What Is an Intravascular Ultrasound (IVUS) Catheter? Specs and Labeling

IVUS catheter explained: FDA class and code, what a label lists (MHz, guide catheter, guidewire, length) and how it differs from other catheters.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

An intravascular ultrasound (IVUS) catheter is a thin catheter with an ultrasound transducer at its tip, used to image the inside of a blood vessel from within. FDA lists it as “Catheter, Ultrasound, Intravascular” (product code OBJ), a Class II device intended “for intracardiac and intraluminal visualization of cardiovascular anatomy and physiology.” This page is written for clinic staff, buyers and students who need to read an IVUS catheter’s label and tell it apart from neighbouring cath-lab devices.

This page describes products and general use; it is not medical advice.

What it is

An IVUS catheter carries an ultrasound transducer on its distal (patient-side) end and connects to an imaging system. Philips, which makes IVUS catheters, describes IVUS as a way to “see inside the coronary arteries” and says the imaging produces “longitudinal and cross-sectional images”; for peripheral procedures it says IVUS “provides a 360-degree view inside the vessel.” It links these images to assessment of “plaque morphology, vessel size and lesion length.” Who uses it: interventional physicians and the cath-lab teams that support them. The catheter is a single product; the console and software that display the image are separate parts of the system.

How it is built

The manufacturer page for one current model, the Philips Eagle Eye Platinum digital IVUS catheter, lists these construction features:

  • A 20 MHz transducer.
  • A soft, tapered tip.
  • A long rapid-exchange lumen, described as improving pushability.
  • GlyDx hydrophilic coating on the distal portion of the catheter, described as providing lubricity and durability.
  • Three radiopaque markers spaced 10 mm apart.

These are the features of one product family. Another manufacturer’s catheter, or another model, can differ, so the label of the exact catheter governs.

Where it sits among regulated catheter types

FDA’s classification database groups IVUS catheters under the intravascular diagnostic catheter regulation, 21 CFR 870.1200. That regulation describes a catheter “used to record intracardiac pressures,” “to sample blood, and to introduce substances into the heart and vessels,” and classifies it as Class II. The table shows how the FDA product-code entries for ultrasound-bearing catheters differ.

FDA product name (code) Regulation / class Intended use as listed by FDA Key difference
Catheter, Ultrasound, Intravascular (OBJ) 21 CFR 870.1200, Class II Intracardiac and intraluminal visualization of cardiovascular anatomy and physiology Imaging catheter
Reprocessed Intravascular Ultrasound Catheter (OWQ) 21 CFR 870.1200, Class II Intracardiac and intraluminal visualization of cardiovascular anatomy and physiology Same use as OBJ, listed as a reprocessed device
Percutaneous Catheter, Ultrasound (PPN) 21 CFR 870.1250, Class II Ultrasound wave enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral vasculature Treatment catheter, not an imaging catheter

Source for the table: FDA product classification database via openFDA, checked 2026-11-01. The general ultrasound imaging regulation, 21 CFR 892.1560, covers “a device intended to project a pulsed sound beam into body tissue” and its accessories; it is a separate regulation from the one FDA lists for IVUS catheters.

Specifications that matter

Every value below comes from the manufacturer’s product page for the Eagle Eye Platinum digital IVUS catheter, checked 2026-11-01. Philips states that it reserves the right to change product specifications without notice, and that availability depends on regulatory clearance in each country.

Specification Published value What it tells you
Frequency 20 MHz Transducer operating frequency
Minimum guide catheter 5F (I.D. 0.056″) Smallest guide catheter the label lists for this catheter
Maximum guidewire 0.014″ Largest guidewire the label lists
Working length 150 cm Usable catheter length
Maximum imaging diameter 20 mm Largest vessel diameter the system lists as imageable

How to check a catheter against the label

The manufacturer’s product page lists the guide-catheter size, guidewire size, working length and imaging diameter shown above, so those are the figures to compare against the rest of the equipment in use. The manufacturer also states: “Always read the label and follow the directions for use.”

Safe use and handling basics

  • Training: Philips states that its medical devices “should only be used by physicians and teams trained in interventional techniques, including training in the use of this device.”
  • Follow the directions for use packaged with the exact catheter. This page does not replace them.
  • Tissue characterization: for the virtual histology (VH) IVUS feature, Philips states that “safety and efficacy of VH IVUS for use in the characterization of vascular lesions and tissue types has not been established.”

Common mistakes

  • Treating one model’s figures as universal. The 20 MHz, 5F and 150 cm values above belong to one catheter family.
  • Confusing imaging with treatment. FDA lists the ultrasound-enhanced balloon catheter (PPN) as a dilatation device, not an imaging catheter.
  • Assuming product availability. The manufacturer says availability depends on country regulatory clearance.

An IVUS catheter is used alongside, not instead of, other consumables. A PTCA (percutaneous transluminal coronary angioplasty) catheter is a balloon dilatation device; FDA’s classification database lists the cutting and scoring type under product code NWX as a Class III device. Balloon inflation syringes, often called indeflators, appear in the same database as “Syringe, Balloon Inflation” (21 CFR 870.1650, Class II). For related device guides on this site, see catheter balloon sizes, extraction balloons, contrast injectors, ultrasound machines and central venous catheters.

Frequently asked questions

What product code does FDA use for an intravascular ultrasound catheter?

FDA lists “Catheter, Ultrasound, Intravascular” under product code OBJ, a Class II device under 21 CFR 870.1200. A reprocessed version is listed separately under code OWQ.

What does an IVUS catheter label list?

For the Philips Eagle Eye Platinum catheter, the product page lists a 20 MHz frequency, a 5F minimum guide catheter, a 0.014″ maximum guidewire, a 150 cm working length and a 20 mm maximum imaging diameter. Other catheters can differ, so the label of the exact product governs.

Is an ultrasound-enhanced balloon catheter the same as an IVUS catheter?

No. FDA lists the ultrasound-enhanced balloon catheter (PPN) as a dilatation device under 21 CFR 870.1250, separate from the imaging catheter under code OBJ.

Is an IVUS catheter available in every country?

The manufacturer states that product availability is subject to country regulatory clearance.

Sources

All sources checked 2026-11-01.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about What Is an Intravascular Ultrasound (IVUS) Catheter? Specs and Labeling

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →