Written and maintained by CASRAI Editorial Board
Published
This page describes products and general use; it is not medical advice.
On this page
Polyethylene gloves are disposable gloves made from polyethylene, a thermoplastic polymer. Whether a given polyethylene glove counts as a medical examination glove is a regulatory question, not a material question: in the United States, an examination glove is a device defined in 21 CFR 880.6250, and a few polyethylene examination gloves have been cleared by FDA under product code LZA. This page explains what separates those from other polyethylene gloves, and lists the specifications from one real FDA-cleared example.
Who uses them: the regulatory definition of a patient examination glove is a device worn on the examiner’s hand or finger to prevent contamination between patient and examiner, so clinic and laboratory staff who handle examination gloves are the audience this page is written for.
What polyethylene gloves are
FDA’s definition of a non-powdered patient examination glove (21 CFR 880.6250(a)) is “a disposable device intended for medical purposes that is worn on the examiner’s hand or finger to prevent contamination between patient and examiner.” The regulation classifies these gloves as Class I, subject to general controls. The definition does not name a material, which is why gloves made from natural rubber latex, vinyl, nitrile and other synthetic polymers are all listed under the same regulation number.
Polyethylene is one of the synthetic polymers used. In FDA’s product classification database, polyethylene examination gloves that have been cleared sit under product code LZA, “Polymer Patient Examination Glove.” For a comparison with the more common materials, see the comparison nitrile vs latex vs vinyl gloves.
How the material is defined
Under 21 CFR 177.1520, the food-contact regulation for olefin polymers, “polyethylene consists of basic polymers manufactured by the catalytic polymerization of ethylene.” The chemistry of the polymer itself is covered on polyolefin vs polyethylene. For gloves, the practical point is that the polymer definition says nothing about whether a finished glove is cleared as a medical device.
Examination glove versus other polyethylene gloves
Two different parts of the FDA’s rulebook can touch a polyethylene glove, and they answer different questions.
| Regulation | What it covers | What it tells a reader |
|---|---|---|
| 21 CFR 880.6250 (non-powdered patient examination glove) | Disposable gloves for medical purposes, worn to prevent contamination between patient and examiner. Class I. | Applies to a glove marketed as an examination glove. Cleared polyethylene examples are listed under product code LZA. |
| 21 CFR 177.1520 (olefin polymers) | Olefin polymers, including polyethylene, permitted in food-contact articles under stated conditions. | Concerns food-contact use of the polymer. It is not an examination-glove clearance. |
The consequence: a glove sold as a “poly glove” is not automatically an examination glove. The label, the product listing and the FDA 510(k) database are where the examination-glove claim would appear.
Examination glove product codes
openFDA lists the examination glove materials below under the same regulation, 880.6250, all Class I (read 2026-11-08).
| Product code | Name in FDA database | Regulation / class |
|---|---|---|
| LZA | Polymer Patient Examination Glove (nitrile or synthetic polymers; this is where cleared polyethylene gloves sit) | 880.6250, Class I |
| LYY | Latex Patient Examination Glove | 880.6250, Class I |
| LYZ | Vinyl Patient Examination Glove | 880.6250, Class I |
| FMC | Patient Examination Glove | 880.6250, Class I |
| OIG | Guayule rubber examination glove | 880.6250, Class I |
| OPC | Polychloroprene examination glove | 880.6250, Class I |
The openFDA 510(k) database returns two polyethylene-named examination glove clearances, both under LZA: K162626, “Thermoplastic Polyethylene Examination (TPE) Glove” (decision date 2017-03-31), and K222531, “Synthetic Polymer Glove – Polyethylene (Black)” (decision date 2023-01-22).
Sizes and specifications: one cleared example
Specifications vary by manufacturer. The values below are from the FDA 510(k) summary for K222531, one manufacturer’s cleared device, and are not a general standard for all polyethylene gloves. The summary describes a powder-free, non-sterile, black glove made from low density polyethylene, intended for over-the-counter use, and states it meets the specifications in ASTM D6319-19 (the standard titled for nitrile examination gloves).
| Property | Value stated in the K222531 summary | Test method named |
|---|---|---|
| Sizes | Small, Medium, Large, X-Large, XX-Large | – |
| Overall length | 230 mm minimum | – |
| Width (tolerance +/-10 mm) | S 80 mm, M 95 mm, L 110 mm, XL 120 mm, XXL 130 mm | – |
| Palm thickness | 0.05 mm minimum (tested average 0.06 mm) | – |
| Residual powder | Less than 2.0 mg per glove | ASTM D6124 |
| Pinholes | AQL 2.5 | ASTM D5151 |
| Tensile properties | Tested | ASTM D412 |
For how thickness and sizing are expressed on other glove types, see the glossary entries nitrile glove thickness (mil) and nitrile glove sizing.
How to check a polyethylene glove
These steps follow directly from the regulatory sources above:
- Look for an examination-glove claim on the labeling. Under 21 CFR 880.6250 the intended purpose is preventing contamination between patient and examiner.
- Look up the product code and 510(k) number in the FDA database. Cleared polyethylene examination gloves appear under LZA.
- Compare the stated size, length, thickness and test-method values with the manufacturer’s own cleared summary, as in the K222531 table.
Safe use basics
The K222531 summary describes the glove as a non-sterile disposable device, and 880.6250 defines examination gloves as disposable. A disposable glove is a single-use item. Follow the manufacturer’s instructions for use for the specific product and your facility’s infection prevention policy, such as the guidance in gloves for health care.
Common mistakes
- Assuming any polyethylene glove is an examination glove. The material alone does not establish clearance; the label and the FDA listing do.
- Treating one manufacturer’s specification as universal. The width, thickness and powder values above belong to one cleared device.
- Mixing up exam-glove and food-contact rules. 21 CFR 177.1520 and 21 CFR 880.6250 govern different questions.
FAQ
Are polyethylene gloves FDA-cleared medical gloves?
Some are. The FDA 510(k) database lists two polyethylene-named examination gloves, K162626 (2017) and K222531 (2023), both under product code LZA. A polyethylene glove without such a clearance is not shown to be an examination glove by that fact alone.
Are examination gloves a Class I device?
Yes. 21 CFR 880.6250 places non-powdered patient examination gloves in Class I, subject to general controls.
Sources
- 21 CFR 880.6250, Non-powdered patient examination glove (read 2026-11-08)
- openFDA device classification API, glove product codes (read 2026-11-08)
- openFDA 510(k) API, polyethylene glove clearances (read 2026-11-08)
- FDA 510(k) summary K222531 (read 2026-11-08)
- 21 CFR 177.1520, Olefin polymers (read 2026-11-08)