Written and maintained by CASRAI Editorial Board
Published
On this page
A Dewar flask is a double-walled, vacuum-insulated vessel built to hold liquefied gases such as liquid nitrogen with as little heat gain as possible. It is named after the Scottish scientist James Dewar, whose vacuum vessel was first shown at the Royal Institution on Christmas Day 1892. Laboratory staff, buyers and students searching for the term usually need to know what a Dewar is for, how the main types differ, and why an ordinary vacuum flask is not a substitute. This page describes equipment and general safe use; it does not replace your institution’s safety rules or the manufacturer’s instructions.
What is the purpose of a Dewar flask?
A Dewar flask keeps a cryogenic liquid cold by reducing heat flow into it, so the liquid can be stored or carried for a useful time. Berkeley Lab Environment, Health and Safety describes two working roles: small, narrow-necked Dewars are filled as needed and used to move liquid to cold vacuum traps or cryostats, while large, wide-necked Dewars act as storage freezers for samples. It describes unpressurized Dewars as suited to mid- to short-term storage of cryogenic liquids. Typical users are laboratory researchers, technicians and facility staff who work with liquid nitrogen. For the temperature itself, see liquid nitrogen temperature.
What does “Dewar” mean?
“Dewar” is a surname turned common noun. The Royal Institution’s object record credits James Dewar with creating the vessel in 1892 as a way to hold liquefied gases, and calls it the forerunner of the Thermos flask. The word is capitalised in many safety documents and written in lower case in vendor catalogues; both refer to the same class of vessel.
How a Dewar is built
The original design, per the Royal Institution record, is two glass flasks nested one inside the other and sealed at the neck, with a partial vacuum between them. The vacuum insulates the inner flask. The record notes that a later version used a narrower neck and a silvered outer coating to reduce heat loss. The museum example is made of glass, tin and wax and measures 350 mm high by 128 mm in diameter.
Berkeley Lab adds two practical features of laboratory Dewars. A relief valve on the vacuum space, often at the bottom centre, vents pressure if the inner reservoir cracks and liquid enters the vacuum. A cap or foam plug that slides into the neck without sealing lets boil-off gas escape, so pressure does not build up.
Types of Dewar flask
| Type | Size given by source | Typical use | Source |
|---|---|---|---|
| Small, narrow-necked Dewar | 5 L or less | Filled as needed; used to move liquid to cold vacuum traps or cryostats | Berkeley Lab EHS |
| Large, wide-necked Dewar | For example 30 L or 50 L | Storage freezers for samples | Berkeley Lab EHS |
Berkeley Lab also advises pouring from transfer Dewars of 5 L or less. For storage equipment built around cryogenic liquids, see cryogenic freezers; for a related low-temperature instrument see what is a cryostat.
Dewar flask vs Thermos or vacuum flask
A consumer vacuum flask works on the same insulation principle but is not a cryogenic vessel. Berkeley Lab states that Thermos and vacuum-flask products are not approved for cryogenic liquid applications: their lids do not vent boil-off gas, which can cause an explosion. Its guidance is to confirm that a container is rated for cryogenic service before use. That also answers “which vacuum flask is the best” for this context: the relevant test is rated cryogenic service, not a ranking of consumer products, and this page makes no brand comparison. Removing limescale from a household flask is a domestic task outside the scope of this page.
Inspecting a Dewar
Berkeley Lab lists these checks:
- Look for corrosion on the casing.
- Check that the safety valve is intact and has not been bypassed or defeated.
- Check that the cap slides easily into the neck and does not seal or stick.
- Look for ice on the outside of the Dewar. In the source’s words, ice “indicates that the Dewar may have lost vacuum”.
- Look for ice on the cap, which suggests the foam plug is broken or removed, giving less insulation and a shorter hold time.
A damaged Dewar, or one with compromised insulation, should be taken out of service and marked defective. Only trained, qualified people should repair it.
Safe use and handling basics
Berkeley Lab identifies liquid transfers as the most dangerous part of Dewar use and advises minimising them. Its general points:
- Keep the original cap, or a loose cover such as foam or foil, on the neck.
- Pour from transfer Dewars of 5 L or less.
- Fill portable flasks on the floor and let the liquid settle.
- Pre-cool flasks and traps, and pour small amounts.
- Use a spare flask for reservoirs above shoulder height.
Berkeley Lab defers protective equipment requirements to Chapter 29 of its ES&H Manual, so follow your own institution’s equivalent. See also liquid nitrogen and cryogen handling safety and glassware safety.
Common mistakes
- Using a household vacuum flask for liquid nitrogen.
- Sealing the neck tightly or defeating the safety valve.
- Ignoring external frost, which can signal lost vacuum.
- Keeping a damaged Dewar in service.
FAQ
What is the purpose of a Dewar flask?
To store or move cryogenic liquids with minimal heat gain, using a vacuum between double walls. Small Dewars serve for transfers; large ones serve as sample storage.
What does “Dewar” mean?
It is named for James Dewar, who created the vacuum vessel in 1892 according to the Royal Institution.
Can a Thermos be used instead?
No. Berkeley Lab states vacuum-flask products are not approved for cryogenic liquids because their lids do not vent boil-off.
How to remove limescale from a Thermos? Which vacuum flask is best?
Both concern household flasks and are outside this laboratory equipment page. The only cryogenic criterion the source gives is rated cryogenic service.
Sources
- Berkeley Lab EHS, Dewars, read 2026-11-08.
- Royal Institution, James Dewar’s vacuum flask (object record), read 2026-11-08.