Particle Size Analyzer: Types, Ranges and Method Choice

What a particle size analyzer measures, how laser diffraction, DLS and sieving differ, stated size ranges, and what to record so results can be reproduced.

Written and maintained by CASRAI Editorial Board

Published

On this page

A particle size analyzer is a laboratory instrument that reports how big the particles in a powder, suspension or emulsion are, usually as a distribution rather than a single number. Most benchtop analyzers use laser diffraction, which infers size from how an ensemble of particles scatters a laser beam, while other designs use sieving, dynamic light scattering for submicron particles in liquid, or image analysis.

Who uses it: researchers and technicians who characterise powders, emulsions, suspensions and formulations, and the staff who write and check the methods those measurements follow. The practical decision is almost never “which analyzer” but “which measurement principle fits my size range, sample form and shape”, so this page is organised around that choice.

What a particle size analyzer is

The term covers several measurement principles that all answer the same question: what is the size distribution of this material? Results are reported as a distribution (the share of material at each size) and are usually summarised with a few percentile or mean values. Because different principles measure different physical quantities, the same sample can give different numbers on different analyzers. That is expected behaviour, not a fault, and it is the main reason a method needs to name its technique, dispersant and settings.

The three techniques with the clearest published method descriptions are sieving (mechanical separation), laser diffraction (static light scattering) and dynamic light scattering. Image-based and counting methods exist too; this site covers two of them separately, in the nanoparticle tracking analysis guide and the Coulter counter guide.

How laser diffraction analyzers work

Microtrac describes laser diffraction as a method “based on the deflection of a laser beam by an ensemble of particles” and notes it is also known as static light scattering. Large particles scatter light preferentially at small angles and small particles at wide angles, so the instrument records the scattering pattern on detectors and software converts it into a size distribution. It is an indirect, ensemble measurement: size is inferred from the pattern, not measured particle by particle (Microtrac laser diffraction page, read 2026-11-14).

Wet and dry dispersion

Laser diffraction analyzers handle both liquids and powders. For suspensions and emulsions, Microtrac says the sample is homogenised by recirculator and pump systems, usually with integrated ultrasonic probes. Dry powders are dispersed by a Venturi nozzle in an air stream. The same page gives typical run times of 1-2 minutes for a wet measurement (depending on ultrasound use and cleaning cycles) and 10-40 seconds for a dry one.

Optical model: Mie and Fraunhofer

Converting a scattering pattern into a size needs an optical model. A Malvern Panalytical whitepaper on Mie theory states that the Mie model uses the refractive-index difference between particle and medium, handles transparent and opaque particles, and is recommended across the complete measurement range (0.1 to 3000 micrometres). It describes the Fraunhofer approximation as needing no refractive-index data, assuming opaque discs, being inaccurate below about 2 micrometres and limited to systems containing particles of 50 micrometres or more. Other vendors present Fraunhofer as a default when optical properties are unknown, so the model choice is a method decision to settle in the written procedure and the instrument’s software guidance, not to leave to a default (Malvern Panalytical whitepaper, read 2026-11-14).

Why shape matters

Laser diffraction does not see shape. Microtrac states that all signals are evaluated as if they came from ideal spheres, and that irregular shapes broaden the distribution because both length and width contribute to the signal. The result is therefore an equivalent-sphere size. This is why a laser diffraction result for needle-like or flaky material may not match a sieve or microscopy result for the same sample.

Types of particle size analyzer and the size ranges vendors state

The table below compares the three techniques that the sources describe in numbers. Every range is a statement by the named source, not a guarantee for any particular instrument model; the instrument’s own specification and your validated method decide what is achievable. The full compiled data table, with sources, appears in the data section after this article.

Technique Stated size range Typical sample Key difference
Sieving Tens to hundreds of micrometres, with poor performance under 45 micrometres (Malvern Panalytical) Dry powders and granular material Separates by whether particles pass openings; flaky or needle-shaped particles can pass openings even if several times larger
Laser diffraction 0.1 to 3500 micrometres in one measurement (Malvern Panalytical); typically 10 nm to 4 mm coverage, about 30 nm to 1,000 micrometres in practice (Microtrac) Dry powders, suspensions, emulsions Ensemble measurement; equivalent-sphere result; governed by ISO 13320
Dynamic light scattering (DLS) Submicron particles in liquid, up to approximately 1 micrometre (Ph. Eur.-harmonised chapter 2.5.13) Dispersions that are free from sedimentation Reports an average hydrodynamic diameter and a polydispersity index, not a full distribution of a broad sample

Sieve analysis

Sieving is the oldest route and still the reference for coarse granular material. Malvern Panalytical’s comparison says sieving is limited to a range of tens to hundreds of microns, performs poorly under 45 microns, takes several minutes per test plus cleaning, and that reproducibility “can vary greatly with different users and over time”. Its figures for laser diffraction by comparison are a typical 15 second test on dry dispersion with 10 mg to 30 g of sample (several grams typically), and vendor-claimed reproducibility errors not exceeding 1% and repeatability within 0.5%. Treat those last figures as the vendor’s claims. For the equipment side of sieving, see the guide to the sieve shaker.

Dynamic light scattering

DLS answers a different question from laser diffraction. The harmonised pharmacopoeial chapter 2.5.13, based on ISO 22412:2017, says DLS determines the average hydrodynamic particle size and the broadness of the size distribution of submicron particles dispersed in a liquid, and is suited to randomly moving particles up to approximately 1 micrometre. It works from Brownian motion: scattered-light intensity fluctuates more slowly for larger particles and more rapidly for smaller ones, and the diffusion coefficient is converted to a hydrodynamic diameter by the Stokes-Einstein equation. The reported outputs are the average diameter (also called the z-average) and the polydispersity index (PI); the chapter defines the average as the intensity-weighted harmonic mean diameter in nanometres. Reading those two numbers correctly is covered in the guide to DLS z-average and PDI interpretation, and a named commercial DLS instrument family is described in the Zetasizer guide.

Specifications that matter

When reading an analyzer’s specification sheet, the numbers that decide whether it can run your method are the following. Each is taken from what the sources above say the technique depends on.

  • Size range. Ask for the range claimed for the technique and the range validated for your material. Vendor pages quote wide theoretical coverage (for example 10 nm to 4 mm) while describing narrower practical use (about 30 nm to 1,000 micrometres).
  • Sample form and dispersion. Wet module, dry module or both; Microtrac describes recirculation with ultrasound for liquids and a Venturi nozzle for powders.
  • Sample quantity. Malvern Panalytical gives 10 mg to 30 g for dry dispersion laser diffraction, with several grams typical.
  • Optical model and refractive-index inputs. Mie needs the refractive-index difference between particle and medium; Fraunhofer does not, but has the limits described above.
  • Temperature control (DLS). The chapter says the sample holder must keep the sample temperature within an appropriate range, for example plus or minus 0.3 degrees Celsius.
  • Standards the software supports. Microtrac names ISO 13320 for laser diffraction methodology and ISO 9276-1 and 9276-2 for the representation of results.

Choosing a measurement principle

Only criteria that the sources state are listed here. They are conditions to check against your material, not a ranking of instruments.

If your situation is… What the sources say
Material mostly finer than 45 micrometres Sieving performs poorly under 45 micrometres (Malvern Panalytical); laser diffraction is stated to cover from 0.1 micrometre.
Particles are needle-like or flaky Such particles can pass sieve openings even if several times larger (Malvern Panalytical); laser diffraction reports an equivalent-sphere size and does not detect shape (Microtrac), so expect the two to differ.
Submicron particles in a liquid, need an average size and PI DLS is described for particles up to approximately 1 micrometre in liquid (chapter 2.5.13).
Both powders and suspensions in the same lab Laser diffraction analyzers handle wet and dry samples (Microtrac), with different run times for each.
Need to see particle shape as well as size Microtrac lists a combined laser diffraction and dynamic image analysis analyzer; image analysis is the route that observes individual particles.

Sample preparation and method basics

Preparation drives the result at least as much as the instrument does. The DLS chapter is the most explicit openly readable procedure, and several of its points apply in spirit to any liquid dispersion measurement.

  • The dispersion medium should not absorb at the laser wavelength, not dissolve, swell or aggregate the particles, have a known refractive index different from the sample, and be free of dust.
  • Concentration must be low enough to avoid multiple scattering. The chapter says to determine the working range from systematically diluted samples until results stop varying significantly.
  • When water is the medium, the chapter recommends fresh distilled or desalted water filtered to a nominal 0.2 micrometre pore size.
  • Air bubbles, which can appear when a refrigerated sample warms to room temperature, are to be avoided, and sediment at the end of a run suggests aggregation, precipitation or an unsuitable sample.
  • Allow the instrument to warm up and the sample to reach temperature equilibrium before measuring.

What to record in the method report

For DLS, the chapter says the test report must include the average particle diameter and PI, the dispersion medium, its refractive index and viscosity, the sample temperature, the measurement principle, the optical configuration, laser wavelength and observation angle, the measuring time or number of acquisitions, the sample nature, concentration and preparation, the dispersion conditions, the instrument settings, the cell type and the data analysis program. It also states that mandatory repeatability limits cannot be set in the chapter, but that a relative standard deviation of not more than 10 per cent (n of at least 3) is good practice for the average diameter. The same discipline, recording the medium, optical model and settings alongside the result, is what makes a laser diffraction result reproducible by someone else.

Checking the instrument

The chapter notes that DLS sizes are absolute values from first principles, so the instrument is not calibrated, but its performance must be checked after installation or when abnormal performance is suspected, using particles of certified diameter, and the check should be repeated at least once a year. It gives polystyrene latex of about 100 nanometres as an example reference, with the measured average within the certified range expanded by 2 per cent on each side, a PI of not more than 0.1 by cumulants and a relative standard deviation of not more than 2 per cent over at least five measurements. These acceptance values come from that chapter and apply to its DLS procedure only; do not carry them over to other techniques.

Common mistakes

  • Comparing results across techniques as if they were the same quantity. Laser diffraction gives an equivalent-sphere distribution and DLS gives a hydrodynamic diameter; neither equals a sieve cut for irregular particles.
  • Using a sieve below its stated range. Malvern Panalytical describes poor performance under 45 micrometres.
  • Accepting the default optical model without checking it. The Fraunhofer approximation is described as inaccurate below about 2 micrometres and limited to particles of 50 micrometres or more in one vendor’s whitepaper, while other vendor guidance differs, so the method should state the model and its basis.
  • Measuring a dusty or over-concentrated liquid sample. The DLS chapter ties both to unreliable results and describes dilution series and dust removal as the fix.
  • Reporting a single size with no settings. Without the dispersant, optical inputs and technique, the number cannot be reproduced.
  • Treating vendor repeatability claims as method validation. Figures such as 1% reproducibility are vendor statements; the lab’s own repeatability on its own material is what counts.

Safe handling basics

These instruments use enclosed lasers and handle powders and dispersions, so follow the manufacturer’s instructions for use for laser safety and for the safety data sheet of every sample and dispersant. The DLS chapter allows rinsing a reusable cell with water or an organic solvent, with a non-abrasive detergent if required, and says cleaning is not necessary with an individually packaged clean disposable cell; the instrument manual defines which cleaning agents are compatible with the wetted parts.

Sources

Data: Particle size measurement techniques: stated size range, sample form, run time and governing standard

Compares sieving, laser diffraction and dynamic light scattering using statements published by two instrument manufacturers and a harmonised pharmacopoeial chapter. Ranges are the sources' stated figures for the technique, not specifications of any instrument model.

Particle size measurement techniques: stated size range, sample form, run time and governing standard
TechniqueStated size rangeSample formTypical run timeStandard named by sourceSource
SievingTens to hundreds of micrometres; poor performance under 45 micrometresDry powderSeveral minutes per test plus cleaningNone namedMalvern Panalytical sieve vs laser diffraction page
Laser diffraction0.1 to 3500 micrometres (Malvern Panalytical); typical coverage 10 nm to 4 mm, about 30 nm to 1,000 micrometres in practice (Microtrac)Dry powder, suspension, emulsionAbout 15 seconds dry (Malvern Panalytical); 10-40 seconds dry, 1-2 minutes wet (Microtrac)ISO 13320; results display per ISO 9276-1 and 9276-2Malvern Panalytical; Microtrac
Dynamic light scatteringSubmicron particles, up to approximately 1 micrometreParticles dispersed in liquid, free from sedimentationNot statedISO 22412:2017 (basis of chapter 2.5.13)Chapter 2.5.13 (Indian Pharmacopoeia Commission copy)

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Particle Size Analyzer: Types, Ranges and Method Choice

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →