A pipette washer is a laboratory appliance that cleans and rinses reusable pipettes, the long, narrow glass or plastic tubes used to measure and transfer liquid, without anyone scrubbing them by hand. The pipettes stand in a holder or basket while water, and in some designs detergent and air, flows through and around them. Two designs are described in the sources below: a simple siphon washer that fills and drains with tap water, and an automated direct-injection washer that pumps water and detergent into each pipette and then dries it.
The device is for people who reuse pipettes: research laboratories, teaching laboratories and analytical or tissue culture facilities. It does not wash pipette tips for micropipettes or the micropipettes themselves; for those, see what a pipette is, the pipette tips guide and the Class A volumetric pipette guide.
What a pipette washer is
Reusable pipettes are hard to clean by hand because the working surface is the inside of a thin tube. A pipette washer solves that by moving cleaning liquid through the tube instead of relying on a brush. The washer is a piece of support equipment: it holds the pipettes upright, controls how liquid reaches them, and drains the used liquid away. Many labs treat it as part of the glassware workflow alongside safe glassware handling.
The search terms “pipette wash”, “pipette washer” and “pipette-washer” all describe the same equipment class. The sections below cover how each design works, how they compare, what the manufacturers and cleaning-product makers say about use, and where the published guidance stops.
How a pipette washer is built and how it works
Siphon washers
A siphon washer is essentially a tall container with a holder for the pipettes. Lab Manager describes this class as plastic jugs that fill with water and siphon it out several times, so the pipettes are cleaned by soaking and flushing. The Alconox technical note describes the working sequence for this kind of washer: a detergent tablet (an Alcotab) is dropped into the bottom of the siphon washer, the pipettes are placed in their holder inside it, and cold or warm water is turned on at a rate that fills the washer and covers the pipettes. The water drains to the bottom during each cycle, and the water keeps running until the tablet has fully dissolved, after which the pipettes are rinsed thoroughly. Alconox says the whole wash and rinse may take an hour.
Automated direct-injection washers
An automated washer pumps cleaning liquid into the pipettes rather than waiting for them to fill and drain. In a laboratory glassware washer, Lab Manager’s vendor-supplied article says each pipette sits in a protective plastic holder in a direct-injection basket, water and detergent are injected, and a deionized (DI) water rinse removes trace residues. Heated water is part of the design: the article says washers can heat water and DI rinse water up to 93 °C. The article also says circulation matters more than pressure, because high pressure can break delicate pipettes. A second Lab Manager piece says that with HEPA-filtered drying, forced hot air circulates through the pipettes to dry them completely.
Both Lab Manager pieces are marked as Miele content (one is a whitepaper from Miele Professional). Their performance claims are vendor claims and are attributed to Miele here.
Dedicated pipette washer-dryers
A third design is a stand-alone washer-dryer made only for pipettes. Labconco’s ScrubAir pipette washer/dryer article describes a steel unit that holds up to 60 pipettes per load and runs three lockable programs in which the user sets wash and dry duration and the number of rinses. After wash and rinse, forced-air drying starts automatically and reaches the pipettes by direct injection. The standard model has one water inlet and one pressurized-air inlet; the heated model adds a second water inlet for purified-water rinsing.
Types and variants
The table below compares how the three designs handle each step, using only statements from the cited pages. The data table at the end of the page lists the published capacities.
| Step | Siphon washer | Direct-injection basket in a glassware washer | Stand-alone washer-dryer (ScrubAir) |
|---|---|---|---|
| Cleaning action | Fills and drains by siphon; detergent tablet dissolves in the water (Alconox) | Water and detergent injected into each pipette (Lab Manager, Miele) | Water, detergent and compressed air (Labconco) |
| Rinse | Continued water rinse; distilled or deionized water for the final rinse in analytical or tissue culture work (Alconox) | Heated DI rinse (Lab Manager, Miele) | Heated model adds a purified-water inlet (Labconco) |
| Drying | Not covered by the Alconox note; Lab Manager says oven drying takes hours and wet handling breaks tips | Optional HEPA-filtered forced hot air (Lab Manager, Miele) | Forced-air drying starts automatically (Labconco) |
| Service connections | Water | The glassware washer’s own water, heat and drain | One water and one air inlet; heated model a second water inlet (Labconco) |
Capacities that are published
Capacities depend entirely on the model and the basket, so there is no single “standard size”. Figures that the sources state are listed here, each with its owner:
- Labconco ScrubAir: up to 60 pipettes per load.
- Miele A303 insert: up to 98 pipettes up to 18 inches long. Miele A202 insert: up to 38 pipettes up to 21 inches long, with space for other glassware (Lab Manager, Miele content).
- Miele E 404 basket: up to 36 pipettes in three rows; E 405 adds a drying connection for HEPA forced-air drying; E 408: up to 96 pipettes for washers with HEPA forced-air drying (Lab Manager, Miele content).
The sources give capacity as a number of pipettes and a maximum length. Because each basket is specific to a washer, the pipette length and the number of pipettes you wash are the two figures to compare against a model’s insert specification.
How to choose between designs
Published guidance supports a few comparison points rather than a ranking. This page does not rank products.
- Time and water use. Labconco states that its ScrubAir uses as little as 12.5 L for wash and rinse, against about 600 L for four hours of manual washing at 2 L/min. It states just under 5 hours for a wash-and-dry load (3.25 hours heated), against more than 23 hours for the manual method. These are manufacturer figures.
- Water per cycle. Miele says its lab washers use about 2.4 gallons per fill (undercounter) up to 9.1 gallons per fill (large chamber), and that this is less than a siphon tank uses. Miele also gives runtimes of 60 to 90 minutes, with a 50-minute pipette wash program and a recommended 30 minutes of drying. These are vendor figures.
- Breakage. Lab Manager says high pressure can break delicate pipettes, and that wet handling after soak-and-siphon makes tips easy to break. Protective holders and high circulation at low discharge rate are the design answer in a direct-injection washer.
- Final rinse water quality. Alconox says to use distilled or deionized water for the final rinse in analytical or tissue culture work. If this is your work, a design with a DI or purified-water rinse stage matches that recommendation; see the lab water purification guide for the water side.
- Detergent. Lab Manager says to use the washer manufacturer’s recommended detergent plus an acid neutralizer in a glassware washer; Alconox describes its own tablet for siphon washers.
Safe use and handling basics
- Pre-soak first. Alconox recommends submerging pipettes immediately after use in 1% Alconox or 1% Liquinox solution, so material does not dry inside the tube.
- Use the cleaning product and detergent the equipment maker specifies. Lab Manager makes this point for glassware washers.
- Handle wet pipettes carefully, because wet pipette tips break easily (Lab Manager).
- Rinse thoroughly after the detergent has dissolved; Alconox says to keep the water running until the Alcotab has fully dissolved and then keep rinsing.
- Heated rinses reach 93 °C in the washers Lab Manager describes. Follow the maker’s instructions when unloading hot glassware, and see glassware safety.
Alconox also notes that there is no CAP-approved procedure for washing pipettes. In a regulated laboratory, the cleaning procedure is therefore one the laboratory writes and validates itself, using the washer maker’s and detergent maker’s instructions.
Common mistakes
- Letting pipettes dry before soaking. The Alconox sequence starts with an immediate pre-soak after use.
- Stopping the rinse early. The siphon washer instructions continue until the detergent tablet is fully dissolved and the pipettes are rinsed thoroughly; the full process may take an hour.
- Using a high-pressure setting meant for other glassware. Lab Manager warns that pressure can break delicate pipettes.
- Using tap water for the last rinse in analytical or tissue culture work. Alconox recommends distilled or deionized water for that step.
- Treating a vendor’s time and water figures as a comparison between brands. The figures above come from individual manufacturers’ pages and use different baselines.
Frequently asked questions
What is a pipette washer used for?
It cleans and rinses reusable pipettes by moving water, and in some models detergent and air, through and around them instead of scrubbing by hand.
What is the difference between a siphon washer and an automated one?
A siphon washer fills with water and drains by siphon. An automated direct-injection washer, such as a glassware washer with a pipette basket or the Labconco ScrubAir, pumps water into each pipette and can dry them with forced air.
Is there an approved method for washing pipettes?
Alconox states that there is no CAP-approved procedure for washing pipettes.
Sources
- Alconox TechNotes: Washing pipettes, checked 2026-11-19
- Lab Manager: Getting the Pipette Cleaning Equation Right Every Time (Miele-supplied content), checked 2026-11-19
- Lab Manager: Effective Cleaning of Pipettes (Miele whitepaper), checked 2026-11-19
- Labconco: New ScrubAir Pipette Washer/Dryer, checked 2026-11-19
Data: Pipette washer designs: published capacities, cycle figures and connections
Compiled from the Alconox technical note, two Lab Manager articles supplied by Miele and the Labconco ScrubAir article. Each row is a named washer or basket with the capacity, length limit, cycle time and water figure stated by that source. Figures are manufacturer claims.
| Washer or basket | Pipettes per load (pipettes) | Maximum pipette length (inches) | Stated figure |
|---|---|---|---|
| Labconco ScrubAir (standard) | 60 | Wash and dry just under 5 h; as little as 12.5 L water | |
| Labconco ScrubAir (heated) | 60 | Wash and dry 3.25 h | |
| Miele A303 insert | 98 | 18 | Miele wash runtimes 60-90 min |
| Miele A202 insert | 38 | 21 | Space for other glassware |
| Miele E 404 basket | 36 | Three rows | |
| Miele E 408 basket | 96 | For washers with HEPA forced-air drying |
Sources
- Lab Manager: Getting the Pipette Cleaning Equation Right Every Time (accessed 2026-11-19)
- Lab Manager: Effective Cleaning of Pipettes (accessed 2026-11-19)
- Labconco: New ScrubAir Pipette Washer/Dryer (accessed 2026-11-19)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.