Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
Transepidermal water loss (TEWL) is the amount of water vapour that passes outward through the epidermis and evaporates from the skin surface. Because it rises when the skin’s barrier is damaged, it is used as a non-invasive index of skin barrier function. In research it is measured with a probe that reads the water-vapour gradient just above the skin, and the numbers are only comparable when device type, room conditions and reporting method are controlled.
Definition and scope
The 2015 international guidelines for assessing skin properties in non-clinical settings describe the skin as a physical barrier that prevents loss of body fluids and the entry of chemicals or infectious agents, with the permeability barrier residing mainly in the stratum corneum, the outermost layer. In that framing, TEWL represents the diffusion of water through the stratum corneum, and it is widely used together with skin hydration as an index of barrier function.
A 2023 systematic review of TEWL devices defines the quantity as the flux in water-vapour density at the skin surface compared with a point further away: water lost from the stratum corneum raises the humidity next to the skin, and the instrument measures how steeply that humidity falls with distance. The same review states that TEWL increases when the epidermis is damaged, which is why it is a commonly used measure of skin barrier integrity.
This page covers TEWL as a laboratory and research measurement: what it is, how instruments obtain it, and what a study has to control and report. It does not interpret an individual person’s reading and is not medical advice. It also does not give reference ranges; the sources used here do not offer a single normal value that applies across devices, and the guidance they give is to compare changes rather than absolute numbers.
How TEWL is measured in practice
All the common instruments estimate the vapour-pressure (or vapour-density) gradient in the thin layer of air above the skin. They differ in how they shield that air layer from the room and how they handle the humidity that builds up when the probe sits on the skin.
Open-chamber instruments
In the open-chamber approach the gradient is calculated from the difference in vapour pressure between two points aligned perpendicularly to the skin. Vapour pressure is derived from relative humidity, read by capacitive sensors, and from temperature, read by fast thermistors, all located in a cylinder with two open ends: one end rests on the skin, the other lets water vapour escape into the room. The 2023 review notes the advantage that this design does not occlude the skin, so the microclimate stays undisturbed, but it is vulnerable to air movement; the EEMCO account it summarises adds that open-chamber measurements vary with temperature, humidity and air movement. The 2015 guideline therefore says a draft shield is required if an open-chamber instrument is used.
Semi-open chambers
A semi-open chamber protects the measuring volume from ambient airflow without letting humidity accumulate the way it can in a fully closed chamber, according to the review’s summary of the EEMCO guidance.
Unventilated closed chambers
A closed chamber is protected from air movement, but humidity builds up inside it. The 2023 review states that the chamber must be lifted from the skin after each reading and that continuous measurement is not possible. The 2015 guideline reports that the measuring time of unventilated closed-chamber instruments is very short (under 10 seconds) and that some are portable; it also notes that some manufacturers claim readings are unaffected by probe angle while several studies have reported an angular dependence.
Condenser-chamber instruments
In the condenser-chamber method, the small measuring cylinder is closed at the top by a condenser held below the freezing point of water. This creates a humidity gradient that allows water vapour to be measured continuously. The 2023 review gives the AquaFlux AF200 as an example: a closed-type device with a condenser cooled to -7.65 °C, a 7 × 7 mm opening and continuous measurement. It also describes GPSkin, a semi-closed device that reports TEWL, hydration and temperature and connects to a smartphone over Bluetooth.
Chamber designs compared
| Design | What it avoids | Trade-off stated in the sources |
|---|---|---|
| Open chamber | Occluding the skin; the microclimate is left undisturbed | Vulnerable to air movement; the 2015 guideline requires a draft shield |
| Semi-open chamber | Ambient airflow, without humidity build-up | Does not allow humidity to build up as a closed chamber can; DermaLab is the semi-open example in the EEMCO guidance |
| Unventilated closed chamber | Air movement; very short measuring time (under 10 s); some units portable | Humidity builds up, so the chamber is lifted after each reading and cannot measure continuously; angular dependence reported in several studies; concerns have been expressed about sensitivity of one portable device |
| Condenser chamber | Humidity build-up, because a sub-freezing condenser sets the gradient | Closed-type; the example device in the review has a fixed 7 × 7 mm opening |
The data table further down lists named instruments with their design and manufacturer as recorded in the 2015 guideline and the 2023 review. The review’s own conclusion is that it remains unclear which chamber type is most valid and reliable.
Measurement conditions that a study has to control
The 2015 guideline groups the sources of variation into endogenous, exogenous, environmental and instrument-related factors, and recommends that investigators minimise or at least recognise each. The conditions it quotes from the EEMCO clinical recommendations can be read as a checklist for a protocol or a standard operating procedure (see standard operating procedure):
- Acclimatisation: at least 15 to 30 minutes at an ambient temperature of 20 to 22 °C and 40 to 60% relative humidity before measuring TEWL; at least 20 minutes for skin hydration measurements.
- Skin exposure: the measurement site exposed to ambient air for at least 10 minutes beforehand.
- Sweating: thermal, emotional and physical sweating raise TEWL and hydration values; adequate acclimatisation and controlled temperature and humidity are how the guideline says to control it.
- Skin and probe temperature: differences between skin and probe temperature can influence TEWL values, so probe heaters (where manufacturers provide them) are recommended, and skin temperature is preferably measured beforehand. The probe should be handled with an insulated glove or other indirect means, because holding it raises the reading.
- Air movement: use a draft shield with open-chamber devices, or choose a closed-chamber device. For workplace studies the guideline recommends a closed-chamber instrument because it is not influenced by air movement.
- Anatomical site: record the site, since the guideline notes studies in which positions varied widely, and record factors that influence values so they can be considered when data are interpreted.
- Calibration and method reporting: report acclimatisation time, calibration of instruments and number of measurements per site. The guideline notes these details were frequently missing in earlier workplace studies, which hindered comparison.
Operationally this is a method-reproducibility problem: another laboratory can only repeat a TEWL result if these conditions are written down (compare methods reproducibility and the broader term reproducibility).
How results should be reported
The 2015 guideline’s four key elements are to recognise the influencing factors, to measure TEWL with a closed-chamber instrument, to report results as a difference or percent change rather than absolute values, and to report any deviations from the guideline. It gives two illustrations: for acute changes caused by an exposure, quantify the difference relative to the same person’s baseline before the exposure; for chronic changes caused by exposure or disease, express the difference between the exposed group and a matched control as a percentage. Results at a given site are expressed as the arithmetic mean and standard deviation.
The reason for relative reporting is stated in the 2023 review’s account of the EEMCO conclusion: measurements taken with different instruments cannot be compared with each other, but each instrument is useful on its own in comparative studies, such as comparing two moisturisers, young with old skin, or diseased with healthy skin. A numeric value from one device does not transfer to another.
TEWL compared with skin hydration
TEWL and hydration are often reported together but measure different things. The 2015 guideline describes TEWL as the diffusion of water through the stratum corneum and skin hydration as the water content of the stratum corneum.
| TEWL | Skin hydration | |
|---|---|---|
| What it reflects | Outward diffusion of water through the stratum corneum | Water content of the stratum corneum |
| Instrument principle | Water-vapour gradient above the skin (open, semi-open, closed or condenser chamber) | Electrical: capacitance (for example Corneometer), conductance (for example Skicon) or impedance (for example the DermaLab Moisture Unit) |
| Recommended acclimatisation | At least 15 to 30 minutes | At least 20 minutes |
For hydration sensors, body hair can interfere with contact between sensor and skin, and the guideline advises avoiding very hairy sites or removing hair and letting the skin recover first. A product can change one without the other, so a study that wants to claim an effect on barrier function and an effect on hydration should measure both and say which instrument produced which number.
Where TEWL matters for laboratories and supply decisions
Cosmetic and topical product testing
The EEMCO guidance for cosmetic sciences (Rogiers, 2001) states that open-chamber water-vapour-gradient TEWL is used to support cosmetic claims such as product mildness, reduction of irritative reactions, skin hydration, skin repair and protection against UV damage. It adds that TEWL can screen ingredients for beneficial effects on the barrier and can monitor topical treatment in vivo without invasive sampling. The same abstract warns that many variables affect results and that working under standardised conditions is of the utmost importance for reliable and reproducible results.
Occupational skin studies
The 2015 guideline was written for non-clinical settings such as workplaces, where it reviews studies of exposed workers. Two of the findings it cites are cautions rather than endorsements: in a study of apprentice hairdressers, a hand TEWL above 15 g/m²/h was associated with higher hand-dermatitis risk but the increase was not statistically significant, and in a study of metalworker trainees no single skin bioengineering method proved valid as a screening test for hand eczema. TEWL is better treated as a research measure of barrier change than as a stand-alone screening cut-off.
Device selection for a protocol
The 2023 review pooled 22 devices across 38 studies. It found the study design was on average rated “adequate”, while the statistical methods for reliability and measurement error were on average rated “doubtful”; the devices showed good reliability and often correlated with one another, and measurement error was highly variable but improved under in vitro conditions. For a protocol writer that supports three practical points taken from the sources: pick a design that suits the setting (closed or shielded where air movement is likely), use the same instrument throughout a study, and state the instrument model in the report. Skin barrier products are also described elsewhere on this site, for example solid skin barriers and silicone adhesives, and in vitro permeation models are covered in diffusion cells.
