Skip to main content
v2026.11,986 entries · CC-BY 4.0
LAC HealthSponsored

Shop salem sump supplies

  • FDA registered
  • ISO 13485
  • BBB Accredited A+

Shop at LAC Health

Salem Sump Tube vs Levin Tube: Key Differences

How a Salem sump tube differs from a Levin tube, how FDA classifies each, and what the open anti-reflux valve recall means for supply checks.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

General product information only, not medical advice and not a procedure guide. Clinical use of any tube follows the clinician’s judgment and the manufacturer’s current instructions for use.

Short answer

A Levin tube is a single-lumen nasogastric tube with several perforations near its tip. A Salem sump is a dual-lumen tube: a suction lumen plus a separate vent lumen. Both fall under the same FDA regulation, 21 CFR 876.5980, but they are filed under different FDA product codes, and the Salem sump now has an open FDA Class 2 recall about its anti-reflux valve. This guide covers the difference, the regulatory identity of each, and what that recall changes for labs and clinics that stock the tube.

What a Salem sump tube is

A peer-reviewed 2026 case report in the Journal of Medical Cases describes the Salem sump as a modern dual-lumen tube, commonly used for nasogastric decompression and medication delivery, that evolved from the earlier Levin design (Hughes, Holland and Tobias, 2026; PMC12885141, checked 2026-10-30). The product names in the FDA recall record use the same wording, for example “Salem Sump Stomach Tube, Dual Lumen,” in sizes from 6 Fr to 18 Fr.

The recall notice names two parts of the sump design: a “blue sump vent lumen” and a “suction lumen,” with a removable anti-reflux valve (ARV) that seats into either one. That is the manufacturer’s own vocabulary. It is why the Salem sump is described as having a vent: the second, smaller lumen is the vent.

Levin tube vs Salem sump

The same case report describes the Levin tube, developed by Dr. Abraham Levin in 1921, as a non-weighted, single-lumen rubber tube with multiple distal perforations, used for nasogastric aspiration, lavage and gavage. The table below keeps to what the sources state: the construction from the case report, and the regulatory identity from FDA records. Nothing in it is a performance, price or “which is better” claim.

Attribute Levin tube Salem sump tube
Lumens (case report) Single lumen, multiple distal perforations Dual lumen
Regulation 21 CFR 876.5980 (named in the rule’s examples: “Levine”) 21 CFR 876.5980 (gastrointestinal tube and accessories)
FDA product code FRQ, “Tube, Levine” KNT, “Tubes, Gastrointestinal (And Accessories)” (as listed in the FDA recall record)
Device class (FDA classification database) Class 2, 510(k) Class 2, 510(k) (code KNT)
FDA recall record for the Salem sump Not applicable Z-0207-2026, Class 2 (anti-reflux valve)

Spelling varies: the regulation and the FDA code both say “Levine,” while many clinical sources write “Levin,” which is the spelling of the physician the tube is named after. For the standard nasogastric product code (BSS, “Tube, Nasogastric”), see What Is a Nasogastric Tube? and Parts of a Nasogastric Tube; BSS is also Class 2 under the same regulation.

Regulatory classification and the anti-reflux valve recall

21 CFR 876.5980 defines a “gastrointestinal tube and accessories” as flexible or semi-rigid tubing used to instill or withdraw fluids, to splint, or to suppress bleeding in the alimentary tract, and its examples include nasogastric, Levine, feeding and gastroenterostomy tubes (eCFR, checked 2026-10-30). The regulation does not describe the vent lumen or the sump principle, so the rule alone does not distinguish the two designs. The FDA product code does: the product-code database lists FRQ (Levine) and BSS (nasogastric) as Class 2, with BSS referencing recognized standards for small-bore enteral connectors (ISO 80369-3 among them).

