Shock Response Spectrum: Meaning, Damping and Test Reports

Interpret shock spectra through oscillator response, damping and peak settings, with a laboratory report checklist and a software settings table.

Written and maintained by CASRAI Editorial Board

Published

On this page

A shock response spectrum (SRS) plots the largest responses of idealized oscillators against their natural frequencies when the same transient acceleration drives their bases. It describes how resonant systems respond to a shock. Interpreting a laboratory report requires the response quantity, damping, peak definition and analysis interval as well as the curve.

Short answer

Read an acceleration SRS as a collection of model responses: at each frequency, the ordinate is the selected acceleration response of an oscillator tuned to that frequency. It is not a measurement of a separate physical component at every plotted point. Siemens explains that software calculates these responses from a measured or predicted acceleration history; a physical bank of springs and masses is an illustration of the calculation.

The practical value is comparison. A measured environment and a test specification can be expressed using the same reference oscillators. However, a curve labeled only “SRS” leaves essential choices unstated. The checklist below connects those choices to measurement validity and retained evidence; the downloadable data table compares settings documented by three software suppliers.

Definition and scope

The usual model is a linear single-degree-of-freedom system, abbreviated SDOF: a mass, spring and viscous damper attached to a moving base. Each model has one natural frequency and a specified damping ratio. Repeating the calculation at different natural frequencies produces the response spectrum. The horizontal axis identifies the oscillator, while the vertical axis identifies its response.

NI distinguishes absolute acceleration of the mass from displacement of the mass relative to its base. These are different response quantities with different engineering meanings. An acceleration SRS is commonly expressed in g or metres per second squared; a relative-displacement spectrum has length units. The word “spectrum” alone does not establish which quantity was calculated.

An SRS condenses a transient record into response values useful for assessing a mechanical shock environment. It does not retain the complete timing and phase information needed to recover the original waveform. This makes it distinct from the transform methods discussed in harmonic analysis. A similar-looking frequency axis does not make the outputs interchangeable.

The underlying linear oscillator also defines the scope of interpretation. Spectral Dynamics warns that shocks sharing an SRS can differ in fatigue loading or nonlinear behavior. Consequently, an SRS match is evidence about the specified reference responses; it is not a universal prediction that every real assembly will behave identically.

How a shock response spectrum is calculated

The calculation takes a base-acceleration time history, evaluates the response of each reference oscillator, extracts the specified response statistic and plots it against natural frequency. The signal is common to the oscillators; their natural frequencies differ. These are computational models, so a single measurement location can support a whole spectrum.

Damping and Q

Damping controls the oscillator response and must accompany the result. For the oscillator convention documented by MathWorks, Q = 1/(2ζ), where Q is quality factor and ζ is damping ratio. Thus Q = 10 corresponds to ζ = 0.05, or 5% of critical damping. SIGVIEW documents the same relationship, and Siemens specifically advises comparing spectra calculated with the same Q.

That reference damping is an analysis choice. A reported Q of 10 does not establish that the actual test article has 5% damping. If an engineer changes Q between analyses, the resulting ordinate values can change even though the measured acceleration input remains the same. Report comparisons therefore need the numerical setting, rather than an assumption about a software default.

Frequency grid and response interval

Octave spacing specifies the distribution of oscillator natural frequencies. A finer grid evaluates more oscillators and can reveal detail between previously plotted points. Siemens notes that increasing points per octave does not change the response of an oscillator at a natural frequency common to both calculations. It changes where the response is sampled across frequency.

The analysis also needs a time interval. An oscillator may keep ringing after the driving pulse ends, and its largest response may occur during that free vibration. MathWorks distinguishes input-only, post-input and full response intervals. Truncating the analysis at the input pulse can therefore omit a relevant peak. This is an example of why signal processing choices belong in the interpretation of a measurement.

Primary, residual and maximax spectra

The primary spectrum concerns response during excitation; the residual spectrum concerns response after excitation. These terms describe time intervals. Positive, negative and absolute peaks describe how amplitude is extracted. NI documents both distinctions. They answer separate questions and should be specified separately in a report.

