A Hudson brace is a hand-held, non-powered holder for cranial drilling and cutting tools. The FDA does not regulate it as a manual drill or burr: the brace handpiece is a separate, lower-risk device from the cutting tools it holds, and the two are classified separately.
What a Hudson brace is
Under 21 CFR 882.4325, a cranial drill handpiece (brace) is “a hand holder, which is used without a power source, for drills, burrs, trephines, or other cutting tools that are used on a patient’s skull.” The Hudson brace is a well-known example of that device type. The brace turns by hand; the cutting tool in its chuck does the work on bone.
Who uses it: surgical teams making one or more holes in the skull (trepanation) before a bone section is removed (craniectomy). The International Red Cross item catalogue notes that familiarity with the technique is important; this page does not describe it.
Parts and burr sizes
The International Red Cross item catalogue lists a complete Hudson cranial drill set (item XSINMEDI576000; the catalogue entry was last updated 2018-12-31) as a brace, an extension piece for the brace and four burrs. The same entry gives the burr sizes below.
| Part | Shape | Size |
|---|---|---|
| Brace (handpiece) | Hand holder, no power source | One per set |
| Extension piece for the brace | Accessory | One per set |
| Burr | Cylindrical | 9 mm |
| Burr | Cylindrical | 14 mm |
| Burr | Spherical | 16 mm |
| Burr | Spherical | 22 mm |
The same entry gives the burr material as martensitic steel and the trephine material as austenitic steel, and says the set is used with a flexible De Martel or Gigli wire saw. Other manufacturers’ sets can differ from this catalogue set.
Hudson brace vs other cranial drills
The words “cranial drill”, “craniotome drill” and “Hudson brace” are often used loosely. The FDA regulations separate the pieces:
| Device | FDA regulation | Product code | Class |
|---|---|---|---|
| Cranial drill handpiece (brace) | 21 CFR 882.4325 | HBD, “Handpiece (brace), drill” | Class I, exempt from 510(k) subject to the limitations in 21 CFR 882.9 |
| Manual cranial drills, burrs, trephines and accessories | 21 CFR 882.4300 | HBG (burrs and trephines) | Class II (performance standards) |
The 882.4300 wording covers “bone cutting and drilling instruments that are used without a power source on a patient’s skull.” Both entries are about instruments used without a power source. A drill that runs from a power source is not what either wording describes. For related instruments see surgical burrs, trephines, the Midas Rex drill and the wider neurosurgical instruments overview.
FDA class and product codes
Class I means general controls apply. For the brace handpiece (HBD), 21 CFR 882.4325 also exempts the device from premarket notification (510(k)), subject to the limitations in 21 CFR 882.9. The cutting tools it holds sit in a higher class (Class II) under 21 CFR 882.4300. Regulation text checked 2026-10-31 (eCFR and Cornell LII, see Sources).
Sterilization and reuse
An AccessGUDID record for a “Hudson Brace with Snap-Lock Chuck” (device identifier 00190776375553, checked 2026-10-31) lists product code HBD, “Handpiece (brace), drill”, and states that the device is reusable (single-use: no), is not supplied sterile, and requires sterilization before use by moist heat or steam. The record includes no MRI information. Follow the manufacturer’s instructions for use for the exact model.
FAQ
Is a Hudson brace the same as a cranial drill?
A Hudson brace is one kind of cranial drill handpiece. Under 21 CFR 882.4325 it is the hand holder (Class I), while the drills, burrs and trephines it holds are regulated separately under 21 CFR 882.4300 (Class II).
Is a Hudson brace powered?
No. The regulatory definition describes a hand holder used without a power source.
What is a craniotome drill?
The term is used for cranial drilling and cutting equipment. The 21 CFR 882.4325 definition covers only a hand holder used without a power source, so it does not describe a drill driven by a power source.
Is a Hudson brace single-use?
The AccessGUDID record for a Hudson brace with snap-lock chuck lists it as reusable and not supplied sterile, so it needs sterilization before use. Other models follow their own instructions for use.
Sources
- 21 CFR 882.4325, cranial drill handpiece (brace): eCFR, checked 2026-10-31.
- 21 CFR 882.4300, manual cranial drills, burrs, trephines and accessories: Cornell LII, checked 2026-10-31.
- International Red Cross item catalogue, Cranial drill, Hudson (XSINMEDI576000), checked 2026-10-31.
- NLM AccessGUDID, Hudson Brace with Snap-Lock Chuck, checked 2026-10-31.