Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
Lead gloves, in the medical sense, are hand-worn radiation shields. They are flexible gloves containing lead (or a lead-based layer) that a wearer uses near an X-ray beam to reduce the dose that scattered radiation delivers to the hands. People also search for them as x ray gloves. The FDA lists them as a regulated device type called a “radiation attenuating medical glove.”
This page describes products and general use; it is not medical advice.
What lead gloves are
A lead glove is protective clothing for staff who work in or near fluoroscopy or radiography. The FDA’s product classification for code OPH describes radiation attenuating medical gloves as gloves used during medical or surgical procedures involving radiation, to shield the wearer’s hands from scattered radiation. The same record says they offer some degree of protection from radiation exposure as well as protection from the transmission of infectious agents.
Who uses them: radiology and imaging staff and other personnel whose hands are near the field during fluoroscopy or radiography. The word “lead” also appears in unrelated products, such as gloves for handling lead-contaminated material or glove boxes; those are different items. For enclosed handling, see what a glove box is.
How lead gloves are built
The International Federation of Red Cross and Red Crescent Societies (IFRC) Items Catalogue lists a “lead gloves for X-ray protection” item built from several flexible lead-rubber or lead-vinyl layers, made from lead with natural rubber or vinyl. The listing says the set is supplied with inner gloves of 100% cotton, along with a multilingual user and maintenance manual, and that the product carries a CE mark. Many ordinary protective gloves are chosen for chemical or barrier performance; the lead layers are what make this glove a radiation device, which is why it is regulated separately from a plain nitrile or latex exam glove.
Regulatory identity and standards
| Item | What the record says | Source |
|---|---|---|
| FDA device name and code | Radiation Attenuating Medical Glove, product code OPH | FDA product classification database |
| Device class | Class 1, General Hospital review panel, 510(k) submission type | FDA product classification database |
| Regulation number | 21 CFR 880.6250 (the regulation for non-powdered patient examination gloves) | FDA product classification database |
| Extra expectation for makers | In addition to meeting the requirements for examination or surgeon’s gloves, manufacturers should have technical data showing that their attenuation claims meet the energy range of x-rays normally used in medical procedures | FDA product classification database |
| Recognized consensus standard | ASTM D7866-23, Standard Specification for Radiation Attenuating Protective Gloves | FDA product classification page for OPH |
| Catalogue example standard | EN 61331, protective devices against diagnostic medical X-radiation | IFRC Items Catalogue |
The regulation text in 21 CFR 880.6250 itself describes a disposable examination glove worn to prevent contamination between patient and examiner; the attenuation language comes from the FDA’s product classification entry, not from the regulation.
Sizes and specifications that matter
| Specification | Value listed | Source |
|---|---|---|
| Standard size | Medium (11-12) | IFRC Items Catalogue |
| Length | About 40-50 cm; the listing recommends a long shaft because it protects most of the lower arm | IFRC Items Catalogue |
| Liner | Inner glove, 100% cotton | IFRC Items Catalogue |
| Lead equivalence | The listing names the item as 0.5 mm Pb in its title and gives 0.35 mm in its specification field, so compare rated values against the product’s own label and test data rather than a catalogue line | IFRC Items Catalogue (marked no longer updated, last updated 2018) |
Lead gloves versus other hand protection
A lead glove is not interchangeable with an exam glove or a cut-resistant glove: the FDA’s statement ties the attenuation claim to the x-ray energy range used in medical procedures, so a rating is meaningful only within the conditions the manufacturer’s data covers. Unlike an apron, which covers the torso (see what an apron is), a glove protects only the hand and wrist area, and dose monitoring is a separate matter covered in how dosimetry badges work.
Safe use and handling basics
- Storage: the IFRC listing gives storage below 45 °C, protected from direct light, not folded (it can be rolled).
- Labelling: the listing says the label shows product name, size, CE mark and manufacturer address.
- Follow the instructions for use: cleaning, disinfection and inspection steps come from the manufacturer’s manual supplied with the product.
Common mistakes
- Folding the gloves for storage, which the catalogue listing says not to do.
- Assuming a stated lead equivalence applies at any x-ray energy; the FDA expects the maker’s technical data to match the energy range normally used.
- Confusing a medical radiation glove with an industrial “lead” glove or a glove box.
Frequently asked questions
What are x ray gloves?
“X ray gloves” is another name for lead or lead-vinyl gloves worn to reduce scatter dose to the hands. The FDA’s formal term is radiation attenuating medical glove.
Are lead gloves regulated by the FDA?
Yes. They are a Class 1 device under product code OPH, listed under 21 CFR 880.6250 in the FDA product classification database, checked 2026-10-31.
Which standard applies?
The FDA product classification page for OPH lists ASTM D7866-23 as a recognized consensus standard. The IFRC catalogue example cites EN 61331 and carries a CE mark.
Sources
- FDA, Product Classification database, product code OPH, radiation attenuating medical glove (accessdata.fda.gov), checked 2026-10-31.
- eCFR, 21 CFR 880.6250 (ecfr.gov), checked 2026-10-31.
- IFRC Items Catalogue, “Gloves, lead, for X-ray protection” (itemscatalogue.redcross.int), checked 2026-10-31.