In its common full-response use, maximax takes the largest absolute response from the primary and residual intervals together. A large negative excursion can therefore control a positive-valued maximax ordinate. Comparing that ordinate with a positive-peak-only spectrum mixes definitions, even if the plots use identical units and damping. Siemens maps maximax to its “Maximum” instance and “Absolute” amplitude settings.

Acceleration SRS and pseudo-velocity shock spectrum

A pseudo-velocity shock spectrum, or PVSS, is calculated from relative displacement and oscillator natural frequency. MathWorks defines pseudo-velocity as ωn times the selected relative-displacement extremum, where ωn is angular natural frequency. It has velocity units, but it is not generally the instantaneous relative velocity of the oscillator during a transient.

This distinction matters when a report changes ordinate type. Acceleration response, relative displacement and pseudo-velocity describe different aspects of the same modeled event. MathWorks discusses PVSS as a useful indicator of modal stress severity for lightly damped, linear elastic response. That interpretation retains its model assumptions; it does not supply a material-independent failure threshold.

Spectrum analysis and shock synthesis

Analysis moves from a time history to oscillator-response values. Synthesis seeks a time history that meets a target spectrum. Because the spectrum does not preserve a unique waveform, more than one synthesized input can meet a target. Spectral Dynamics describes using this freedom to produce waveforms compatible with shaker velocity and displacement limits.

That freedom also creates a review obligation: inspect requirements beyond the response curve. A synthesized pulse and a measured event can share spectral peaks while differing in oscillatory character or duration. Those differences matter whenever the governing test specification constrains them.

Where SRS matters in laboratory testing

SRS is useful when a mechanical-test laboratory must represent a field shock at an equipment interface and compare it with a controlled test. Siemens describes calculating the spectrum at a component base and placing the laboratory control accelerometer at a corresponding location. The measurement location is part of the environment definition, rather than an incidental plotting detail.

NASA-STD-7003A, the 2011 Revision A copy of Pyroshock Test Criteria, supplies a concrete aerospace example. Section 4.4.2 specifies Q = 10 maximax spectra for externally induced pyroshock tests. Sections 4.4.4 and 4.4.4.1 also address waveform oscillatory character and duration. Those requirements illustrate why matching spectral levels alone does not describe the complete test.

The intended survival outcome also needs definition. The MathWorks practical introduction distinguishes avoiding permanent damage from avoiding performance degradation during or after the event. A transient interruption can matter even when hardware remains physically intact. An SRS curve describes the loading environment; functional observations describe what the test article did under that loading.

For research equipment, these distinctions help separate measurement, environmental qualification and instrument condition. Calibration provides a related measurement concept, but the shock report must still identify its own response model and acceptance specification. Neither a calibrated measurement chain nor a plotted response curve supplies the missing functional acceptance criteria.

A checklist for reviewing an SRS report

The following checklist combines documented software parameters with the NASA Revision A measurement provisions. It is a guide to interpreting the evidence in a report; project acceptance limits remain those of the governing specification.

  • Identify the ordinate. Establish whether the spectrum shows absolute acceleration, relative displacement, relative velocity or a pseudo quantity. NI and MathWorks document distinct models; identical numerical values with different response definitions are not equivalent results.
  • Record damping explicitly. Obtain Q or damping ratio for the reference oscillators. Compare spectra on a consistent damping basis, as Siemens advises, and distinguish that setting from measured damping of the hardware.
  • Match the peak and interval definitions. Check the selected amplitude statistic and whether post-pulse response is included. A primary positive spectrum cannot stand in for full-response maximax without changing the quantity being compared.
  • Check the actual frequency band. Record the minimum and maximum oscillator frequencies and their spacing. Software defaults are implementation choices; they do not establish that the complete project test band has been evaluated.
  • Review acquisition and baseline validity. NASA Revision A requires analog anti-alias filtering before conversion and warns that small acceleration zero shifts can substantially distort the computed spectrum. Its section 4.7 recommends integration and comparison of positive and negative spectra as validity checks.
  • Retain the waveform and analysis identity. Section 4.8.1.3 requires preserving the laboratory acceleration history for comparison with flight data. Section 4.7 also identifies algorithm choice as a source of variability. A spectrum image alone cannot reproduce those checks.

Illustrative composite: a frequency-band mismatch

Suppose a laboratory has a record sampled at 48,000 Hz and uses the documented default frequency range of the MathWorks srs function. Its upper oscillator frequency is Fs/24, so the calculation gives 48,000/24 = 2,000 Hz. The record’s Nyquist frequency is 48,000/2 = 24,000 Hz. Those two limits describe different things.

If an illustrative project specification asks for results through 5,000 Hz, that default spectrum does not reach the requested upper frequency. The existence of a sampled record with a higher Nyquist frequency does not fill the missing plotted band. The reviewer needs an explicitly defined analysis range and a justified acquisition and calculation method for it.

This is arithmetic derived from a published software default, not measured test data or a recommended sampling ratio. MathWorks explains that the Fs/24 default aims to limit high-frequency peak-estimation error. The example shows why a report should state the frequencies actually evaluated instead of relying on the sample rate to imply them.

The data table below brings together MathWorks defaults, SIGVIEW spacing options and Siemens parameter mappings. It supports translation between report vocabularies; it does not rank the packages or establish their equivalence under every configuration.

Common misunderstandings

A high SRS ordinate is not necessarily the input’s peak acceleration. The curve describes oscillator responses, which can be amplified. Spectral Dynamics shows that the high-frequency absolute-acceleration response approaches the input peak, while intermediate-frequency oscillators can respond more strongly. Reading every ordinate as a directly measured input peak confuses response with excitation.

A matched spectrum does not recreate the original event uniquely. Phase and detailed timing are lost. Different waveforms can share the same SRS, so the stored time history remains important when duration, waveform character or repeated response cycles affect the question being tested.

Maximax does not identify a particular response time or direction. It selects an absolute extreme. Primary, residual and signed spectra can help explain that extreme, but they must be labeled so that diagnostic outputs are not confused with the specified comparison quantity.

A corrected baseline is not automatically proof of valid data. Siemens describes offset corrections, while NASA Revision A calls for data-validity evaluation and warns about acquisition errors. The defensible interpretation depends on the measurement record, the processing applied and the response model together. A clean-looking curve is only one part of that evidence.

Sources

Data: Shock response spectrum settings across three software references

Original compilation of documented damping, frequency-grid and peak-selection settings from MathWorks srs, the SIGVIEW manual and Siemens Simcenter Testlab guidance. Entries describe the cited implementations; options and defaults are labeled separately.

Shock response spectrum settings across three software references
Software referenceDamping documentationFrequency-grid documentationPeak or interval documentationSource
MathWorks srs (R2026b)Default Q = 10; damping ratio 0.05Default 1/6 octave; lower Fs/(3*2^15), upper Fs/24Default full response and maximax; primary/residual segments and signed peaks are optionshttps://www.mathworks.com/help/signal/ref/srs.html
SIGVIEW SRSQ = 10 corresponds to 5% relative dampingOptions: 1/24, 1/12, 1/6, 1/3 octave or linear; configurable minimum and maximum frequenciesPeak-detection options control detection of instantaneous acceleration peakshttps://sigview.com/help/ShockResponseSpectrumSRS.html
Siemens Simcenter TestlabUser-set Q; typical Q = 10 corresponds to 5% critical dampingPoints per octave range 1 to 48; typical value 6Maximax mapping: Instance = Maximum, Amplitude = Absolutehttps://community.sw.siemens.com/articles/en_US/Knowledge/shock-response-spectrum-srs

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Shock Response Spectrum: Meaning, Damping and Test Reports

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →