of those methods or documented in an appropriate manner. FDA’s recalls page for devices likewise says the letter and envelope should be marked “urgent” for class I and class II recalls.
Product and distribution table
The model letter’s product table is where recipients match their stock against the notice. Columns follow the model: product name, identifier (catalog or model, serial or code), lot number, dates, and quantity. Section 7.49(c)(1)(ii) is the underlying requirement: enough descriptive detail for accurate and immediate identification. For how lot and code fields are laid out on labels, see how to read a lot number.
Response form
The response form is the “ready means” required by 7.49(c)(1)(v). In the FDA model it asks the recipient to acknowledge the notice and report adverse events, implants and inventory. 7.49(c)(2) adds that where necessary, follow-up communications should be sent to those who fail to respond to the initial communication.
Recall classes and urgency
FDA’s recalls page for devices defines the classes by the health consequence of using the product.
| Class | FDA definition |
|---|---|
| Class I | Reasonable probability of serious adverse health consequences or death |
| Class II | Temporary or medically reversible adverse health consequences |
| Class III | Not likely to cause adverse health consequences |
See class III recall for the lowest class, and FDA enforcement reports for where classified recalls are listed.
Who sends it and when to talk to FDA first
The recalling firm is responsible for promptly notifying each affected direct account. FDA’s recalls page says the firm is encouraged to discuss the recall letter with its FDA Division of Recall Coordinators (DRC) before issuing the notification, and says to be brief and to the point and to explain concisely the reason for the recall and the hazard. Manufacturers and importers must report a correction or removal that involves a risk to health to FDA within 10 working days; see 21 CFR Part 806 and 21 CFR Part 7 recalls.
Recipient checklist
21 CFR 7.49(d) says consignees that receive a recall communication should immediately carry out the instructions in it and, where necessary, extend the recall to their own consignees. A practical reading of that duty:
- Check the product table against what you hold: product, size, lot or serial numbers, codes.
- Note whether the letter tells you to cease distribution or use of remaining product, as 7.49(a)(2) provides for.
- Carry out the letter’s instructions for the product, as 7.49(d) provides.
- Return the response form to the recalling firm, including when you hold none of the product.
- If you supplied the product to others, notify them where the letter calls for it.
- Report adverse events as the letter and FDA MedWatch direct.
The checklist is an editorial summary of 7.49(d) and the model response form, not a substitute for the letter you receive.
Common mistakes the rule warns against
- Burying the message. 7.49(c)(2) says the communication should not contain irrelevant qualifications, promotional materials or any other statement that may detract from the message.
- Vague identification. Without lot, code or serial detail, recipients cannot tell whether they hold affected product.
- No reply route. The letter needs a way for recipients to report whether they have the product.
- No follow-up. Non-responders should be followed up where necessary.
- Unconfirmed phone calls. Telephone contacts should be confirmed in writing or documented.
For an example of a recall that recipients had to check against their own stock, see glucose meter recall; for the corrective-action process behind a recall, see CAPA and FDA.
FAQ
Does FDA provide a model medical device recall letter?
Yes. FDA publishes a Medical Device Model Recall Notification Letter as a Word document, with a return response form. The structure above follows it.
What is the difference between a recall letter and a field safety notice?
Both notify customers of a safety action. A field safety notice is the EU term used with a field safety corrective action; the FSCA guide sets out the cross-walk.
How should a recall letter be marked?
Per 21 CFR 7.49(b): conspicuously, preferably in bold red type, with “recall” or “correction” wording, and “urgent” for class I and class II recalls.