Common misunderstandings
- “TEWL is the same as sweating.” No. The guideline treats sweating as a confounder that raises TEWL and hydration readings and that acclimatisation is meant to remove.
- “TEWL is the same as skin hydration.” No. They are different measures with different instruments (see the comparison above).
- “A TEWL value from one instrument can be compared with another instrument’s value.” The EEMCO conclusion is that measurements from different instruments cannot be compared.
- “There is a single reading that separates healthy from damaged skin.” The sources reviewed give no such universal threshold, and the occupational studies above show why a fixed cut-off is a poor basis for conclusions.
- “Open-chamber, closed-chamber and condenser devices are interchangeable.” They manage airflow and humidity build-up differently, which changes both how they are used and how they should be reported.
- “A handheld probe reading is just a probe reading.” Handling matters: holding the probe raises the reading, and angle dependence has been reported for closed-chamber types.
Frequently asked questions
What does transepidermal water loss mean?
It is the outward movement of water through the epidermis to the skin surface, where it evaporates. In research it is quantified as a flux, with the 2015 guideline’s example values given in g/m²/h.
What does a higher TEWL indicate?
According to the 2023 systematic review, TEWL increases when the epidermis is damaged, so a rise relative to baseline is read as reduced barrier integrity. Sweating and uncontrolled room conditions can also raise it, which is why measurement conditions are standardised.
How is TEWL measured?
With a probe held on the skin that reads the water-vapour gradient in the air just above it, using an open, semi-open, closed or condenser chamber design. Participants first acclimatise for at least 15 to 30 minutes at 20 to 22 °C and 40 to 60% relative humidity.
Is TEWL the same as skin moisture?
No. TEWL is water leaving through the stratum corneum; hydration is the water content of the stratum corneum and is measured electrically, for instance by capacitance.
Can I compare TEWL values from different devices?
The EEMCO conclusion summarised in the 2023 review is that values from different instruments cannot be compared. Use one instrument per study and report change from baseline or percent difference from a control.
Which TEWL device is best?
The 2023 review states that it is unclear which chamber type is most valid and reliable, so the sources do not support a general ranking. The choice should follow the study setting and the conditions you can control.
Sources
- Sources checked 2026-11-12: International guidelines for the in vivo assessment of skin properties in non-clinical settings: Part 2. Transepidermal water loss and skin hydration (Skin Research and Technology, 2015).
- Devices measuring transepidermal water loss: a systematic review of measurement properties (2023).
- Rogiers V. EEMCO guidance for the assessment of transepidermal water loss in cosmetic sciences, Skin Pharmacology and Applied Skin Physiology 2001;14:117-128 (abstract).
Data: TEWL measurement instruments: chamber design and manufacturer
Compilation of named transepidermal water loss instruments with their chamber design and manufacturer, taken from the 2015 international non-clinical skin-property guideline instrument table and the 2023 COSMIN systematic review of TEWL devices.
Period covered: 2015/2023.
| Instrument | Chamber design | Manufacturer | Note |
|---|---|---|---|
| Evaporimeter EP1 / EP2 | Open chamber | ServoMed, Stockholm, Sweden | Listed in 2015 guideline instrument table |
| Tewameter TM210 / TM300 | Open chamber | Courage & Khazaka, Cologne, Germany | Listed in 2015 guideline instrument table |
| DermaLab | Open chamber (2015 table); semi-open example in EEMCO 2018 as summarised in 2023 review | Cortex Technology, Hadsund, Denmark | Classification differs between the two sources |
| VapoMeter SWL3 | Unventilated closed chamber | Delfin Technologies, Kuopio, Finland | Portable handheld; concerns about sensitivity noted in 2023 review |
| H4300 | Unventilated closed chamber | Nikkiso-YSI, Tokyo, Japan | No longer manufactured per 2015 guideline |
| AquaFlux AF200 | Condenser chamber (closed type) | Biox Systems, London, UK | Condenser cooled to -7.65 C; continuous measurement; 7 x 7 mm opening |
| GPSkin | Semi-closed | GPower, Seoul, South Korea | Reports TEWL, hydration and temperature; Bluetooth to smartphone |
Sources
- International guidelines for the in vivo assessment of skin properties in non-clinical settings: Part 2. TEWL and skin hydration (accessed 2026-11-12)
- Devices measuring transepidermal water loss: a systematic review of measurement properties (accessed 2026-11-12)