Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
General laboratory and product information, not medical advice or a validated procedure. Follow your laboratory’s SOP and the manufacturer’s instructions for use.
Short answer
The starch hydrolysis test checks whether a bacterial isolate secretes amylase, an exoenzyme that breaks starch down. Organisms are grown on starch agar, then the plate is flooded with Gram’s iodine: a colourless zone around the growth means starch was hydrolysed, while a blue or purple zone means it was not (HiMedia technical data M107S, checked 2026-10-30).
Definition and scope
Starch agar is a nutrient agar with soluble starch added, sold for the “detection of starch hydrolysing microorganisms” (HiMedia M107S). The test is a single biochemical reaction, not an identification by itself: it is read alongside other tests. In the United States such media are regulated as differential culture media, which 21 CFR 866.2320 describes as devices used to cultivate and identify types of microorganisms “by a visible change (e.g., a color change) in a specific biochemical component” (law.cornell.edu, checked 2026-10-30).
How the test works in practice
Starch is too large to cross the bacterial cell membrane. An organism that makes the exoenzyme amylase releases it into the medium, where it breaks starch into smaller sugars such as glucose and maltose. Starch is invisible in agar until iodine is added, and iodine with starch gives a dark blue to black colour (Ohio State University Microbiology Lab, LibreTexts 4.03, checked 2026-10-30).
The reading rule in the manufacturer’s data sheet: flood the surface of a 24 to 48 hour culture with Gram’s iodine; hydrolysis appears as a colourless zone around the colonies, and a blue or purple zone indicates starch was not hydrolysed (HiMedia M107S).
What the clear zone must surround
The clearing has to surround the growth itself. LibreTexts cautions that a strong hydrolyser can clear starch around a neighbouring streak, so a clear halo next to another organism’s growth is not a positive result for the organism beside it.
The medium as a manufacturer specifies it
For lab staff who stock or prepare the plate, one manufacturer’s published specification (HiMedia Starch Agar M107S) gives the following:
| Item | Stated specification |
|---|---|
| Meat extract | 3.0 g/L |
| Peptic digest of animal tissue | 5.0 g/L |
| Starch, soluble | 2.0 g/L |
| Agar | 15.0 g/L |
| Dehydrated medium | 25 g suspended in 1000 ml |
| Final pH | 7.2 ± 0.1 at 25°C |
| Sterilisation | Autoclave at 15 lbs (121°C) for 10 minutes |
| Storage | Powder below 30°C, tightly closed; prepared medium at 2-8°C; use before expiry |
The data sheet marks the composition as adjusted and standardised to suit performance parameters, so other manufacturers’ formulations may differ. For background on the sterilisation step, see our guide to what an autoclave is; for another widely used plate medium, see Mueller-Hinton agar.
Quality control strains and expected results
The same data sheet lists the quality control organisms for the medium, incubated at 35-37°C for 18-48 hours with an inoculum of 50-100 CFU. Expected growth is luxuriant with recovery of at least 70%.
| QC strain | Expected starch result |
|---|---|
| Bacillus cereus ATCC 10876 | Positive: clearing around the colony |
| Bacillus subtilis ATCC 6633 | Positive: clearing around the colony |
| Escherichia coli ATCC 25922 | Negative: no clearing |
| Staphylococcus aureus ATCC 25923 | Negative: no clearing |
The purpose of the panel is to run a known positive and a known negative alongside test isolates. General expectations for laboratory quality control are covered in our guide to CLIA quality control requirements, and incubation conditions depend on the incubator and the medium’s data sheet.
Where it matters for labs and supply decisions
Regulatory placement of the medium
Under 21 CFR 866.2320, a differential culture medium is Class I (general controls) and exempt from premarket notification (510(k)), subject to the limitations in section 866.9. The openFDA device classification database (record updated 2026-10-19, checked 2026-10-30) lists product code JSF, “Culture Media, Single Biochemical Test”, under regulation number 866.2320 in the Microbiology specialty.
Standards history
HiMedia states the medium was formulated by Vedder in 1915 for Neisseria (J Infect Dis 1915;16:385) and that its present formulation is accepted by the Bureau of Indian Standards (IS 5887 Part IV, 1976) for detecting starch hydrolysis by Bacillus cereus.
Common misunderstandings
- A clear zone anywhere on the plate is a positive. It must surround the organism’s own growth (see above).
- Starch clearing is visible before iodine is added. Starch is not visible until iodine is flooded onto the surface (LibreTexts 4.03).
- A blue or purple zone means the organism did not grow. Per the data sheet it means starch was not hydrolysed; growth and hydrolysis are separate observations.
- One result identifies the organism. It is one differential biochemical reaction.
FAQ
What does a positive starch hydrolysis test look like?
A colourless zone around the colonies after the plate is flooded with Gram’s iodine (HiMedia M107S).
What does a negative result look like?
A blue or purple zone, meaning starch remains (HiMedia M107S).
Which enzyme is being detected?
Amylase, an exoenzyme that breaks starch into smaller sugars such as glucose and maltose (LibreTexts 4.03).
Is starch agar a regulated device?
In the US, a differential culture medium falls under 21 CFR 866.2320 as a Class I device exempt from 510(k) premarket notification, with the limitations in section 866.9 (checked 2026-10-30).
Sources
- HiMedia Laboratories, Starch Agar technical data M107S: himedialabs.com/media/TD/M107S.pdf (checked 2026-10-30)
- 21 CFR 866.2320, Differential culture medium: law.cornell.edu/cfr/text/21/866.2320 (checked 2026-10-30)
- openFDA device classification, regulation 866.2320: api.fda.gov (checked 2026-10-30)
- Ohio State University Microbiology Lab, 4.03 Starch Hydrolysis: bio.libretexts.org (checked 2026-10-30)