Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
General product and regulatory information for laboratory, clinic and supply staff. It is not medical advice and does not describe how surgery is performed.
Short answer
A stapedectomy prosthesis, also called a stapes piston or stapes prosthesis, is an implant that replaces the stapes bone in the middle ear so sound vibrations can still reach the inner ear. Published studies describe two common families: fluoroplastic (Teflon) pistons with a steel or platinum wire loop, and titanium pistons that attach by a clip or a crimped loop. In the US the FDA regulates these devices under 21 CFR 874.3450 as Class II.
Definition and scope
The stapes is the smallest bone of the middle-ear chain. A stapedectomy prosthesis is the implant that takes over its job of passing vibration from the incus to the inner ear. This page covers what the device is, how its main designs differ in published studies, and how it is classified in the US. It does not cover surgical technique, patient selection or individual treatment decisions; those belong to the treating surgeon.
How stapes prostheses are classified by FDA
The regulation is titled “Partial ossicular replacement prosthesis” (21 CFR 874.3450, checked 2026-10-31 on Cornell LII). It describes a device implanted for the functional reconstruction of segments of the ossicular chain that helps conduct sound from the tympanic membrane to the inner ear. It lists materials such as stainless steel, tantalum, polytetrafluoroethylene, polyethylene, a polytetrafluoroethylene with carbon fibers composite, absorbable gelatin material and porous polyethylene, or combinations of them. The device is Class II.
The openFDA classification database (checked 2026-10-31) shows three product codes under this one regulation, all Class II:
| Product code | FDA device name |
|---|---|
| LBP | Replacement, Ossicular (Stapes) Using Absorbable Gelatin Material |
| ETB | Prosthesis, Partial Ossicular Replacement |
| LBM | Porous Polyethylene Ossicular Replacement |
LBP is flagged as an implant and not as life-sustaining or life-supporting. openFDA lists only LBP among device names containing “stapes”, so the code should not be read as covering every piston-style prosthesis. The regulation text itself never uses the words “stapes” or “piston”. For the related partial and total ossicular replacement devices, see the CASRAI guide to TORP and PORP.
Types: piston, clip and crimp designs
Two peer-reviewed comparisons, both checked 2026-10-31, describe the designs below. The studies are small or single-centre and report outcomes for their own patients; they do not rank products.
| Design | What the source says | Source |
|---|---|---|
| Teflon (fluoroplastic) piston with steel or platinum wire loop | Described as one of the two most common piston-style types. | Iran J Otorhinolaryngol 2017 (PMC5307301) |
| Titanium soft-clip piston | Titanium has a lower weight and is much softer than the steel or platinum wire used with Teflon pistons. | Iran J Otorhinolaryngol 2017 (PMC5307301) |
| Titanium CliP Piston | Titanium CliP Piston or crimp prostheses are the most widely used today; the CliP gives a stiff connection to the incus. | Eur Arch Otorhinolaryngol 2023 (PMC10220147) |
| Crimp prosthesis | A bent loop that imitates the incudostapedial joint. | Eur Arch Otorhinolaryngol 2023 (PMC10220147) |
Stapes piston
In the 2017 study the Teflon pistons were paired with wire loops and the titanium soft-clips with a clip attachment, and the 2023 study compared a CliP Piston with a crimp prosthesis, so the studies compare attachment style as well as material.
Stapes prosthesis types, in the two head-to-head studies
- In a 2017 study of 57 ears with a titanium soft-clip and 63 ears with a Teflon piston, the speech reception threshold gain did not differ (P=0.919). The Teflon-piston group showed a larger air-bone gap improvement at 0.25 and 0.5 kHz (P=0.007 and P=0.001).
- In a 2023 retrospective single-centre study of 190 patients (2016-2019, one surgeon), the mean air-bone gap was under 10 dB in most cases whichever prosthesis was used. The crimp prosthesis closed the gap better at 0.5 kHz, a significant difference.
The dimensions in these papers are study parameters, not a product range: 0.4 mm piston diameter in the 2023 study; in the 2017 study Teflon pistons were 4-6 mm long and 0.6 mm wide, soft-clips 4.25-4.75 mm long and 0.5 mm wide.
Where it matters for labs, clinics and supply decisions
Because the device is an implant under a Class II regulation, records matter. A checklist of what staff can log for any stapes implant record:
- Material and design as stated on the manufacturer’s labelling (for example Teflon piston with wire loop, titanium clip, titanium crimp).
- FDA regulation number (874.3450) and the product code shown for that specific device in the FDA database, rather than assuming LBP.
- The unique device identifier, where the label carries one; see the CASRAI guide to UDI for medical devices.
For how Class II devices reach the US market, see FDA 510(k) premarket notification and FDA device classification and pathways. For the general definition, see what is a medical device.
Common misunderstandings
- “LBP covers every stapes prosthesis.” LBP is the code named for absorbable gelatin material; openFDA lists only that code for device names containing “stapes”.
- “Material alone decides the hearing result.” In the 2023 study, mean air-bone gap was under 10 dB in most cases regardless of prosthesis; differences appeared at specific frequencies.
- “The sizes in a paper are standard sizes.” They are what each study used.
FAQ
What is a stapes piston?
It is a piston-style stapes prosthesis, made in designs such as Teflon with a wire loop or titanium with a clip or crimp, that replaces the stapes bone in the middle-ear chain.
What FDA class is a stapes prosthesis?
Class II under 21 CFR 874.3450; product code LBP (gelatin material) is Class II, as are ETB and LBM under the same regulation.
Which stapes prosthesis types are most used?
A 2023 study reports titanium CliP Piston or crimp prostheses are most widely used, with platinum wire/Teflon implants less so.
Sources
- openFDA device classification database, regulation 874.3450 and product code LBP, api.fda.gov, checked 2026-10-31.
- 21 CFR 874.3450, Partial ossicular replacement prosthesis, Cornell LII, checked 2026-10-31.
- Audiological outcome after stapes surgery in relation to prosthesis type, Eur Arch Otorhinolaryngol 2023, PMC10220147, checked 2026-10-31.
- Comparison of Fluoroplastic Causse Loop Piston and Titanium Soft-Clip in Stapedotomy, Iran J Otorhinolaryngol 2017, PMC5307301, checked 2026-10-31.