Protein Assays: Bradford, BCA, Lowry and UV Compared

Protein assays compared: Bradford, BCA, Lowry, A280 and fluorescent methods, with a sample-first selection table and protein-to-protein variability data.

Written and maintained by CASRAI Editorial Board

Published

On this page

Short answer

Protein assays are laboratory methods that estimate how much protein is in a sample. The main families are UV absorbance at 280 nm, copper-based colorimetric assays (BCA, Lowry), dye-binding assays (Bradford), and fluorescent dye assays. The right one depends mostly on what else is in your buffer: detergents, reducing agents and expected concentration decide the chemistry.

Definition and scope

A protein assay converts protein content into a measurable signal, usually absorbance or fluorescence, and reads that signal against a standard curve of a reference protein. The result is an estimate relative to the standard, not an absolute mass. This page compares methods at the level of chemistry and selection; step-by-step bench protocols are covered in the linked method guides.

How protein assays work in practice

Thermo Fisher’s overview of protein assays groups the methods into four categories (page read 2026-11-12):

  • UV absorbance at 280 nm: simple and reagent-free, but described as highly error prone with protein mixtures or complex samples such as cell lysates. See UV-Vis spectrophotometer basics.
  • Copper-based (biuret-type) colorimetric assays: BCA and Lowry rely on protein-copper chelation and detect the reduced copper.
  • Dye-based colorimetric assays: the Bradford assay is fast and run at room temperature.
  • Fluorescent dye assays: EZQ, Qubit, NanoOrange and CBQCA Plus offer high sensitivity with less protein but need specialised equipment.

Whichever method you use, Thermo recommends running a standard curve in every assay, with at least two replicates per point, and confirming that your spectrophotometer or plate reader suits the detection method. See what a plate reader is.

Types of protein assay

Bradford assay (protein quantification)

Bio-Rad describes the Bradford assay as based on Coomassie Brilliant Blue G-250 dye (page read 2026-11-12). The dye’s red form absorbs maximally at 470 nm; the protein-bound blue anionic form absorbs at 595 nm; a neutral green form absorbs at 650 nm. The dye must be doubly protonated to bind, and it binds charged groups, with sulfonic acid groups binding positively charged amines and van der Waals forces also contributing. Bio-Rad notes the standard curve flattens at high concentration, so samples above the linear range are underestimated and must be diluted and re-assayed. For the calculation, see Bradford standard curve calculation and Coomassie blue.

BCA assay

BCA is a copper-chelation method. Thermo lists Standard BCA, Micro BCA and Rapid Gold BCA among its variants. See BCA standard curve calculation and the Micro BCA assay.

Lowry assay

The Lowry assay is also copper-based. See the Lowry assay guide.

UV absorbance and fluorescent assays

These sit at opposite ends: A280 needs no reagent but tolerates complex samples poorly, while fluorescent assays give high sensitivity at the cost of specialised equipment. See Qubit fluorometer.

Where it matters: choosing a chemistry from your sample

This table restates Thermo Fisher’s selection guidance (read 2026-11-12).

If your sample contains or needs Thermo guidance
Detergents Copper-based assays are not inhibited by low to moderate amounts; dye-based assays are incompatible with surfactants.
Reducing agents Coomassie dye assays are compatible; copper methods are incompatible with DTT.
Dilute protein Micro BCA, NanoOrange or CBQCA Plus.
More than 2 mg/mL Rapid Gold BCA or Qubit Protein BR.

Protein-to-protein variability

Thermo reports how much different pure proteins respond differently at 1000 µg/mL. Copper chelation assays showed less protein-to-protein variability than dye binding.

Assay CV (%) Average response ratio to BSA
Standard BCA 20.52 0.97
Micro BCA 17.78 1.11
Rapid Gold BCA 20.67 1.10
Coomassie 41.57 0.69
Coomassie Plus 37.62 0.75

A ratio of 0.69 means the average protein gave about 69% of the signal BSA gives, so a Coomassie result read against a BSA curve will on average understate those proteins.

Common misunderstandings

  • “Coomassie is Coomassie.” Bradford assays use G-250. Bio-Rad states R-250 is for gel staining and is not used in protein assays.
  • “BSA is always the right standard.” Thermo uses BSA as the default because it is pure, available and inexpensive, but suggests bovine gamma globulin (BGG) for antibody work because its curve resembles IgG, and says the most accurate standard is the most abundant protein in your sample. Bio-Rad likewise says the ideal standard is the same protein as the sample, which is often impractical.
  • “A reading is an absolute amount.” It is an estimate relative to the chosen standard, which is why the variability table matters.
  • “Any reading on the curve is valid.” Readings above the linear range are underestimated and need dilution and re-assay.

FAQ

What is the Bradford assay?

A dye-binding protein assay using Coomassie Brilliant Blue G-250, read at 595 nm for the protein-bound blue form (Bio-Rad).

Which protein assay tolerates detergent?

Per Thermo, copper-based assays such as BCA are not inhibited by low to moderate detergent; dye-based assays are incompatible with surfactants.

Which protein assay tolerates reducing agents?

Coomassie dye assays are compatible; copper methods are incompatible with DTT (Thermo).

Why does the same protein give different results in different assays?

Each chemistry responds differently to each protein; see the variability table above.

Sources

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Protein Assays: Bradford, BCA, Lowry and UV Compared

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →