Stericup Vacuum Filtration: Types, Specs & Use

Stericup vacuum filtration systems provide sterile clarification of media and buffers using 0.22 µm PES or PVDF membranes under regulated vacuum pressure.

Written and maintained by CASRAI Editorial Board

Published

On this page

A Stericup is a single-use, disposable vacuum filtration system engineered for the sterile filtration and clarification of cell culture media, biological fluids, and laboratory aqueous buffers. Manufactured by MilliporeSigma (Merck KGaA), the system integrates a top filtration funnel containing a microporous membrane with an ergonomic receiver storage flask via a threaded collar adapter, allowing researchers to rapidly process heat-labile liquids without autoclaving or multi-step decanting.

This page describes products and general laboratory use; it is not medical advice.

What is a Steri cup?

A Stericup (commonly searched as a Steri cup) is an integrated, vacuum-driven filtration device widely used in biomedical research, pharmaceutical development, and analytical testing laboratories. The system consists of four principal components: an upper solution reservoir (the filter funnel), an integrated membrane support base, a vacuum port equipped with a hydrophobic barrier filter, and a receiver storage bottle that collects the sterilized filtrate. Once filtration finishes, the receiver bottle detaches to serve as a standalone, sterile storage container, sealed using an included airtight cap.

From a regulatory standpoint, general laboratory vacuum filtration units fall under the framework for general laboratory equipment as referenced in 21 CFR § 862.2050 by the U.S. Food and Drug Administration (FDA). Unless specifically labeled, cleared, or validated for in vitro diagnostic (IVD) specimen preparation or human clinical infusion, these units are classified for research and routine analytical bench applications. They are manufactured under controlled cleanroom conditions from medical-grade virgin polystyrene and radio-sterilized via gamma irradiation to ensure they are sterile and non-pyrogenic.

In modern cell biology workflows, sterile disposable filter assemblies eliminate the labor-intensive assembly, cleaning, and autoclaving historically required for reusable glass Büchner funnels or stainless steel filter holders. By providing an entirely pre-sterilized fluid path, Stericup units minimize the risk of bioburden introduction and cross-contamination across sensitive primary cell strains and continuous cell culture lines.

How does vacuum filtration work?

Vacuum filtration works by generating a pressure differential between the atmosphere above an open or vented reservoir and the lower pressure inside an evacuated collection vessel. Liquid poured into the upper funnel rests upon a microporous polymer membrane supported by a rigid grid or frit. When an external vacuum source connects to the side port of the collar, air is evacuated from the receiver bottle. The resulting differential pressure forces liquid downward through the microscopic pore network of the membrane, while particles, precipitates, and microbial cells exceeding the effective pore diameter are retained on the upstream surface.

The driving force governing flow through porous membranes is described hydrodynamically by the Hagen-Poiseuille relationship, where liquid flux is directly proportional to the applied differential pressure and inversely proportional to the liquid viscosity and membrane flow resistance. Under standard laboratory conditions, the safe operating differential vacuum should not exceed 700 mmHg (~0.93 bar or 13.5 psi) at 25 °C. Subjecting plastic filtration assemblies to excessive vacuum pressure does not indefinitely accelerate filtration; instead, it risks mechanical collapse of the plastic receiver, membrane pore compaction, and rapid foaming of protein-containing fluids.

Benchtop laboratory vacuum filtration follows a structured operational sequence to preserve sterility and hardware integrity:

  1. Inspection and Aseptic Unpacking: In a laminar flow biosafety cabinet, inspect the sterile packaging for tears, punctures, or seal compromise. Remove the Stericup assembly and verify that the receiver bottle and funnel collar are firmly engaged.
  2. Vacuum Line Connection: Attach flexible vacuum tubing to the side hose barb on the filter collar. Ensure an in-line hydrophobic trap and a calibrated vacuum regulator protect the vacuum pump from accidental liquid overflow or aerosol ingestion.
  3. Sample Introduction: Uncap the upper funnel and pour the liquid reagent or media gently down the funnel wall to minimize bubbling. Avoid touching the inner rim or membrane surface with pipettes or bottle necks.
  4. Vacuum Application and Flow Control: Initiate vacuum gradually, adjusting the line pressure to approximately 300 to 500 mmHg. Monitor the filtrate stream entering the receiver bottle. Maintain vacuum until the entire liquid volume has passed through the membrane and the filter runs dry.
  5. Disconnection and Secure Storage: Turn off the vacuum supply and vent the line before pulling the tubing off the collar. Unscrew the funnel from the receiver bottle, promptly transfer the sterile cap from its packaging onto the receiver bottle, and thread it down until the seal clicks firmly into position.

Single-use vacuum filtration unit types and membrane architectures

Single-use vacuum filtration devices are engineered in multiple mechanical configurations and polymer chemistries to accommodate diverse laboratory liquid volumes, chemical compatibilities, and sustainability goals. The traditional Stericup Quick Release platform couples an upper funnel to a receiver bottle via a quick-turn threaded collar. The receiver bottles feature indented, curved grip profiles and low centers of gravity to prevent tipping inside biosafety hoods. Standard receiver capacities include 150 mL, 250 mL, 500 mL, and 1,000 mL volumes.

For laboratories seeking to reduce plastic consumption, funnel-less eco-designs such as Stericup E thread directly onto pre-existing commercial media or reagent glass and plastic bottles with 38 mm or 45 mm neck threads. The user adds reagents directly into the manufacturer base bottle, screws the compact filter collar into place, couples the receiver, and inverts the assembly to filter. Similarly, bottle-top units (such as Steritop) consist solely of the upper filter funnel and threaded collar, allowing direct filtration onto standard laboratory GL45 threaded glass bottles.

The core determinant of filtration performance is the membrane polymer. Manufacturers formulate vacuum filtration units with distinct polymers, summarized in the comparative reference table below:

Membrane Polymer Common Pore Sizes Aqueous Flux Rate Protein Binding Profile Primary Laboratory Use
Polyethersulfone (PES / Millipore Express PLUS) 0.1 µm, 0.22 µm Very High Low Cell culture media, serum-supplemented broths, biological buffers
Polyvinylidene Fluoride (PVDF / Durapore) 0.22 µm, 0.45 µm Moderate to High Extremely Low Antibody purification, dilute protein solutions, high-value biologics
Mixed Cellulose Esters (MCE) 0.22 µm, 0.45 µm High Moderate to High General aqueous buffers, biological clarification, microbial analysis
Hydrophilic Nylon 0.2 µm, 0.45 µm Moderate High Aqueous and mixed-solvent HPLC mobile phases, tissue culture without serum

Polyethersulfone (PES) is the dominant choice for tissue culture because its asymmetric pore geometry provides exceptionally high volumetric throughput and low non-specific protein adsorption. In contrast, polyvinylidene fluoride (PVDF) is selected when processing dilute, high-value cytokines or therapeutic proteins where even slight adsorption to the filter polymer would compromise experimental recovery. Understanding these differences distinguishes modern vacuum filtration from basic gravity filtration techniques that rely on uncalibrated paper filters.

Which filter is used for sterile filtration?

For liquid sterilization, the benchmark standard is a 0.22 µm (or 0.2 µm) sterilizing-grade membrane filter. Under the U.S. FDA Guidance for Industry on Sterile Drug Products Produced by Aseptic Processing, a sterilizing-grade filter is formally defined as a membrane capable of retaining a minimum concentration of 107 colony-forming units per square centimeter (CFU/cm²) of effective filter area of the diminutive test bacterium Brevundimonas diminuta (ATCC 19146) under validated operating pressure conditions. This microbial retention standard is standardized across the filtration industry by ASTM International Standard F838-20, requiring zero detectable organisms in the effluent filtrate, which represents a log reduction value (LRV) of 7 or greater.

In specialized cell culture applications, standard 0.22 µm filtration may prove insufficient. Mycoplasmas—such as Acholeplasma laidlawii and Mycoplasma orale—are pleomorphic, wall-less bacteria that lack a peptidoglycan envelope, allowing them to deform and traverse the pores of a standard 0.22 µm membrane during prolonged vacuum exposure. For bovine serum and cell culture stocks susceptible to mycoplasma contamination, laboratories utilize 0.1 µm pore-rated PES membranes. Conversely, 0.45 µm membranes are classified as clarification or bioburden-reduction filters; they retain larger bacteria and airborne debris but cannot be relied upon for complete liquid sterilization.

In industrial pharmaceutical production and cGMP compounding, filter integrity must be certified before and after use via non-destructive testing, such as bubble point, diffusive flow, or water intrusion testing. The bubble point test measures the minimum gas pressure required to overcome liquid surface tension and force air through the largest wet pores of the membrane, as defined by the Young-Laplace equation. While single-use disposable Stericup units are pre-validated by lot at the factory and cannot be non-destructively bubble-point tested on a benchtop without breaching downstream sterility, researchers can refer to technical standards for sterilizing-grade membrane filters to understand the correlation between membrane bubble point specifications and bacterial challenge retention.

What are the disadvantages of vacuum filtration?

While vacuum filtration offers speed and convenience, it possesses distinct physical, biochemical, and operational disadvantages that laboratory researchers must manage carefully:

  • Protein Denaturation and Foaming: Applying strong negative pressure to solutions containing proteins, bovine serum albumin (BSA), or fetal bovine serum (FBS) causes vigorous aeration and frothing in the receiver flask. The rapid expansion of air bubbles creates high shear stress at gas-liquid interfaces, which can denature delicate globular proteins, cause antibody aggregation, and induce loss of enzymatic activity.
  • Volatile Component Evaporation and pH Drift: Under reduced pressure, the boiling point of volatile solutes decreases. Vacuum filtration can rapidly degas dissolved carbon dioxide (CO₂) from bicarbonate-buffered cell culture media (such as DMEM or RPMI 1640), driving an immediate alkaline pH shift. Volatile supplements such as ethanol, acetone, or mercaptoethanol will also evaporate preferentially into the vacuum exhaust.
  • Membrane Fouling and Cake Resistance: Vacuum filtration operates in a dead-end mode, meaning the entirety of the fluid stream passes perpendicularly through the filter. Colloidal particles, cellular fragments, and macromolecular aggregates accumulate directly on the membrane surface, forming an impermeable filter cake. In viscous solutions, such as dense bacterial lysates or whole serum, flow rates can drop precipitously within minutes, stranding unsterilized liquid in the funnel.
  • Fixed Pressure Ceilings: Unlike positive-pressure filtration systems (such as peristaltic pump-driven capsule filters or syringe filters) that can generate 2 to 3 bar of driving head, vacuum filtration is physically capped by atmospheric pressure (approximately 1 bar / 14.7 psi absolute). Consequently, highly viscous liquids cannot be forced through tight 0.1 µm or 0.22 µm pores by simply increasing vacuum power.
  • Implosion and Aerosol Hazards: Evacuated vessels store mechanical strain energy. Glass or rigid plastic receiver bottles harboring hairline fractures, manufacturing defects, or deep scratches can implode violently under vacuum, projecting sharp debris and aerosolizing potentially biohazardous fluids across the laboratory.

Laboratories that prepare large volumes of aqueous solutions often weigh vacuum units against alternatives, such as positive-pressure filtration or dedicated systems evaluated in guides to filtration sterilization mechanisms and water purification protocols.

Laboratory workflow, handling practices, and safety standards

Operating vacuum filtration equipment safely and reliably requires strict adherence to institutional biosafety procedures and chemical hygiene standards. The World Health Organization (WHO) Laboratory Biosafety Manual (4th edition) highlights vacuum filtration as a potential source of laboratory aerosol generation and liquid aspiration accidents. When filtering biological samples or recombinant cell cultures, all vacuum manipulation must occur within a certified Class II biosafety cabinet to protect the operator and room environment from fugitive droplets.

To safeguard facility infrastructure, vacuum connections must never run directly from a benchtop filtration funnel into a building vacuum line or house pump. The Occupational Safety and Health Administration (OSHA) laboratory safety standard, codified at 29 CFR 1910.1450, requires chemical hygiene controls to prevent hazardous substances from entering laboratory atmospheres. A robust vacuum station incorporates a dual-flask overflow trap assembly charged with an appropriate disinfectant (such as 10% sodium hypochlorite) positioned between the receiver bottle and the vacuum tap. A hydrophobic 0.2 µm PTFE in-line disc filter must be inserted downstream of the trap to physically arrest aqueous aerosols and foaming liquid before they can reach the house vacuum plumbing or an electric diaphragm pump.

Furthermore, operators must inspect all receiver flasks prior to evacuation. Polystyrene vessels must never be cleaned with aggressive organic solvents (such as acetone, ethyl acetate, or chloroform) that induce environmental stress cracking. When reagent solutions require high-purity water, researchers must confirm the quality of their water feedstocks, as analyzed in comparative studies of distilled vs sterile water preparation in laboratory environments.

Common misunderstandings in sterile vacuum filtration

Misconceptions regarding membrane capabilities and equipment limitations frequently compromise laboratory experiments. Addressing these operational errors ensures protocols achieve true sterility without unexpected biochemical artifacts:

  • Sterile Filtration Eliminates Endotoxins: A persistent misunderstanding is that passing a buffer through a 0.22 µm Stericup unit removes bacterial endotoxins (lipopolysaccharides, LPS). In reality, bacterial endotoxin monomers and micelles range from 10,000 to 1,000,000 Daltons (typically under 0.05 µm in diameter). While the membrane retains intact Gram-negative bacterial cells, free endotoxins pass unimpeded into the filtrate. True depyrogenation requires ultrafiltration membranes with a molecular weight cutoff of 10 kDa or specialized positively charged endotoxin-adsorption matrices.
  • Polystyrene Filter Units Can Be Autoclaved: Some laboratories mistakenly attempt to re-sterilize used Stericup funnels or receiver bottles in an autoclave. Medical-grade polystyrene exhibits a glass transition temperature of approximately 90 °C to 100 °C and softens rapidly above 50 °C. Autoclaving at standard cycle parameters (121 °C at 15 psi) will completely melt the housing, distort the threaded neck, and delaminate the filter membrane. Stericup units are strictly single-use disposables.
  • Higher Vacuum Always Accelerates Slow Filtration: When a filter membrane begins to foul with particulate matter, operators often turn the vacuum regulator to maximum negative pressure. However, increased vacuum compresses flexible, gelatinous biological filter cakes, dramatically decreasing cake porosity. The resulting compaction elevates flow resistance and can completely arrest fluid passage while increasing the probability of membrane micro-tears.
  • All 0.22 µm Membranes Behave Identically with Proteins: Membrane chemical composition dictates surface adsorption. Untreated nylon or mixed cellulose ester membranes possess significant binding capacities for proteins and nucleic acids, leading to substantial yield loss when filtering low-concentration recombinant proteins. Low-binding polymers such as PVDF (Durapore) or surface-modified PES must be deliberately selected when protein recovery is a critical experimental variable.

The data table below compares technical specifications across common single-use vacuum filtration systems, including volume formats, membrane polymer chemistries, effective filtration areas, and maximum differential vacuum limits.

Frequently asked questions

What is a Steri cup?

A Stericup is a disposable, gamma-irradiated vacuum filtration unit manufactured by MilliporeSigma that combines a filter funnel, a microporous membrane (typically PES or PVDF), and a receiver storage flask into a single pre-sterilized assembly for preparing sterile cell culture media, serum, and aqueous laboratory reagents.

What are the disadvantages of vacuum filtration?

Key disadvantages of vacuum filtration include protein foaming and shear-induced denaturation, rapid evaporation of volatile solvents and buffer degassing leading to pH shifts, dead-end membrane clogging from particulate cakes, fixed pressure limits capped by atmospheric pressure, and the risk of vessel implosion under high vacuum.

Which filter is used for sterile filtration?

Sterile filtration standardly utilizes a 0.22 µm (or 0.2 µm) pore-size membrane filter that meets the ASTM F838-20 retention standard by completely retaining a minimum challenge of 107 CFU/cm² of Brevundimonas diminuta. For mycoplasma clearance in critical bovine serum or cell lines, a tighter 0.1 µm membrane is used.

How does vacuum filtration work?

Vacuum filtration operates by applying a mechanical vacuum to an airtight receiver flask below a microporous membrane. The atmospheric pressure above the fluid column forces liquid downward through the membrane pores into the collection vessel while particulate contaminants and microorganisms larger than the pore rating are retained on the filter surface.

Can a 0.22 micron vacuum filter remove endotoxins?

No. A standard 0.22 µm membrane filter retains whole bacterial cells but does not remove bacterial endotoxins (lipopolysaccharides). Endotoxins are sub-micron biomolecules that pass freely through 0.22 µm pores; their removal requires ultrafiltration with a 10 kDa molecular weight cutoff or specialized affinity adsorption media.

Can Stericup vacuum filtration units be autoclaved or reused?

No. Stericup units are manufactured from polystyrene, which has a maximum temperature threshold of 50 °C (122 °F). Exposing the unit to standard autoclave temperatures (121 °C) melts the plastic housing, warps the threaded seals, and destroys membrane integrity. The units are engineered strictly for single-use applications.

Sources

Data: Single-Use Vacuum Filtration Unit Specifications and Operational Parameters

Original compilation comparing single-use sterile vacuum filtration systems and membrane architectures across volume capacity, polymer chemistry, pore size, effective filtration area, and vacuum limits, sourced from FDA guidance, ASTM F838 standards, and manufacturer technical datasheets.

Period covered: 2024/2026. Coverage: Global.

Single-Use Vacuum Filtration Unit Specifications and Operational Parameters
System ConfigurationMembrane PolymerPore Size (µm)Funnel Volume (mL)Filtration Area (cm²)Max Differential Vacuum (mmHg)Primary Laboratory Application
Stericup Quick Release GPPolyethersulfone (PES)0.2250040700Sterile filtration of cell culture media and buffers
Stericup Quick Release GVPolyvinylidene Fluoride (PVDF)0.2250040700Sterile clarification of dilute protein and antibody solutions
Stericup Quick Release VPPolyethersulfone (PES)0.150040700Mycoplasma reduction in serum and critical cell lines
Stericup Quick Release HVPolyvinylidene Fluoride (PVDF)0.4550040700Coarse clarification and biological fluid pre-filtration
Stericup E-GP Funnel-LessPolyethersulfone (PES)0.22100040700Eco-friendly high-volume media preparation with direct bottle coupling
Steritop Quick Release GPPolyethersulfone (PES)0.22100040700Bottle-top sterile filtration onto standard GL45 glass storage bottles

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Stericup Vacuum Filtration: Types, Specs & Use

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →