A qPCR plate is a multi-well microplate made of polypropylene that holds the reactions of a quantitative (real-time) PCR run in a thermal cycler, so many samples and assays are amplified and read together. The plate is built for thermal cycling and optical reading: thin-walled wells transfer heat, the format fits the instrument block, and the top is closed with an optical sealing film so fluorescence can be measured without evaporation. The two common formats are 96-well and 384-well plates, and the choices that matter are format, skirt, profile, well volume, colour and the seal.
This guide explains how a qPCR plate is built, the variants you will meet in catalogues, the specifications that vendors quote, and the handling basics that follow from vendor documentation. It is written for bench scientists, lab managers and people ordering consumables for a qPCR instrument. It sits beside the general guide to the microplate, the 96-well plate dimensions reference and the PCR tube page, and the assay side is covered in the qPCR and RT-qPCR guide.
What a qPCR plate is
A qPCR plate is the reaction vessel for real-time PCR in array form. Instead of loading individual tubes, the operator pipettes sample and master mix into numbered wells, seals the whole plate and places it in the thermal cycler. Vendors describe the same product family as a PCR plate or a qPCR plate; the product pages read for this guide list plates for standard PCR and for real-time PCR together. For example, Analytik Jena lists a 96-well plate (article 844-70292-0) for standard and real-time PCR, and Thermo Fisher Scientific lists a 96-well polypropylene PCR plate (AB0800SS) with a full skirt and low profile.
Who uses it: researchers and technicians running gene-expression, genotyping and similar assays on a real-time PCR instrument. The plate is consumable: it is used for one run and then discarded.
How a qPCR plate is built
Material and well design
The plates described by Analytik Jena are made of medical-grade polypropylene that meets USP class VI. The same vendor lists a raised well rim, alphanumeric well coding, and holes on the sides for robot handling. Thermo Fisher describes thin well walls that reduce dead space, which is the unused volume left in a well, and eight holes in the skirt that help position the plate and remove it from the block.
Single-moulded plates and two-component plates
Most plates are moulded in one piece. A second construction is described by Azenta for its FrameStar range: individual polypropylene tubes held in a rigid polycarbonate frame. According to Azenta, the frame resists distortion, and the design reduces evaporation-related inaccuracy and allows smaller reaction volumes. These are the manufacturer’s claims, not independent test results.
Footprint
Thermo Fisher states that its plate has an ANSI:SBS footprint and is stackable. ANSI/SLAS microplate standards define the outline dimensions that let plates fit instruments and automation: SLAS lists ANSI/SLAS 1-2004 (footprint dimensions), 2-2004 (height dimensions), 3-2004 (bottom outside flange dimensions), 4-2004 (well positions) and 6-2012 (well bottom elevation). The numeric values are in the standards themselves; the 96-well plate dimensions guide linked above covers the footprint.
Types and variants
Catalogues describe qPCR plates along a few independent axes. The table below lists the variants stated in the vendor pages and the Azenta article. Each choice is independent of the others, so a catalogue entry is a combination such as a 96-well, full-skirt, low-profile plate.
| Variant | Options stated by vendors | What differs | Source |
|---|---|---|---|
| Number of wells | 96 and 384 | Analytik Jena lists 0.1 mL wells for its 96-well plate and 40 μL wells for its 384-well plate; Azenta offers both formats. | Analytik Jena; Azenta |
| Skirt | Non-skirted, semi-skirted, fully skirted | The skirt is the frame around the wells. Thermo Fisher notes that holes in the skirt aid positioning and removal from the block. | Azenta; Thermo Fisher |
| Profile | Low profile and standard profile | Height of the plate; both Analytik Jena and Thermo Fisher list low-profile 96-well plates. | Azenta; Analytik Jena; Thermo Fisher |
| Colour | White (Analytik Jena), natural with clear well bottom (Thermo Fisher) | Vendors list different colours for plates in the same format. | Analytik Jena; Thermo Fisher |
| Construction | Single moulded or tubes in a frame | The FrameStar design places polypropylene tubes in a polycarbonate frame. | Azenta |
| Pre-spotted | Wells already contain assays | Each well is pre-seeded with an assay target or reference gene. | Thermo Fisher blog |
Specifications that matter
The data table below this guide compares three plates from two manufacturers: number of wells, volume as the vendor quotes it, skirt, profile and colour. Two points stand out.
First, well volume is not quoted the same way by every maker. Analytik Jena gives 0.1 mL for its 96-well plate and 40 μL for its 384-well plate. Thermo Fisher gives “0.2 mL maximum well volume (when used with adhesive or heat seals)” for its 96-well plate. A higher maximum volume is therefore not the same as a larger recommended reaction volume, and the volume depends on the seal used. Read what each number refers to in the vendor’s own document before comparing plates.
Second, the plate has to match the instrument. The 384-well plate from Analytik Jena has corner cut-outs at positions A24 and P24, and its 96-well low-profile plate has a cut-out at H1. Check the position of these cut-outs against the instrument block before a run.
Sealing a qPCR plate
A plate is only complete with its seal. Thermo Fisher lists adhesive films and foils, heat seals, and flat and domed cap strips as seals for its 96-well plate. Analytik Jena lists an optical sealing foil (844-70046-0) measuring 79 x 142 mm that is adhesive, transparent, peelable and withstands -40 °C; the vendor describes it as low-evaporation sealing. Vendors call these films optical sealing foils because fluorescence has to be read through them.
The Thermo Fisher maximum well volume applies only when the plate is used with adhesive or heat seals, so the seal and the volume should be chosen together.
Pre-spotted plates
A pre-spotted plate has each well already seeded with an assay target or reference gene, so one sample combined with master mix is added to every well, for example with a multichannel pipette. Thermo Fisher describes this format as suited to medium-sized projects with more than 10 samples, fewer than 48 assay targets, or both. It lists four types: Inventoried (more than 100), Flexible, Custom and Specialty, plus Endogenous Control plates with 32 assays in triplicate. The setup file for the plate layout is downloaded from the vendor’s TaqMan Files page.
Handling and care basics
- Certified free of contaminants. Both vendors state that their plates are certified free of DNase, RNase, human DNA and PCR inhibitors; Analytik Jena also lists DNA and endotoxins. Handling with gloves and keeping the plate sealed or covered until use are standard laboratory practice to preserve that state.
- Autoclaving. Analytik Jena states its plates are autoclavable at 121 °C for 15 minutes; Thermo Fisher also lists its plate as autoclavable. Plates are normally single use, so autoclaving is relevant mainly for waste handling under local laboratory rules.
- Storage of sealed plates. The sealing foil from Analytik Jena withstands -40 °C, a temperature figure that is relevant if a sealed plate must be held cold between setup and the run.
- Labelling. The alphanumeric well coding on the plate lets wells be identified by row and column; a plate map recorded alongside the run avoids mix-ups. See the MIQE reporting checklist for what a qPCR report should record.
Common mistakes
- Treating the maximum well volume as the reaction volume to use. The vendor figure depends on the seal, and plate makers quote volumes differently.
- Ordering a plate whose skirt or profile does not match the instrument block or the adapters in use. The skirt options and low or standard profiles are separate choices.
- Choosing a seal without checking that it is listed for the plate and for optical reading.
- Loading the plate in the wrong orientation. Check the cut-out position (for example H1 on the Analytik Jena 96-well plate, A24 and P24 on its 384-well plate) against the instrument block.
- Reading the run without understanding how amplification data become results; see Ct values in qPCR for how Ct values are interpreted.
Frequently asked questions
What is a qPCR plate?
It is a multi-well polypropylene microplate, usually 96 or 384 wells, used to hold reactions during real-time PCR in a thermal cycler. It is closed with an optical seal so fluorescence can be read during cycling.
What is the difference between a PCR plate and a qPCR plate?
Vendors often sell one plate for both uses. Analytik Jena lists its 96-well plate for standard and real-time PCR. The qPCR-specific points are the optical seal and the colour and skirt options.
Are qPCR plates reusable?
The vendor pages read for this guide describe plates sold in packs, such as 50 plates, for single runs. They do not describe a reuse procedure, so treat them as consumables.
Sources
- Analytik Jena, qPCR and PCR plates (checked 2026-11-17)
- Thermo Fisher Scientific, PCR plate AB0800SS (checked 2026-11-17)
- Azenta, qPCR results improved by novel microplate design, SelectScience, 29 August 2024 (checked 2026-11-17)
- Thermo Fisher Scientific, qPCR simplified: start with a pre-spotted plate (checked 2026-11-17)
- SLAS, ANSI/SLAS microplate standards (checked 2026-11-17)
Data: Vendor-stated specifications of three qPCR and PCR plates (Analytik Jena and Thermo Fisher)
Compares three plates from two manufacturers on well count, volume as each vendor states it, skirt, profile and colour. Compiled from the manufacturers' own product pages; values are quoted as the vendors give them and are not harmonised.
Period covered: 2026.
| Plate | Wells | Volume as stated by vendor | Skirt | Profile | Colour | Orientation cut-out |
|---|---|---|---|---|---|---|
| Analytik Jena 844-70292-0 | 96 | 0.1 mL | Full skirt | Low profile | White | H1 |
| Analytik Jena 844-70239-0 | 384 | 40 μL | Full skirt | Not stated | White | A24 and P24 (corners) |
| Thermo Fisher AB0800SS | 96 | 0.2 mL maximum (with adhesive or heat seals) | Full skirt | Low profile | Natural, clear well bottom | Not stated |
Sources
- Analytik Jena, qPCR and PCR plates (accessed 2026-11-17)
- Thermo Fisher Scientific, PCR plate AB0800SS (accessed 2026-11-17)