Recall Z-0207-2026

The FDA recall database record (checked 2026-10-30) lists a Class 2 recall of “Salem Sump” dual-lumen stomach tubes from Cardinal Health 200, LLC, covering 11 product codes and all lots, distributed worldwide:

  • Initiated by the firm: September 18, 2025. Posted by FDA: October 17, 2025.
  • Reason given: complaints of anti-reflux valve breakage “due to excessive force during use.” The firm is providing additional guidance for use. FDA recorded the cause as use error.
  • Status: open and classified; the record’s footnote states that the manufacturer has initiated the recall and not all products have been corrected or removed.
  • Relation to an earlier notice: the record says the 2025 action updates a July 2024 notice (RES 95214) and adds product codes.

The firm’s correction letter, as reproduced in the FDA record, supplies additional guidance for use of the valve; that guidance is written for trained staff and is not reproduced here. The current instructions for use are the reference, and the notice says they are on the manufacturer’s electronic IFU site.

Where it matters for labs, clinics and supply decisions

Every item below traces to the records cited above; none is a purchasing recommendation.

  • Product identity on a listing. A listing that says only “NG tube” can mean a single-lumen Levine-type tube or a dual-lumen sump. The lumen configuration, and the FDA product code if the listing gives one, are the facts that separate them.
  • Recall scope. The Salem sump recall covers all lots of the 11 listed product codes, so the question for a facility that holds these tubes is whether a stocked item’s product code appears in the FDA record, not whether a lot number matches.
  • Labeling version. The notice points users to updated electronic instructions for use. A printed insert or a shelf label in stock may predate the 2025 correction.
  • Connector standard. FDA lists ISO 80369-3 as a recognized standard for the nasogastric tube product code, and FDA has published guidance on misconnection risks for small-bore enteral connectors.

For the wider regulatory context, see FDA recall classification: Class I vs II vs III, field safety corrective actions and FDA unique device identifier requirements, which explain how a product code or UDI-DI is used to match a device to a recall record.

Common misunderstandings

Is “Levin” and “Levine” the same tube?

Yes, they are two spellings of the same tube name. FDA’s regulation and product code use “Levine”; the case report and many clinical texts use “Levin.”

Is the vent lumen a second suction or feeding channel?

The recall notice labels it the “sump vent lumen” and the other lumen the “suction lumen,” so the sources treat them as separate lumens with separate roles. FDA’s records do not describe how the vent works, so this page does not explain the airflow.

Does the recall mean Salem sump tubes were banned?

No. The record describes a Class 2 correction with additional guidance for use. FDA’s recall database footnote explains the record is created when a firm initiates a correction or removal action.

FAQ

What is the main difference between a Salem sump and a Levin tube?

The Levin is a single-lumen tube with distal perforations; the Salem sump is a dual-lumen tube with a separate vent lumen (Hughes et al., 2026).

What does “Salem sump vs Levin tube” mean for classification?

Both are covered by 21 CFR 876.5980. FDA lists the Levine tube as product code FRQ, Class 2; the Salem sump recall record lists code KNT.

Is there a recall on Salem sump tubes?

FDA’s recall database lists Class 2 recall Z-0207-2026, initiated September 18, 2025, for anti-reflux valve breakage, covering all lots of 11 product codes (checked 2026-10-30).

Sources

  • eCFR, 21 CFR 876.5980, Gastrointestinal tube and accessories: ecfr.gov/current/title-21/part-876/section-876.5980, checked 2026-10-30.
  • FDA Product Classification database, codes FRQ and BSS, and openFDA device classification for KNT (data updated 2026-10-19), checked 2026-10-30.
  • FDA Medical Device Recalls database, Salem Sump recall Z-0207-2026, checked 2026-10-30.
  • Hughes M, Holland K, Tobias JD. Misplaced and Knotted Nasogastric Tubes in Infants and Children: Report of Two Cases. J Med Cases, 2026 Mar. doi:10.14740/jmc5272, PMC12885141, checked 2026-10-30.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Salem Sump Tube vs Levin Tube: Key Differences

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →